RT Journal Article SR Electronic T1 Measurement of myocardial oxygen extraction fraction using bolus inhalation of 15O-oxygen and dynamic PET JF Journal of Nuclear Medicine JO J Nucl Med FD Society of Nuclear Medicine SP 652 OP 652 VO 50 IS supplement 2 A1 Lubberink, Mark A1 Wong, Yeun Ying A1 Knaapen, Paul A1 Raijmakers, Pieter A1 Luurtsema, Gert A1 Vonk-Noordegraaf, Anton A1 Lammertsma, Adriaan YR 2009 UL http://jnm.snmjournals.org/content/50/supplement_2/652.abstract AB 652 Objectives The aim of this study was to determine accuracy of oxygen extraction fraction (OEF) measurements using either a single scan after bolus inhalation of 15O2 to determine OEF, flow (MBF), and perfusable tissue fraction (PTF) or a 15O2 scan combined with a separate H215O scan for determination of MBF and PTF. Methods Simulations were performed to assess accuracy and precision of OEF when determining (A) all parameters from 15O2, (B) OEF from 15O2 and MBF and PTF from H215O, and (C) all parameters from 15O2, but restricting MBF and PTF to within ±10% of values derived from H215O to account for minor changes between the scans. Three patients underwent a 10 min scan on an ECAT Exact HR+ scanner after bolus injection of 1.1 GBq H215O, two 10 min scans after bolus inhalation of ~ 1.4 GBq 15O2, and a 6 min static scan after bolus inhalation of ~ 0.8 GBq C15O for ROI definition. Results Simulations showed highest correlation between simulated and true OEF for model B (r2 0.96 vs. 0.56 and 0.88 for A and C, respectively), with no bias. If small differences in parameters were present between H215O and 15O2 scans, correlation was best for model C (r2 0.88 vs 0.56 and 0.82 for A and B). In patients best fits were obtained for model B. Whole myocardium test-retest accuracy of OEF was 14±7%. Conclusions Best accuracy and precision for myocardial OEF measurements following bolus inhalation of 15O2 are obtained when other necessary parameters, such as MBF, are derived from an additional H215O scan.