TABLE 2.

Resolution Estimates Obtained Using MFA and RERC Equation for a Range of PET and SPECT Radionuclides

Radionuclides (collimator)ModelSBRResolution modelingFilterVoxel size (xy × z mm)RERC (FWHM mm)*MFA resolution (FWHM mm)Relative difference (%)
18FQuadraYesN/A1.652 × 1.653.3 (3.0–3.6)3.711.0
18FQuadraNoN/A1.652 × 1.654.3 (4.0–4.7)4.54.2
18FQuadra8:1YesN/A1.652 × 2.004.8 (4.4–5.4)4.90.9
18FQuadra8:1NoN/A1.652 × 2.005.5 (5.0–5.9)5.50.2
68GaQuadra8:1Yes3 mm1.652 × 2.006.2 (5.9–6.6)6.42.9
68GaQuadra8:1NoN/A1.652 × 2.006.5 (6.3–6.9)6.84.9
68GaBio mCT8.5:1Yes5 mm4.072 × 2.007.7 (7.2–8.3)8.26.5
99mTc (LEHR)Intevo 68.5:1NoN/A4.802 × 4.8012.0 (9.8–13.6)12.97.0
177Lu (MELP)Intevo 68.8:1NoN/A4.802 × 4.8018.1 (15.0–19.4)17.4−4.1
177Lu (MEGP)GE 870 DR8.5:1YesN/A4.422 × 4.4211.1 (8.3–14.6)10.9−1.9
  • * Values are mean with continuous data being minimum to maximum range.

  • N/A = not applicable; LEHR = low-energy high-resolution; MELP = medium-energy low-penetration; MEGP = medium-energy general-purpose.