TABLE 1

Change of STR and BKG Activity with Attenuation and Scatter Correction in Phantom Experiment

Observed STR/BKG ratioActivity (kBq/mL)
STRBKG
TrueAC and SC (% bias)AC and no SC (% bias)TrueAC and SC (% bias)AC and no SC (% bias)
Zero-BKG168.0164.9 (−1.8%)96.7 (−42.4%)000.4
19.7:1168.0162.8 (−3.1%)104.5 (−37.8%)8.68.7 (+1.2%)9.0 (+4.7%)
9.7:1168.0157.7 (−6.1%)108.8 (−35.2%)17.317.9 (+3.5%)18.2 (+5.2%)
4.8:1168.0154.6 (−8.0%)112.7 (−32.9%)34.835.3 (+1.4%)35.3 (+1.4%)
1.9:1168.0149.1 (−11.3%)124.9 (−25.7%)87.984.9 (−3.4%)84.6 (−3.8%)
1:1168.0145.4 (−13.5%)138.5 (−17.6%)177.0165.7 (−6.4%)164.6 (−7.0%)
Total1008.0934.5 (−7.3%)686.1 (−31.9%)325.6312.5 (−4.0%)312.1 (−4.1%)
  • AC = attenuation correction; SC = scatter correction.