Abstract
There has been an evolutionary leap in SPECT imaging with the advent of camera systems that use solid-state crystals and novel collimator designs configured specifically for cardiac imaging. Solid-state SPECT camera systems have facilitated dramatic reductions in both imaging time and radiation dose while maintaining high diagnostic accuracy. These advances are related to simultaneous improvement in photon sensitivity due to the collimator and imaging geometry, as well as image resolution due to the improved energy resolution of the new crystals. Improved photon sensitivity has facilitated fast or low-dose myocardial perfusion imaging (MPI), and early dynamic imaging has emerged as a technique for assessing myocardial blood flow with SPECT. Lastly, general-purpose solid-state camera systems and hybrid SPECT/CT systems have also been developed that may have important clinical roles in cardiac imaging. This review summarizes state-of-the-art solid-state SPECT MPI technology and clinical applications, including emerging techniques for SPECT MPI flow estimation. We also discuss imaging protocols used with the new cameras, potential imaging pitfalls, and the latest data providing large-scale validation of the diagnostic and prognostic value of this new technology.
Myocardial perfusion imaging (MPI) with SPECT is widely used for the diagnosis and management of patients with known or suspected coronary artery disease, providing crucial information regarding both myocardial perfusion and function. Since being introduced in the 1970s, SPECT camera system designs have undergone several incremental improvements. More recently, however, there has been an evolutionary leap with dedicated cardiac SPECT camera systems that use solid-state crystals and novel collimator designs, such as multipinhole and locally focusing collimators configured specifically for cardiac imaging (1). These innovations have facilitated dramatic reductions in both imaging time and radiation dose while maintaining high diagnostic accuracy. These features are related to simultaneous improvement in sensitivity (5–8 times higher (2,3) due to the collimator and imaging geometry) and image resolution (up to 2 times higher (3) due to the improved energy resolution of the new crystals). Additionally, the physical space requirements for solid-state camera systems have been significantly reduced because photomultiplier tubes are no longer required. New solid-state scanners can also be coupled with diagnostic-quality CT systems suitable for calcium scanning or coronary CT angiography. With the growing clinical experience using solid-state SPECT camera systems, there is also emerging evidence demonstrating comparable or superior diagnostic and prognostic utility.
NOTEWORTHY
This review summarizes the technologic advancements in commercially available solid-state camera systems and a few systems still in development.
Emerging data regarding MBF measurements with solid-state SPECT camera systems are summarized.
Evidence providing large-scale validation of the diagnostic and prognostic utility of solid-state SPECT camera systems for MPI is outlined.
This review summarizes current solid-state SPECT technology and clinical applications, including emerging techniques for myocardial blood flow (MBF) quantitation. We will also discuss imaging protocols, potential imaging pitfalls, and the validation efforts for this new technology regarding diagnosis and prognosis.
Several technologic advancements are responsible for the improved imaging characteristics of solid-state SPECT camera systems. No single change can be attributed to the significant improvements in both photon sensitivity and spatial resolution realized with these systems. We review these advances within the context of the overall camera system.
SOLID-STATE DETECTORS
The radiation detectors are made of cadmium-zinc-telluride (CZT) crystals, which are more expensive than NaI(Tl) crystal detectors but have important imaging advantages. CZT detectors have an improved energy response, which reduces the scatter component of measured data by 30% (4), and superior intrinsic spatial resolution compared with conventional Anger cameras. Removing the need for photomultiplier tubes has resulted in a more compact detector size, which has facilitated more efficient camera designs with innovative imaging geometries and high-sensitivity collimation. Solid-state detectors can also be combined with scintillation crystals to indirectly detect photons. In these detectors, thallium-doped cesium iodide CsI(Tl) scintillation crystals are coupled with photodiodes (instead of photomultipliers) (5). CsI(Tl) has superior spectral light response characteristics compared with NaI(Tl) when used with silicon photodiodes.
