Medical radioisotopes produce Cerenkov luminescence (CL) from charged subatomic particles (β+/−) traveling faster than light in dielectric media (e.g., tissue). CL is a blue-weighted and continuous emission, decreasing proportionally to increasing wavelength. CL imaging (CLI) provides an economic PET alternative with the advantage of also being able to image β− and α emitters. Like any optical modality, CLI is limited by the optical properties of tissue (scattering, absorption, and ambient photon removal). Shortwave-infrared (SWIR, 900–1700 nm) CL has been detected from MeV linear accelerators but not yet from keV medical radioisotopes. Methods: Indium-gallium-arsenide sensors and SWIR lenses were mounted onto an ambient light-excluding preclinical enclosure. An exposure and processing pipeline was developed for SWIR CLI and then performed across 6 radioisotopes at in vitro and in vivo conditions. Results: SWIR CL was detected from the clinical radioisotopes 90Y, 68Ga, 18F, 89Zr, 131I, and 32P (biomedical research). SWIR CLI’s advantage over visible-wavelength (VIS) CLI (400–900 nm) was shown via increased light penetration and decreased scattering at depth. The SWIR CLI radioisotope sensitivity limit (8.51 kBq/μL for 68Ga), emission spectrum, and ex vivo and in vivo examples are reported. Conclusion: This work shows that radioisotope SWIR CLI can be performed with unmodified commercially available components. SWIR CLI has significant advantages over VIS CLI, with preserved VIS CLI features such as radioisotope radiance levels and dose response linearity. Further improvements in SWIR optics and technology are required to enable widespread adoption.
Subatomic relativistic particles generate Cerenkov luminescence (CL) in dielectric media (1). The traveling particle polarizes surrounding molecules, which generate luminescence upon relaxation (2), with CL being ultraviolet-weighted. CL production decreases exponentially with increasing wavelength (1/λ2) (3). CL’s intensity, but not the spectrum profile, correlates with the energy of the emitted particle and has been used in astrophysics, nuclear physics, and recently in biomedical imaging (4,5). CL imaging (CLI) is a cost- and time-effective PET alternative for surface-weighted imaging, for example, for triaging patients into those who do not need PET and those who do (2,6,7). CLI has focused on CL detection from clinical β- (β+/−) emitting radioisotopes and linear accelerators (LINAC) (4,8,9). Preclinical discoveries and the development of novel targeted radiotracers, dosimetry, and radiotherapy-based treatments have been aided by CLI (10,11). Clinical CLI has also found applications in image-guided surgery for detecting margins, determining the clinical uptake of radiotracers, and making real-time dosimetry readings (7,12–15).
Single-photon-sensitive devices responsive to the CL spectrum are readily available, including low-dark-noise and low-read-noise optical devices (2). However, visible wavelength (VIS) CLI (400–900 nm) suffers significant drawbacks in preclinical and clinical settings. Endogenous chromophores limit the achievable VIS CL penetration depths, with scattering further reducing the resolution, contrast, and sensitivity of VIS CLI at depth (16,17). Optical imaging has shifted to longer wavelengths at which light absorption and scattering are reduced (18). Near-infrared (NIR) imaging (>650 nm) reduces absorption by approximately 2 orders of magnitude over VIS imaging (16,19). Studies have used dyes and nanoparticles for the red-shifted conversion of CL (20–22). Shortwave-infrared (SWIR) imaging (900–1,700 nm) has shown that tissue absorption, scattering, and autofluorescence are of negligible levels (23). SWIR’s advantages for resolution and contrast have been shown via the clinically approved indocyanine green (24). Förster resonance energy transfer--based SWIR CL has been achieved via x-ray-excited nanoprobes and LINAC-excited emission of quantum dots (20,25–27). LINAC SWIR CLI was demonstrated without secondary emitters, showing improvements over VIS-NIR CLI (28). However, radioisotopes produce CL that is an order of magnitude dimmer than LINACs and are therefore more demanding to image (LINAC, 6–24 MeV; 68Ga, 0.836 MeV) (29–32). Radioisotope CLI requires complete exclusion of ambient light combined with efficient optical imaging systems and cannot be pulse-synchronized in its acquisition as LINAC CLI can (30,33). The already low CL radiance of radioisotopes is further reduced at SWIR wavelengths, magnifying the difficulty of SWIR CL detection (3). Nevertheless, in this work SWIR CLI from radioisotopes was achieved via unmodified and commercially available imaging components and revealed defined advantages over VIS CLI.
