Abstract
Despite the known influence of anatomic variability on internal dosimetry, dosimetry for 18F-FDG and other diagnostic radiopharmaceuticals is routinely derived using reference phantoms, which embody population-averaged morphometry for a given age and sex. Moreover, phantom format affects dosimetry estimates to a varying extent. Here, we applied newly developed mesh format reference phantoms and a patient-dependent phantom library to assess the impact of height, weight, and body contour variation on dosimetry of 18F-FDG. We compared the mesh reference phantom dosimetry estimates with corresponding estimates from common software to identify differences related to phantom format or software implementation. Our study serves as an example of how more precise patient size–dependent dosimetry methodology could be performed. Methods: Absorbed dose coefficients were computed for the adult mesh reference phantoms and for a derivative patient-dependent phantom series by Monte Carlo simulation using the Particle and Heavy Ion Transport Code System (PHITS) within the software called PARaDIM (PHITS-Based Application for Radionuclide Dosimetry in Meshes). The dose coefficients were compared with reference absorbed dose coefficients obtained from International Commission on Radiological Protection publication 128 or were generated using software including OLINDA 2.1, OLINDA 1.1, and IDAC-Dose 2.1. Results: Differences in dosimetry arising from anatomic variations were shown to be significant, with detriment-weighted dose coefficients for the percentile-specific phantoms varying by up to ±40% relative to the corresponding reference phantom effective dose coefficients, irrespective of phantom format. Similar variations were seen in the individual organ absorbed dose coefficients for the percentile-specific phantoms relative to the reference phantoms. The effective dose coefficient for the mesh reference adult was 0.017 mSv/MBq, which was 5% higher than estimated by a corresponding voxel phantom and 10% lower than estimated by the stylized phantom format. Conclusion: We observed notable variability in 18F-FDG dosimetry across morphometrically different patients, supporting the use of patient-dependent phantoms for more accurate dosimetric estimations relative to standard reference dosimetry. The methodology employed may help in optimizing imaging protocols and research studies, in particular when longer-lived isotopes are used.
The most widely used diagnostic radiopharmaceutical in oncology is 18F-FDG, with 2.2 million clinical PET scans performed across U.S. facilities in 2019 (1). The typical effective dose from 18F-FDG imaging is approximately 10mSv and is considered low (2,3). Nevertheless, the increasing use of radiation-based medical imaging over the past several decades has led to concerns about possible detriment, and these concerns have precipitated campaigns such as Image Wisely (adult patients) and Image Gently (pediatric patients) (4,5). A cornerstone of those campaigns is attention to optimization. Accurate dosimetry is fundamental to understanding relationships between radiation dose and patient detriment and is needed to provide objective recommendations for administered activity for different patient populations (6).
Dosimetry estimates in nuclear medicine are derived using Monte Carlo–based dose calculations incorporating computerized representations of anatomy known as computational phantoms. The accuracy of these estimates depends on the degree to which anatomic features of the patient are modeled within the phantom. Virtually all routine dosimetry estimates for 18F-FDG in adult patients have used standard reference (i.e., non–patient-specific) phantoms to represent patient anatomy. Reference phantoms serve to define an average patient based on only age and sex criteria; they do not consider differing body sizes and organ dimensions, which may influence absorbed dose calculations by 20%–60% (7). The geometric format (e.g., mesh, voxel, stylized, or hybrid) used to define the phantom anatomy further influences dose calculations. Despite the availability of more modern formats, stylized adult phantoms (8,9) as described in International Commission on Radiological Protection (ICRP) publication 23 (10) continue to be routinely used in nuclear medicine dosimetry software and provide the underpinning for dose coefficients promulgated in currently accepted sources of radiopharmaceutical reference data (7). Limitations of stylized phantoms include reliance on simple linear and quadratic surface equations, which coarsely approximate human anatomy and generally underestimate organ cross-dose contributions for low-energy photons. Progress in computational phantom and Monte Carlo code development (11–18) now enables more accurate dosimetry protocols, which will support dosimetry research, optimization of clinical imaging protocols, and planning of research studies.
