TABLE 5

Image Quality Phantom Results for 4:1 Lesion-to-Background Ratio

Sphere diameter (mm)Contrast recovery coefficientCoefficient of variation
3 min6 min12 min3 min6 min12 min
100.056 ± 0.0370.038 ± 0.0310.035 ± 0.02313.00 ± 1.109.77 ± 1.078.23 ± 1.63
130.126 ± 0.0320.115 ± 0.0110.119 ± 0.02513.40 ± 1.1010.00 ± 1.308.34 ± 1.71
170.212 ± 0.0820.192 ± 0.0330.196 ± 0.01111.20 ± 0.908.41 ± 0.917.27 ± 1.53
220.272 ± 0.0260.264 ± 0.0180.265 ± 0.01210.80 ± 1.208.38 ± 0.617.41 ± 1.28
280.494 ± 0.0200.515 ± 0.0430.515 ± 0.0209.56 ± 0.757.23 ± 0.526.54 ± 1.12
370.542 ± 0.0450.571 ± 0.0320.569 ± 0.0178.49 ± 0.636.61 ± 0.446.20 ± 0.81
Foam insert0.605 ± 0.0250.596 ± 0.0050.596 ± 0.0208.02 ± 0.676.51 ± 0.416.17 ± 0.68
  • Data are mean ± SD.