TABLE 5

PET/CT Scanners: Main Technical Parameters Relative to CT Scanners Used (Factory Data)

ParameterPhilips GEMINICTI REVEAL/CPS BIOGRAPHGeneral Electric DISCOVERY LS
PET scannerALLEGROECAT EXACT HR+ or ECAT ACCELADVANCE Nxi
CT scannerMX 8000 DUALSOMATOM EMOTION DUOLIGHTSPEED PLUS
 Maximum power (kW)604060
 Maximum tube voltage (kV)140130140
 Maximum tube current (mA)500240440
 Selectable kV values90, 120, 14080, 100, 13080, 100, 120, 140
 mA range30–50030–24010–440
 Heat capacity (MHU)6.53.56.3
 Slices/rotation224
 Slice number × thickness (mm)2 × 0.625 (axial only)
2 × 0.52 × 11 and 4 × 1.25
2 × 12 × 1.54 × 2.5
2 × 2.52 × 2.54 × 3.75
2 × 52 × 44 × 5
2 × 82 × 52 × 7.5
2 × 102 × 10
 Pitch minimum0.3750.250.75
 Pitch maximum221.5
 Angular speed (s/rotation)0.5, 0.75, 1, 1.5, 20.8, 1, 1.50.5, 0.6, 0.7, 0.8, 0.9, 1, 2, 3, 4
 Maximum scan time (s)60100120
 Detectors array672 × 2672 × 2912 × 16
(1,344 elements)(1,344 elements)(14,592 elements)
 Transverse FOV (mm)500500500
  • MHU = Mega Heat Units.