TABLE 3

State-of-the-Art PET Scanners: Physical Performance (Factory Data)

PerformanceALLEGRO (Philips-ADAC)C-PET (Philips-ADAC)ECAT ART (CTI-Siemens)ECAT EXACT (CTI-Siemens)ECAT EXACT HR+ (CTI-Siemens)ECAT ACCEL (CTI-Siemens)ADVANCE/ADVANCE Nxi (General Electric)
Transaxial resolution
 FWHM (mm) at 1 cm4.6 (2D)6.2 (2D)
4.8*5.0*6.26.0 (2D/3D)4.5 (3D)*6.3 (3D)*4.8 (2D/3D)
 FWHM (mm) at 10 cm5.4 (2D)6.7 (2D)
5.9*6.4*6.96.7 (2D/3D)5.6 (3D)*7.4 (3D)*5.4 (2D/3D)
Axial resolution
 FWHM (mm) at 0 cm4.5 (2D)4.2 (2D)4.3 (2D)4.0 (2D)
5.4*5.5*4.94.6 (3D)4.2 (3D)*5.8 (3D)*4.7 (3D)
 FWHM (mm) at 10 cm5.9 (2D)5.0 (2D)6.0 (2D)5.4 (2D)
6.5*5.9*6.66.5 (3D)5.7 (3D)*7.1 (3D)*6.3 (3D)
System sensitivity4.9 (2D)5.4 (2D)5.4 (2D)5.4 (2D)
(net trues) (cps/Bq/mL)19.0*12.1*7.321.1 (3D)24.3 (3D)27.0 (3D)31.0 (3D)
Scatter fraction (%)16 (2D)17 (2D)16 (2D)10 (2D)
25253636 (3D)36 (3D)36 (3D)35 (3D)
  • * Assessed according to NEMA NU 2-2001 (11).

  • Measured at 300-keV LLD in high sensitivity mode.

  • Measured at 300-keV LLD.

  • All parameters were measured following NEMA NU 2-1994 (10) whenever not specified.