Quantitative Total-Body Imaging of Blood Flow with High-Temporal-Resolution Early Dynamic 18F-FDG PET Kinetic Modeling ====================================================================================================================== * Kevin J. Chung * Abhijit J. Chaudhari * Lorenzo Nardo * Terry Jones * Moon S. Chen, Jr. * Ramsey D. Badawi * Simon R. Cherry * Guobao Wang ## Abstract Past efforts to measure blood flow with the widely available radiotracer 18F-FDG were limited to tissues with high 18F-FDG extraction fraction. In this study, we developed an early dynamic 18F-FDG PET method with high-temporal-resolution (HTR) kinetic modeling to assess total-body blood flow based on deriving the vascular phase of 18F-FDG transit and conducted a pilot comparison study against a 11C-butanol flow-tracer reference. **Methods:** The first 2 min of dynamic PET scans were reconstructed at HTR (60 × 1 s/frame, 30 × 2 s/frame) to resolve the rapid passage of the radiotracer through blood vessels. In contrast to existing methods that use blood-to-tissue transport rate as a surrogate of blood flow, our method directly estimated blood flow using a distributed kinetic model (adiabatic approximation to tissue homogeneity [AATH] model). To validate our 18F-FDG measurements of blood flow against a reference flow-specific radiotracer, we analyzed total-body dynamic PET images of 6 human participants scanned with both 18F-FDG and 11C-butanol. An additional 34 total-body dynamic 18F-FDG PET images of healthy participants were analyzed for comparison against published blood-flow ranges. Regional blood flow was estimated across the body, and total-body parametric imaging of blood flow was conducted for visual assessment. AATH and standard compartment model fitting was compared using the Akaike information criterion at different temporal resolutions. **Results:** 18F-FDG blood flow was in quantitative agreement with flow measured from 11C-butanol across same-subject regional measurements (Pearson correlation coefficient, 0.955; *P* < 0.001; linear regression slope and intercept, 0.973 and –0.012, respectively), which was visually corroborated by total-body blood-flow parametric imaging. Our method resolved a wide range of blood-flow values across the body in broad agreement with published ranges (e.g., healthy cohort values of 0.51 ± 0.12 mL/min/cm3 in the cerebral cortex and 2.03 ± 0.64 mL/min/cm3 in the lungs). HTR (1–2 s/frame) was required for AATH modeling. **Conclusion:** Total-body blood-flow imaging was feasible using early dynamic 18F-FDG PET with HTR kinetic modeling. This method may be combined with standard 18F-FDG PET methods to enable efficient single-tracer multiparametric flow-metabolism imaging, with numerous research and clinical applications in oncology, cardiovascular disease, pain medicine, and neuroscience. * total-body PET * blood-flow/perfusion imaging * high-temporal resolution dynamic imaging * tracer kinetic modeling * distributed kinetic modeling Imaging blood flow has garnered considerable interest over the past 50 y, as its dysfunction is characteristic in many diseases (1–3). PET imaging with a blood-flow–specific radiotracer, such as 11C-butanol or 15O-water, is widely considered the gold standard for blood-flow imaging (4–6). These flow radiotracers are freely diffusible across capillary membranes (4–6); accordingly, the measured PET signal is closely proportional to blood flow. Blood flow can then be quantified by a standard 1-tissue compartment (S1TC) model because of the complete or near-complete extraction of these freely diffusible flow radiotracers (4–6). Importantly, these flow radiotracers are highly extracted in tissue across the entire body and allow total-body imaging of blood flow (4,5,7). However, the short half-lives of the radioisotopes in flow radiotracers create practical challenges that hinder their broader accessibility. 15O-water has a half-life of 2.04 min, which necessitates an on-site cyclotron and a dose-delivery system (8). 11C-butanol has a longer half-life (20.40 min) but still requires nearby production, thus limiting its access to urban or research PET centers. Other flow radiotracers, such as 82RbCl and 13N-ammonia, similarly have short radioisotope half-lives and high costs in addition to nonlinear uptake with flow (9). A blood-flow imaging method using a widely available radiotracer, such as 18F-FDG, may mitigate these challenges and open opportunities for imaging of blood flow and glucose metabolism with a single-tracer dynamic scan. Early dynamic 18F-FDG PET has been used to measure blood flow in select tissues, such as tumors (10), liver (11), and myocardium (12), where 18F-FDG is moderately to highly extracted. The first 2- to 3-min dynamic 18F-FDG PET signal is principally weighted toward the initial delivery of the radiotracer to the tissue (13), and the higher regional extraction fraction makes the analysis amenable to simplified modeling, like that of freely diffusible flow radiotracers. However, these approaches are not generally applicable to other regions, such as the brain, with lower 18F-FDG extraction fractions (14,15). An intravenously injected tracer is delivered to local tissue vasculature at a rate equal to blood flow. Standard compartmental models neglect this transient process, but distributed kinetic models explicitly model the blood flow and transit time associated with the radiotracer traversing the blood vessels (16,17). Although described several decades ago, distributed kinetic models had limited application in PET due to the poor temporal resolution and statistical quality of time–activity curves measured with conventional PET scanners (18,19). Total-body PET has substantially greater sensitivity (20–22) compared with conventional PET systems and allows high-temporal resolution (HTR) dynamic imaging (21,23) and kinetic modeling (13,24,25). This may revitalize opportunities to apply distributed kinetic models for blood-flow estimation with 18F-FDG in various tissues. Here, we describe the development of an early dynamic 18F-FDG PET method for total-body blood-flow imaging with HTR kinetic modeling and its validation against an 11C-butanol reference in a subset of participants scanned with both radiotracers. ## MATERIALS AND METHODS ### Total-Body Dynamic PET Two human cohorts were pooled in this study, each separately approved by the institutional review board at the University of California, Davis. Written informed consent was obtained for all participants. All participants received total-body dynamic imaging on the uEXPLORER PET/CT system (United Imaging Health Care), with the scan commencing immediately before bolus injection of the radiotracer. The first cohort comprised 6 participants (4 women; mean age, 67 ± 15 y) with chronic low-back myofascial pain who underwent total-body dynamic PET, receiving bolus injections of both 18F-FDG (98 ± 9 MBq) and 11C-butanol PET (268 ± 6 MBq) at 2 scanning sessions within 14 d (Clinicaltrials.gov identifier [NCT05876858](http://jnm.snmjournals.org/lookup/external-ref?link_type=CLINTRIALGOV&access_num=NCT05876858&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom)). The median interval between scans was 9 d (range, 0–14 d). Two participants were scanned on the same day with 11C-butanol scanning commencing first, followed by an interval of at least 3 h before the 18F-FDG PET to allow 11C to decay to negligible levels. The second cohort comprised 34 healthy participants (21 women; mean age, 51 ± 13 y) with no self-reported history of cancer or myocardial infarction in the past 5 y (26). Participants were scanned with total-body dynamic 18F-FDG PET (mean injected activity, 358 ± 33 MBq, bolus injection), and their data were used for methodologic development and validation against literature blood-flow ranges. Two participants in the first cohort and 20 participants in the second cohort self-identified as belonging to racial or ethnic minorities (26). For all dynamic scans, the first 2 min were reconstructed at HTR (60 × 1 s, 30 × 2 s) using reconstruction software provided by the vendor. This involved a time-of-flight ordered-subset expectation-maximum algorithm-based reconstruction without point-spread function modeling and with 4 iterations, 20 subsets, and standard corrections for attenuation, scatter, randoms, dead time, and decay (22). We used a matrix size of 150 × 150 × 486 and an isotropic voxel size of 4 mm. ### Tracer Kinetic Modeling of Blood Flow from Dynamic 18F-FDG Data Existing methods to measure blood flow with 18F-FDG have been limited to select tissue with high extraction fraction such that the blood-to-tissue transport rate (*K*1) approximates blood flow directly (10,11) or by nonlinear calibration (12). *K*1 is defined as the product of blood flow (*F*) and extraction fraction (*E*):![Formula][1] Eq. 1 Equation 1 shows that *K*1 is a good approximation of blood flow only when *E* is close to 1. 