The ICRP recently developed mesh format reference computational phantoms for the adult male and female, and these phantoms provide anatomically more realistic representations of the human body (Fig. 1) (19). For example, these phantoms define numerous additional source and target regions and define small, thin, or complex radiosensitive target regions impractical to treat using voxel or stylized phantom formats. More recently, the adult mesh reference phantoms were reshaped and scaled to match Caucasian patient population-dependent parameters extracted from anthropometric survey data (20,21)––culminating in an additional 9 male and 9 female phantoms representing the 10th, 50th, and 90th percentiles of standing height and weight for the adult Caucasian population. Termed the percentile-specific phantom series, these 18 phantoms are morphometrically considered patient-dependent phantoms and foster research into dosimetric implications of adult patient size variability across population subsets (22).
Comparison of phantom formats. (Left) MIRD 5 era stylized adult anthropomorphic phantom (1969; ∼25 source/target organs) used in OLINDA 1.0 and ICRP 128. (Middle) ICRP 110 adult male voxel phantom (2009; 79 source regions, 43 target regions) used in IDAC-Dose 2.1. (Right) ICRP mesh reference adult male (2020; 190 source regions, 153 target regions) archetypal example used in this work.
A primary objective of this work was to assess differences in 18F-FDG dose coefficient estimations using a patient-dependent dosimetry paradigm, in comparison to standard one-size-fits-all reference dosimetry. We therefore calculated dosimetry estimates across a series of current-generation mesh patient-dependent phantoms to evaluate the differences arising from anatomic variability. We also investigated differences due to phantom format by comparing dose coefficients derived using the mesh reference phantoms with those obtained using stylized reference phantoms, voxel reference phantoms, or hybrid reference phantoms.
MATERIALS AND METHODS
Biokinetic Data for 18F-FDG
Biokinetic data for 18F-FDG, in the form of time-integrated activity coefficients (TIACs) (23) for the brain, heart wall, lungs, liver, urinary bladder contents, and rest of body, were obtained from the ICRP 128 (7) radiopharmaceutical data compendium (Table 1). The TIAC for the lung region provided in ICRP 128 applies to both lungs jointly; however, in the mesh phantoms, the lungs are modeled as separate regions, with each requiring an input TIAC. Therefore, the lung TIAC was divided between the right and left lung regions by mass fraction (relative to the total lung mass). Similarly, the rest-of-body TIAC was divided among the regions it comprises by tissue mass fraction (relative to the total tissue mass of the rest of the body).
ICRP 128 Reference Organ TIACs Used in Present Study
Absorbed Dose Calculations with Reference- and Patient-Dependent Mesh Phantoms
Absorbed dose coefficients (in units of mGy/MBq administered activity) for the explicitly defined mesh reference and percentile-specific phantom regions were computed by direct Monte Carlo simulation using the Particle and Heavy Ion Transport Code System (PHITS) (c)–based (14) PHITS Application for Radionuclide Dosimetry in Meshes (PARaDIM) software (17). Each simulation was configured using the multisource absorbed dose mode of PARaDIM, which generates both region-level mean absorbed dose coefficients and a 3-dimensional voxelized dose map (here, sampled at 1.0-cm isotropic spatial resolution). The material surrounding the phantom was defined as void. In total, 107 histories were simulated for each phantom. PARaDIM defaults for physical models were used, which specified the PHITS–electron γ-shower 5 method for treatment of multiple scattering, explicit treatment of fluorescent x-rays, consideration of Rayleigh and incoherent scattering, and consideration of electron-impact ionization. Sampling was used for determination of bremsstrahlung polar angles, pair electron polar angles, and distribution of photoelectrons. Cutoff energies of 1.0 keV were used for both positrons and photons.
