Journal of Nuclear Medicine
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI
  • Subscribe
  • My alerts
  • Log in
  • My Cart
Journal of Nuclear Medicine

Advanced Search

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Corporate & Special Sales
    • Journal Claims
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Permissions
    • Advertisers
    • Continuing Education
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • Follow JNM on Twitter
  • Visit JNM on Facebook
  • Join JNM on LinkedIn
  • Subscribe to our RSS feeds
Research ArticlePhysics/Instrumentation

Hybrid Surgical Guidance: Does Hardware Integration of γ- and Fluorescence Imaging Modalities Make Sense?

Gijs H. KleinJan, Daan Hellingman, Nynke S. van den Berg, Matthias N. van Oosterom, Kees Hendricksen, Simon Horenblas, Renato A. Valdes Olmos and Fijs WB van Leeuwen
Journal of Nuclear Medicine April 2017, 58 (4) 646-650; DOI: https://doi.org/10.2967/jnumed.116.177154
Gijs H. KleinJan
1Interventional Molecular Imaging Laboratory, Department of Radiology, Leiden University Medical Hospital, Leiden, The Netherlands
2Department of Nuclear Medicine, The Netherlands Cancer Institute—Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Daan Hellingman
2Department of Nuclear Medicine, The Netherlands Cancer Institute—Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands
3Oncovision (GEM Imaging S.A.), Centro de Investigación Príncipe Felipe, Valencia, Spain; and
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Nynke S. van den Berg
1Interventional Molecular Imaging Laboratory, Department of Radiology, Leiden University Medical Hospital, Leiden, The Netherlands
4Department of Urology, The Netherlands Cancer Institute—Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Matthias N. van Oosterom
1Interventional Molecular Imaging Laboratory, Department of Radiology, Leiden University Medical Hospital, Leiden, The Netherlands
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Kees Hendricksen
4Department of Urology, The Netherlands Cancer Institute—Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Simon Horenblas
4Department of Urology, The Netherlands Cancer Institute—Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Renato A. Valdes Olmos
1Interventional Molecular Imaging Laboratory, Department of Radiology, Leiden University Medical Hospital, Leiden, The Netherlands
2Department of Nuclear Medicine, The Netherlands Cancer Institute—Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Fijs WB van Leeuwen
1Interventional Molecular Imaging Laboratory, Department of Radiology, Leiden University Medical Hospital, Leiden, The Netherlands
4Department of Urology, The Netherlands Cancer Institute—Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Supplemental
  • Info & Metrics
  • PDF
Loading

Abstract

The clinically applied hybrid tracer indocyanine green-99mTc-nanocolloid enables combined radio- and fluorescence image guidance during sentinel node (SN) biopsy procedures. To provide optimal surgical guidance, this tracer requires the presence of both γ- and fluorescence modalities in the operating room. We reasoned that the combination or integration of these modalities could further evolve the hybrid surgical guidance concept. To study this potential, we clinically applied 2 setups that included the combination of γ-detection modalities and an open surgery fluorescence camera. Methods: To attach the fluorescence camera (VITOM) to either a γ-ray detection probe (GP; VITOM-GP) or a portable γ-camera (GC; Vitom GC), clip-on brackets were designed and printed in 3-dimensional sterilizable RC31. Both combined modalities were evaluated in, respectively, 5 and 6 patients with penile cancer during an SN biopsy procedure using indocyanine green-99mTc-nanocolloid. Intraoperatively, radio- and fluorescence-guided SN detection rates were scored at working distances of 0, 10, 20, and 30 cm for both combinations. Results: Using the VITOM-GP combination, we evaluated 9 SNs. γ-tracing rates were shown to be 100%, 88.9%, 55.6%, and 55.6% at a respective working distance of 0, 10, 20, and 30 cm. Detection rates for the fluorescence imaging–based detection were found to be 100%, 77.8%, and 77.8%, at respective working distances of 10, 20, and 30 cm. When the VITOM-GC setup was used, all 10 intraoperatively evaluated SNs could be visualized with the γ-camera independent of the working distance. Fluorescence detection rates were 90%, 80%, and 80% at 10-, 20-, and 30-cm working distances. The integrated detection modalities were shown to work synergistically; overall the, GC was most valuable for rough localization (10- to 30-cm range) of the SNs, the GP for providing convenient real-time acoustic feedback, whereas fluorescence guidance allowed detailed real-time SN visualization. Conclusion: Our findings suggest that full integration of a fluorescence camera with γ-detector (GP or GC) can be of value when a hybrid, radioactive and fluorescent tracer is used.