18F-FDG *K*1 can be measured with early dynamic imaging and an S1TC model, as the phosphorylation and dephosphorylation of 18F-FDG is not identifiable during the first few minutes of the dynamic scan (13,27). The impulse response function, ![Formula][2], of the S1TC is determined using Equation 2.![Formula][3] Eq. 2where ![Formula][4] is the blood volume and ![Formula][5] and ![Formula][6] are the blood-to-tissue transport and clearance rates, respectively (Fig. 1). Here, the value of ![Formula][7] at ![Formula][8] reflects the compartmental assumption that radiotracer instantaneously and uniformly mixes in regional blood vessels. ![FIGURE 1.](http://jnm.snmjournals.org/https://jnm.snmjournals.org/content/jnumed/early/2025/04/30/jnumed.124.268706/F1.medium.gif) [FIGURE 1.](http://jnm.snmjournals.org/content/early/2025/04/30/jnumed.124.268706/F1) FIGURE 1. Modeling intravascular delivery of PET tracers. (A) PET voxel partly comprises arteries, arterioles, capillaries, venules, and veins. PET tracers are initially delivered to and circulate through these vascular volumes via blood flow (*F*). Tracer transport from blood into (*K*1) and out of tissue (*k*2) occurs almost exclusively at capillary level. S1TC model (B) assumes that tracer instantaneously mixes in vascular volume, and effectively mean vascular transit time (*Tc*) is zero. AATH model (C) accounts for plug flow with single transit time (*Tc*) for tracer to traverse total vascular volume via blood flow. In reality, the radiotracer requires a nonzero transit time to traverse the length of the blood vessels at a rate equal to blood flow. This can be explicitly modeled in distributed parameter models (16,17). Here, we used the adiabatic approximation to tissue homogeneity (AATH) model (17), a distributed kinetic model with a closed-form time-domain solution that explicitly models intravascular blood flow and a mean vascular transit time. The impulse response function, ![Formula][9] can be determined using Equation 3.![Formula][10] Eq. 3where *F* is intravascular blood flow and ![Formula][11] is the mean vascular transit time for the radiotracer to pass through the total vascular blood volume (![Formula][12] in a voxel. ![Formula][13] is equal to the ratio of the blood volume to blood flow (![Formula][14]) (Fig. 1). Here, *F* is modeled separately from ![Formula][15]. This method differs from conventional methods using extraction fraction correction (12) of ![Formula][16] to approximate blood flow. ![Formula][17] describes a finite-time vascular phase (![Formula][18]) and a tissue phase (![Formula][19]). During the vascular phase, the radiotracer traverses the total vascular blood volume (arteries, arterioles, capillaries, venules, and veins) residing in a voxel at a uniform rate equal to the blood flow (plug flow). Capillary transit occurs for a fraction of this time during which a fraction (*E*) of the radiotracer is extracted to the extravascular tissue compartment. At the start of the tissue phase (![Formula][20]), the unextracted fraction (1 – *E*) is removed from the voxel volume through the venous outlet by blood flow. The extracted radiotracer then returns from the extravascular tissue compartment to the intravascular space and is cleared through the venous outlet by blood flow (![Formula][21]). Accordingly, the tissue phase of ![Formula][22] follows an exponential decay, like the S1TC model, with their response functions mainly differing by the presence of a nonzero-length vascular phase in the AATH model. We expect that the AATH and S1TC fittings may perform similarly at high extraction fractions as blood flow becomes similar to *K*1. For a general arterial input, ![Formula][23], the tissue time–activity curve, ![Formula][24], can be derived using Equation 4.![Formula][25] Eq. 4where ![Formula][26] is the time delay between radiotracer arrival at the measured arterial input location and local tissue vasculature. We used a basis function method with time-delay correction for least-squares parameter estimation using parametric forms of each model as described previously (28,29) and detailed in the supplemental materials, available at [http://jnm.snmjournals.org](http://jnm.snmjournals.org) (30–33). The AATH model was applied for both 18F-FDG and 11C-butanol. ### Image Analysis Total-body PET enabled noninvasive measurement of an image-derived input function for kinetic analysis. The ascending aorta was used for kinetic modeling of all tissue except the lungs, for which a right ventricle input function was used (24,34,35). Early 18F-FDG kinetics were quantified by analyzing regional time–activity curves obtained from tissue segmentations in 10 regions of interest (supplemental materials) (36–40). Total-body parametric images of early kinetics were generated by voxelwise kinetic modeling on 4-mm isotropic reconstructions. The dynamic and parametric images were smoothed by the kernel method, which is analogous to nonlocal means denoising (41,42). Prior composite images were derived from multiple static PET images (18F-FDG, 0–5, 5–20, 20–40, and 40–60 min; 11C-butanol, 0–1, 1–2, and 2–3 min), using 49 nearest neighbors within a 9 × 9 × 9 voxel neighborhood, as in our previous work (41,42). ### Evaluating Time–Activity Curve Fitting We compared the quality of the AATH and S1TC model time–activity curve fits using the Akaike information criterion (AIC) (supplemental materials) (43). The difference in AIC (AATH minus S1TC) was computed for each region of interest. A lower AIC indicated better fitting after accounting for the number of model parameters and the residual fitting error. Practical identifiability analysis was also performed as in previous work (27) to determine the reliability of AATH parameter estimates. To evaluate the effect of temporal resolution on the suitability of the AATH model, we frame-averaged each measured regional time–activity curve in the healthy 18F-FDG PET cohort at 1-, 2-, 3-, 5-, and 10-s/frame intervals. The resampled data were fitted with the AATH and S1TC models, and AIC differences were compared for each region and frame interval. ### Validating 18F-FDG Blood-Flow Quantification The mean and SD of regional blood-flow values estimated with the AATH model were computed for all participants. In participants with both 18F-FDG and 11C-butanol PET images, we performed correlation and Bland–Altman analyses (44) of regional blood-flow estimates between radiotracers. Total-body blood-flow parametric images were visually compared between radiotracers. For the healthy 18F-FDG PET cohort, we compared their average regional values against published ranges (Supplemental Table 1) (4,6,18,45–66). ## RESULTS ### Time–Activity Curve Fitting and Model Selection An example HTR 18F-FDG time–activity curve fitting in the cortical gray matter with the S1TC and AATH models