MATERIALS AND METHODS
Radioisotope SWIR Setup
SWIR CLI was achieved via commercially available indium-gallium-arsenide focal plane arrays (NIRVana 640 TE [Teledyne Princeton] or ZephIR 1.7 [Photon etc. Inc.]) and a SWIR lens set to ƒ/1.4 (SWIR-16 [Navitar] or 8-mm SWIR lens 83-815 [Edmund Optics]) mounted on an enclosure (Fig. 1). Figures 1B–1E were acquired without optical filters (NIRVana 640 TE, 900–1,700 nm) (34). All other figures used a 650- or 900-nm optical density 4.0 long-pass filter (Edmund Optics 84-759 and 84-764). Acquisition (ninety 10s frames [15 min]) was controlled via respective acquisition software. Equivalent dark noise was recorded and subtracted for each sensor. White-light images were acquired with room lights on and enclosure door open as the room lights had SWIR spectral emission (Pentron 3000K; Osram Sylvania) (35). Radioisotopes were imaged in suitable containers within a lead pig. Black posterboard (TB5; Thorlabs) was used to block CL and allow γ particles to pass.
SWIR CL radioisotope imaging setup and characterization. (A) A dark enclosure was used for all imaging. Mounted on the camera was a 16 or 8 mm ƒ/1.4 SWIR lens. (©2022 Memorial Sloan Kettering Cancer Center. All rights reserved.) (B) Linearity assessment of the camera via 68Ga imaged at positions (P1, P2) across the field of view, with P1 + cardboard confirming that detection was not as a result of γ strikes. (C) SWIR gray levels and corresponding 68Ga activity (MBq). Linear regression and 95% CIs are shown (Pearson R2 = 0.9839, 2-tailed P < 0.0001). SWIR CLI is linearly responsive and quantitative as found with VIS CLI. (D) 32P SWIR gray-value intensity in relation to exposure time changes. (E) Graphical representation of radioisotope (32P) SWIR CL emission spectrum from 900 to 1,500 nm, with the blue line representing a 1-phase exponential decay function (R2 = 0.9812). Inherent system noise, low photon production, and water absorption prevent detection > 1,400 nm. (F) Descending radioisotope radiance (32P, 68Ga, 89Zr, 18F, and 131I) corrected for concentration (kBq/μL) and spatial field of view (cm2). Student t test (unpaired, 2-sided) P values are shown. In all cases, mean measurements (red line), SDs (black lines), and individual measurements (n = 90 technical replicates, gray dots) are shown, for radiance comparison negative values were excluded.
Determining the Radioisotope SWIR CLI Temporal Resolution and Emission Spectrum
The SWIR CL temporal resolution and emission spectrum were determined using 370 MBq of 32P in 1 mL of water (β− average, 0.695 MeV; half-time, 14.3 d; NEX060005MC [Perkin Elmer]). Ninety-frame acquisitions at respective exposure times were performed (Fig. 1D; Supplemental Fig. 1; supplemental materials are available at http://jnm.snmjournals.org). The limit of detection was defined as when SWIR CL was not determinable from the noise. Long-pass filters (1,000–1,500 nm in 100-nm steps (FELH1000:100:1500, SM1L03, and SM1A57; Thorlabs) elucidated the SWIR radioisotope spectrum.
Image Processing and Statistical Analysis
Fiji (ImageJ, version 2.0 (36)) was used for image processing (Supplemental Fig. 2). Dark noise was subtracted from the data followed by binning (8 × 8, to improve sensitivity), median filtering (outlier, γ strike removal), fast-Fourier-transform bandpass transformations (artifact removal), and rolling-ball background subtraction (37). Images were resized for white-light overlays. Statistical analysis graphing was performed in Prism, version 9 (GraphPad Software LLC). Statistical analyses and replicate information are shown in all cases. Radiance was calculated via gray values corrected for isotope concentration (kBq/μL) and field of view (cm2). Full width at half maximum was measured in MATLAB (version 2020b; MathWorks Inc.).