Solid-state detectors are integrated with innovative collimator designs that increase photon sensitivity in the myocardial region. The high-sensitivity characteristics of solid-state SPECT cameras are often mistakenly attributed to the CZT crystal. Although the density of CZT is higher than NaI(Tl) (5.8 vs. 3.7 g/cm3), because of cost considerations CZT detectors are thinner (∼5 mm) than typical NaI(Tl) crystals (9.5 mm), such that the intrinsic detection efficiency of these systems is comparable to conventional Anger detectors.
DEDICATED CARDIAC DETECTORS
CZT SPECT camera systems specifically designed for cardiac use are commercially available from 2 vendors. One system (D-SPECT; Spectrum Dynamics) uses pixelated CZT detector arrays. Nine vertical columns are arranged in a 90° gantry geometry. Each column consists of 1,024 CZT elements (2.46 × 2.46 × 5 mm thick), arranged in a 16 × 64 element array with an overall size of 40 × 160 mm. Since CZT crystals are expensive, the vendor offers a system with 6 detector columns, creating a trade-off between cost and image quality. The other camera design (Discovery NM 530c; GE Healthcare) features a curved array of 19 CZT pixelated detector arrays coupled to pinhole collimators with geometry optimized for cardiac imaging. Each detector array consists of 4 detectors, each with 246 CZT detector elements (2.46 × 2.46 × 5 mm thick) arranged in 16 × 16 elements.
The indirect solid-state detector has been used in traditional 2- or 3-detector configurations (Digirad), allowing for a more compact design than conventional Anger cameras. Each detector head contains an array of 768 (6.1 × 6.1 × 6 mm thick) CsI(Tl) crystals coupled to individual silicon photodiodes that convert the crystal scintillations to electrical pulses, which are then processed digitally.
DEDICATED CARDIAC COLLIMATORS AND GEOMETRIES
The improved sensitivity of solid-state SPECT cameras is achieved by novel collimator designs and dedicated cardiac imaging configurations. Increasing photon sensitivity is important since a conventional Anger camera detects less than 0.01% of incoming photons (6). Diagrams of these novel collimator technologies, for cardiac solid-state cameras and a general-purpose solid-state camera, are shown in Figure 1.
The D-SPECT features 1 small, high-sensitivity tungsten collimator for each detector column (as described above). These collimators can be positioned close to the patient’s body, which increases the number of photons detected and improves spatial resolution. The collimators have square, parallel holes with dimensions of 2.26 × 2.26 mm and wall thickness of 0.2 mm matched to the dimensions of a single detector element. The hole dimensions are larger than for conventional high-resolution collimators, resulting in higher sensitivity at the expense of image resolution. However, the loss in spatial resolution resulting from this design is more than compensated for by the use of the small, pixelated CZT detectors and software-based resolution recovery applied during image reconstruction (2). During acquisition, each of the 9 detectors is focused on the heart and rotates over 45°, with a rotation time of 3 s. Patients are positioned on a chair rather than on a bed, increasing patient comfort. Patients are typically imaged in both semisupine and upright positions.
The Discovery NM 530/570c camera systems use pinhole collimation. The design includes 19 pinhole collimators with multiple focal depths and 1 hole (4.75 mm in diameter) in each collimator. The CZT detectors and collimators do not move, and photons are acquired simultaneously through all the pinholes. Simultaneously acquired pinhole views allow consistent tomographic sampling, reducing motion artifacts. Patients are typically imaged in both supine and prone positions, with their arms placed above their head. Higher sensitivity is achieved by the wider acceptance angle of photons with multipinhole collimation and the cardiocentric acquisition, which limits photon detection to the myocardial region.
SOLID-STATE SPECT/CT SYSTEMS
MPI can benefit from CT attenuation correction (AC) using either dedicated SPECT/CT systems or separate acquisitions and provides a new opportunity to enhance MPI with additional CT information; for example, by obtaining the coronary calcium score. Initially, only multipinhole solid-state camera systems (GE Healthcare 570c) were offered with a hybrid diagnostic CT configuration; however, there are only a few such systems operating worldwide.
