Depth (cm) | % Source dose | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
10-g tumor | 20-g tumor | 40-g tumor | |||||||||
131I | 90Y | 131I | 90Y | 131I | 90Y | ||||||
Source | 100 | 100 | 100 | 100 | 100 | 100 | |||||
0.0025 | 46 | 61 | 46 | 58 | 47 | 56 | |||||
0.01 | 32 | 56 | 33 | 54 | 34 | 52 | |||||
0.02 | 22 | 52 | 24 | 50 | 24 | 48 | |||||
0.03 | 17 | 48 | 18 | 46 | 19 | 44 | |||||
0.04 | 13 | 44 | 14 | 42 | 15 | 41 | |||||
0.05 | 10 | 42 | 11 | 39 | 12 | 38 | |||||
0.06 | 8 | 39 | 9 | 37 | 10 | 36 | |||||
0.07 | 7 | 36 | 8 | 35 | 9 | 34 | |||||
0.08 | 6 | 34 | 7 | 33 | 8 | 31 | |||||
0.09 | 5 | 32 | 6 | 31 | 8 | 30 | |||||
0.10 | 5 | 30 | 6 | 28 | 7 | 28 | |||||
0.15 | 4 | 22 | 5 | 21 | 6 | 20 | |||||
0.25 | 3 | 11 | 4 | 11 | 6 | 11 | |||||
0.40 | 3 | 4 | 4 | 4 | 5 | 4 | |||||
0.60 | 2 | 1 | 3 | 1 | 4 | 1 | |||||
S-value in source (μGy/MBq·s) | 3.15 | 12.00 | 1.58 | 6.38 | 0.83 | 3.30 |