Silica Nanoparticle Radiolabeling and Injection
Silica nanoparticles (SiNPs) were labeled with 68Ga or 90Y as previously described (38). SiNPs were incubated with free isotope at a pH of 8.8 for 60 min, on a thermomixer at 70°C and 500 rpm. Radiolabeled SiNPs were resuspended in 30 μL of saline for footpad injection.
Preclinical SWIR CLI
All experiments were performed in accordance with Institutional Animal Care and Use Committee guidelines at Memorial Sloan Kettering Cancer Center and with National Institutes of Health guidance on the care and use of laboratory animals. Three percent isoflurane in 100% O2 v/v was followed by 1%–2% isoflurane in 100% O2 v/v for maintenance of anesthesia; euthanasia was performed using CO2 in accordance with approved protocols. All mice (10 in total, FoxN1NU, stock no. 069; Envigo) were suitably housed with food and water ad libitum. Four xenograft mice were generated by injecting 0.3 × 106 4T1 cells (ATCC, CRL-2539) suspended in 30 μL of Matrigel (catalog no. 354234; Corning) into the fourth mammary fat pad (Supplemental Fig. 3). Three euthanized xenografted mice were injected (mice were randomized to receive a range of activity from 0–166.5 MBq) with up to 166.5 MBq of 18F-FDG, with 1 mouse acting as a negative control. One non-xenografted euthanized mouse was injected into the footpad with 68Ga-labeled SiNPs. Four non-xenografted mice were injected with 90Y-labeled SiNPs, with an additional mouse acting as a negative control for in vivo experiments.
RESULTS
Confirmation of SWIR CL Detection from Radioisotopes
In Figure 1A, the SWIR CLI radioisotope setup is shown detecting 68Ga suspended in 0.1 M HCl (39,40). Figure 1B shows the processed SWIR CL image overlaid onto the white-light image. This demonstrates that SWIR signal is coming from 68Ga and that the processing steps sufficiently removed γ strikes while retaining CL signals. CL detection is further confirmed by a lack of signal when cardboard was placed over the sample, blocking light but not the highly energetic photons from 68Ga’s decay (e.g., 511 and 1,077 keV). The 68Ga source was further moved and detected around the field of view, decaying in the process, visible via a decreasing SWIR signal, following the decay half-life of the isotope. Manual regions of interest were measured to determine the signal intensity (gray values) of the sensor for each position and time point. SWIR CLI showed quantitative linearity to 68Ga levels, as previously reported for VIS CLI (4). We then successfully detected 4 additional radioisotopes under comparative conditions: 32P,18F, 89Zr, and 131I (Fig. 1; Supplemental Fig. 4). The SWIR radiance of each was calculated as before (Fig. 1), with values corrected for concentration (kBq/μL) and spatial field of view (cm2). SWIR CLI readily differentiated the tested radioisotopes in line with VIS CLI, as shown by the P values in Figure 2B (41,42). Theoretically 68Ga has a higher radiance than 32P; however, the increased sensor strikes from 511- and 1,077-keV photons by 68Ga prevent SWIR CLI from confirming this, as the setup did not incorporate lead shielding (42).
Reduced scattering via SWIR CLI (900–1,700 nm) over VIS CLI (400–900 nm). (A) Left, normalized VIS CLI line intensity profiles of a 90Y source (Eppendorf, 55.5 MBq in 200 mL of saline) at increasing scattering tissue depths (chicken breast: 0, 10, and 15 mm). Full widths at half maximum increase with scattering tissue depth at 0, 10, and 15 mm with respective full widths at half maximum of 6.38, 13.05, and 33.64 mm. Right, representative VIS CLI images. (B) Left, normalized SWIR CLI line intensity profiles from the phantom setup with full widths at half maximums of 6.24, 6.40, and 7.04 mm. Right, representative SWIR CL images. In all cases, 3 separate line measurements are made from images at each depth (dotted lines), with mean shown (solid line). FWHM = full width at half maximum.