In the last 2 years, general-purpose solid-state SPECT cameras coupled with diagnostic-quality CT scanners—capable of coronary calcium scoring—have been introduced by 2 vendors. Although these systems are designed for general imaging use, they can also be used effectively for MPI. Spectrum Dynamics introduced the Veriton-CT system with 8 CZT modules within each of 12 detector blocks, each with individual detector arms with swivel motion, which allow precise and close 3-dimensional contouring around the patient during imaging. The field-of-view length is 320 mm and can be focused for single-organ acquisitions or used for whole-torso scanning. The optimized collimator design results in higher volumetric sensitivity (0.0192%) than dual-head camera systems with conventional collimators (<0.01%). Additionally, with image resolution recovery, this system achieves high spatial resolution (4.3 mm). Initial clinical studies of the prototype design with a detector design similar to Veriton demonstrate potential improvements in image quality compared with a conventional camera (7). The Veriton-CT system is configured with up to a 128-slice CT camera, allowing both calcium scoring and CT angiography studies. GE Healthcare introduced a general-purpose solid-state camera, the Discovery NM/CT 870 CZT, equipped with conventional collimators. This camera system offers similar photon sensitivity to the conventional Anger system (as the collimation system is the same) but allows better energy resolution and image resolution due to the CZT detectors and has demonstrated improved image quality in early clinical implementation (7,8).
Hybrid SPECT/CT solid-state systems are not yet in widespread clinical use. However, separate noncontrast CT coronary artery calcium scans are increasingly combined with SPECT-MPI studies. It has been shown that separately acquired coronary artery calcium scans can be used for AC of SPECT MPI studies (9). The use of separately obtained CT for AC is supported by vendors for routine clinical use (10,11).
RECONSTRUCTION INCLUDING RESOLUTION RECOVERY AND ANATOMIC CONSTRAINTS
In parallel with improvements in technical aspects of SPECT scanner hardware, enhanced image reconstruction techniques have been implemented. New reconstruction algorithms include modeling of the collimators and detectors, as well as resolution recovery. Most systems compensate for spatial resolution losses using iterative reconstruction. The aggressive resolution recovery is needed to compensate for the effects of high-sensitivity collimation, which reduces spatial resolution by accepting photons from a wider angle. Therefore, software modeling the collimator response during reconstruction is a critical component needed to achieve the simultaneous increase in spatial resolution and count sensitivity with the solid-state SPECT camera systems.
Further improvements in spatial resolution can be achieved using iterative reconstruction with myocardial anatomic priors, which constrain the shape and thickness of myocardial walls to fit within an expected range. This approach can further improve image quality, especially for noisy images. However, if the actual patient anatomy differs significantly from the expected pattern, the convergence of the algorithm may be falsely biased by the assumed constraints. CT scan data can further enhance image reconstruction approaches. With the hybrid SPECT/CT camera systems, the CT can be used during the reconstruction not only for AC but also as a prior to improve the anatomic constraints.
PERFORMANCE
The clinical solid-state SPECT systems offer improved sensitivity and spatial resolution compared with conventional Anger cameras. The actual technical performance of solid-state SPECT cameras was compared with conventional dual-head scanners by Imbert at al (3). Additionally, similar results are available from one other vendor. Comparison of count sensitivity (expressed in %) with PET are shown in Figure 2. The solid-state SPECT cameras offer up to 8 times higher count sensitivity and more than double the effective image resolution of the conventional Anger cameras. Imaging times with conventional doses of radioactivity are in the range of 2–3 min for systems with dedicated solid-state detectors (12,13), compared with 15 min on systems with a standard collimator. The count sensitivities for dedicated cardiac CZT systems can approach those of 2-dimensional PET systems, which to date have been used in most of the PET-MPI literature.
CURRENT CLINICAL USE
It is estimated that over 820 solid-state dedicated cardiac CZT SPECT camera systems are currently in clinical use (Yoel Zilbertein and Gil Kovalski, oral communication, March 1, 2019). The improved technical performance of solid-state SPECT camera systems was initially used to dramatically reduce imaging time, improving patient comfort and minimizing motion artifacts. However, clinical protocols with these systems can also use very low-dose radiopharmaceutical injections (148–222 MBq [4–6 mCi] of the 99mTc perfusion agents) (14–16) with standard imaging times. When using stress-only scans with 10- to 12-min acquisitions, effective radiation doses under 1 mSv are achieved (17,18). By comparison, the average annual background radiation exposure in the United States is approximately 3.1 mSv (19).
