TABLE 4

Image-Quality Phantom Results for 8:1 Lesion-to-Background Ratio

Sphere diameter (mm)Contrast recovery coefficientCoefficient of variation
3 min6 min12 min3 min6 min12 min
100.060 ± 0.0180.059 ± 0.0200.059 ± 0.01314.0 ± 0.9011.3 ± 1.009.13 ± 0.63
130.194 ± 0.0480.189 ± 0.0130.190 ± 0.00514.4 ± 1.0011.6 ± 0.909.35 ± 0.56
170.155 ± 0.0070.153 ± 0.0020.153 ± 0.00112.0 ± 0.909.72 ± 0.848.03 ± 0.62
220.342 ± 0.0250.352 ± 0.0250.353 ± 0.01611.4 ± 0.909.09 ± 0.727.62 ± 0.58
280.391 ± 0.0700.446 ± 0.0710.446 ± 0.04810.0 ± 0.608.02 ± 0.676.88 ± 0.60
370.624 ± 0.0480.615 ± 0.0320.617 ± 0.0218.25 ± 0.506.84 ± 0.736.10 ± 0.62
Foam insert0.677 ± 0.0190.693 ± 0.0230.694 ± 0.0157.40 ± 0.516.35 ± 0.715.78 ± 0.58
  • Data are mean ± SD.