RT Journal Article SR Electronic T1 Routine Quality Control of Clinical Nuclear Medicine Instrumentation: A Brief Review JF Journal of Nuclear Medicine JO J Nucl Med FD Society of Nuclear Medicine SP 1114 OP 1131 DO 10.2967/jnumed.107.050203 VO 49 IS 7 A1 Pat Zanzonico YR 2008 UL http://jnm.snmjournals.org/content/49/7/1114.abstract AB This article reviews routine quality-control (QC) procedures for current nuclear medicine instrumentation, including the survey meter, dose calibrator, well counter, intraoperative probe, organ (“thyroid”) uptake probe, γ-camera, SPECT and SPECT/CT scanner, and PET and PET/CT scanner. It should be particularly useful for residents, fellows, and other trainees in nuclear medicine, nuclear cardiology, and radiology. The procedures described and their respective frequencies are presented only as general guidelines.