2-POSITION IMAGING: UPRIGHT/SUPINE OR SUPINE/PRONE
Most solid-state SPECT cameras in clinical use are not in hybrid SPECT/CT configurations and therefore do not offer CT-based AC. At the same time, very low-dose protocols with reduced image counts challenge the accuracy of visual interpretation, especially in the presence of attenuation artifacts. To address this issue, most clinical protocols use 2 sequential scans performed in different patient positions (supine/upright or supine/prone). Since soft tissue shifts in different patient positions, comparing results from 2 positions can allow differentiation of true perfusion defects from artifacts (20–22). Two-position protocols add only modestly to the total time of single-view protocols with high-sensitivity scanners, with no added radiation exposure.
LOW-DOSE PROTOCOLS
The technical advancements of solid-state SPECT cameras can be used to dramatically reduce radiation exposure during a patient’s scan. The list-mode capabilities of SPECT equipment can be used to evaluate retrospectively the efficacy of low-dose imaging protocols. One such study simulated reduced radiation protocols in 79 patients, who were imaged with a solid-state scanner for 14 min with 802.9 ± 199.8 MBq (21.7 ± 5.4 mCi) of 99mTc injected at stress (18). There was no significant difference in quantitative perfusion or functional measurements even with simulated activity corresponding to an effective radiation dose of less than 1 mSv, as outlined in Figure 3. The clinical implementation of low-dose protocols was studied in patients who underwent repeat rest scans with split radiation doses. The image quality for very low-dose rest scans (mean effective dose, 1.15 mSv) was superior to conventional SPECT imaging (mean effective dose, 2.39 mSv) (23). Einstein et al. reported stress-only imaging in 69 patients with a mean effective radiation dose of 0.99 mSv (24). In a study of 284 patients undergoing SPECT MPI, Sharir et al. demonstrated similar diagnostic accuracy with low-dose imaging compared with standard dose imaging, but with 50% radiation reduction (25). This was confirmed by another group who used 160-MBq injected doses for rest and stress acquisitions of 15 min (26). Palyo et al. showed that regional perfusion defect size and functional parameters were similar despite 50% radiation reductions using a hybrid SPECT/CT system with CT-based AC (27). With more aggressive radiation reduction (75%), there was a tendency to overestimate perfusion defect size in patients with abnormal results (27). Importantly, to preserve reproducibility between high-count and low-count data, the quantification of low-dose protocols requires normal limits specifically calibrated with low-dose images because of higher variability of low-count data (28).
As noted above, the lowest total radiation exposure with current SPECT MPI (∼1 mSv) can be accomplished clinically by performing stress-only imaging with a solid-state camera system. Stress-only MPI can reduce radiation exposure by up to 60% (29) and also provides a significant reduction in radiation exposure to staff performing the scan (30). Additionally, stress-only protocols dramatically shorten total study times to as low as 30–45 min, including the stress test, improving the patient experience and laboratory efficiency. Since most SPECT MPI scans are normal (31), an ultralow-dose (1 mSv) stress-only scan could potentially be performed in most patients undergoing SPECT MPI with a solid-state camera system. Further research is needed to demonstrate the diagnostic accuracy and prognostic value of this ultralow-dose, stress-only approach.
SIMULTANEOUS DUAL-ISOTOPE MPI
Solid-state cameras with CZT detectors have superior energy resolution, which allows better separation of photons from different isotopes and improved cross-talk correction (32). In addition, stationary detector systems eliminate projection inconsistencies that confound dual-isotope imaging. One possible clinical application is a combined rest/stress protocol with simultaneous imaging of 99mTc and 201Tl as proposed for conventional cameras (33), which was not implemented clinically on standard cameras because of technical difficulties and concerns regarding higher radiation with 201Tl.
