Accuracy of 18F-FDG PET/CT Vs. 68Ga-FAPI PET/CT for Nodal Staging in 18 Patients
Tracer | Malignant | Benign | Statistics |
---|---|---|---|
18F-FDG | |||
Positive | True positive, n = 4 | False positive, n = 6 | Positive predictive value, 56% |
Negative | False positive, n = 2 | True negative, n = 6 | Negative predictive value, 89% |
Statistics | Sensitivity, 66% | Specificity, 50% | Accuracy, 55% |
68Ga-FAPI | |||
Positive | True positive, n = 6 | False positive, n = 1 | Positive predictive value, 86% |
Negative | False positive, n = 0 | True negative, n = 11 | Negative predictive value, 100% |
Statistics | Sensitivity, 100% | Specificity, 92% | Accuracy, 94% |
Accuracy was tabulated on per-patient basis (1 node per patient).