TABLE 1

Parameters for 2 Phantom Experiments

ParameterUnitsGoalExperiment 1*Experiment 2
TrueMeasured§TrueMeasured§
LungskBq/mL4.53.331.313.79.62
Heart chamberkBq/mL1813.017.115.215.3
Heart wallkBq/mL5230.724.066.649.2
LiverkBq/mL2114.216.822.822.6
Phantom backgroundMBq48.1122
HeadMBq81.4100
Head positioncm38.015.0
BladderMBq9551,200
Bladder positioncm3315
True eventsMegacounts1034.3
Random eventsMegacounts61.814.4
  • * Experiment was started 70 min late and so corresponds to 1070 MBq injected rather than 1480 MBq.

  • Values determined from recorded counting rates or ROIs from 2D scanning (1480 MBq 82Rb) of typical human subject.

  • True radioactivity used to fill phantom inserts; decay corrected to first PET acquisition.

  • § Radioactivity determined by drawing ROIs on first phantom PET images.

  • Distance from end of thoracic phantom to center of 3-L cylinder used to simulate head and bladder radioactivity.

  • ROI = region of interest.