To date, 3 studies have demonstrated the feasibility of simultaneous imaging of 201Tl (stress) and 99mTc (rest) SPECT MPI with solid-state cameras. One study proposed a single stress/rest 20-min acquisition for the 2 isotopes (34). A separate study demonstrated the feasibility of simultaneous dual-isotope imaging in 24 patients with 80 MBq of 201Tl injected at rest and 250 MBq of 99mTc sestamibi injected during adenosine infusion (35). For this application, dedicated scatter correction methods are required because triple-energy window corrections overestimate scatter and cross-talk in CZT-based systems (36). These dedicated corrections have been shown to improve defect contrast and myocardium–to–blood-pool ratios compared with triple-energy window corrections (36). Diagnostic performance and image quality of simultaneous dual-isotope MPI with solid-state systems have been demonstrated to be comparable to conventional SPECT with separate stress/rest acquisitions (32).
Although use of dual-isotope MPI with 99mTc and 201Tl is now discouraged, simultaneous dual-isotope SPECT may become increasingly used for other imaging protocols. As an example, simultaneous dual-isotope imaging for cardiac sarcoidosis with 99mTc-pyrophosphate and 201Tl has recently been described, with 201Tl allowing precise definition of myocardial boundaries (37). This dual-isotope approach has importance for dedicated cardiac systems in which the tight gantry and imaging focused on the heart make it is harder to correctly position the 99mTc-pyrophosphate acquisitions. Notably, CT attenuation imaging could play a similar role in hybrid SPECT/CT systems.
Simultaneous dual-isotope imaging can also be used in combined SPECT perfusion/metabolism protocols. Many novel metabolism agents, either in clinical or research use, only show abnormal portions of the myocardium, which are better understood within the context of perfusion imaging. As an example, 123I-betamethyl-p-iodophenyl-pentadecanoic acid has diagnostic and prognostic potential after acute myocardial infarction or in patients with Takotsubo cardiomyopathy (38). Solid-state cameras have been shown to simultaneously acquire 99mTc-perfusion-agent and 123I-betamethyl-p-iodophenyl-pentadecanoic acid images with image quality similar to that achieved with separate acquisitions (39). Simultaneous metaiodobenzylguanidine and 99mTc (40), or 123I-metaiodobenzylguanidine and 201Tl imaging can also be acquired (41). In phantom studies, 123I and 99mTc images can be acquired simultaneously with image quality similar to separate acquisitions with 2 commercially available solid-state cameras (42). Simultaneous dual-isotope acquisitions, which are facilitated by the high-energy resolution of CZT crystals, reduce imaging time, ensure precise image coregistration, and reduce artifacts related to patient or camera motion between acquisitions.
NORMAL PERFUSION LIMITS FOR SOLID-STATE CAMERAS
Solid-state SPECT scanners have significantly different regional count distributions than standard Anger cameras—primarily due to different types of collimation and detector geometry. Early studies of CZT camera systems demonstrated these differences, and dedicated normal limits were subsequently recommended (13), with dedicated normal databases created for the Discovery NM 530c (GE Healthcare) (22,43) and D-SPECT (13,20) camera systems. Importantly—for optimal quantification—separate databases are required for prone and supine acquisitions.
COMBINED QUANTIFICATION FROM 2 POSITIONS
Physicians use multiple imaging positions to visually differentiate attenuation artifact from perfusion deficits; however, this approach has only more recently been applied to quantitative analysis. Combined quantitative analysis of upright and supine acquisitions on the solid-state D-SPECT scanner has been evaluated (20). Quantification performed jointly from coregistered images obtained in 2 positions offered better identification of attenuation artifacts and reduced false-positive defects (Fig. 4). Compared with a reference of invasive coronary angiography, combined quantitative analysis showed higher diagnostic accuracy than separate upright or supine acquisitions. Similar results have been shown with 2-position imaging in obese patients using the D-SPECT system (21) and for the Discovery NM530c scanner with supine and prone acquisitions (22). Two-position imaging may also detect position-related truncation artifacts, which can occur because of the limited field of view on solid-state SPECT cameras (44). New approaches applying deep learning with combined 2-position imaging on solid-state cameras have demonstrated further improvements (45).
