TABLE 2.

Measured Contrast Recovery Coefficient and Background Variability for 3 and 7 Iterations of OSEM Reconstruction

Sphere size
Parameter10 mm13 mm17 mm22 mm28 mm37 mmLung residual error
Contrast recovery coefficient
 3-iteration OSEM
  Mean74.582.182.686.189.492.61.7
  Maximum76.184.384.787.089.893.11.7
  Minimum72.680.381.585.088.991.61.7
 7-iteration OSEM
  Mean83.385.487.290.292.394.80.9
  Maximum85.687.990.391.792.995.50.9
  Minimum80.982.284.988.291.693.40.9
Background variability
 3-iteration OSEM
  Mean3.12.41.91.61.41.31.4
  Maximum3.42.62.01.71.51.31.4
  Minimum2.62.11.81.51.41.31.4
 7-iteration OSEM
  Mean5.03.62.82.21.91.61.9
  Maximum5.43.93.02.42.01.71.9
  Minimum4.23.12.52.11.81.51.8
  • Results are reported as mean, minimum, and maximum percentages over 3 measurements.