SWIR CLI Radioisotope Temporal Detection Limit and Emission Spectrum
The system was characterized using 32P (β−), with concentrations used to facilitate SWIR CLI at a temporal resolution of 0.25 s (Fig, 1; Supplemental Fig, 1). 32P’s radiance and half-life enabled determination of the radioisotope SWIR CL emission spectrum (42). No filter was used (>920 nm, spectral response of the sensor), and long-pass filtered acquisitions (1,000–1,500 nm, 100-nm steps) were performed. This sensor has a nonthinned indium phosphide substrate bandgap (1.35 eV) preventing detection of light less than 920 nm, and the bandgap of indium-gallium-arsenide (0.75 eV) results in a short-pass cutoff at 1,700 nm (43–45). The radioisotope SWIR CL spectrum is shown in Figure 1E, with intensity exponentially decreasing (1-phase exponential decay, R2 = 0.9812), as expected and reported for VIS CLI and LINAC SWIR CLI (2,28). Detection above 1,400 nm is challenging because of system noise, lens inefficiency, and water absorption (Figs. 1E and 1F; Supplemental Figs. 1 and 7) (19,46).
Reduced Scattering via SWIR CLI over VIS CLI
A 55.5-MBq quantity of 90Y in an Eppendorf tube (β− emitter; average, 0.94 MeV; half-time, 64 h in 200 μL of saline) was imaged on both the SWIR (900–1,700 nm) and the IVIS (PerkinElmer) (VIS CLI, 400–900 nm) imaging systems to assess the advantages of SWIR CLI. 90Y is clinically administered from 0.500 to 5.291 GBq to treat primary liver cancer (hepatocellular carcinoma) and metastases in the liver (47,48). Scattering medium (raw chicken breast: 0, 10, and 15 mm) was placed over the source before CLI was performed. Respective exposure times without scattering medium (VIS CLI, 10 s; SWIR CLI, 900 s) were maintained at all tissue depths. SWIR CLI shows an improvement in resolution compared with VIS CLI when imaging through scattering tissue (Fig. 2). In SWIR CLI, the Eppendorf tube shape is consistent up to 15 mm of tissue, whereas the shape is distorted and enlarged over 5 times from scatter in VIS CLI. In this investigation SWIR CLI provided greater resolution at depth than VIS CLI for radioisotope location. This increased accuracy is demonstrated by the full widths at half maximum (6.38, 13.05, and 33.64 mm for VIS CLI and 6.24, 6.40, and 7.04 mm for SWIR CLI at 0, 10, and 15 mm of tissue, respectively) (Fig. 2).
Sensitivity Limits of SWIR CLI Radioisotope In Vitro and Ex Vivo
Commercial SWIR sensors are insensitive to light of less than 920 nm without indium phosphide cap thinning. In systems with indium phosphide thinning, the sensor has a largely increased range but is rather insensitive compared with silicon-based electron-multiplying charge-coupled devices. One such SWIR sensor, the ZephIR 1.7x [Photon etc. Inc.]), was used with significantly reduced quantum efficiency in the VIS range (∼25% at 600 nm). The sensor enabled a comparison of the spatial localization of radioisotope VIS-SWIR (400–1,700 nm), NIR-SWIR (650–1,700 nm), and SWIR (900–1,700 nm) CLI. 68Ga-conjugated SiNPs (in 30 μL of saline) opposite a nonradiolabeled SiNP control were imaged at VIS-SWIR, NIR-SWIR, and SWIR spectra for numerous half-lives via appropriate long-pass filters. The detected CL signals are localized to the radiolabeled SiNPs throughout the spectra (Supplemental Figs. 5A–5D). The phantom detection limit of radioisotope SWIR CLI was 259 kBq (8.51 kBq/μL) for 68Ga-labeled SiNPs (Figs. 3A–3B). Theoretical limits for radioisotopes tested here in SWIR CLI are shown in Supplemental Table 1 (42).