MOTION CORRECTION ON SOLID-STATE CAMERAS
Respiratory and patient motion can degrade SPECT image quality, with several recent studies investigating techniques to mitigate these effects. Because of improvements in spatial resolution, heart motion has become the dominant motion-related degrading factor in solid-state SPECT imaging. The feasibility of using list-mode data, with registration of consecutive short-time image frames, to detect and correct for respiratory motion has been demonstrated on these new systems (46). Respiratory gating can impact functional cardiac parameters, resulting in reduced volume estimates and differences in regional left ventricular wall motion (47). End-expiratory gating can be used to improve contrast differences between the myocardium and blood pool (48). Respiratory-phase–matched CT attenuation maps may further improve image quality and have been shown to significantly improve interobserver agreement compared with uncorrected images (49). Patient motion can also be corrected using a similar approach to reduce the rate of false-positive perfusion defects (50). Dual respiratory/cardiac gated MPI has been demonstrated in phantoms and patients, showing progressive improvement of the myocardium–to–blood-pool contrast with the application of respiratory and then dual gating (48).
POTENTIAL PITFALLS
Although the novel designs of dedicated cardiac solid-state SPECT camera systems offer important clinical benefits, there are potential imaging pitfalls that need to be considered. The camera geometry is optimized for cardiac acquisitions, so there is nonuniform count sensitivity in the field of view (51). Therefore, patient positioning is critical and patients with significantly abnormal cardiac anatomy may not be well imaged. Importantly, the typical attenuation defect with supine imaging is more lateral with solid-state camera designs than with conventional cameras (52). For example, because of increased space between the chest wall and cardiac border, obese patients may require repositioning or multiple-position imaging to ensure diagnostic quality (44). In patients with hypertrophic cardiomyopathy, iterative reconstruction algorithms using a heart-shaped model significantly underestimate wall thickness compared with ordered-subsets expectation maximization, which does not use a shape constraint (53). Two-position protocols may allow for detection of position-related artifacts and potentially address some of these concerns (21).
SPECT MBF
Absolute MBF and myocardial flow reserve (MFR) provide clinically meaningful data beyond that obtained by assessment of relative perfusion deficit alone. Absolute measurements can identify patients with multivessel coronary artery disease and predict the extent of disease more accurately than relative perfusion quantification or visual interpretation (54). MFR is also a useful prognostic marker, significantly improving risk stratification beyond relative perfusion defects (55,56). Although myocardial flow measurements were initially developed and validated in PET imaging, SPECT cameras may be capable of obtaining similar measurements. The improved count sensitivity related to collimator and camera geometry advances with solid-state SPECT cameras may make them particularly well suited for flow measurements.
Human dynamic SPECT flow studies have been demonstrated on high-efficiency dedicated SPECT, with factor analysis and a 2-compartment model used for quantification (57). The clinical feasibility was demonstrated in 95 patients, in whom the data were reconstructed to 3-s frames resulting in 60–70 frames for each stress and rest study. The authors demonstrated that MFR derived from such studies was higher in patients with normal scans (as established by standard relative quantification) and that lower MFR (both regional and global) was associated with a static-imaging stress perfusion defect, age, and smoking. The MFR also showed stepwise reduction related to the severity of the obstructive disease in a subset of 20 patients on whom invasive coronary angiography was performed. To date, at least 7 studies have been performed on humans demonstrating the SPECT MBF measurements with 99mTc and solid-state cameras (57–63). Importantly, Wells et al. have demonstrated close agreement between SPECT MBF and MFR measurements, with or without AC, compared with PET measurements (Fig. 5) (59). This finding has important implications, since most solid-state scanners are currently operating without AC capabilities. The clinical protocol for SPECT flow has gained significant attention recently, as the Food and Drug Administration approved dedicated clinical software with both solid-state scanners, with several sites commencing clinical SPECT flow imaging. Examples of MPS flow results are shown in Figure 6.
