TABLE 1

Comparison of Clinical Performance Results for Conventional and Pico-3D Digital Electronics

MetricConventionalPico-3D
1R2R1R2R
Peak patient NECR (at 70 kg; kcps)29.419.359.637.3
NECR at optimal dose (at 70 kg; kcps)26.517.448.330.2
NECR at 370 MBq (at 70 kg; kcps)24.717.633.925.9
Mean optimal dose (MBq)464366717524
Mean scatter fraction0.530.530.440.44
Mean ratio of randoms to trues1.871.541.501.14