Abstract
Molecular breast imaging (MBI) using 99mTc-sestamibi has advanced rapidly over the past decade. Technical advances allow lower-dose, higher-resolution imaging and biopsy capability. MBI can be used for supplemental breast cancer screening with mammography for women with dense breasts, as well as to assess neoadjuvant therapy response, evaluate disease extent, and predict breast cancer risk. This article highlights the current state of the art and future directions in MBI.
The term molecular breast imaging (MBI) may broadly refer to all nuclear medicine and advanced MRI techniques for breast imaging. However, MBI usually refers to dedicated breast imaging systems using 99mTc-sestamibi with dual-head solid-state cadmium-zinc-telluride (CZT) detectors. Compared with breast-specific γ-imaging that uses single-head scintillation detectors, MBI uses dual-head CZT detectors and advanced collimator designs to improve spatial resolution, shorten imaging times, and facilitate lower-dose imaging. This article summarizes the current state of the art and the future uses of MBI.
CURRENT STATE OF THE ART
There are 2 commercially available MBI systems: LumaGem (CMR Naviscan) and Eve Clear Scan (SmartBreast Corp.). Both use CZT semiconductor detectors that directly convert γ-photons into electrons. Compared with single-head breast-specific γ-imaging systems, dual-head CZT detectors improve detection of small lesions (1).
The primary radiopharmaceutical used for MBI is 99mTc-sestamibi. 99mTc-sestamibi uptake in breast tumors results from increased mitochondrial density, increased blood flow, and negative membrane potentials of tumor cells (2). 99mTc-sestamibi has been used in hundreds of thousands of patients with a favorable safety profile and few contraindications, namely pregnancy and a prior allergic reaction to 99mTc-sestamibi (2). Patients who are breast feeding may continue to do so, unless otherwise specified by institutional guidelines. 99mTc-sestamibi clears from the bloodstream rapidly within 2–3 min; thus, image acquisition can begin within 5 min after injection (3). Having the patients fast, warming them with a blanket, and having them keep still may reduce background breast uptake (4).
Standard craniocaudal and mediolateral oblique projections of each breast are acquired using gentle compression, with 7–10 min per view (4). Breast positioning is comparable to mammography, aiding correlation of findings between mammography and MBI. MBI detector size is 20 × 16 cm or 24 × 16 cm, compared with 24 × 29 cm for mammography. Tiled acquisitions may be required with MBI if the breast size exceeds the field of view. Lesions near the chest wall and axillary lymph nodes may be difficult to visualize with MBI, because of increased dead space at the edge of the field of view (8 mm) compared with mammography (4 mm) (4). Furthermore, MBI detectors are opaque, making maximal posterior positioning difficult to confirm (4). The patient can breathe normally during imaging, with instructions to otherwise remain still to reduce motion blur. Performing screening MBI during the follicular phase may minimize background parenchymal uptake (BPU), and menstrual cycle phase may be included in the report (5). However, no studies have examined menstrual cycle timing and MBI diagnostic performance, and recent data from screening breast MRI outcomes indicate that scheduling based on menstrual cycle phase is not necessary (6). MBI should not be scheduled within 4 half-lives of other 99mTc-based studies, PET/CT, or targeted radionuclide therapies because of photon energy window overlap.
RADIATION EXPOSURE
The absorbed radiation dose during MBI is proportional to the administered activity of 99mTc-sestamibi. Optimizations in detector design, patient preparation, and tracer delivery have reduced the administered activity from 740–1,100 MBq (20–30 mCi) to the current off-label standard of 240–300 MBq (6.5–8 mCi) (7,8). Weight-based adjustments for injected activity are not necessary for MBI (9). The average absorbed radiation dose to the breast from 300 MBq (8 mCi) of 99mTc-sestamibi is estimated to be 1.1 mGy, compared with 3.0–4.5 mGy with mammography and tomosynthesis.
Since 99mTc-sestamibi is systemically distributed, tissues outside the breast receive the largest radiation dose. The estimated effective (whole-body) dose for 300 MBq (8 mCi) of 99mTc-sestamibi is 2.1–2.6 mSv, which is at, or lower than, annual natural background levels (∼3 mSv) (10). For reference, the effective dose of chest CT can approach 6 mSv (11). Tissues with the highest exposures include the colon (7.1 mGy), urinary bladder (3.2 mGy), and gallbladder (11.5 mGy) (12). Various organizations assessing radiation risk and radiation protection (Health Physics Society, American Association of Physicists in Medicine, International Organization for Medical Physics, and United Nations Scientific Committee on the Effects of Atomic Radiation) state that risks from radiation doses of less than 100 mSv are not significantly different from zero (10). Thus, current MBI radiation exposure is deemed to pose negligible risk to the patient, with minimal theoretic risk of inducing cancer in any of these organs (10).