is shown in Figure 2. The first-pass peak, which was accurately measured with HTR dynamic imaging, was better fitted with the AATH model compared with the S1TC model. Furthermore, the peak of the intravascular component (dashed red line) of the AATH-fitted curve better aligned with the peak of the measured curve. The intravascular component is a scaled version of the arterial input function (Supplemental Eq. 4) in the S1TC model fit, whereas it is a smoothed, widened version in the AATH fit due to the nonzero *Tc* (Supplemental Eq. 2). Accordingly, the intravascular distribution of the S1TC-fitted curve was smaller than that with the AATH model fit; to compensate, the extravascular distribution of the S1TC-fitted curve grew larger than that of the AATH model fit. In all regions of interest investigated, the AATH model was preferred on average over the S1TC model across 34 HTR dynamic 18F-FDG scans of healthy participants (Fig. 3A). Similarly, for 11C-butanol, S1TC model fitting was worse on average (Supplemental Fig. 1), and this justified our use of the AATH model for blood-flow comparisons against 18F-FDG. ![FIGURE 2.](http://jnm.snmjournals.org/https://jnm.snmjournals.org/content/jnumed/early/2025/04/30/jnumed.124.268706/F2.medium.gif) [FIGURE 2.](http://jnm.snmjournals.org/content/early/2025/04/30/jnumed.124.268706/F2) FIGURE 2. Time–activity curve fits in cortical gray matter using S1TC model (A) and AATH model (B) at HTR (60 × 1 s/frame, 30 × 2 s/frame). Dashed red and green lines represent intravascular and extravascular components of fitted curve, respectively. Black arrows indicate areas where S1TC fitting was poor. ![FIGURE 3.](http://jnm.snmjournals.org/https://jnm.snmjournals.org/content/jnumed/early/2025/04/30/jnumed.124.268706/F3.medium.gif) [FIGURE 3.](http://jnm.snmjournals.org/content/early/2025/04/30/jnumed.124.268706/F3) FIGURE 3. Difference in AIC between AATH and S1TC models using original HTR data (60 × 1 s/frame, 30 × 2 s/frame) (A) and at different simulated frame intervals (B). Negative AIC indicates preference toward AATH model. GM = gray matter. Figure 3B illustrates the difference in AIC between the AATH and S1TC models at different temporal resolutions and in different tissue regions for the 18F-FDG cohort of 34 healthy participants. Using the AIC, the AATH model had improved fitting over the S1TC model at 1- to 2-s/frame intervals, though the average magnitude of AIC differences was lower at 2 s/frame. Beyond 3-s/frame intervals, the S1TC model was clearly preferred by the AIC. ### Validation of 18F-FDG PET Blood Flow Against 11C-Butanol PET 11C-butanol and 18F-FDG blood flow were compared across all 6 participants, each with 10 tissue regions, using correlation and Bland–Altman analyses (Figs. 4A and 4B). 18F-FDG blood flow estimated with our proposed method had strong quantitative agreement with the 11C-butanol reference measurement, with a Pearson correlation coefficient of 0.955 (*P* < 0.001) and a linear regression slope and intercept of 0.973 and –0.012, respectively. The mean difference in blood flow (18F-FDG minus 11C-butanol) was –0.031 mL/min/cm3, indicating that our 18F-FDG blood-flow measures marginally underestimated that of 11C-butanol, on average. The Bland–Altman 95% limits of agreement were –0.445 to 0.383 mL/min/cm3, with the larger differences mainly driven by tissues with higher blood flow. One participant had severe intraframe respiratory motion during the 11C-butanol scan, which prevented accurate lung blood-flow quantification and substantial overestimation (>1.0 mL/min/cm3) with our 18F-FDG PET method. Further analyses stratified by regions with similar blood flow are shown in Supplemental Figures 2 and 3. In general, S1TC 18F-FDG *K*1 did not strongly agree with 11C-butanol blood flow (Fig. 4B). ![FIGURE 4.](http://jnm.snmjournals.org/https://jnm.snmjournals.org/content/jnumed/early/2025/04/30/jnumed.124.268706/F4.medium.gif) [FIGURE 4.](http://jnm.snmjournals.org/content/early/2025/04/30/jnumed.124.268706/F4) FIGURE 4. Correlation (left) and Bland–Altman (right) plots comparing 11C-butanol blood flow against 18F-FDG blood flow with AATH model (A) and S1TC model (B) *K*1 in same participants. GM = gray matter; MD = mean difference. ### Total-Body Parametric Imaging of Blood Flow with 18F-FDG Total-body parametric images of blood flow generated with 18F-FDG and 11C-butanol in the same participant are shown in Figure 5. Parametric images appeared similar both visually and in quantitative ranges across the body. A notable difference observed between the 2 blood-flow maps was the absence of sagittal and transverse sinuses in the 11C-butanol parametric image. This is likely due to the high extraction fraction of 11C-butanol in the brain, resulting in lower venous concentration of the tracer. A representative parametric image of the *Tc* from a healthy participant is shown in Supplemental Figure 4. One participant had substantial differences in cerebral blood flow between 11C-butanol and 18F-FDG (Supplemental Fig. 5) (67,68), which may be due to a combination of physiologic and methodologic factors and warrants a future test–retest study. ![FIGURE 5.](http://jnm.snmjournals.org/https://jnm.snmjournals.org/content/jnumed/early/2025/04/30/jnumed.124.268706/F5.medium.gif) [FIGURE 5.](http://jnm.snmjournals.org/content/early/2025/04/30/jnumed.124.268706/F5) FIGURE 5. Total-body parametric imaging of blood flow with early dynamic 18F-FDG method compared with 11C-butanol flow-tracer PET reference in same participant. White arrows indicate sagittal and transverse sinuses in brain. ### Regional 18F-FDG Early Kinetics in Healthy Participants The distribution of blood-flow estimates with our 18F-FDG method in 34 healthy participants is plotted in Figure 6. On average, all tissue regions were within the expected range except the subcortical gray matter and lungs, which were slightly below and above the upper range of average blood-flow values reported in literature, respectively (Fig. 6; Supplemental Table 1) (4). Estimates of *Tc* (Table 1) reasonably agreed with those reported in the CT/MR perfusion and dual-tracer PET literature (Supplemental Table 1). The identifiability of regional blood-flow estimates with our proposed method was excellent overall across tissue regions (absolute mean error, <5%; SD, <15%) except the skeletal muscle (mean overestimation, 6.4%; Supplemental Table 2). ![FIGURE 6.](http://jnm.snmjournals.org/https://jnm.snmjournals.org/content/jnumed/early/2025/04/30/jnumed.124.268706/F6.medium.gif) [FIGURE 6.](http://jnm.snmjournals.org/content/early/2025/04/30/jnumed.124.268706/F6) FIGURE 6. Regional blood flow in 34 healthy participants estimated with early dynamic 18F-FDG method. Plots are separated by range of blood-flow values. Average estimates mostly fell within range of average blood-flow values reported in literature (Supplemental Table 1), indicated by green boxes. GM = gray matter. View this table: [TABLE 1](http://jnm.snmjournals.org/content/early/2025/04/30/jnumed.124.268706/T1) TABLE 1 18F-FDG Early Kinetics Across Healthy Participants with AATH Model (*n* = 34) Regional 18F-FDG extraction fraction values in the healthy cohort are summarized in Table 1. The 18F-FDG extraction fraction varied greatly among tissue regions across the body. Accordingly, S1TC 18F-FDG *K*1 was in general agreement with 11C-butanol blood flow only in tissues with a high extraction fraction (Supplemental Fig. 6). ## DISCUSSION We developed an early dynamic 18F-FDG PET method for total-body blood-flow imaging with HTR kinetic modeling and validated it against a 11C-butanol reference in a subset of participants scanned with both radiotracers. Conventional methods for 18F-FDG blood-flow imaging have been limited to tissues with relatively high extraction fraction where blood-to-tissue transport rate *K*1 can approximate blood flow. Our proposed method uses a distributed kinetic model (16,17) that explicitly accounts for the blood-flow delivery rate of radiotracer to blood vessels, which was resolved with HTR dynamic imaging. 