RT Journal Article SR Electronic T1 Multiple-Hospital Survey of Ejection-Fraction Variability Using a Cardiac Phantom JF Journal of Nuclear Medicine JO J Nucl Med FD Society of Nuclear Medicine SP 81 OP 84 VO 26 IS 1 A1 Makler, P. T. A1 McCarthy, David M. A1 Bergey, Philip A1 Marshall, Kenneth A1 Bourne, Mark A1 Velchik, Michael A1 Alavi, Abass YR 1985 UL http://jnm.snmjournals.org/content/26/1/81.abstract AB A dynamic cardiac phantom was used to provide identical input data at 11 different nuclear medicine laboratories throughout the Philadelphia area, and the variability in the resulting calculations of ejection fraction (EF) was assessed. The variability observed between different operators using the same computer system averaged 3 EF units, which was similar to that between different observers using different types of computers. In the range of low ejection fractions, however, there was a suggestion that EFs calculated with an MDS computer were slightly lower than those from a DEC computer.