The administered activity of 99mTc-sestamibi for MBI continues to decrease with technologic advancement. Tao et al. showed that new image-processing algorithms maintain lesion conspicuity with a simulated half-dose (150 MBq [4 mCi]) injection (13). Continued advances in CZT module design will improve sensitivity and should allow further dose reduction.
INDICATIONS AND CLINICAL APPLICATIONS
Supplemental Screening
Approximately 43% of the 40 million individuals in the U.S. screening population have dense breast tissue (14–16). Screening mammography has relatively reduced sensitivity for breast cancer detection in individuals with dense breasts (15). Furthermore, breast cancer risk is higher for individuals with dense breasts (15,17). Supplemental screening may therefore benefit 16–20 million U.S. individuals with dense breasts who are at risk for false-negative results on screening mammography. Breast cancer detection with supplemental screening methods such as whole-breast ultrasound, breast MRI, contrast-enhanced mammography, and MBI is not limited by dense breast tissue (15).
The ideal supplemental screening modality would have a high breast cancer detection rate and a high negative predictive value while maintaining safety, affordability, ease of access, quick imaging acquisition and interpretation times, and acceptable rates of false-positive findings. Although MBI addresses some of these issues, it also has some limitations as a screening test. MBI excels with high lesion-to-background contrast in individuals with dense breasts or breast implants (18), with an incremental cancer detection rate of 8.8 per 1,000 exams on the first MBI screening (8). For comparison, the baseline detection rate of screening mammography is approximately 3.2 cancers per 1,000 individuals screened (19). The incremental cancer detection rate per 1,000 exams in individuals with dense breasts is approximately 1.7 for tomosynthesis, 2.7 for ultrasound, 15 for full-protocol MRI, and 10 for abbreviated MRI (15). MBI also provides favorable false-positive rates (20), favorable recall rates (8), rapid interpretation times (18), low cost (15), few contraindications (21), and a high negative predictive value (22). Disadvantages of MBI for supplemental screening include an imaging exam time of 28–40 min (15), the need for nuclear medicine licensing and coordination between nuclear medicine and breast imaging sections, lack of widespread availability, variable insurance coverage, and a higher whole-body radiation exposure than for other supplemental screening options (10,13,15).
When used for supplemental screening, the estimated benefit of MBI in terms of deaths averted is 5–9 times greater than the estimated risk of lives lost (23), and the estimated benefit-to-radiation risk ratio now approaches that of mammography (23). However, concerns regarding MBI radiation risks, though disputed, have delayed widespread use for supplemental screening (10,21,23). The American College of Radiology appropriateness criteria for supplemental breast cancer screening based on breast density currently do not recommend MBI (24).
The prospective multicenter Density MATTERS trial compares MBI with tomosynthesis for supplemental screening in women with dense breasts. This is the first trial to examine incidence screening MBI in which performance is evaluated not only for the initial scan but also at a subsequent screening round. Preliminary results show an incremental cancer detection rate for MBI beyond tomosynthesis of 9.3 per 1,000 screened, with 6 invasive cancers seen only on MBI (median size, 1 cm; 5 of 6 lymph node–negative) among 537 of the planned 3,000 participants (25). By comparison, the Eastern Cooperative Oncology Group/American College of Radiology Imaging Network EA1141 trial of abbreviated breast MRI for screening women with dense breasts found that abbreviated breast MRI after tomosynthesis had an incremental invasive cancer detection rate of of 7 per 1,000 screened (26). Thus, MBI may detect invasive breast cancers occult on tomosynthesis in women with dense breasts at a rate similar to that of abbreviated breast MRI. Studies of women invited to undergo screening breast MRI showed that 41%–42% declined participation for reasons including lower socioeconomic status, self-reported contraindications to MRI (27), claustrophobia, and financial concerns (28). MBI may be considered an alternative for such individuals.
Local Tumor Extent
For newly diagnosed breast cancer, accurate determination of disease extent is important for defining clinical stage and for planning therapy and surgery. MBI can be used for local tumor staging to detect multifocal, multicentric, or contralateral malignancy, especially for patients who cannot obtain preoperative breast MRI (Fig. 1). A study comparing extent of disease using MBI, contrast-enhanced mammography, and MRI found that MBI and contrast-enhanced mammography were effective for local staging, with similar visualization of the index cancers and increased specificity for additional cancers compared with MRI (20). However, unlike MRI, MBI is limited for evaluation of axillary and chest wall disease. In a study of 90 patients with breast cancer before starting neoadjuvant chemotherapy (NAT), MBI detected 16 of 20 tumors smaller than 1 cm, compared with 17 of 20 tumors for MRI (29). An additional limitation of MBI for surgical planning regards invasive lobular carcinomas, which have less intense sestamibi uptake than invasive ductal carcinomas, resulting in lower detection (30).