The ICRP defines several target regions that comprise multiple uniquely defined target regions within the phantoms (e.g., lungs, comprising left and right lung; kidneys, comprising left and right renal pelvis, cortex, and medulla). The mean dose to such regions was computed as a mass-weighted combination of absorbed doses to each subregion:
Eq. 1
where
is the absorbed dose coefficient for a multiregion target mreg;
is the mass of a region
comprising the multiregion target;
is the total mass of the multiregion target; and
is the absorbed dose coefficient for a subregion.
The active marrow and endosteum are implicitly defined within the spongiosa and marrow cavity regions of the mesh reference and percentile-specific phantoms; the absorbed dose coefficients for these specific tissues were derived from those for the corresponding spongiosa and marrow cavities per the recommendations in ICRP 116, namely:
Eq. 2
Eq. 3
where
and
are the skeletal-averaged absorbed dose coefficients for the active marrow and endosteal cells, respectively;
and
are the masses of these tissues in skeletal target region
;
and
are the total masses of active marrow and endosteal cells contained within the entire skeleton; and
and
are the absorbed dose coefficients for the spongiosa and medullary marrow regions that encompass the active marrow and endosteal cells in skeletal region
.
The effective dose quantity is derived from absorbed dose calculations and is a widely used concept in diagnostic radiation dosimetry; it provides a strategy for combining the variable organ doses into a single stochastic risk-relevant number (24–26). Effective dose coefficients (27) were computed using the methodology and tissue-specific weighting factors promulgated in ICRP 103 and ICRP 133, namely:
Eq. 4
where e is the effective dose coefficient,
is the tissue-specific weighting factor for tissue
T, and
is the radiation weighting factor for radiation type R (here, taken as unity for 18F positrons and photons).
By definition, the effective dose is relevant only for the reference person (i.e., it does not formally apply to height, weight, or otherwise constrained subsets of the patient population) (23,24). It is also restricted to the use of sex-averaged organ absorbed doses as per Equation 4. To gain insight into a similar single-value risk characterization, a related quantity—the detriment-weighted dose (28–30)—has been previously introduced; this quantity applies ICRP tissue and radiation weighting factors but removes the requirement of sex-averaging of organ absorbed doses, as well as the restriction of nonreference phantom dose values. The detriment-weighted dose provides a basis for consideration of risk from stochastic effects of radiation, as relevant to population subsets (as used here) or potentially individuals, and is used with the justification that absorbed dose, not patient geometry, is the relevant seat for quantifying risk:
Eq. 5
where
is the detriment-weighted dose coefficient for a phantom of sex S, height
H, and total-body mass M.
Absorbed Dose Calculations with Reference Voxel, Stylized, and Hybrid Phantoms
The ICRP 89 adult male and female hybrid reference phantoms (18) were used in OLINDA 2.1. The adult male and female stylized reference phantoms were used in OLINDA 1.1, and the ICRP 110 adult voxel reference phantoms were used in IDAC-Dose 2.1 software. The default organ masses were used for each phantom in all software. Further details on the phantoms used are provided in Table 2.
Characteristics of Phantoms as Implemented in Software Used in Present Study
Hardware and Software Specification
An HP Z8 workstation running the Microsoft Windows 10 operating system and using a 3.6-GHz Intel Xeon 5122 processor was used for all Monte Carlo calculations.
3D Slicer (version 4.11; www.slicer.org) was used for dose map visualization, and Paraview (version 5.6.2; www.paraview.org) was used for 3-dimensional rendering of phantom geometry. Statistical analysis was performed with GraphPad Prism (version 8.3.1; GraphPad Software, Inc.).
