TABLE 3

Estimates for Operator-Received Doses Caused by Different Radioisotopes for Coronary Radiation Therapy: Comparison with Fluoroscopy

Source activity (GBq)γ-Energy (keV)Quality of primary radiationDwell time (min)Operator absorbed dose (μSv)
Positron emitter1.85511β+46
188Re1.85172β−, γ42
192Ir7.4380γ30195*
X-ray (PTCA)X-rayNot reported50–160
  • * Measured at neck collar level for coronary angiography, including single-vessel angioplasty (15).

  • Dose calculated under assumption that operator would remain beside patient during entire dwell time (not current practice).