RT Journal Article SR Electronic T1 Initial Characterization of a Dedicated Breast PET/CT Scanner During Human Imaging JF Journal of Nuclear Medicine JO J Nucl Med FD Society of Nuclear Medicine SP 1401 OP 1408 DO 10.2967/jnumed.109.064428 VO 50 IS 9 A1 Spencer L. Bowen A1 Yibao Wu A1 Abhijit J. Chaudhari A1 Lin Fu A1 Nathan J. Packard A1 George W. Burkett A1 Kai Yang A1 Karen K. Lindfors A1 David K. Shelton A1 Rosalie Hagge A1 Alexander D. Borowsky A1 Steve R. Martinez A1 Jinyi Qi A1 John M. Boone A1 Simon R. Cherry A1 Ramsey D. Badawi YR 2009 UL http://jnm.snmjournals.org/content/50/9/1401.abstract AB We have constructed a dedicated breast PET/CT scanner capable of high-resolution functional and anatomic imaging. Here, we present an initial characterization of scanner performance during patient imaging. Methods: The system consisted of a lutetium oxyorthosilicate–based dual–planar head PET camera (crystal size, 3 × 3 × 20 mm) and 768-slice cone-beam CT. The position of the PET heads (separation and height) could be adjusted for varying breast dimensions. For scanning, the patient lay prone on a specialized bed and inserted a single pendent breast through an aperture in the table top. Compression of the breast as used in mammography is not required. PET and CT systems rotate in the coronal plane underneath the patient sequentially to collect fully tomographic datasets. PET images were reconstructed with the fully 3-dimensional maximum a posteriori method, and CT images were reconstructed with the Feldkamp algorithm, then spatially registered and fused for display. Phantom scans were obtained to assess the registration accuracy between PET and CT images and the influence of PET electronics and activity on CT image quality. We imaged 4 women with mammographic findings highly suggestive of breast cancer (breast imaging reporting and data system, category 5) in an ongoing clinical trial. Patients were injected with 18F-FDG and imaged for 12.5 min per breast. From patient data, noise-equivalent counting rates and the singles-to-trues ratio (a surrogate for the randoms fraction) were calculated. Results: The average registration error between PET and CT images was 0.18 mm. PET electronics and activity did not significantly affect CT image quality. For the patient trial, biopsy-confirmed cancers were visualized on dedicated breast PET/CT on all patient scans, including the detection of ductal carcinoma in situ in 1 case. The singles-to-trues ratio was found to be inversely correlated with breast volume in the field of view, suggesting that larger breasts trend toward increased noise-equivalent counting rates for all other things equal. Conclusion: Scanning of the uncompressed breast with dedicated breast PET/CT can accurately visualize suspected lesions in 3 dimensions.