NAT
NAT is often the standard of care for locally advanced breast cancer. NAT can reduce tumor burden, allowing less extensive surgery. MBI is relatively accurate for prediction of pathologic response and evaluation of residual disease after NAT (Fig. 2). A metaanalysis of 14 studies with 529 breast cancer patients found that MBI had a sensitivity of 70.3% (95% CI, 56.5%–81.3%) and a specificity of 90.1% (95% CI, 77.5%–96.0%) for residual disease (31). A retrospective report of 114 patients who underwent NAT found slightly lower sensitivity for MBI than for MRI (70% vs. 83%) but higher specificity (90% vs. 60%) for residual disease, with similar overall performance (respective κ-values, 0.47 vs. 0.41 [P < 0.001]) (32). A prospective study of 104 patients undergoing NAT found that MBI had lower sensitivity than MRI (58.9% vs. 82.8%) but improved specificity (82.4% vs. 69.4%) in evaluating residual disease (29). The accuracy of MBI for residual disease assessment depends on tumor molecular subtype, with the highest accuracy for triple-negative and human epidermal growth factor receptor 2–positive subtypes and the lowest accuracy for luminal subtypes (33). However, no imaging technique is currently able to definitively determine complete pathologic response to therapy without surgical confirmation, in part because of inherent resolution limitations.
PRESENT AND FUTURE DEVELOPMENTS
MBI-Guided Biopsy
For suggestive MBI findings, biopsy is necessary to confirm malignancy. For MBI-detected masses, targeted ultrasound and ultrasound-guided biopsy are performed, paralleling MRI workflows. In a study of 1,585 examinations performed before the availability of MBI biopsy, 115 MBI-positive findings were detected, resulting in 50 biopsy recommendations. Of these 50 lesions recommended for biopsy on the basis of MBI, 38 (76%) were visible sonographically, but 12 (24%) required MBI biopsy (34). The billed expense of MRI-guided biopsy is estimated at $3,500, whereas MBI-guided biopsy is roughly half this amount (35).
MBI-directed biopsy is currently available for 1 system (Eve Clear Scan). This unit is a self-contained accessory mounted on the dual-head system and includes an angled pair of CZT detectors for obtaining stereotactic views, allowing specimen imaging to confirm adequate sampling. A breast-specific γ-imaging–guided biopsy system has also been described (36). MRI-guided biopsy can be used if an MRI correlate is identified, but it does not allow specimen imaging.
Breast Cancer Risk Assessment
Numerous models exist for predicting an individual’s breast cancer risk (37). These models inform whether and when individuals should undergo high-risk screening and potentially benefit from risk-reducing endocrine therapy or prophylactic mastectomy.
Imaging biomarkers are increasingly used for risk stratification (38). Breast density is an anatomic imaging biomarker associated with increased breast cancer risk (15,39). Growing evidence suggests that functional imaging modalities, such as MBI and MRI, can predict breast cancer risk (40–42), perhaps more accurately than risk models alone (42). BPU on MBI is a reliable quantitative and qualitative biomarker (41,43) that describes the amount of radiotracer uptake in normal breast tissue relative to subcutaneous fat. BPU is assessed qualitatively as photopenic, minimal to mild, moderate, or marked (Fig. 3) (44). BPU is analogous to background parenchymal enhancement on MRI, which describes the amount of enhancement in normal breast tissue and is similarly assessed qualitatively as minimal, mild, moderate, or marked (45). Beyond predicting breast cancer risk, BPU varies with menopausal status and dose of risk-reduction endocrine therapy (46). A large single-center study found that BPU is an independent risk factor for breast cancer in postmenopausal women with elevated BPU and dense breasts at highest risk, well above the threshold for considering use of risk-reducing therapy (47). Future studies incorporating functional imaging biomarkers into breast cancer risk models may facilitate personalized screening and risk management.
Novel Radiotracers
MBI systems can also be used for research using γ-emitting radionuclides targeting more specific aspects of tumor biology beyond 99mTc-sestamibi uptake. Radiolabeled antibodies, peptides, and receptor ligands may broaden future applications of MBI for tumor phenotyping, neoadjuvant treatment selection, and therapy response prediction. Use of MBI for αVβ3 integrin–targeted imaging of tumor angiogenesis has been explored in 2 small studies (48,49). Other 99mTc-labeled radiotracers targeting specific receptors such as estrogen receptor and human epidermal growth factor receptor 2 have been developed and could use MBI technology but have not been evaluated yet.
CONCLUSION
MBI is an important emerging breast imaging technology for supplemental screening, local tumor staging, neoadjuvant therapy response assessment, and breast cancer risk assessment.
DISCLOSURE
Matthew Covington is a consultant for Invicro, LLC, for medical imaging review. Ephraim Parent receives research funding from Blue Earth Diagnostics. Gaiane Rauch receives research funding from GE Healthcare. Amy Fowler receives research support from GE Healthcare and book chapter royalties from Elsevier, Inc. All disclosures are outside the scope of this project. No other potential conflict of interest relevant to this article was reported.