  • fluorescence guidance
  • radioguidance
  • hybrid modality
  • sentinel node biopsy
  • image-guided surgery

The radioguided surgery technique has strongly evolved since its introduction in the early 1960s (1,2) whereby the sentinel node (SN) biopsy procedure can be considered the best-known example. Basically this technique uses a radiotracer to (specifically) label lesions that require surgical removal. In the case of SN biopsy, after the injection of a radiocolloid (often 99mTc-nanocolloid) in or around the primary tumor, on drainage of the radiocolloid through the lymphatic system the primary tumor draining lymph nodes, so-called SNs can be identified (3). Acoustic (and numeric) feedback generated by a γ-ray detection probe (GP) can then be used to provide directional guidance during the intervention (4). Alternatively, portable γ-imaging modalities can be used to provide an intraoperative image of the nodal uptake (5,6). Despite research-oriented efforts, commercially available portable γ-detection modalities do not yet allow depiction of detailed anatomic information (7,8).

To provide high-resolution optical identification of the SNs, small dyes such as patent blue (V), fluorescein, and indocyanine green (ICG) have been used (9,10). Unfortunately, when radiocolloids and fluorescent agents are used separately, superficial intraoperative optical findings (<1 cm) can deviate from the (in-depth) observations made using the radiocolloid (11). Such shortcomings can be overcome using a hybrid tracer, for example, ICG- 99mTc-nanocolloid, a tracer that contains both a radioisotope and an optical fluorescent dye. The clinical value of ICG-99mTc-nanocolloid has already been extensively demonstrated by us (12–14) and others (11,15) whereby it was shown that using the hybrid tracer allows for direct translation of (in-depth) preoperative imaging findings into the operation theater. The radiolabel of the hybrid tracer allows preoperative SN mapping in a fashion similar to the conventional radioguided approach (12), meaning it also provides directional information for the placement of the incisions. During the operation, the combination of radio- and fluorescence guidance can be used to refine this even further by confirming the actual location and removal of the SNs defined at preoperative imaging (12–14). This combined approach significantly outperformed blue dye–based SN identification (12–14).

The increase in the combined clinical use of radio- and fluorescence-guidance methodologies, either using separate radioactive and fluorescent tracers or using hybrid tracers, has generated a need for multiple detection modalities in the operating room. This in turn has driven the development of integrated (or rather hybrid) navigation setups wherein the fluorescence camera is positioned using SPECT-based findings (16,17). To date, only 1 true hybrid modality has been reported in clinical use, namely a γ-probe–based device that was extended with fiber-based acoustic fluorescence tracing capabilities (18). On the basis of this success, one may wonder if it is also possible to accomplish the reverse in engineering by extending a known surgical fluorescence camera technology with the ability to detect γ-rays. To this end, we combined a small-sized fluorescence camera (VITOM) (19,20) with a GP and a portable γ-camera (GC). Using these 2 setups, we assessed the clinical logistics during the intervention and identified the engineering challenges that needed to be resolved to make an optimal hybrid detection apparatus for clinical use.

MATERIALS AND METHODS

Hardware

Near-infrared-fluorescence imaging was performed using a lightweight fluorescence camera (VITOM; KARL STORZ GmbH & Co. KG). This open surgery camera provides real-time on-screen fluorescence images (fluorescence signal in blue) with respect to the anatomy of the patient (20) and enables switching between the white light and the fluorescence imaging modus.

