TABLE 5

Interobserver and Intraobserver Variability for V/Q Scintigraphy with Technegas

VariabilityConsensusObserver 2Observer 1
MismatchMatchMismatchMatchMismatchMatch
Local
    Mismatch375356355
    Match241235531
Observer 1
    Mismatch38638522
    Match236235222
Observer 2
    Mismatch365221
    Match43925
Technegas V/Q scintigraphynκ (95% CIs)McNemar
Local vs. consensus850.84  (0.72–0.95)0.45
Observer 1 vs. consensus820.81  (0.68–0.93)0.29
Observer 2 vs. consensus840.79  (0.65–0.92)1.0
Observer 1 vs. local760.74  (0.58–0.89)1.0
Observer 2 vs. local780.80  (0.66–0.93)0.29
Observer 1 vs. observer 2800.83  (0.70–0.95)0.45
Observer 1 vs. observer 1460.91  (0.80–1.0)0.50
Observer 2 vs. observer 2480.96  (0.88–1.0)1.0
  • Number of patients varies because of variation in indication for ventilation. Interobserver agreement is given in lightface type. Numbers in boldface type indicate intraobserver agreement.