TABLE 1.

Key Parameters of Total-Body PET/CT Acquisition Protocols at SYSUCC

CT-AC parametersCT-Dx parametersReconstruction parameters of PET
AgeProtocolDose (MBq/kg)Time (min)Tube voltage (kV)Fixed tube (mAs)Tube voltage (kV)Ref.mAs of TCM (mAs)Rotation time (s)IterationsSubsetsMatrixFOV (mm)
>20 y, torso or TB routineTorso routine3.70512010120800.5320256 × 256600
Torso NPC3.7031205120800.5220256 × 256600
TB routine1.858NANA1201200.5320256 × 256600
TB NPC1.856NANA1201200.5320256 × 256600
≤20 y, TB routine only0- to 4-y nonlymphoma1.8510NANA100750.3320512 × 512600
0- to 4-y lymphoma1.8510NANA100700.3320512 × 512600
0- to 4-y lymphoma follow-up1.8510NANA100550.3320512 × 512600
4- to 7-y nonlymphoma1.8510NANA100800.3320512 × 512600
4- to 7-y lymphoma1.8510NANA100700.3320512 × 512600
4- to 7-y lymphoma follow-up1.8510NANA100600.3320512 × 512600
7- to 15-y nonlymphoma1.8510NANA120600.3320512 × 512600
7- to 15-y lymphoma1.8510NANA120550.3320512 × 512600
7- to 15-y lymphoma follow-up1.8510NANA120500.3320512 × 512600
15- to 20-y nonlymphoma3.706NANA120650.5320360 × 360600
15- to 20-y lymphoma3.706NANA120600.5320360 × 360600
15- to 20-y lymphoma follow-up3.706NANA120550.5320360 × 360600
>45-kg nonlymphoma3.70/1.856/10NANA120600.5320256 × 256600
>45-kg lymphoma3.70/1.856/10NANA120500.5320256 × 256600
>45-kg lymphoma follow-up3.70/1.856/10NANA120400.5320256 × 256600