TABLE 2

Errors in Activity Estimates and in BP Values for Brain Simulation

CorrectionMethod% Error in estimate for:
ActivityBP
PutamenCaudate nucleiBackgroundPutamenCaudate nuclei
None20% + OSEM−84−87−67−60−66
AC20% + OSEM-A−51−57+5−61−65
SACTEW + OSEM-A−46−52+3−54−60
PRIM + OSEM-A−45−53+4−54−60
AC-PVC20% + OSEM + PVC1−3−3+5−9−9
20% + OSEM + PVC2−3−3+5−9−9
SAC-PVCTEW + OSEM-A + PVC1+8+8+3+5+5
PRIM + OSEM-A + PVC1+8+8+4+6+5
TEW + OSEM-A + PVC2+8+8+3+5+5
PRIM + OSEM-A + PVC2+8+8+4+6+5
  • Putamen and caudate nucleus percent errors were averaged over right and left structures.