TABLE 2

Scatter Correction in ROIs Measured as Scatter Fraction Coefficient k Under Uniform and Nonuniform Radiolabeled Aerosol Distributions

SourceProjectionLung regionCentral region
UniformAnterior0.400 ± 0.0230.349 ± 0.020
UniformPosterior0.465 ± 0.0340.451 ± 0.016
UniformArithmetic mean0.433 ± 0.0280.400 ± 0.017
NonuniformAnterior0.429 ± 0.0350.345 ± 0.017
NonuniformPosterior0.467 ± 0.0310.470 ± 0.005
NonuniformArithmetic mean0.448 ± 0.0320.408 ± 0.008
  • Values are mean ± SD.