RESULTS
Figure 2 shows voxel-level absorbed dose coefficients for the mesh reference and percentile-specific phantom series. These 3-dimensional dose maps are volume-rendered as maximum-intensity projections, and as they were derived under the assumption of uniformly distributed activity within whole organs, they are intended to reflect spatial nonuniformity of dose deposition averaged across a patient subpopulation matching the specified height and weight. Region-level mean absorbed dose coefficients for selected target organs of the phantoms are presented in Figures 3 and 4. A comprehensive list of dose coefficients for all defined target regions is provided in Supplemental Tables 1–8 (supplemental materials are available at http://jnm.snmjournals.org). The highest absorbed dose coefficients were observed in the heart wall, followed by the urinary bladder wall and brain, for the male and female mesh phantoms and their voxel analogs implemented in IDAC-Dose 2.1 (Fig. 5). In contrast, the urinary bladder wall was identified as the critical organ in the stylized and hybrid reference phantom dose estimates, with the bladder wall absorbed dose coefficients being approximately 2-fold higher than those for the heart wall (the next most irradiated tissue) and were approximately 2.5-fold higher than estimated with the mesh reference phantoms. This discrepancy has been noted previously (17), and here, was the largest relative deviation in the dose coefficient estimated by the mesh reference phantoms in comparison to the corresponding stylized and hybrid phantom estimates. The disagreement can be partially explained by the methodology applied for determination of specific absorbed fractions used in derivation of self-irradiation S values (and correspondingly, absorbed dose coefficients) for the urinary bladder of the stylized phantoms (31,32). Absorbed dose coefficients for the basal cell layer of the urinary bladder of the mesh reference adults (0.089 mGy/MBq and 0.091 mGy/MBq for adult male and female, respectively; Supplemental Tables 1 and 2) were more consistent with the stylized and hybrid phantom bladder dose estimates.
Voxel-level absorbed dose coefficients for 18F-FDG. (A) Adult female reference and percentile-specific phantom series juxtaposed with maximum-intensity projections of their corresponding 18F-FDG voxel dose coefficient maps. (B) Same as A but for adult male. Hx = xth percentile standing height; Wy = yth weight percentile within preceding standing height percentile.
(A) Organ-level absorbed dose coefficients for 18F-FDG for female mesh reference computational phantom (MRCP) and comparison with percentile-specific phantom series. (B) Percentile specific dose coefficients presented as fractional differences relative to reference female. Hx = xth percentile standing height; Wy = yth weight percentile within preceding standing height percentile.
(A) Organ-level absorbed dose coefficients for 18F-FDG for male mesh reference computational phantom and comparison with percentile-specific phantom series. (B) Percentile-specific dose coefficients presented as fractional differences relative to reference male. Hx = xth percentile standing height; MRCP = mesh reference computational phantom; Wy = yth weight percentile within preceding standing height percentile.
Comparison of organ-level dosimetry computed with reference phantoms in modern and legacy dosimetry software. Note overall improved agreement between mesh reference adults and output of IDAC-Dose 2.1, where anatomically equivalent phantoms were represented by different formats (i.e., mesh vs. voxel). MRCP = mesh reference computational phantom.
Relative to the reference mesh phantoms of the corresponding sex, absorbed dose coefficients for most target organs of the percentile-specific phantom series deviated by 20%–30% for the 10th and 90th weight percentile extremes. The reference and percentile-specific estimates converged near the 50th standing-height percentile/50th weight percentile within the 50th standing-height percentile; this finding was expected because of the similar body morphometry of the reference phantoms (176.0 cm/73 kg male; 163 cm/60.0 kg female) and the 50th standing-height percentile/50th weight percentile within the 50th standing-height percentile analogs (176.5 cm/79.3 kg male; 163.3 cm/64.1 kg female). Of all target tissues, absorbed dose coefficients for the brain varied the least (in relative terms) among phantoms of the same sex, primarily because of the weak dependence of brain mass on body morphometry—particularly, body mass for a given height.
The effective dose coefficient was calculated as 0.017 mSv/MBq using the male and female mesh reference phantoms in combination with the tissue-specific weighting factors of ICRP 103. This represents an approximately 10% decrease relative to the ICRP 128 estimate of 0.019 mSv/MBq and reflects geometric differences in the phantoms, tissue-specific weighting factors (Table 2), and lack of sex-averaging of the ICRP 128 organ doses used to derive the effective dose. The effective dose coefficient for the ICRP 110 voxel phantoms was 0.016 mSv/MBq (sex-averaged), which agrees within approximately 5% with the mesh reference phantom estimates; of note, these phantoms were designed to represent identical geometry but in different formats.