A hand-held GP (Europrobe 2; Eurorad), providing both a numeric and an acoustic read-out, was used for intraoperative γ-tracing. A mobile GC (Sentinella; Oncovision) was used for intraoperative γ-imaging (21). The latter is equipped with a hand-controlled laser pointer that enables marking of the focus point/location of γ-imaging on the skin or in the wound.

Engineering of Clip-on Devices

To connect the VITOM to the GP or GC, custom clip-on brackets were designed using 3-dimensional (3D) computer-aided engineering software (Solidworks; Dassault Systèmes) and subsequently 3-dimensional printed (VDM Kunststoftechniek) in sterilizable RC31 (Envisiontec) (Supplemental Fig. 1; supplemental materials are available at http://jnm.snmjournals.org).

The clip-on brackets were engineered in such a way that the focal points of γ-detection devices and the fluorescence camera were aligned (Fig. 1; Supplemental Fig. 1). For the fluorescence camera and γ-probe (VITOM-GP) combination, the focal point was approximately 11 cm, with an angle in alignment of both detection modalities of 7.24° (Fig. 1A). For the fluorescence camera and γ-camera (VITOM-GC) combination, the VITOM was placed under a 33.8° angle. The focal point was again localized at approximately 11 cm (measured from the tip of the VITOM) and at 6.4 cm from the collimator of the GC (9.5 cm from the detector; Fig. 1B).

FIGURE 1.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 1.

Schematic overview of VITOM-GP and VITOM-GC combination. (A) VITOM-GP combination. (B) VITOM-GC combination. Because of angle in alignment, devices had to be moved horizontally to compensate for misalignment at distances beyond focal plane.

FIGURE 2.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 2.

Fluorescence imaging versus γ-detection. On left side, SN detection rates (%) of VITOM-GP combination are shown in relation to the evaluated working distances (A–F). On right, SN detection rates (%) of VITOM-GC are presented in relation to the different evaluated working distances (G–L).

Patients

Eleven patients with cT1–cT3N0 penile cancer scheduled for SN biopsy and subsequent treatment of the primary tumor were prospectively included after obtaining written informed consent. Patient characteristics are shown in Table 1.

View this table:
  • View inline
  • View popup
TABLE 1

Patient Characteristics

Followed procedures were in accordance with the ethical standards of the responsible committee on human experimentation (institutional and national) and with the Helsinki Declaration of 1975, as revised in 2008. This study was approved by the Institutional Review Board of The Netherlands Cancer Institute—Antoni van Leeuwenhoek hospital (NKI-AVL, Amsterdam, The Netherlands).

Injection Procedure and Preoperative Imaging

Hybrid tracer injection and preoperative SN mapping were performed 1 d before surgery or on the morning of the day of surgery as previously described (13). In brief, 4 deposits of 0.1 mL of hybrid tracer ICG-99mTc-nanocolloid was injected per patient around the primary lesion. Dynamic (during the first 10 min after injection) and static (at 15 min and 2 h after injection) planar lymphoscintigraphy was performed. Lymphoscintigrams at 2 h after injection were followed by SPECT fused with CT (SPECT/CT) imaging. The number and location of SNs were determined on the basis of preoperative imaging and marked on the skin.

System Evaluation

The current study was performed by experienced surgical personnel that regularly uses intraoperative fluorescence- and γ-tracing modalities, even in combination with the hybrid tracer (13).

Before incision, the ability of the VITOM-GP or the VITOM-GC to detect the SN transcutaneously, via fluorescence imaging (22) and γ-detection, was evaluated when the probe was placed in direct contact with the skin (0 cm; VITOM-GP only) and at the optimal focal point (≈11 cm) of the VITOM. After tissue preparation and full exposition of the SN, the sensitivity of fluorescence imaging and γ-detection (visual or acoustic [numeric]) of the VITOM-GP and VITOM-GC was evaluated at different working distances namely: 0 (VITOM-GP only), 10 (≈VITOM focal plane), 20, and 30 cm (Fig. 2).

