TABLE 3

Quantitative Evaluation of Different Gyroid Structures Inside Cuboid Phantom

IterationsNo postprocessing filter10-mm gaussian20-mm gaussian
Activity concentration G1 (1:1) gyroid (HPGe-based): 0.64
 120.810.740.58
 240.790.740.58
 480.760.710.57
Activity concentration G2 (1:3.5) gyroid (HPGe-based): 0.50
 120.590.550.46
 240.570.550.46
 480.540.520.45
Activity concentration G3 (1:7.5) gyroid (HPGe-based): 0.29
 120.330.310.26
 240.320.310.26
 480.290.290.25
  • Vertically are shown different numbers of iterations; horizontally are shown postfilters of different full widths at half maximum. All activity concentrations are given in MBq/mL.