RT Journal Article SR Electronic T1 Data Acquisition Using a Scintillation Detector Interfaced to a Personal Microcomputer JF Journal of Nuclear Medicine JO J Nucl Med FD Society of Nuclear Medicine SP 85 OP 87 VO 26 IS 1 A1 Lawrence Davis A1 John P. Wexler A1 Ari Rabinowitz A1 Thierry LeJemtel A1 Mi-Kyung Lee A1 L. Rao Chervu A1 M. Donald Blaufox YR 1985 UL http://jnm.snmjournals.org/content/26/1/85.abstract AB A method is described for interfacing a cadmium telluride semiconductor nonimaging detector to a personal microcomputer in order to store and display nuclear medicine data. There was virtual identity between the count rates stored in the computer and those recorded from the detector's display, demonstrating that the computer accurately acquired data from the probe without erroneous loss or addition of data. Interfacing a nonimaging detector to a microcomputer may provide an extremely versatile method of acquiring, storing, and displaying nuclear medicine data.