18F-FDG blood-flow estimates were in quantitative agreement with 11C-butanol in direct comparisons for the same subjects. We further validated our method with 34 healthy participants, showing that regional blood-flow values across the body were broadly within published ranges (Fig. 6). We report data on 18F-FDG extraction fraction values across the body, which varied between tissue types (4%–95%; Table 1). To our knowledge, this is the first study to perform total-body blood-flow imaging with 18F-FDG and compare the results with 11C-butanol flow-tracer PET in the same subjects. The key difference between the AATH model and the S1TC model is that the former does not assume that tracer instantaneously mixes in the vascular volume. In the AATH model, blood flow represents the intravascular flow rate through the entire vascular volume residing in a voxel volume. This subtly differs from conventional 11C-butanol or 15O-water PET measurements of tracer clearance into tissue (*K*1), which occurs at the capillaries and is directly proportional to capillary blood flow when *E* approximates 1. However, because our blood-flow estimates were in close agreement with published ranges, it appears that regional blood flow may be tightly coupled with capillary blood flow. This is further supported by the agreement of our AATH regional blood-flow estimates with ranges reported in the literature (Fig. 6). Similarly, AATH *Tc* represents the mean vascular transit time through the entire vascular volume in a voxel and not only the capillaries. Our *Tc* estimates generally agreed with those from previously published CT/MR perfusion and dual-tracer PET data (e.g., 3–7 s in the brain) (Supplemental Table 1). Our data agree with prior work suggesting the need for HTR imaging for distributed kinetic modeling (16–19,25,69). Here, a temporal resolution of 1 to 2 s was required for the AATH model to be preferred over the S1TC model based on the AIC metric (Fig. 3). The temporal resolution may need to be closer to 1 s/frame for tissues such as lung, where the right ventricle input function often has a very fast, sharp bolus (24). Total-body PET now allows the requisite temporal resolution for blood-flow imaging across the body using the widely available 18F-FDG radiotracer. Ongoing advancements in image reconstruction may further improve signal-to-noise for HTR imaging (70). Whether these improvements allow lower-dose 18F-FDG blood-flow imaging and voxel-level AIC model selection (42) will be considered in future work. Our method is generally applicable across the body, in contrast to conventional 18F-FDG blood-flow estimation methods that require a high extraction fraction. Directly using 18F-FDG *K*1 as a surrogate of blood flow does not generalize across all tissue types (Fig. 4B). HTR 18F-FDG PET has also been used by Larsson et al. (25) to measure cerebral blood flow with model-independent deconvolution analysis, despite the low cerebral extraction fraction of 18F-FDG (Table 1). Our estimates of 18F-FDG extraction fraction in the cortical gray matter were similar to their estimates in the thalamus (0.19 ± 0.05) as well as those found in another dual-tracer PET study (14). However, our estimates of 18F-FDG extraction in other brain regions were larger than these values. Variability in parameter estimates due to differences in estimation methods and modeling has been recognized in MR and CT perfusion (31). The number and characteristics of study participants may have also contributed to this discrepancy. Future work is warranted to compare model-independent deconvolution and AATH parameter estimates and conduct test–retest studies. Intravenously injected tracers are delivered to tissue vasculature by blood flow, and distributed kinetic models explicitly account for this process (16,18,19). Historically, the AATH model has been used for blood-flow imaging using inert contrast agents and HTR dynamic CT or MRI (30,31). Here, we showed that distributed modeling is applicable to a noninert metabolic radiotracer (18F-FDG) as well as a freely diffusible flow radiotracer (11C-butanol), provided the requisite temporal resolution is used. Our method may be generally applicable to a wide range of tracers, allowing single-tracer blood-flow imaging. This has been demonstrated in the brain with several PET tracers using the current approach (28) and model-independent deconvolution (25) on HTR dynamic PET, though further validation is required beyond the brain. Numerous opportunities can be enabled for single-tracer multiparametric imaging of physiologic parameters, such as flow-metabolic imaging (1–3) with 18F-FDG or joint quantification of blood flow and amyloid burden with amyloid PET tracers (71). This initial work demonstrated the feasibility of using the AATH model for estimating total-body blood flow with HTR early dynamic 18F-FDG PET, but further modeling improvements may be considered at the organ level. First, dual-input modeling with the portal vein is required to enable liver blood-flow measurements (72). However, existing methods (11,72) using 18F-FDG *K*1 as a surrogate of hepatic blood flow may be sufficient because of the high permeability of liver sinusoids (73). Also, our initial analysis of the myocardium suggested that spillover from the right and left ventricles was substantial at HTR, resulting in substantial blood-flow overestimation. Spillover and cardiac motion correction will be investigated in the future. For the lungs, we used a single right-ventricle input function and neglected the normally small contribution from the bronchial circulation. Dual-input modeling may be required in lung tumors, where the bronchial fraction is nonnegligible (74). The kidneys have complex vasculature, and additional technical considerations, such as partial-volume effect and blood-flow heterogeneity (55), may be required to better model their unique anatomy and vascular transport mechanisms. Lastly, dispersion correction may be required for tissues that are considerably distal to the image-derived arterial input function. This study had limitations. First, the sample size of participants scanned with both 18F-FDG and 11C-butanol was small in this pilot study. Instead, the validity of our 18F-FDG blood-flow measurements was supported by additional analyses of 34 additional healthy participants. Additional subjects who underwent PET scans with both radiotracers will be analyzed in future studies. Second, participants were not recruited specifically for validation of 18F-FDG blood flow. One participant from the dual-tracer group was suspected of having physiologic differences between 11C-butanol and 18F-FDG scans. Future studies will better account for physiologic confounds by measuring CO2 partial pressure, O2 partial pressure, and heart rate, among others. Lastly, validation is required in patients with major diagnosed blood-flow defects, such as peripheral, carotid, or coronary artery disease. ## CONCLUSION We developed an early dynamic 18F-FDG PET method with HTR kinetic modeling for total-body blood-flow imaging. Using the ubiquitous 18F-FDG radiotracer for blood-flow imaging may mitigate the need for a costly and practically challenging flow-tracer PET scan. In combination with standard 18F-FDG PET methods for glucose metabolic imaging, our proposed method may allow efficient single-tracer imaging of blood flow and metabolism, resulting in lower radiation exposure to the patient, shorter scan times, fewer infrastructural requirements, and lower costs. Our method may be generally applicable to other radiotracers, broadening the possibility of single-tracer multiparametric imaging from a single dynamic PET scan. ## DISCLOSURE