The efficiency of detection was scored and presented in percentage detected SNs relative to the total amount of SNs pursued (Table 1). The true value of a possible future hybrid device was evaluated in the focal plane because here both signals were detected synchronically in the same focal spot.

After SN excision, the wound bed was checked for the presence of residual radioactivity and fluorescence with either the VITOM-GC or the VITOM-GP. The SNs that were removed without evaluation of the respective hybrid modality were located with intermittent use of the GP and fluorescence detection (VITOM).

Pathology

Excised tissue specimens were processed as described previously (13).

RESULTS

Preoperative Imaging and Pathology

Details on the injected dose of the hybrid tracer, preoperative SN mapping, and pathology results are stated in Table 1.

Intraoperative Evaluation of VITOM-GP Combination

Preoperatively, 14 SNs were identified on SPECT/CT in the VITOM-GP patient group. For practical reasons, in some patients the limited available time prevented us from evaluating all SNs using the VITOM-GP combination. Therefore, only 9 of these 14 preoperatively identified SNs were included in the evaluation of the VITOM-GP combination.

Before placement of the incision, the γ-detection rate was 100% in all cases. In contrast, via fluorescence detection none of the SNs could be detected transcutaneously (Supplemental Fig. 2). Here, the VITOM was unable to fully focus at this short distance (0 cm). On SN exposure, the detection rates in the focal plane (≈11 cm from the SN) were 88.9% for the γ-detection and 100% for fluorescence detection. At 20 and 30 cm working distance, γ-tracing identified 55.6% of the pursued SNs, whereas fluorescence imaging enabled detection of 88.9% of the SNs (Fig. 2). At the larger distances, SN detection using the numeric/acoustic nature of the read-out of the GP was sometimes hindered by emissions coming from the nearby primary injection site.

The VITOM-GP combination was relatively small in size and could easily be handled with one hand. It also allowed real-time interpretation of both the γ-signal (acoustic/numeric) and the fluorescence signal (on-screen, visual). When the device was moved, the γ-tracing option provided directional feedback for the placement of the VITOM-GP. As a result of the angulation of the different modalities, dual imaging was possible only at the focal point (≈11 cm). Beyond that distance, placement of the 2 modalities had to be corrected horizontally to accommodate fluorescence detection (Fig. 1).

Intraoperative Evaluation of VITOM-GC Combination

Again, for practical reasons only 10 of the 15 SNs identified using SPECT/CT were surgically pursued with the VITOM-GC combination. Before initiation of the surgical procedure the detection rate of the GC was 100%, irrespective of the evaluated working distance (10, 20, or 30 cm). Transcutaneous fluorescence-based SN detection was not possible. On SN exposure, the GC detection rate remained 100%. Fluorescence imaging provided a 90% detection rate at the focal plane, which dropped to 80% at 20 and 30 cm; in 1 patient with a body mass index over 40 fluorescence, detection failed as a likely result of tissue attenuation. When γ-imaging beyond the 20-cm distance was performed, the injection site also became visible. Other than with the GP (see above), here the GC signal could be used to aid in the orientation (Figs. 2J and 2K).

The VITOM-GC combination was large and heavy, and required a retractable (external) supporting device for placement. This device was considered convenient as it enabled accurate camera placement for longer periods of time, while leaving the surgeon’s hands free to perform the resection under image guidance. The laser pointer on the GC could be used to highlight the area of interest during γ-imaging (acquisition times up to 30 s). However, as the pointer interfered with the fluorescence findings, it had to be switched off before fluorescence imaging was initiated (Supplemental Figs. 2E and 2F). Data interpretation in the focal plane was limited by the fact that both signals were visually displayed and the operating surgeon had to look at 2 screens to determine the relation between the 2 findings. Beyond the focal point, the large angle between both modalities (33.8°) meant that the spatial displacement between fluorescence- and γ-imaging findings was even more critical in this setup.