Notwithstanding this early enthusiasm about SPECT MPI flow, there are well-known limitations of this use with 99mTc-sestamibi or tetrofosmin. Notably, the myocardial extraction of these radiotracers from the blood is substantially reduced at high flow rates, which requires mathematic compensation. However, even after compensation, this roll-off phenomenon increases result variability. Despite these concerns, good agreement between PET and SPECT flows has been shown in head-to-head studies (58,62). In a study of 28 consecutive patients undergoing SPECT with AC and 13N-ammonia PET, K1 and MFR were computed without using any extraction fraction correction for 99mTc-tetrofosmin (62). The authors found that the hyperemic SPECT MBF values were lower than expected and, consequently, that MFR was lower than PET values (62). Agostini et al. compared 99mTc-sestamibi SPECT–derived MBF and MFR, using a solid-state camera and a correction for 99mTc extraction fraction, with values obtained with 15O-water PET in 30 patients (58). SPECT MBF measurements were lower than PET values on stress and rest acquisitions. However, SPECT-derived MFR was similar to PET values and had high diagnostic accuracy for PET-defined ischemia (>93%) or ischemia defined by invasive fractional flow reserve (>81%) (58). Therefore, although SPECT flow measurements may not be as robust as PET, they are emerging as an important ancillary measure. Notably, no additional tracer injection is needed, thus minimizing additional costs and adding no radiation. Clinical validation studies for MBF measurement with solid-state MPI are summarized in Table 1.
EARLY LEFT VENTRICULAR EJECTION FRACTION
SPECT measurements of MBF and MFR require an acquisition at the time of radiotracer injection. This facilitates measurement of stress left ventricular ejection fraction (LVEF) and LVEF reserve (defined as stress LVEF – rest LVEF)—which are important ancillary PET measurements. Although poststress LVEF changes are useful with SPECT, they are obtained at least 15 min after stress and are less sensitive than measurements during stress. High LVEF reserve has been shown to significantly reduce the probability of severe coronary artery disease during dipyridamole stress 82Rb PET (64). The feasibility of LVEF reserve measurements using early poststress acquisitions with a dedicated cardiac solid-state SPECT scanner has been demonstrated in a study of 50 patients (65). Significantly lower LVEF reserve was obtained in the fifth and the ninth minutes after regadenoson bolus in patients with significant ischemia compared with patients without ischemia. In the future, it may be feasible to measure LVEF reserve and MBF from the same early SPECT acquisitions, providing critical ancillary diagnostic and prognostic markers.
LARGE-SCALE CLINICAL VALIDATION
The increasing clinical use of solid-state SPECT camera systems has led to a large body of evidence validating their prognostic and diagnostic utility. Major studies of the diagnostic or prognostic utility of solid-state SPECT cameras are outlined in Table 2. One of the largest such efforts is a large, multicenter, international registry—the Registry of Fast Myocardial Perfusion Imaging with Next Generation SPECT (REFINE SPECT) (66). The novel collaborative design of this registry features the contribution of clinical data and image datasets from 9 (to date) investigative centers around the world into a comprehensive clinical-imaging database established at the central core laboratory at Cedars-Sinai Medical Center with centralized automatic quantitative analysis (66). To date, 20,418 solid-state MPI scans with collected prognostic data have been included in the registry. Initial results from the registry have shown that quantitative perfusion analysis, measured as total perfusion deficit (TPD), provides fine gradation of risk (67). Increasing TPD, even at very low levels, is associated with increasing risk of death, nonfatal myocardial infarction, unstable angina, or late revascularization (67). Notably, patients with more than 0 but less than 1% TPD had a 40% higher risk of events than patients with a TPD of 0, demonstrating the importance of even small deficits, which may be missed without the higher spatial resolution of new solid-state SPECT cameras (Fig. 7) (67).
