TABLE 3

Stress Studies: Comparison of SPECT 99mTc-MIBI Uptake with 13N-Ammonia PET Perfusion (n = 23)

SegmentPETSPECT NCSPECT AC
Flow (mL/100 mL · min)%NCMD*P*ACMDP
Anteroapical195.6 ± 56.967.9 ± 12.088.7 ± 10.520.8<0.000180.6 ± 8.812.7<0.0001
Inferoapical191.0 ± 59.965.0 ± 12.883.3 ± 9.818.6<0.000181.8 ± 7.817.1<0.0001
Apical anterolateral213.6 ± 75.073.8 ± 14.194.0 ± 6.020.3<0.000189.2 ± 8.915.4<0.0001
Apical anterior207.0 ± 58.372.0 ± 12.090.5 ± 9.418.5<0.000181.8 ± 9.59.80.0011
Apical anteroseptal218.0 ± 69.475.4 ± 14.688.6 ± 8.113.10.000187.6 ± 9.311.90.0007
Apical inferoseptal219.3 ± 96.178.1 ± 16.189.0 ± 9.711.0NS95.0 ± 7.515.60.0041
Apical inferior215.6 ± 76.677.2 ± 16.578.9 ± 11.0−1.0NS87.7 ± 8.48.00.0406
Apical inferolateral212.9 ± 73.473.1 ± 12.383.2 ± 8.210.10.000186.4 ± 8.213.3<0.0001
Mid anterolateral223.5 ± 69.678.2 ± 14.792.4 ± 9.914.2<0.000190.1 ± 10.911.90.0002
Mid anterior208.4 ± 70.173.2 ± 14.188.3 ± 6.715.6<0.000185.7 ± 8.813.10.0003
Mid anteroseptal221.7 ± 77.380.6 ± 11.984.9 ± 9.04.1NS88.9 ± 8.58.30.0358
Mid inferoseptal257.6 ± 140.977.7 ± 19.179.7 ± 10.12.2NS90.4 ± 8.312.70.0305
Mid inferior255.1 ± 99.184.4 ± 12.772.7 ± 12.0−10.60.002286.9 ± 10.73.6NS
Mid inferolateral239.1 ± 86.882.0 ± 13.578.6 ± 10.7−3.2NS86.2 ± 11.75.4NS
Basal anterolateral222.3 ± 77.276.9 ± 15.181.7 ± 10.54.7NS82.4 ± 10.25.40.0421
Basal anterior221.1 ± 85.274.8 ± 14.577.1 ± 8.51.8NS79.4 ± 7.64.4NS
Basal anteroseptal200.6 ± 75.164.4 ± 13.362.6 ± 11.33.9NS65.8 ± 10.25.3NS
Basal inferoseptal216.4 ± 102.864.4 ± 17.657.8 ± 10.1−1.9NS68.1 ± 9.76.5NS
Basal inferior245.9 ± 117.077.1 ± 23.864.0 ± 11.7−12.90.015679.0 ± 12.01.0NS
Basal inferolateral191.1 ± 59.980.0 ± 13.471.2 ± 12.7−9.10.031479.1 ± 12.9−1.0NS
  • * Comparison of PET and NC.

  • Comparison of PET and AC.

  • NC = nonattenuation-corrected SPECT stress study; AC = attenuation-corrected SPECT stress study; % = studies normalized to segment with maximum flow (PET) or uptake (SPECT); MD = mean difference; NS = not significant.

    Data are given as mean of 23 studies ± SD.