TABLE 2.

Diagnostic Performance of 99mTc-Sestamibi Breast Imaging as Function of Patients' Age

ParameterSensitivity (%)Specificity (%)PPV (%)NPV (%)Accuracy (%)
≤50 y old>50 y old≤50 y old>50 y old≤50 y old>50 y old≤50 y old>50 y old≤50 y old>50 y old
Overall75.375.578.788.358.187.389.177.177.881.7
(61/81)(117/155)(163/207)(128/145) (61/105)(117/134)(163/183)(128/166)(224/288)(245/300)
Palpable abnormalities
 Institutional84.987.575.776.964.388.690.775.078.884.0
(45/53)(70/80) (78/103)(30/39)(45/70)(70/79)(78/86)(30/40)(103/156)(100/119)
 Reader 170.879.277.588.263.093.483.166.775.282.1
(34/48)(57/72)(69/89)(30/34)(34/54)(57/61)(69/83)(30/45)(103/137) (87/106)
 Reader 272.378.781.991.469.495.284.066.778.582.7
(34/47)(59/75)(68/83)(32/35)(34/49)(59/62)(68/81)(32/48)(102/130) (91/110)
 Reader 371.477.983.889.568.693.885.666.779.781.7
(35/49)(60/77)(83/99)(34/38)(35/51)(60/64)(83/97)(34/51)(118/148) (94/115)
Nonpalpable abnormalities
 Institutional57.162.781.792.545.785.587.677.876.580.1
(16/28)(47/75) (85/104) (98/106)(16/35)(47/55)(85/97) (98/126)(101/132)(145/181)
 Reader 450.052.888.898.153.895.087.275.780.780.0
(14/28)(38/72) (95/107)(106/108)(14/26)(38/40) (95/109)(106/140)(109/135)(144/180)
 Reader 533.345.594.299.157.196.885.874.682.778.6
 (8/24)(30/76) (97/103)(106/107) (8/14)(30/31) (97/113)(106/142)(105/127)(136/173)
 Reader 644.856.687.493.950.086.084.976.478.078.9
(13/29)(43/76) (90/103)(107/114)(13/26)(43/50) (90/106)(107/140)(103/132)(150/190)