TABLE 3

Diagnostic PET/CT Acquisition Parameters for Whole-Body and Combined Head/Neck–Torso Protocol

ProtocolStandard whole-bodyNeck–Torso
Imaging rangeWhole-bodyTorsoNeck
Topogram1024 mm756 mm256 mm
CT contrast140 mL: 90 mL at 3 mL/s, 50 mL at 1.5 mL/s90 mL: 60 mL at 3 mL/s, 30 mL at 1.5 mL/s60 mL: 60 mL at 3 mL/s
CT scanSingle-spiral 130 kVp, 110 mAs 0.8-s rotation time 5-mm slice width 8-mm table feedSingle-spiral 130 kVp, 100 mAs 0.8-s rotation time 5-mm slice width 8-mm table feedSingle-spiral 130 kVp, 160 mAs 0.8-s rotation time 3-mm slice width 5-mm table feed
Emission scan3.5 min per bed 2 iterations and 8 subsets on 128 matrix with 5-mm Gaussian3 min per bed 2 iterations and 8 subsets on 128 matrix with 5-mm Gaussian6 min per bed 4 iterations and 8 subsets on 256 matrix with 3-mm Gaussian, zoom 2
CT postprocessingLung windowLung windowZoomed head