The detriment-weighted dose coefficient for each of the percentile-specific phantoms was computed by applying ICRP 103 tissue-specific weighting factors directly to the associated tissue equivalent dose coefficients (i.e., not using reference male/female equivalent dose averaging; Eq. 5). The patient dependence of the detriment-weighted dose is provided in Figure 6 for the percentile-specific phantoms. Though the relationships depend on height, weight, sex, and underlying physical factors (linear energy transfer, attenuation), the detriment-weighted dose negatively correlates strongly with body mass (Spearman r = −0.9505; P < 0.0001) but less strongly with BMI (Spearman r = −0.7049; P = 0.0011). Notably, given a particular sex, the detriment-weighted dose-versus-mass curves for different height classifications were observed to decrease monotonically. For a given mass, detriment-weighted dose coefficients for females were larger than for males. Together, these results suggest that consideration of body morphometry, rather than only body mass, will further improve dose estimates and optimization strategies.
Dependence on body mass (A), and dependence on body mass index (BMI) (B), of detriment-weighted dose coefficients (eDW) for male and female percentile-specific phantom series. Hx = xth percentile standing height; Wy = yth weight percentile within preceding standing height percentile.
DISCUSSION
Diagnostic dose measurement has roles in optimization, comparison of techniques and modalities, quality assurance, research and development, and support of regulatory standards and public health protection (24). Understanding the variability and uncertainty in dose estimates is of heightened importance considering repeat oncologic imaging sessions in which uncertainties must be propagated across multiple imaging sessions throughout disease management. The contributing factors for uncertainty in organ-level absorbed dose calculations (33–35) can include uncertainty in radionuclide physical decay data, system calibration, Monte Carlo calculation of energy deposition, body morphometry of the patient in comparison to the phantom representing the patient, and radiopharmaceutical biokinetics and intraorgan distribution. Of these, the uncertainty in radionuclide decay data is negligible in most cases, and uncertainty in Monte Carlo simulations can be minimized with adequate sampling. In contrast, body morphometry and radiopharmaceutical biokinetics and intraorgan distribution represent the principal organ-level dosimetric uncertainties.
Our study addresses uncertainty and variability in body morphometry for specific subsets of the Caucasian patient population, and we provide patient-dependent dose estimates (with a broad approximation of the anatomy of height- and weight-defined subsets of the patient population), in contrast to patient-specific calculations (which account for the unique anatomy of a specific patient) (22). In doing so, we have used reference 18F-FDG biokinetic data in order to ensure that the variability in the results originates plainly from the underlying anatomic differences. The biokinetics of 18F-FDG may vary across patients, such as because of dietary conditions, presence of tumors, and kidney dysfunction. To our knowledge, the biokinetic dependence on adult body morphometry has not been comprehensively investigated. Considering the uncertainty in biokinetics, as well as the impracticality of creating patient-specific phantoms for routine imaging procedures, there is some rationale for applying these dose estimates to specific patients by matching them to the appropriate height and weight percentiles. Such strategies could support scan-specific dose reporting in nuclear medicine in the near future.
Role of Patient-Dependent Dose Coefficients in Diagnostic Nuclear Medicine
Optimization strategies for 18F-FDG should continue to prioritize the delivery of images of diagnostic quality. The Image Wisely and Image Gently initiatives (4,5) espouse safe and efficacious imaging by striving to deliver diagnostic images while maintaining radiation doses as low as reasonably achievable. Our calculations provide supporting data for patient size–based 18F-FDG administration and related optimization strategies. At 18F-FDG dose levels expected clinically, the CIs in risk–dose relationships are currently too broad to justify the consideration of risk in management of individual patients (36,37). Therefore, the dose coefficients we have provided should be used primarily for improved scan-specific dose reporting, which will enable better elucidation of dose–risk relationships in epidemiologic studies. Incorporating patient size–dependent variation in 18F-FDG biokinetics would allow further refinement of the present dose coefficient estimates and would allow them to be more confidently used in setting guidelines for administered activity as a function of adult body morphometry. A greater impact of morphometry-dependent variation in absorbed dose and detriment-weighted dose coefficients can be expected with longer-half-life diagnostic radiopharmaceuticals.
