TABLE 6

Reproducibility of Image Quality Results: CRC

Diameter (cm)Scan 1Scan 2Average% Diff.
1.00.0470.0760.06148
1.30.160.190.1718
1.70.240.220.238.7
2.20.420.440.434.6
2.80.480.450.466.5
3.70.480.470.482.1
  • Data were acquired on ADAC UGM C-PET scanner. Percentage difference (% Diff.) was calculated as absolute difference between 2 scans, divided by average value.