In vitro and ex vivo SWIR CLI (900–1,700 nm) radioisotope sensitivity limit for 68Ga radiolabeled SiNPs. (A) SWIR CLI radioisotope in vitro detection limit for 68Ga-SiNPs after multiple half-lives. (B) SWIR CLI decay tracking to the limit of detection, linear regression (solid black line, R2 = 0.9882) and 95% CIs (solid gray lines) are shown. (C) Ex vivo SWIR CLI limit of detection for 68Ga-labeled SiNPs. The detection limit slightly worsens in tissue compared with in vitro imaging (∼140 kBq less sensitive). (D) Linear regression analysis (R2 = 0.9923) of ex vivo SWIR CLI of 68Ga-labeled SiNPs to limit of detection (403.3 kBq, C), 95% CIs are shown (solid gray lines). Detection limit paw is labeled with a white arrow.
Next, the SWIR CLI detection limit in tissue was assessed. A euthanized mouse received 30 μL of respective paw injections of varying 68Ga-SiNP activities. As can be seen in Figures 3C–3D, 68Ga-SiNPs were detected via SWIR CLI down to 403.3 kBq, within clinically administered levels (∼148 MBq) (49).
In Vivo Detection of Radioisotope SWIR CLI
90Y conjugated to SiNPs was used for in vivo SWIR CLI radioisotope detection (38). 90Y has been shown to improve the SNR of CL detection over, for example, 18F or 68Ga because of reduced γ sensor strikes (50). During in vivo imaging, the long exposure times for SWIR CLI contaminated the SWIR CL image with thermal signatures (removed in postprocessing via rolling-ball background subtraction), complicating SWIR CLI over VIS CLI. Mice were injected with approximately 7.4 MBq of 90Y-labeled SiNPs into a single footpad and imaged 3 h later. 90Y is administered clinically for radioembolization at activities ranging from 500 to 5,291 MBq (47,48). The resulting SWIR CLI signal was readily detected over background thermal signatures present in vivo (4 mice; Fig. 4; Supplemental Fig. 6). Footpad-injected radiolabeled SiNPs slowly migrate through the lymphatic system (48 h), preventing CL contamination of the endogenous thermal signature (38). Respective residual thermal signatures remaining after background subtraction from each mouse were used to divide the image, producing measurements in terms of signal-to-thermal-background ratio (Supplemental Fig. 2). Signal-to-background ratios from injected mice ranged from 1.68 to 4.63, with a mean of 3.07 (Fig. 4B).
In vivo SWIR CLI detection of 90Y-labeled SiNPs 3 h after injection into footpad. (A, left) Representative image of mouse injected (+) with 90Y-labeled SiNPs (∼7.4 MBq). (A, right) Image of control mouse without any injection. (B) Quantified signal-to-background ratios of injected (n = 4) vs. control mice (n = 1). All images are shown in respective signal-to-background ratios. SBR = signal-to-background ratio.
DISCUSSION
This work aimed to detect the theoretical SWIR CL emission from clinical radioisotopes via commercially available components (Supplemental Fig. 7). To date, SWIR CL has been detected only from LINAC sources, which produce CL that is an order of magnitude brighter (25,28). The devised setup and enclosure provided ambient light-free imaging (51,52). The state-of-the-art thermoelectric cooled (TEC) SWIR sensors used in our study produce dark noise 2–3 orders of magnitude higher than cameras based on silicon electron-multiplying charge-coupled devices. This intrinsic noise within the SWIR setup, an obstacle not only for SWIR CLI, has been the main limiting factor throughout this work (53). In addition, the even further reduced light output at longer wavelengths necessitated longer exposure times (minutes for SWIR CLI vs. seconds for VIS CLI). Nevertheless, the radioisotopes we explored produced detectable SWIR CL (Fig. 1), with relative radiances in line with VIS CLI (3,41,42). Importantly, SWIR CLI increased linearly with radioisotope levels (Fig. 1; Fig. 3). SWIR CLI radioisotope sensitivity was found to be 259 kBq (8.51 kBq/μL) in vitro and 403.3 kBq in tissue with 68Ga. VIS CLI is 4 orders of magnitude more sensitive, with 68Ga detection reported at 0.00333 kBq/μL (54). The insensitivity of SWIR CLI limits applications with the current generation of SWIR cameras. However, its advantage of reduced scattering in combination with the increased transmission of CL in the SWIR region (Fig. 2; Supplemental Fig. 7), potentially paired with SWIR-void room lighting, could open new applications for radioisotope SWIR CLI (28) without light-excluding enclosures that also benefit from improved radioisotope resolution at depth. Combining the strengths of indium-gallium-arsenide–based (900–1,700 nm) and silicon-based (400–900 nm) sensors would open further applications for CLI and is worth investigating (16,19,46,55,56). Assuming sufficient radioisotope levels, SWIR CLI can be performed close to video rates (0.25 s; Fig. 1).
