TABLE 1.

Spatial Resolution of 22Na Point Source Measured with Filtered Backprojection

Axial positionRadial off-center position (cm)Full width at half maximum (mm)
RadialTangentialAxial
½ AFOV11.631.682.55
102.102.062.97
202.172.872.82
⅛ AFOV11.761.472.91
101.992.082.80
202.202.783.05