TABLE 1

Main CT Technical Characteristics of B-HIREZ Scanner

CharacteristicValue
Scan modeHelical, axial, scout
Aperture (cm)70
Transverse scan field (cm)50
Maximum no. of CT slices16
Nominal slice width (mm)0.6, 0.75, 1, 1.5, 2, 3, 4, 5, 6, 7, 8, 10
Tube voltage (kV)80, 100, 120, 140
Tube current (mA)28–500
Detector materialSolid state (ultrafast ceramic [Siemens Medical Solutions])
No. of elements along z-axis24
No. of detectors/row672
Total effective length of detector array (mm)24
Rotation time (s)/360° rotation0.42, 0.5, 0.75, 1.0, 1.5
Pitch factor (volume pitch)0.5–2.0 (1–20)
Spiral scan time (s)100
Helical interpolation algorithm availableAdaptive multiplane reconstruction
Heat capacity (heat units × 106)5.3
Maximum power (kW)60