The University of California, Davis has a research agreement and a revenue sharing agreement with United Imaging Health Care. This work was supported in part by National Institutes of Health (NIH) grants R01 EB033435 and R61 AT012187. The image data of healthy participants were acquired under the support of NIH R01 CA206187 and P30 CA093373. No other potential conflict of interest relevant to this article was reported. #### KEY POINTS **QUESTION:** Can HTR early dynamic 18F-FDG PET kinetic analysis be used for total-body blood-flow imaging? **PERTINENT FINDINGS:** Blood-flow estimates between 18F-FDG and gold standard 11C-butanol PET in the same participants showed good quantitative agreement across the body. Regional blood-flow measurements with the proposed early dynamic 18F-FDG PET method in 34 healthy participants were within well-established reference ranges in tissues across the body. **IMPLICATIONS FOR PATIENT CARE:** Total-body blood-flow imaging can be performed with the widely available 18F-FDG radiotracer, possibly mitigating the need for a dedicated flow radiotracer and expanding opportunities to efficiently study blood flow and glucose metabolism in combination with standard 18F-FDG metabolic imaging methods. ## ACKNOWLEDGEMENTS We gratefully acknowledge Dr. Benjamin A. Spencer and Dr. Yasser G. Abdelhafez for helpful discussions and technical support, as well as the technologists and staff at the EXPLORER Molecular Imaging Center, particularly Lynda E. Painting, for their assistance in patient consent and data acquisition. Figure 1 and the graphical abstract were created in part using [Biorender.com](http://Biorender.com) and Servier Medical Art ([https://smart.servier.com](https://smart.servier.com)). ## Footnotes * Published online Apr. 30, 2025. * © 2025 by the Society of Nuclear Medicine and Molecular Imaging. Immediate Open Access: Creative Commons Attribution 4.0 International License (CC BY) allows users to share and adapt with attribution, excluding materials credited to previous publications. License: [https://creativecommons.org/licenses/by/4.0/](https://creativecommons.org/licenses/by/4.0/). Details: [https://jnm.snmjournals.org/page/permissions](https://jnm.snmjournals.org/page/permissions). ## REFERENCES 1. 1.Miles KA, Williams RE. Warburg revisited: imaging tumour blood flow and metabolism. Cancer Imaging. 2008;8:81–86. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1102/1470-7330.2008.0011&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=18390391&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) [Web of Science](http://jnm.snmjournals.org/lookup/external-ref?access_num=000207775000011&link_type=ISI) 2. 2.Verfaillie SCJ, Adriaanse SM, Binnewijzend MAA, et al. Cerebral perfusion and glucose metabolism in Alzheimer’s disease and frontotemporal dementia: two sides of the same coin? Eur Radiol. 2015;25:3050–3059. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1007/s00330-015-3696-1&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=25899416&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 3. 3.Anagnostopoulos C, Georgakopoulos A, Pianou N, Nekolla SG. Assessment of myocardial perfusion and viability by positron emission tomography. Int J Cardiol. 2013;167:1737–1749. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1016/j.ijcard.2012.12.009&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=23313467&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 4. 4.Li EJ, López JE, Spencer BA, et al. Total-body perfusion imaging with [11C]-butanol. J Nucl Med. 2023;64:1831–1838. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6Njoiam51bWVkIjtzOjU6InJlc2lkIjtzOjEwOiI2NC8xMS8xODMxIjtzOjQ6ImF0b20iO3M6NDc6Ii9qbnVtZWQvZWFybHkvMjAyNS8wNC8zMC9qbnVtZWQuMTI0LjI2ODcwNi5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 5. 5.Knuuti J, Tuisku J, Kärpijoki H, et al. Quantitative perfusion imaging with total-body PET. J Nucl Med. 2023;64(suppl 2):11S–19S. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6Njoiam51bWVkIjtzOjU6InJlc2lkIjtzOjE5OiI2NC9TdXBwbGVtZW50XzIvMTFTIjtzOjQ6ImF0b20iO3M6NDc6Ii9qbnVtZWQvZWFybHkvMjAyNS8wNC8zMC9qbnVtZWQuMTI0LjI2ODcwNi5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 6. 6.Herscovitch P, Raichle ME, Kilbourn MR, Welch MJ. Positron emission tomographic measurement of cerebral blood flow and permeability-surface area product of water using [15O]water and [11C]butanol. J Cereb Blood Flow Metab. 1987;7:527–542. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1038/jcbfm.1987.102&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=3498732&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) [Web of Science](http://jnm.snmjournals.org/lookup/external-ref?access_num=A1987K466100002&link_type=ISI) 7. 7.Jochumsen MR, Christensen NL, Iversen P, Gormsen LC, Sørensen J, Tolbod LP. Whole-body parametric mapping of tumour perfusion in metastatic prostate cancer using long axial field-of-view [15O]H2O PET. Eur J Nucl Med Mol Imaging. 2024;51:4134–4140. [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=38940842&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 8. 8.Slart RHJA, Martinez-Lucio TS, Boersma HH, et al. [15O]H2O PET: potential or essential for molecular imaging? Semin Nucl Med. 2023;54:761–773. [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=37640631&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 9. 9.Dewey M, Siebes M, Kachelrieß M, et al.; Quantitative Cardiac Imaging Study Group. Clinical quantitative cardiac imaging for the assessment of myocardial ischaemia. Nat Rev Cardiol. 2020;17:427–450. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1038/s41569-020-0341-8&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=32094693&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 10. 10.Mullani NA, Herbst RS, O’Neil RG, Gould KL, Barron BJ, Abbruzzese JL. Tumor blood flow measured by PET dynamic imaging of first-pass 18F-FDG uptake: a comparison with 15O-labeled water-measured blood flow. J Nucl Med. 2008;49:517–523. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6Njoiam51bWVkIjtzOjU6InJlc2lkIjtzOjg6IjQ5LzQvNTE3IjtzOjQ6ImF0b20iO3M6NDc6Ii9qbnVtZWQvZWFybHkvMjAyNS8wNC8zMC9qbnVtZWQuMTI0LjI2ODcwNi5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 11. 11.Winterdahl M, Munk OL, Sørensen M, Mortensen FV, Keiding S. Hepatic blood perfusion measured by 3-minute dynamic 18F-FDG PET in pigs. J Nucl Med. 2011;52:1119–1124. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6Njoiam51bWVkIjtzOjU6InJlc2lkIjtzOjk6IjUyLzcvMTExOSI7czo0OiJhdG9tIjtzOjQ3OiIvam51bWVkL2Vhcmx5LzIwMjUvMDQvMzAvam51bWVkLjEyNC4yNjg3MDYuYXRvbSI7fXM6ODoiZnJhZ21lbnQiO3M6MDoiIjt9) 12. 12.Zuo Y, López JE, Smith TW, et al. Multiparametric cardiac 18F-FDG PET in humans: pilot comparison of FDG delivery rate with 82Rb myocardial blood flow. Phys Med Biol. 2021;66:155015. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1088/1361-6560/ac15a6&link_type=DOI) 13. 13.Feng T, Zhao Y, Shi H, et al. Total-body quantitative parametric imaging of early kinetics of 18F-FDG. J Nucl Med. 2021;62:738–744. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6Njoiam51bWVkIjtzOjU6InJlc2lkIjtzOjg6IjYyLzUvNzM4IjtzOjQ6ImF0b20iO3M6NDc6Ii9qbnVtZWQvZWFybHkvMjAyNS8wNC8zMC9qbnVtZWQuMTI0LjI2ODcwNi5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 14. 14.Huisman MC, Van Golen LW, Hoetjes NJ, et al. Cerebral blood flow and glucose metabolism in healthy volunteers measured using a high-resolution PET scanner. EJNMMI Res. 2012;2:63. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1186/2191-219X-2-63&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=23168248&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 15. 15.Hasselbalch SG, Knudsen GM, Holm S, Hageman LP, Capaldo B, Paulson OB. Transport of D-glucose and 2-fluorodeoxyglucose across the blood-brain barrier in humans. J Cereb Blood Flow Metab. 1996;16:659–666. