RT Journal Article SR Electronic T1 Performance Evaluation of a New Dedicated Breast PET Scanner Using NEMA NU4-2008 Standards JF Journal of Nuclear Medicine JO J Nucl Med FD Society of Nuclear Medicine SP 1198 OP 1203 DO 10.2967/jnumed.113.131565 VO 55 IS 7 A1 Kanae K. Miyake A1 Keiichi Matsumoto A1 Mika Inoue A1 Yuji Nakamoto A1 Shotaro Kanao A1 Tae Oishi A1 Shigeto Kawase A1 Keishi Kitamura A1 Yoshiyuki Yamakawa A1 Ayako Akazawa A1 Tetsuya Kobayashi A1 Junichi Ohi A1 Kaori Togashi YR 2014 UL http://jnm.snmjournals.org/content/55/7/1198.abstract AB The aim of this work was to evaluate the performance characteristics of a newly developed dedicated breast PET scanner, according to National Electrical Manufacturers Association (NEMA) NU 4-2008 standards. Methods: The dedicated breast PET scanner consists of 4 layers of a 32 × 32 lutetium oxyorthosilicate–based crystal array, a light guide, and a 64-channel position-sensitive photomultiplier tube. The size of a crystal element is 1.44 × 1.44 × 4.5 mm. The detector ring has a large solid angle with a 185-mm aperture and an axial coverage of 155.5 mm. The energy windows at depth of interaction for the first and second layers are 400–800 keV, and those at the third and fourth layers are 100–800 keV. A fixed timing window of 4.5 ns was used for all acquisitions. Spatial resolution, sensitivity, counting rate capabilities, and image quality were evaluated in accordance with NEMA NU 4-2008 standards. Human imaging was performed in addition to the evaluation. Results: Radial, tangential, and axial spatial resolution measured as minimal full width at half maximum approached 1.6, 1.7, and 2.0 mm, respectively, for filtered backprojection reconstruction and 0.8, 0.8, and 0.8 mm, respectively, for dynamic row-action maximum-likelihood algorithm reconstruction. The peak absolute sensitivity of the system was 11.2%. Scatter fraction at the same acquisition settings was 30.1% for the rat-sized phantom. Peak noise-equivalent counting rate and peak true rate for the ratlike phantom was 374 kcps at 25 MBq and 603 kcps at 31 MBq, respectively. In the image-quality phantom study, recovery coefficients and uniformity were 0.04–0.82 and 1.9%, respectively, for standard reconstruction mode and 0.09–0.97 and 4.5%, respectively, for enhanced-resolution mode. Human imaging provided high-contrast images with restricted background noise for standard reconstruction mode and high-resolution images for enhanced-resolution mode. Conclusion: The dedicated breast PET scanner has excellent spatial resolution and high sensitivity. The performance of the dedicated breast PET scanner is considered to be reasonable enough to support its use in breast cancer imaging.