Methodologic Considerations
In contrast to the orthodox method of computing dose coefficients from specific absorbed fraction–derived S-value lookup tables, here we have computed task-specific dose coefficients via direct Monte Carlo simulation. The derivation of a complete set of phantom organ–specific absorbed fractions requires a monumental computational effort. For example, ICRP 133 provides specific absorbed fractions for the ICRP 110 series of adult phantoms in 79 unique source regions across 3 corpuscular radiation types plus photons, and approximately 20 discrete energies are necessary for interpolation—effectively requiring over 6,000 individual Monte Carlo simulations per phantom—but thereafter enabling dose from any radionuclide or biokinetic dataset to be immediately computed. For the direct approach used here, a single phantom requires a new Monte Carlo simulation for every biokinetic dataset, but the simulation can be run on a standard desktop computer on a clinically acceptable timescale. Thus, future availability of specific absorbed fractions for the mesh reference–specific and percentile-specific phantoms will increase flexibility and computation speed, but their current unavailability should not present a barrier to implementation of these phantoms.
Some organ-level dosimetry software, including OLINDA, allows users to modify reference phantom organ masses to approximate patient-specific dose calculations. These approaches typically involve linearly scaling the self-S contributions from weakly penetrating radiations (α-particles and electrons) with inverse mass, whereas photon self-S contributions are scaled inversely with the two-thirds power of the mass (32). This approach assumes that source–target proximity is invariant on changes in organ mass or patient shape. This approach is reasonable when the phantom already recapitulates the salient morphometry of the patient or patient population subset (e.g., for small mass variation in larger spheric organs) but may be subject to considerable errors when, for example, scaling reference phantom S values to match patients of much smaller or much larger overall size. As the percentile-specific phantoms account for gross size variation across patients, they would serve as a better basis for patient-specific organ S-value scaling.
CONCLUSION
Using newly developed computational phantoms, we provide patient-dependent dose coefficients and revised reference dose coefficients for 18F-FDG. Differences in patient morphometry imparted large deviations in dose coefficients for the upper and lower 10th percentile extremes of patient height and weight (up to ±40% relative to the reference patient). Although our work was limited to 18F-FDG, greater impact is expected for longer-lived isotopes in imaging. These data pertain to stochastic risk estimates and should not be used for individual patient management, but the methodology described may support optimization of administered activities in clinical practice.
DISCLOSURE
This research was funded in part through the NIH/NCI Cancer Center support grant P30 CA008748 and by NIH U01 EB028234. Lukas Carter acknowledges support from the Ruth L. Kirschstein NRSA Postdoctoral Fellowship (NIH F32 EB025050). Simone Krebs was supported in part by the NIH/NCI Paul Calabresi Career Development Award for Clinical Oncology (K12 CA184746). No other potential conflict of interest relevant to this article was reported.
ACKNOWLEDGMENTS
We are enormously grateful for technical support from the PHITS team of the Japanese Atomic Energy Administration, including Drs. Tatsuhiko Sato and Takuya Furuta.
KEY POINTS
QUESTION: Can the influence of patient size on 18F-FDG dosimetry be elucidated using newly developed patient-dependent phantoms?
PERTINENT FINDINGS: Dose coefficients for 18F-FDG were shown to vary by up to ±40% across the 10th–90th percentiles for standing height and weight across the Caucasian adult patient population.
IMPLICATIONS FOR PATIENT CARE: Implementation of the current state-of-the-art phantom libraries into modernized dosimetry software will foster greater accuracy in dosimetry and ultimately support optimized strategies for personalized nuclear medicine.