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1097/00004647-199607000-00017&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=8964806&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 16. 16.Larson KB, Markham J, Raichle ME. Tracer-kinetic models for measuring cerebral blood flow using externally detected radiotracers. J Cereb Blood Flow Metab. 1987;7:443–463. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1038/jcbfm.1987.88&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=3611204&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 17. 17.St Lawrence KS, Lee T-Y. An Adiabatic approximation to the tissue homogeneity model for water exchange in the brain: I. Theoretical derivation. J Cereb Blood Flow Metab. 1998;18:1365–1377. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1097/00004647-199812000-00011&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=9850149&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 18. 18.Quarles RP, Mintun MA, Larson KB, Markham J, MacLeod AM, Raichle ME. Measurement of regional cerebral blood flow with positron emission tomography: a comparison of [15O]water to [11C]butanol with distributed-parameter and compartmental models. J Cereb Blood Flow Metab. 1993;13:733–747. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1038/jcbfm.1993.94&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=8360280&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 19. 19.Muzic RF, Saidel GM. Distributed versus compartment models for PET receptor studies. IEEE Trans Med Imaging. 2003;22:11–21. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1109/TMI.2002.806576&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=12703756&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 20. 20.Cherry SR, Jones T, Karp JS, Qi J, Moses WW, Badawi RD. Total-body PET: maximizing sensitivity to create new opportunities for clinical research and patient care. J Nucl Med. 2018;59:3–12. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6Njoiam51bWVkIjtzOjU6InJlc2lkIjtzOjY6IjU5LzEvMyI7czo0OiJhdG9tIjtzOjQ3OiIvam51bWVkL2Vhcmx5LzIwMjUvMDQvMzAvam51bWVkLjEyNC4yNjg3MDYuYXRvbSI7fXM6ODoiZnJhZ21lbnQiO3M6MDoiIjt9) 21. 21.Badawi RD, Shi H, Hu P, et al. First human imaging studies with the EXPLORER total-body PET scanner. J Nucl Med. 2019;60:299–303. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6Njoiam51bWVkIjtzOjU6InJlc2lkIjtzOjg6IjYwLzMvMjk5IjtzOjQ6ImF0b20iO3M6NDc6Ii9qbnVtZWQvZWFybHkvMjAyNS8wNC8zMC9qbnVtZWQuMTI0LjI2ODcwNi5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 22. 22.Spencer BA, Berg E, Schmall JP, et al. Performance evaluation of the uEXPLORER total-body PET/CT scanner based on NEMA NU 2-2018 with additional tests to characterize PET scanners with a long axial field of view. J Nucl Med. 2021;62:861–870. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6Njoiam51bWVkIjtzOjU6InJlc2lkIjtzOjg6IjYyLzYvODYxIjtzOjQ6ImF0b20iO3M6NDc6Ii9qbnVtZWQvZWFybHkvMjAyNS8wNC8zMC9qbnVtZWQuMTI0LjI2ODcwNi5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 23. 23.Zhang X, Cherry SR, Xie Z, Shi H, Badawi RD, Qi J. Subsecond total-body imaging using ultrasensitive positron emission tomography. Proc Natl Acad Sci USA. 2020;117:2265–2267. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6NDoicG5hcyI7czo1OiJyZXNpZCI7czoxMDoiMTE3LzUvMjI2NSI7czo0OiJhdG9tIjtzOjQ3OiIvam51bWVkL2Vhcmx5LzIwMjUvMDQvMzAvam51bWVkLjEyNC4yNjg3MDYuYXRvbSI7fXM6ODoiZnJhZ21lbnQiO3M6MDoiIjt9) 24. 24.Wang Y, Spencer BA, Schmall J, et al. High-temporal-resolution lung kinetic modeling using total-body dynamic PET with time-delay and dispersion corrections. J Nucl Med. 2023;64:1154–1161. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6Njoiam51bWVkIjtzOjU6InJlc2lkIjtzOjk6IjY0LzcvMTE1NCI7czo0OiJhdG9tIjtzOjQ3OiIvam51bWVkL2Vhcmx5LzIwMjUvMDQvMzAvam51bWVkLjEyNC4yNjg3MDYuYXRvbSI7fXM6ODoiZnJhZ21lbnQiO3M6MDoiIjt9) 25. 25.Larsson HBW, Law I, Andersen TL, et al. Brain perfusion estimation by Tikhonov model-free deconvolution in a long axial field of view PET/CT scanner exploring five different PET tracers. Eur J Nucl Med Mol Imaging. 2024;51:707–720. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1007/s00259-023-06469-w&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=37843600&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 26. 26.Calabró A, Abdelhafez YG, Triumbari EKA, et al. 18F-FDG gallbladder uptake: observation from a total-body PET/CT scanner. BMC Med Imaging. 2023;23:9. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1186/s12880-022-00957-5&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=36627570&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 27. 27.Zuo Y, Sarkar S, Corwin MT, Olson K, Badawi RD, Wang G. Structural and practical identifiability of dual-input kinetic modeling in dynamic PET of liver inflammation. Phys Med Biol. 2019;64:175023. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1088/1361-6560/ab1f29&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=31051490&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 28. 28.Chung KJ, Abdelhafez YG, Spencer BA, et al. Quantitative PET imaging and modeling of molecular blood-brain barrier permeability. Nat Commun. 2025;16:3076. [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=40159510&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 29. 29.Gunn RN, Lammertsma AA, Hume SP, Cunningham VJ. Parametric imaging of ligand-receptor binding in PET using a simplified reference region model. Neuroimage. 1997;6:279–287. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1006/nimg.1997.0303&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=9417971&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) [Web of Science](http://jnm.snmjournals.org/lookup/external-ref?access_num=000071052500006&link_type=ISI) 30. 30.Sourbron SP, Buckley DL. Tracer kinetic modelling in MRI: estimating perfusion and capillary permeability. Phys Med Biol. 2012;57:R1. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1088/0031-9155/57/2/R1&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=22173205&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 31. 31.Chung KJ, De Sarno D, Lee T-Y. Quantitative functional imaging with CT perfusion: technical considerations, kinetic modeling, and applications. Front Phys. 2023;11:1246973. 32. 32.Lawson CL, Hanson RJ. Solving Least Squares Problems. Philadelphia: SIAM; 1995:158–173. 33. 33.Knapp WH, Helus F, Oberdorfer F, et al. 11C-butanol for imaging of the blood-flow distribution in tumor-bearing animals. Eur J Nucl Med. 1985;10:540–548. [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=3928384&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 34. 34.Volpi T, Maccioni L, Colpo M, et al. An update on the use of image-derived input functions for human PET studies: new hopes or old illusions? EJNMMI Res. 2023;13:97. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1186/s13550-023-01050-w&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=37947880&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 35. 35.Sari H, Mingels C, Alberts I, et al. First results on kinetic modelling and parametric imaging of dynamic 18F-FDG datasets from a long axial FOV PET scanner in oncological patients. Eur J Nucl Med Mol Imaging. 2022;49:1997–2009. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1007/s00259-021-05623-6&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=34981164&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 36. 36.Fedorov A, Beichel R, Kalpathy-Cramer J, et al. 3D Slicer as an image computing platform for the Quantitative Imaging Network. Magn Reson Imaging. 2012;30:1323–1341. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1016/j.mri.2012.05.001&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=22770690&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) [Web of Science](http://jnm.snmjournals.org/lookup/external-ref?access_num=000309946000013&link_type=ISI) 37. 