Results by Subject
Subject no. | Prior lung therapy | Time between PET and CT scans (d) | PET EM tumor volume (mm3) | CT tumor volume (mm3) | EM centroid loci by | CT centroid (reference) | Centroid error regarding CT (mm) | Distance between methods A and B (mm) | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Registration of TR scan* | Direct registration of EM scan† | |||||||||||||||||||||||
z (mm) | y (mm) | x (mm) | z (mm) | y (mm) | x (mm) | z (mm) | y (mm) | x (mm) | Method A | Method B | ||||||||||||||
1 | None | 0 | 225 | 626 | −86.1 | 247.2 | 172.5 | −92.4 | 246.7 | 172.2 | −89.8 | 251.2 | 169.7 | 6.2 | 5.7 | 6.3 | ||||||||
2 | None | 0 | 579 | 675 | −75.0 | 25.1 | −27.2 | −75.0 | 21.7 | −26.0 | −80.2 | 24.5 | −27.2 | 5.3 | 6.1 | 3.6 | ||||||||
3 | None | 0 | 2,789 | 5,758 | −117.8 | 246.9 | 163.8 | −117.2 | 251.0 | 160.1 | −115.6 | 230.9 | 167.9 | 16.7 | 21.6 | 5.5 | ||||||||
4 | None | 0 | 5,708 | 5,951 | −128.4 | −40.4 | 74.7 | −131.1 | −39.3 | 72.8 | −129.7 | −39.8 | 77.0 | 2.6 | 4.4 | 3.5 | ||||||||
5 | None known | 0 | 6,342 | 4,888 | −56.7 | −36.7 | −56.7 | −55.8 | −34.7 | −59.8 | −58.4 | −38.6 | −65.9 | 9.6 | 7.7 | 3.8 | ||||||||
6 | None known | 0 | 6,932 | 9,406 | −146.4 | −3.3 | 45.0 | −154.2 | −8.4 | 45.3 | −147.3 | −9.9 | 52.2 | 9.8 | 9.9 | 9.2 | ||||||||
7 | None | 0 | 8,138 | 5,645 | −33.0 | −28.1 | 42.3 | −24.3 | −37.8 | 41.7 | −38.2 | −30.3 | 43.8 | 5.8 | 16.0 | 13.1 | ||||||||
8 | None | 0 | 9,400 | 3,827 | −166.5 | 246.3 | 112.9 | −175.2 | 246.0 | 111.1 | −169.3 | 246.1 | 110.1 | 3.9 | 6.0 | 8.9 | ||||||||
9 | None | 5 | 9,629 | 5,339 | −60.2 | −18.8 | −47.1 | −60.9 | −22.5 | −43.1 | −56.9 | −20.7 | −48.0 | 3.9 | 6.6 | 5.5 | ||||||||
10 | None | 0 | 12,771 | 6,197 | −29.4 | 272.1 | 185.4 | −41.4 | 260.7 | 192.3 | −27.7 | 268.3 | 188.6 | 5.2 | 16.1 | 17.9 | ||||||||
11 | None | 1 | 14,844 | 9,720 | −69.8 | 16.4 | 53.8 | −69.5 | 18.5 | 50.8 | −73.2 | 12.3 | 54.0 | 5.4 | 7.9 | 3.7 | ||||||||
12 | None known | 0 | 15,386 | 8,448 | −52.3 | 0.1 | −74.5 | −52.8 | −0.1 | −76.6 | −50.6 | −0.8 | −73.2 | 2.2 | 4.0 | 2.1 | ||||||||
13 | None | 0 | 18,025 | 9,551 | 22.3 | −27.5 | −100.1 | 12.2 | −59.9 | −97.0 | 22.6 | −28.2 | −92.5 | 7.6 | 33.6 | 34.0 | ||||||||
14 | None | 0 | 120,008 | 154,867 | −75.6 | −13.8 | 52.1 | −77.7 | −13.1 | 52.0 | −77.2 | −14.5 | 54.3 | 2.8 | 2.7 | 2.2 |