TABLE 3.

Mean Quantification Error for Emission Versus Transmission Template Matching for Perfusion and Receptor Ligand Studies Based on Count Difference Registration Accuracy

Parameter[99mTc]ECD[123I]IBZM[123I]β-CIT[123I]R91150
ECTTCTECTTCTECTTCTECTTCT
Translation
 Δx1.61.70.90.93.01.42.01.9
 Δy1.41.62.72.89.81.42.42.1
 Δz1.21.11.41.619.83.23.63.2
 All Δ2.42.63.23.322.33.84.84.3
Rotation
 Φxy0.30.20.60.75.51.51.90.7
 Φxz0.20.10.80.710.51.61.30.3
 Φyz0.30.21.20.811.21.64.61.5
 All Φ0.50.31.61.316.32.75.11.7
Scaling
 Σx0.31.00.71.32.91.51.11.0
 Σy0.80.50.71.07.00.81.71.0
 Σz0.70.51.21.44.23.55.83.0
 All Σ1.11.21.62.28.03.96.13.3
All dimensions2.72.93.94.228.96.19.35.7
Significance (paired t test)NS (P = 0.9)NS (P = 0.2)P = 0.006P = 0.021
  • NS = not significant.

  • All values are expressed as percentages.