37.Sundar LKS, Yu J, Muzik O, et al. Fully automated, semantic segmentation of whole-body 18F-FDG PET/CT images based on data-centric artificial intelligence. J Nucl Med. 2022;63:1941–1948. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6Njoiam51bWVkIjtzOjU6InJlc2lkIjtzOjEwOiI2My8xMi8xOTQxIjtzOjQ6ImF0b20iO3M6NDc6Ii9qbnVtZWQvZWFybHkvMjAyNS8wNC8zMC9qbnVtZWQuMTI0LjI2ODcwNi5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 38. 38.Hammers A, Allom R, Koepp MJ, et al. Three‐dimensional maximum probability atlas of the human brain, with particular reference to the temporal lobe. Hum Brain Mapp. 2003;19:224–247. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1002/hbm.10123&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=12874777&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) [Web of Science](http://jnm.snmjournals.org/lookup/external-ref?access_num=000184532600002&link_type=ISI) 39. 39.Otsu N. A threshold selection method from gray-level histograms. IEEE Trans Syst Man Cybern. 1979;9:62–66. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1109/TSMC.1979.4310076&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=20007788&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) [Web of Science](http://jnm.snmjournals.org/lookup/external-ref?access_num=A1979GE96000010&link_type=ISI) 40. 40.Modat M, Cash DM, Daga P, Winston GP, Duncan JS, Ourselin S. Global image registration using a symmetric block-matching approach. J Med Imaging (Bellingham). 2014;1:024003. [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=26158035&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 41. 41.Wang G, Qi J. PET image reconstruction using kernel method. IEEE Trans Med Imaging. 2015;34:61–71. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1109/TMI.2014.2343916&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=25095249&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 42. 42.Wang G, Nardo L, Parikh M, et al. Total-body PET multiparametric imaging of cancer using a voxelwise strategy of compartmental modeling. J Nucl Med. 2022;63:1274–1281. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6Njoiam51bWVkIjtzOjU6InJlc2lkIjtzOjk6IjYzLzgvMTI3NCI7czo0OiJhdG9tIjtzOjQ3OiIvam51bWVkL2Vhcmx5LzIwMjUvMDQvMzAvam51bWVkLjEyNC4yNjg3MDYuYXRvbSI7fXM6ODoiZnJhZ21lbnQiO3M6MDoiIjt9) 43. 43.Akaike H. A new look at the statistical model identification. IEEE Trans Autom Control. 1974;19:716–723. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1109/TAC.1974.1100705&link_type=DOI) 44. 44.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327:307–310. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1016/S0140-6736(86)90837-8&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=2868172&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 45. 45.Leenders KL, Perani D, Lammertsma AA, et al. Cerebral blood flow, blood volume and oxygen utilization: normal values and effect of age. Brain. 1990;113:27–47. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1093/brain/113.1.27&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=2302536&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) [Web of Science](http://jnm.snmjournals.org/lookup/external-ref?access_num=A1990CN01200002&link_type=ISI) 46. 46.Berridge MS, Adler LP, Nelson AD, et al. Measurement of human cerebral blood flow with [15O]butanol and positron emission tomography. J Cereb Blood Flow Metab. 1991;11:707–715. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1038/jcbfm.1991.127&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=1874804&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) [Web of Science](http://jnm.snmjournals.org/lookup/external-ref?access_num=A1991GD34400001&link_type=ISI) 47. 47.Herzog H, Seitz RJ, Tellmann L, et al. Quantitation of regional cerebral blood flow with 15O-butanol and positron emission tomography in humans. J Cereb Blood Flow Metab. 1996;16:645–649. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1097/00004647-199607000-00015&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=8964804&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) [Web of Science](http://jnm.snmjournals.org/lookup/external-ref?access_num=A1996UU34800015&link_type=ISI) 48. 48.Martin WR, Raichle ME. Cerebellar blood flow and metabolism in cerebral hemisphere infarction. Ann Neurol. 1983;14:168–176. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1002/ana.410140203&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=6605110&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) [Web of Science](http://jnm.snmjournals.org/lookup/external-ref?access_num=A1983RD09700002&link_type=ISI) 49. 49.Gaillard WD, Zeffiro T, Fazilat S, DeCarli C, Theodore WH. Effect of valproate on cerebral metabolism and blood flow: an 18F-2-deoxyglusose and 15O water positron emission tomography study. Epilepsia. 1996;37:515–521. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1111/j.1528-1157.1996.tb00602.x&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=8641226&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) [Web of Science](http://jnm.snmjournals.org/lookup/external-ref?access_num=A1996UN42000002&link_type=ISI) 50. 50.Warnert EA, Harris AD, Murphy K, et al. In vivo assessment of human brainstem cerebrovascular function: a multi-inversion time pulsed arterial spin labelling study. J Cereb Blood Flow Metab. 2014;34:956–963. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1038/jcbfm.2014.39&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=24594624&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 51. 51.Kahn D, Weiner G, Ben-Haim S, et al. Positron emission tomographic measurement of bone marrow blood flow to the pelvis and lumbar vertebrae in young normal adults. Blood. 1994;83:958–963. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6MTI6ImJsb29kam91cm5hbCI7czo1OiJyZXNpZCI7czo4OiI4My80Lzk1OCI7czo0OiJhdG9tIjtzOjQ3OiIvam51bWVkL2Vhcmx5LzIwMjUvMDQvMzAvam51bWVkLjEyNC4yNjg3MDYuYXRvbSI7fXM6ODoiZnJhZ21lbnQiO3M6MDoiIjt9) 52. 52.Bertoldo A, Peltoniemi P, Oikonen V, Knuuti J, Nuutila P, Cobelli C. Kinetic modeling of [18F]FDG in skeletal muscle by PET: a four-compartment five-rate-constant model. Am J Physiol Endocrinol Metab. 2001;281:E524–E536. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1152/ajpendo.2001.281.3.E524&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=11500308&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) [Web of Science](http://jnm.snmjournals.org/lookup/external-ref?access_num=000170569900014&link_type=ISI) 53. 53.Taniguchi H, Yamaguchi A, Kunishima S, et al. Using the spleen for time-delay correction of the input function in measuring hepatic blood flow with oxygen-15 water by dynamic PET. Ann Nucl Med. 1999;13:215–221. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1007/BF03164895&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=10510876&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 54. 54.Oguro A, Taniguchi H, Koyama H, et al. Quantification of human splenic blood flow (quantitative measurement of splenic blood flow with H2 15O and a dynamic state method: 1). Ann Nucl Med. 1993;7:245–250. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1007/BF03164705&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=8292450&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 55. 55.Kudomi N, Koivuviita N, Liukko KE, et al. Parametric renal blood flow imaging using [15O]H2O and PET. Eur J Nucl Med Mol Imaging. 2009;36:683–691. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1007/s00259-008-0994-8&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=19050876&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 56. 56.Hopkins SR, Wielpütz MO, Kauczor H-U. Imaging lung perfusion. J Appl Physiol. 2012;113:328–339. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1152/japplphysiol.00320.2012&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=22604884&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 57. 