DISCUSSION

This study describes 2 hybrid imaging setups (VITOM-GP and VITOM-GC) and their potential to provide surgical guidance in combination with the hybrid tracer ICG-99mTc-nanocolloid. In both setups, the in-depth information of the γ-signal was used to determine the site of incision and provided directional guidance for positioning of the VITOM. When the SN was exposed, fluorescence imaging could be used to visually detect the SNs with respect to the anatomic context. These findings clearly underline the strengths and weaknesses of the 2 individual imaging signatures. Moreover, the reported detection percentages show that the find rates with the hybrid modality are highly similar to those previously reported for combined use of the individual modalities (13). The current prototype setup was not optimized from an efficacy point of view, and its use influenced operating room logistics and the operation time. More extensive future studies, however, are required to evaluate whether the use of an optimized hybrid modality can in fact result in reduced morbidity, operation time, and postoperative infection rates.

The findings of the current study allow us to derive several engineering challenges that have to be faced before imaging-based hybrid surgical guidance modalities will become efficient tools for routine use. The clip-on design was considered convenient because it efficiently combined 2 clinically available modalities but still allowed for their individual use. Unfortunately, the angled alignment was not optimal for distances smaller or larger than the focal plane. In both combined devices, the angle between the modalities dictated spatial placement to get the information from the SN for both modalities (Fig. 1). Most likely future hybrid modalities will require both modalities to be placed perpendicular to each other and with an identical field of view. Though this will require full hardware integration. The findings of the current study also suggest that placement of both image guidance modalities on a stable, but retractable, arm could be of value.

In the current setup, the optimal detection plane of the individual modalities was shown to differ for the VITOM (>11 cm), GP (0–10 cm), and GC (0–30 cm). Furthermore, intraoperative adaptation of the fluorescence focus was considered inconvenient. The concept of hybrid imaging hardware could only be accurately evaluated in the focal plane of the VITOM (≈11 cm). In future hybrid devices, the focus of the 2 modalities should be matched, at least in a certain range. Possibly an autofocus option should also be included for the optical imaging.

The largest and final hurdle to be tackled, however, will be the difference in detection sensitivity between the radioguidance and fluorescence modalities used. Although full integration is possible from an engineering point of view, its realization is not straightforward when combination of state-of-the art detector sensitivity for both fluorescence and radioactivity is envisioned. Especially when considering that the fluorescence modality could benefit from higher fluorescence sensitivity at longer distances. This would require both a stronger excitation light-source and a higher detection sensitivity. The white light option should remain included because this information created directional feedback and provided anatomic context. The difference in sensitivity between the GP and GC modalities influenced the acquisition time, collimation, and field of view (2,6), all of which can be improved further.

Overall the VITOM-GP combination was preferred by the surgeons over the VITOM-GC combination. The reason for this was that the acoustic/numeric feedback provided by the GP supplied information that could be processed simultaneously with the fluorescence imaging findings. Hereby 2 sensory organs, ears and eyes, work in conjunction, providing directional guidance for the placement of the fluorescence camera. In contrast, parallel display of the VITOM and GC findings relied on a single sensory organ, making it more difficult to process. Having the display of the 2 different imaging findings on 2 different screens was sometimes considered confusing. Hereby especially the lack of anatomic information in the GC images made it difficult to relate them to the fluorescence imaging findings. The value of the VITOM-GC combination could potentially increase when both imaging findings are integrated in a single (video screen) display, a technology that we previously successfully applied in a navigation setup (23).

CONCLUSION

In this study, we demonstrated that combined radio- and fluorescence imaging modalities have the potential to make sense in the hybrid surgical guidance concept. The integrated detection modalities were shown to work synergistically; overall the GC was most valuable for rough localization (10- to 30-cm range) of the SNs and the GP for providing convenient real-time acoustic feedback, whereas fluorescence guidance allowed detailed real-time SN visualization.