The detected radioisotope SWIR CL spectrum was found to be in line with the SWIR LINAC spectrum (28). Similarly to SWIR LINAC CL, theoretically emitted CL above 1,400 nm could not be reliably detected, likely because of sensor noise and water absorption (Supplemental Fig. 7). However, the advantage of wavelengths above 1,400 nm compared with 900–1,300 nm is unclear for SWIR CLI in any case because of the increased absorption of water above 1,300 nm (19,46).
The preclinical applicability of SWIR CL was investigated and initially focused on the ex vivo SWIR CLI of intratumorally injected 18F-FDG (Supplemental Fig. 3). The weak CL of 18F (26 times dimmer than 68Ga) required 1h acquisition times for accurate signal detection (42,54). 90Y was used to overcome the limitations of 68Ga and 18F for this investigation (negligible γ emission, brighter CL, longer half-life than 68Ga, 64.2 h vs. 1.13 h) (2,41). 90Y-labeled SiNPs were injected into the footpad of live mice for in vivo SWIR CLI detection (38). Mice were imaged 3 h after injection (∼7.4 MBq of 90Y-SiNPs) together with a noninjected control mouse. Exposure times of 15 min along with suitable image processing steps provided a reliable SWIR CL signal that was detected over the endogenous thermal signature and inherent noise (Fig. 4).
CONCLUSION
Our study presents the first example and proof-of-principle of in vivo radioisotope SWIR CLI. Considering the optical properties of tissue, it has been shown that most CL emitted from tissue at depth is above 600 nm (3,16). Therefore, the ideal radioisotope CL camera would be one that combines the spectral range of thinned SWIR sensors (600–1,700 nm, see Supplemental Fig. 7) with the photon sensitivity of sensors based on silicon electron-multiplying charge-coupled devices. Future iterations of SWIR CLI should aim to tackle its main limitations via improved lenses (f/0.95) and a dark-noise–reduced camera sensor to improve overall sensitivity, along with lead shielding to further increase sensitivity via γ strike reduction. Such components would be highly customized and outside the scope of this proof-of-principle work. Human eyes respond to light from approximately 400 to 700 nm, and by changing ambient lighting to non–SWIR-emitting light-emitting diodes, radioisotope SWIR CLI might be performed in a well-lit room and without the need for a dark enclosure, as achieved for VIS and SWIR LINAC CLI (7,30). This would directly impact radioisotope preclinical CLI, which is a common, fast, and cost-effective PET alternative for novel radiotracers and treatment tracking (57,58). However, significant improvements are required in SWIR optics and technology before this can be realized.
DISCLOSURE
Funding was in part provided by National Institutes of Health NCI grants R01CA183953 (to Jan Grimm) and P30 CA08748 (to Craig Thompson/Memorial Sloan Kettering Cancer Center). No other potential conflict of interest relevant to this article was reported.
KEY POINTS
QUESTION: SWIR CLI (900–1,700 nm) has so far only been detected from LINACs. Is it possible to detect SWIR CLI from radioisotopes, orders of magnitude dimmer?
PERTINENT FINDINGS: This work detected SWIR CLI via commercially available components from numerous preclinical and clinical radioisotopes, with the radiances and spectrum performing in a manner consistent with VIS CLI (400–900 nm).
IMPLICATIONS FOR PATIENT CARE: SWIR CLI has distinct advantages over VIS CLI in terms of reduced scattering and absorption at depth; however, significant technologic and optical improvements are required for comparable sensitivity.
ACKNOWLEDGMENTS
We thank Craig Wall, Manjul Shah, Sabbir Liakat, and Princeton Teledyne for providing the NIRvana 640 TE camera and lens; Photon etc. for the Zephir 1.7x camera; and Valerie Longo and Pat Zanzonico of Memorial Sloan Kettering Cancer Center for animal experiment assistance and logistics of equipment setup.