uMI Panorama GS System Specifications
Parameter | Specification |
---|---|
Crystal size | 2.76 × 2.76 × 18.1 mm3 |
Microblock | 3 × 3 crystal array coupled to 2 × 2 SiPM with pitch size of 2.85 mm |
Detector cell | 8 × 8 microblocks (24 × 24 crystal array) |
Detector unit | 7 cells along axial direction |
34 cells in transaxial direction | |
System | 3 detector units |
PET transaxial FOV | 76.0 cm |
PET AFOV | 148.2 cm |
Energy window | 430–650 keV |
Time window | 4.7, 6.2, 8.3 ns for unit difference of 0, 1, 2 |
CT slices | 160 |
Bore length with CT | 2,263 mm |
Maximal patient weight | 250 kg |
CT generator power | 100 kW |
CT minimal slice spacing | 0.5 mm |
Total system length | 6,739 mm |
System weight | 6,100 kg |
SiPM = silicon photomultiplier.