57.Matsunaga K, Yanagawa M, Otsuka T, et al. Quantitative pulmonary blood flow measurement using 15O-H2O PET with and without tissue fraction correction: a comparison study. EJNMMI Res. 2017;7:102. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1186/s13550-017-0350-8&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=29274016&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 58. 58.Schuster DP, Kaplan JD, Gauvain K, Welch MJ, Markham J. Measurement of regional pulmonary blood flow with PET. J Nucl Med. 1995;36:371–377. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6Njoiam51bWVkIjtzOjU6InJlc2lkIjtzOjg6IjM2LzMvMzcxIjtzOjQ6ImF0b20iO3M6NDc6Ii9qbnVtZWQvZWFybHkvMjAyNS8wNC8zMC9qbnVtZWQuMTI0LjI2ODcwNi5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) 59. 59.Ibaraki M, Ito H, Shimosegawa E, et al. Cerebral vascular mean transit time in healthy humans: a comparative study with PET and dynamic susceptibility contrast-enhanced MRI. J Cereb Blood Flow Metab. 2007;27:404–413. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1038/sj.jcbfm.9600337&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=16736045&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 60. 60.Biffar A, Sourbron S, Schmidt G, et al. Measurement of perfusion and permeability from dynamic contrast‐enhanced MRI in normal and pathological vertebral bone marrow. Magn Reson Med. 2010;64:115–124. [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=20572150&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 61. 61.Kershaw LE, Buckley DL. Precision in measurements of perfusion and microvascular permeability with T1‐weighted dynamic contrast‐enhanced MRI. Magn Reson Med. 2006;56:986–992. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1002/mrm.21040&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=16986107&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 62. 62.Zhu B, Wang C, Gao J, Liu H, Li N, Teng Y. CT perfusion imaging of the liver and the spleen can identify severe portal hypertension. Abdom Radiol (NY). 2024;49:1084–1091. [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=38416165&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 63. 63.Sauter AW, Feldmann S, Spira D, et al. Assessment of splenic perfusion in patients with malignant hematologic diseases and spleen involvement, liver cirrhosis and controls using volume perfusion CT (VPCT). Acad Radiol. 2012;19:579–587. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1016/j.acra.2012.01.007&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=22326669&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 64. 64.Sourbron SP, Michaely HJ, Reiser MF, Schoenberg SO. MRI-measurement of perfusion and glomerular filtration in the human kidney with a separable compartment model. Invest Radiol. 2008;43:40–48. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1097/RLI.0b013e31815597c5&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=18097276&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) [Web of Science](http://jnm.snmjournals.org/lookup/external-ref?access_num=000251992900006&link_type=ISI) 65. 65.Eikefjord E, Andersen E, Hodneland E, et al. Use of 3D DCE-MRI for the estimation of renal perfusion and glomerular filtration rate: an intrasubject comparison of FLASH and KWIC with a comprehensive framework for evaluation. AJR. 2015;204:W273–W281. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.2214/AJR.14.13226&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=25714312&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 66. 66.Ohno Y, Hatabu H, Murase K, et al. Quantitative assessment of regional pulmonary perfusion in the entire lung using three‐dimensional ultrafast dynamic contrast‐enhanced magnetic resonance imaging: preliminary experience in 40 subjects. J Magn Reson Imaging. 2004;20:353–365. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1002/jmri.20137&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=15332240&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) [Web of Science](http://jnm.snmjournals.org/lookup/external-ref?access_num=000223522200001&link_type=ISI) 67. 67.Giardino ND, Friedman SD, Dager SR. Anxiety, respiration, and cerebral blood flow: implications for functional brain imaging. Compr Psychiatry. 2007;48:103–112. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1016/j.comppsych.2006.11.001&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=17292699&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 68. 68.Van Den Bergh O, Zaman J, Bresseleers J, Verhamme P, Van Diest I. Anxiety, pCO2 and cerebral blood flow. Int J Psychophysiol. 2013;89:72–77. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1016/j.ijpsycho.2013.05.011&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=23727628&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 69. 69.Kershaw LE, Cheng HM. Temporal resolution and SNR requirements for accurate DCE‐MRI data analysis using the AATH model. Magn Reson Med. 2010;64:1772–1780. [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=20715059&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 70. 70.Li S, Wang G. Deep kernel representation for image reconstruction in PET. IEEE Trans Med Imaging. 2022;41:3029–3038. [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=35584077&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 71. 71.Ottoy J, Verhaeghe J, Niemantsverdriet E, et al. 18F‐FDG PET, the early phases and the delivery rate of 18F‐AV45 PET as proxies of cerebral blood flow in Alzheimer’s disease: validation against 15O‐H2O PET. Alzheimers Dement. 2019;15:1172–1182. [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=31405824&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 72. 72.Wang G, Corwin MT, Olson KA, Badawi RD, Sarkar S. Dynamic PET of human liver inflammation: impact of kinetic modeling with optimization-derived dual-blood input function. Phys Med Biol. 2018;63:155004. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1088/1361-6560/aac8cb&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=29847315&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 73. 73.Poisson J, Lemoinne S, Boulanger C, et al. Liver sinusoidal endothelial cells: physiology and role in liver diseases. J Hepatol. 2017;66:212–227. [CrossRef](http://jnm.snmjournals.org/lookup/external-ref?access_num=10.1016/j.jhep.2016.07.009&link_type=DOI) [PubMed](http://jnm.snmjournals.org/lookup/external-ref?access_num=27423426&link_type=MED&atom=%2Fjnumed%2Fearly%2F2025%2F04%2F30%2Fjnumed.124.268706.atom) 74. 74.Wang Y, Abdelhafez YG, Spencer BA, et al. High-temporal-resolution kinetic modeling of lung tumors with dual-blood input function using total-body dynamic PET. J Nucl Med. 2024;65:714–721. [Abstract/FREE Full Text](http://jnm.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6Njoiam51bWVkIjtzOjU6InJlc2lkIjtzOjg6IjY1LzUvNzE0IjtzOjQ6ImF0b20iO3M6NDc6Ii9qbnVtZWQvZWFybHkvMjAyNS8wNC8zMC9qbnVtZWQuMTI0LjI2ODcwNi5hdG9tIjt9czo4OiJmcmFnbWVudCI7czowOiIiO30=) * Received for publication September 7, 2024. * Accepted for publication April 8, 2025. [1]: /embed/mml-math-1.gif [2]: /embed/mml-math-2.gif [3]: /embed/mml-math-3.gif [4]: /embed/mml-math-4.gif [5]: /embed/mml-math-5.gif [6]: /embed/mml-math-6.gif [7]: /embed/mml-math-7.gif [8]: /embed/mml-math-8.gif [9]: /embed/mml-math-9.gif [10]: /embed/mml-math-10.gif [11]: /embed/mml-math-11.gif [12]: /embed/mml-math-12.gif [13]: /embed/mml-math-13.gif [14]: /embed/mml-math-14.gif [15]: /embed/mml-math-15.gif [16]: /embed/mml-math-16.gif [17]: /embed/mml-math-17.gif [18]: /embed/mml-math-18.gif [19]: /embed/mml-math-19.gif [20]: /embed/mml-math-20.gif [21]: /embed/mml-math-21.gif [22]: /embed/mml-math-22.gif [23]: /embed/mml-math-23.gif [24]: /embed/mml-math-24.gif [25]: /embed/mml-math-25.gif [26]: /embed/mml-math-26.gif