The registry has also allowed researchers to perform various collaborative projects, including the development of novel machine learning methods tailored for the analysis of solid-state SPECT MPI and subsequent integration of quantitative image data variables with prescan clinical information. A machine-learning score derived with deep-convolutional neural networks has been shown to have superior diagnostic accuracy for obstructive coronary artery disease compared with quantitative ischemia assessments (45,68). A separate machine-learning algorithm had superior prognostic accuracy compared with either quantitative ischemia assessment or physician assessment (45). Other validation studies have been performed for solid-state MPI; these include validation of diagnostic accuracy including 2-position imaging and AC (21), determination of the transient ischemic dilation limits (69), and prognostic performance (67). Given the rich data collected, further validation studies are expected from this large multicenter cohort in the future.
FUTURE HARDWARE DESIGNS
Although these recent SPECT hardware improvements have proven to be clinically important, further advancements are still possible. One possible area for improvement is additional gains in photon sensitivity. Although more sensitive than conventional Anger cameras, even dedicated cardiac SPECT scanners detect less than 0.1% of emitted photons. Preclinical studies have demonstrated that higher efficiency may be possible. In a simulation, Zeng et al. (70) demonstrated that multidivergent beam collimators could double the photon sensitivity of current SPECT camera systems without sacrificing spatial resolution. Simulations of multipinhole designs with paired curved paraboloid or hemi-ellipsoid detectors could improve photon sensitivity by up to 300% (71,72). In phantom studies, these designs have improved photon sensitivity and spatial resolution compared with published characteristics or commercially available solid-state camera systems (72). Nevertheless, these published reports are still preclinical, and it remains to be seen if these designs will be cost-effective and adequate for clinical implementation.
SUMMARY
Solid-state cardiac SPECT offers substantial improvements in photon sensitivity and spatial resolution over conventional Anger cameras. These advances are due to novel collimator and geometric designs as well as incorporation of CZT crystals. Solid-state camera systems have facilitated dramatic reductions in both imaging time and radiation dose while maintaining high diagnostic accuracy. Growing clinical experience has validated the excellent diagnostic and prognostic utility attained with these camera systems. With the CZT systems, simultaneous dual-isotope SPECT is feasible and has been shown to be useful in selected applications. The improved technical performance has also facilitated measurement of ancillary high-risk markers including MBF and LVEF reserve. Although these advances have already provided a substantial improvement over the use of standard NaI crystal–based systems, the full clinical potential of solid-state cardiac SPECT camera systems has not yet been realized.
DISCLOSURE
No potential conflict of interest relevant to this article was reported.
Footnotes
Published online Aug. 2, 2019.
- © 2019 by the Society of Nuclear Medicine and Molecular Imaging.
REFERENCES
- 1.↵
- 2.↵
- 3.↵
- 4.↵
- 5.↵
- 6.↵
- 7.↵
- 8.↵
- 9.↵
- 10.↵
- 11.↵
- 12.↵
- 13.↵
- 14.↵
- 15.
- 16.↵
- 17.↵
- 18.↵
- 19.↵
- 20.↵
- 21.↵
- 22.↵
- 23.↵
- 24.↵
- 25.
- 26.↵
- 27.↵
- 28.↵
- 29.↵
- 30.↵
- 31.↵
- 32.↵
- 33.↵
- 34.↵
- 35.↵
- 36.↵
- 37.↵
- 38.↵
- 39.↵
- 40.↵
- 41.↵
- 42.↵
- 43.↵
- 44.↵
- 45.↵
- 46.↵
- 47.↵
- 48.↵
- 49.↵
- 50.↵
- 51.↵
- 52.↵
- 53.↵
- 54.↵
- 55.↵
- 56.↵
- 57.↵
- 58.↵
- 59.↵
- 60.
- 61.
- 62.↵
- 63.↵
- 64.↵
- 65.↵
- 66.↵
- 67.↵
- 68.↵
- 69.↵
- 70.↵
- 71.↵
- 72.↵
- 73.
- 74.
- 75.
- 76.
- 77.
- 78.
- 79.
- 80.
- 81.
- 82.
- 83.
- 84.
- 85.
- 86.
- Received for publication April 19, 2019.
- Accepted for publication June 18, 2019.