DISCLOSURE

This work was partially supported by an NWO-STW-VIDI grant (no. STW BGT11272) and a European Research Council under the European Union’s Seventh Framework Program (FP7/2007-2013) grant (no. 2012-306890). No other potential conflict of interest relevant to this article was reported.

Acknowledgments

We gratefully acknowledge the staff of the nuclear medicine department and the surgical staff of the Netherlands Cancer Institute—Antoni van Leeuwenhoek Hospital for their support and assistance.

Footnotes

  • Published online Sep. 29, 2016.

  • © 2017 by the Society of Nuclear Medicine and Molecular Imaging.

REFERENCES

  1. 1.↵
    1. KleinJan GH,
    2. Bunschoten A,
    3. Brouwer OR,
    4. Berg NS,
    5. Valdés-Olmos RA,
    6. Leeuwen FWB
    . Multimodal imaging in radioguided surgery. Clin Transl Imaging. 2013;1:433–444.
    OpenUrl
  2. 2.↵
    1. Povoski SP,
    2. Neff RL,
    3. Mojzisik CM,
    4. et al
    . A comprehensive overview of radioguided surgery using gamma detection probe technology. World J Surg Oncol. 2009 27;7:11.
    OpenUrlCrossRefPubMed
  3. 3.↵
    1. Nieweg OE,
    2. Tanis PJ,
    3. Kroon BB
    . The definition of a sentinel node. Ann Surg Oncol. 2001;8:538–541.
    OpenUrlCrossRefPubMed
  4. 4.↵
    1. Heller S,
    2. Zanzonico P
    . Nuclear probes and intraoperative gamma cameras. Semin Nucl Med. 2011;41:166–181.
    OpenUrlCrossRefPubMed
  5. 5.↵
    1. Vermeeren L,
    2. Valdés Olmos RA,
    3. Klop WMC,
    4. Balm AJM,
    5. van den Brekel MWM
    . A portable gamma-camera for intraoperative detection of sentinel nodes in the head and neck region. J Nucl Med. 2010;51:700–703.
    OpenUrlAbstract/FREE Full Text
  6. 6.↵
    1. Tsuchimochi M,
    2. Hayama K
    . Intraoperative gamma cameras for radioguided surgery: technical characteristics, performance parameters, and clinical applications. Phys Med. 2013;29:126–138.
    OpenUrlCrossRefPubMed
  7. 7.↵
    1. Hellingman D,
    2. Vidal-Sicart S,
    3. de Wit-van der Veen LJ,
    4. Paredes P,
    5. Valdés Olmos RA
    . A new portable hybrid camera for fused optical and scintigraphic imaging: first clinical experiences. Clin Nucl Med. 2016;41:e39–e43.
    OpenUrl
  8. 8.↵
    1. Lees JE,
    2. Bassford DJ,
    3. Blake OE,
    4. Blackshaw PE,
    5. Perkins AC
    . A hybrid camera for simultaneous imaging of gamma and optical photons. J Instrum. 2012;7:06009–P06009.
    OpenUrl
  9. 9.↵
    1. Morton DL,
    2. Wen DR,
    3. Wong JH,
    4. et al
    . Technical details of intraoperative lymphatic mapping for early stage melanoma. Arch Surg. 1992;127:392–399.
    OpenUrlCrossRefPubMed
  10. 10.↵
    1. van Leeuwen FWB,
    2. Hardwick JCH,
    3. van Erkel AR
    . Luminescence-based imaging approaches in the field of interventional molecular imaging. Radiology. 2015;276:12–29.
    OpenUrlCrossRefPubMed
  11. 11.↵
    1. Stoffels I,
    2. Leyh J,
    3. Pöppel T,
    4. Schadendorf D,
    5. Klode J
    . Evaluation of a radioactive and fluorescent hybrid tracer for sentinel lymph node biopsy in head and neck malignancies: prospective randomized clinical trial to compare ICG-99mTc-nanocolloid hybrid tracer versus 99mTc-nanocolloid. Eur J Nucl Med Mol Imaging. 2015;42:1631–1638.
    OpenUrl
  12. 12.↵
