TABLE 2

Impact of Inadequate Axial Sampling on Axial Resolution Measurement

Axial positionFWHM (mm)
Meas. 1Meas. 2
A8.36.8
B8.57.8
C6.16.5
D6.06.2
Average (A, B)8.47.3
Average (C, D)6.056.35
Average (A, B, C, D)7.2 ± 1.46.8 ± 0.7
  • Results from 2 separate measurements (Meas. 1 and Meas. 2), taken on GE Advance scanner in 3D acquisition mode, are shown. Axial position A was at center of axial FOV, whereas position B was ¼ of axial FOV from center. Positions C and D were ½ slice separation from positions A and B, respectively.