TABLE 3.

MRD% of PVC Activity Concentrations Relative to Nominal Ground Truth Activity Concentration (100 kBq/mL) for 99mTc and 177Lu Inserts and Average RERC for Each Shape Using RECOVER-EM Model

RadionuclideShapeUncorrected (MRD%)EANM 2-PL (MRD%)RECOVER-EM (MRD%)RECOVER-GM (MRD%)RERC average resolution (mm)
99mTcSphere20.66.62.12.19.8
Prolate56.36.855.19.9
Oblate67.420.53.73.69.6
177LuSphere37.15.93.53.515
Prolate81.59.95.55.215.2
Oblate89.621.75.45.913.8