    1. Brouwer OR,
    2. Buckle T,
    3. Vermeeren L,
    4. et al
    . Comparing the hybrid fluorescent-radioactive tracer indocyanine green-99mTc-nanocolloid with 99mTc-nanocolloid for sentinel node identification: a validation study using lymphoscintigraphy and SPECT/CT. J Nucl Med. 2012;53:1034–1040.
    OpenUrlAbstract/FREE Full Text
  13. 13.↵
    1. Brouwer OR,
    2. van den Berg NS,
    3. Mathéron HM,
    4. et al
    . A hybrid radioactive and fluorescent tracer for sentinel node biopsy in penile carcinoma as a potential replacement for blue dye. Eur Urol. 2014;65:600–609.
    OpenUrlCrossRefPubMed
  14. 14.↵
    1. KleinJan GH,
    2. van den Berg NS,
    3. Brouwer OR,
    4. et al
    . Optimisation of fluorescence guidance during robot-assisted laparoscopic sentinel node biopsy for prostate cancer. Eur Urol. 2014;66:991–998.
    OpenUrlCrossRefPubMed
  15. 15.↵
    1. Christensen A,
    2. Juhl K,
    3. Charabi B,
    4. et al
    . Feasibility of real-time near-infrared fluorescence tracer imaging in sentinel node biopsy for oral cavity cancer patients. Ann Surg Oncol. 2016;23:565–572.
    OpenUrl
  16. 16.↵
    1. Brouwer OR,
    2. Buckle T,
    3. Bunschoten A,
    4. et al
    . Image navigation as a means to expand the boundaries of fluorescence-guided surgery. Phys Med Biol. 2012;57:3123–3136.
    OpenUrlCrossRefPubMed
  17. 17.↵
    1. KleinJan GH,
    2. van den Berg NS,
    3. van Oosterom MN,
    4. et al
    . Towards (hybrid) navigation of a fluorescence camera in an open surgery setting. J Nucl Med. 2016;57:1650–1653.
    OpenUrlAbstract/FREE Full Text
  18. 18.↵
    1. van den Berg NS,
    2. Simon H,
    3. Kleinjan GH,
    4. et al
    . First-in-human evaluation of a hybrid modality that allows combined radio- and (near-infrared) fluorescence tracing during surgery. Eur J Nucl Med Mol Imaging. 2015;42:1639–1647.
    OpenUrlCrossRefPubMed
  19. 19.↵
    1. Mamelak AN,
    2. Drazin D,
    3. Shirzadi A,
    4. Black KL,
    5. Berci G
    . Infratentorial supracerebellar resection of a pineal tumor using a high definition video exoscope (VITOM®). J Clin Neurosci. 2012;19:306–309.
    OpenUrlPubMed
  20. 20.↵
    1. Buda A,
    2. Dell’Anna T,
    3. Vecchione F,
    4. Verri D,
    5. Di Martino G,
    6. Milani R
    . Near-infrared sentinel lymph node mapping with indocyanine green using the VITOM II ICG Exoscope for open surgery for gynecologic malignancies. J Minim Invasive Gynecol. 2016;23:628–632.
    OpenUrl
  21. 21.↵
    1. Sánchez F,
    2. Benlloch JM,
    3. Escat B,
    4. et al
    . Design and tests of a portable mini gamma camera. Med Phys. 2004;31:1384–1397.
    OpenUrlPubMed
  22. 22.↵
    1. van den Berg NS,
    2. Miwa M,
    3. KleinJan GH,
    4. et al
    . (Near-infrared) fluorescence-guided surgery under ambient light conditions: a next step to embedment of the technology in clinical routine. Ann Surg Oncol. 2016;23:2586–2595.
    OpenUrl
  23. 23.↵
    1. KleinJan GH,
    2. Bunschoten A,
    3. van den Berg NS,
    4. et al
    . Fluorescence guided surgery and tracer-dose, fact or fiction? Eur J Nucl Med Mol Imaging. 2016;43:1857–1867.
    OpenUrl
  • Received for publication April 20, 2016.
  • Accepted for publication September 8, 2016.
PreviousNext
Back to top

In this issue

Journal of Nuclear Medicine: 58 (4)
Journal of Nuclear Medicine
Vol. 58, Issue 4
April 1, 2017
  • Table of Contents
  • Table of Contents (PDF)
  • About the Cover
  • Index by author
Print
Download PDF
Article Alerts
Sign In to Email Alerts with your Email Address
Email Article

Thank you for your interest in spreading the word on Journal of Nuclear Medicine.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Hybrid Surgical Guidance: Does Hardware Integration of γ- and Fluorescence Imaging Modalities Make Sense?
(Your Name) has sent you a message from Journal of Nuclear Medicine
(Your Name) thought you would like to see the Journal of Nuclear Medicine web site.
Citation Tools
Hybrid Surgical Guidance: Does Hardware Integration of γ- and Fluorescence Imaging Modalities Make Sense?
Gijs H. KleinJan, Daan Hellingman, Nynke S. van den Berg, Matthias N. van Oosterom, Kees Hendricksen, Simon Horenblas, Renato A. Valdes Olmos, Fijs WB van Leeuwen
Journal of Nuclear Medicine Apr 2017, 58 (4) 646-650; DOI: 10.2967/jnumed.116.177154

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Share
Hybrid Surgical Guidance: Does Hardware Integration of γ- and Fluorescence Imaging Modalities Make Sense?
Gijs H. KleinJan, Daan Hellingman, Nynke S. van den Berg, Matthias N. van Oosterom, Kees Hendricksen, Simon Horenblas, Renato A. Valdes Olmos, Fijs WB van Leeuwen
Journal of Nuclear Medicine Apr 2017, 58 (4) 646-650; DOI: 10.2967/jnumed.116.177154
Twitter logo Facebook logo LinkedIn logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One
Bookmark this article

Jump to section

  • Article
    • Abstract
    • MATERIALS AND METHODS
    • RESULTS
    • DISCUSSION
    • CONCLUSION
    • DISCLOSURE
    • Acknowledgments
    • Footnotes
    • REFERENCES
  • Figures & Data
  • Supplemental
  • Info & Metrics
  • PDF

Related Articles

  • This Month in JNM
  • PubMed
  • Google Scholar

Cited By...

  • Synthesis of Patient-Specific Transmission Data for PET Attenuation Correction for PET/MRI Neuroimaging Using a Convolutional Neural Network
  • Dynamic 18F-FDG PET Lymphography for In Vivo Identification of Lymph Node Metastases in Murine Melanoma
  • Google Scholar

More in this TOC Section

  • Comparison of PET/CT with Sequential PET/MRI Using an MR-Compatible Mobile PET System
  • Zero-Echo-Time and Dixon Deep Pseudo-CT (ZeDD CT): Direct Generation of Pseudo-CT Images for Pelvic PET/MRI Attenuation Correction Using Deep Convolutional Neural Networks with Multiparametric MRI
  • PET Imaging Stability Measurements During Simultaneous Pulsing of Aggressive MR Sequences on the SIGNA PET/MR System
Show more Physics/Instrumentation

Similar Articles

Keywords

  • Fluorescence guidance
  • Radioguidance
  • Hybrid modality
  • sentinel node biopsy
  • image-guided surgery
SNMMI

© 2023 Journal of Nuclear Medicine

Powered by HighWire