Abstract
Interest in nuclear breast imaging is increasing because of technical improvements in dedicated devices that allow the use of relatively low doses of radiotracers with high sensitivity for even small breast cancers. For women with newly diagnosed cancer, primary chemotherapy is often recommended, and improved methods of assessing treatment response are of interest. With widespread breast density notification, functional rather than anatomic methods of screening are of increasing interest as well. For a cancer imaging technology to be adopted, several criteria must be met that will be discussed: evidence of clinical benefit with minimal harm, standardized interpretive criteria, direct biopsy guidance, and acceptable cost-effectiveness.
Two radiotracers have been used widely to depict breast cancer: the γ-emitting 99mTc-sestamibi (99mTc-methoxyisobutylisonitrile), 140 keV, half-life of 6 h, originally developed as a myocardial perfusion agent; and the positron-emitting glucose analog 18F-FDG, 511 keV, half-life of 2 h. Standard lead shielding as is used in a mammography suite is sufficient when using 99mTc-sestamibi. Using 18F-FDG requires much more extensive shielding and may not be feasible because of workspaces above or below the planned unit. State licensing requirements for nuclear medicine must be considered; typically, a hot lab with oversight by a trained nuclear medicine technologist and a physicist is required (1).
Both 99mTc-sestamibi and 18F-FDG were initially studied for breast cancer using whole-body scanners. Low sensitivity to invasive cancers smaller than 1 cm was observed with whole-body γ-camera imaging in a prospective multicenter series at 48.2%, compared with 74.2% for larger tumors (2); detection of small cancers, particularly small invasive cancers, is a major goal of breast imaging. Similar results were observed with whole-body PET, with sensitivity to invasive tumors 2 cm or smaller of only 30 of 44 (68%) compared with 57 of 62 (92%) for larger cancers (3).
Dedicated breast γ-camera imaging can be performed with single-detector scintillating-crystal systems such as breast-specific γ-imaging (BSGI) (Dilon Technologies) with a 20 × 15 cm field of view, 3.3-mm pixel size; or using dual-head cadmium–zinc–telluride detector systems, often referred to as molecular breast imaging (MBI) (GE Healthcare, with a 24 × 16 cm field of view and 2.5-mm pixel size; or Gamma Medica, Inc., with a 20 × 16 cm field of view and 1.6 mm pixel size). Both approaches use positioning similar to that of mammography, with the breast gently stabilized between a compression paddle and the detector (BSGI) or between 2 detectors (MBI). It is important to include mammographic technologists in the positioning of the patients, at least for a minimum number of cases (e.g., 25), to ensure full inclusion of posterior tissues for both a craniocaudal and a mediolateral oblique (MLO) acquisition. Particularly with the dual-head systems, it can be difficult to prevent skin folds or to ensure that the nipple is in profile because the opaque detectors block direct visualization of the breast during positioning.
Typically, imaging begins within 5 min after intravenous radiotracer injection and includes 10-min acquisitions in both craniocaudal and MLO projections of each breast, for a total minimum of 40 min to examine both breasts. Additional views such as a laterally exaggerated craniocaudal view or axillary tail view may be needed to fully include all breast tissue, and even then the extreme posterior tissues are difficult to fully include when positioning the breast between 2 rather bulky detectors on MBI or positron emission mammography (PEM) (CMR Naviscan). In a series of 149 invasive cancers imaged by BSGI (4), all cancers 8 mm or larger were visualized, as were all cancers grade 2 or higher; only 3 of 6 grade 1 cancers 7 mm or smaller were seen. In prototype studies of lesions imaged before biopsy using a single- versus dual-head MBI system (5), only 2 of 7 (29%) cancers 5 mm or smaller and 24 of 28 (86%) of those 6–10 mm were detected on the single-head system compared with 11 of 16 (69%) 5 mm or smaller and 41 of 45 (91%) 6–10 mm on the dual-head system. Dual-head systems are generally now used for clinical MBI. In a more recent series using MBI (6), among 357 invasive cancers for which pathology size was available, 310 (87%) were seen; sensitivity was reduced for smaller tumors, with 45 of 63 (71%) cancers 5 mm or smaller seen, 64 of 84 (76%) of those 6–10 mm, 75 of 81 (93%) of those 11–15 mm, 43 of 45 (96%) of those 16–20 mm, and 83 of 84 (99%) of those larger than 2 cm identified on MBI (Amy L. Conners, personal communication, August 2015; P < 0.0001 by χ2 test). Of 49 MBI-occult cancers, 8 (16%) were thought to have been outside the field of view (which corresponded to 8 of 360, 2.2% of all tumor foci) (6). Of note, 99mTc-sestamibi was first used as a cardiac radiotracer, and it is possible to perform dedicated breast γ-camera imaging after cardiac imaging using the same dose of radiotracer (7).
Several dedicated breast PET devices have been developed; and again, imaging can be performed after PET/CT or PET/MRI using the same radiotracer dose, although insurance reimbursement may be denied for the breast PET in that circumstance. The first PEM system (CMR Naviscan) uses positioning similar to that of mammography, with the breast gently stabilized between a transparent compression paddle and the detector, and 10-min craniocaudal and MLO acquisitions. In direct comparison studies, PEM has shown greater sensitivity to small cancers than whole-body PET or PET/CT, although only 10 of 14 (71%) index cancers 6–10 mm in size and 5 of 12 (42%) 5 mm or smaller were seen on PEM in one series (8), as were 11 of 15 (73%) cancers 1 cm or smaller in another series (9). Despite direct involvement of mammography technologists in positioning or at least the initial training of nuclear medicine technologists in positioning at least the first 25 patients per site, inadequate positioning was prospectively reported in 36 of 388 (9.3%) participants in a multicenter PEM trial; another 11 (2.8%) participants had large breasts, requiring tiled images to include all breast tissue (10). In part because of the coincidence requirement for counting positron emissions (at 180° from each other), approximately 1 cm of tissue at the extreme posterior of the breast is not well evaluated on PEM.
Dedicated prone breast PET systems, which should improve visualization of posterior tissues, have also been developed: Mammi-PEM (Oncovision; Gem Imaging SA), Clear PEM (Crystal Clear Collaboration; CERN), and O-scanner (Shimadzu Medical Systems), as well as a C-scanner in which the seated patient leans into a C-shaped detector system (Shimadzu). With 18F-FDG–based imaging, the radiotracer must circulate for at least an hour before imaging is started, and longer circulation times of up to 2 h seem to improve lesion-to-background uptake.
Both 99mTc-sestamibi– and 18F-FDG–based breast imaging benefit from the patient fasting, with greater radiotracer uptake by the breast tissue in a fasting state (11). For 18F-FDG, a minimum of a 4-h and preferably a 6-h fast is recommended. The patient should be kept resting and warm to minimize muscle uptake of either radiotracer (11). The injection site (typically the antecubital fossa) can be imaged to ensure lack of infiltration or muscle uptake.
For all medical imaging, it is important to minimize the dose of ionizing radiation while still providing appropriate diagnostic information. This is particularly true when imaging younger individuals who have at least a 10- to 20-y life expectancy during which radiation-induced cancers could be observed. Women who are pregnant or lactating should not be imaged with nuclear breast techniques. Most published articles about BSGI reported injected doses of 555–1,110 MBq (15–30 mCi). With dual-head cadmium–zinc–telluride detector systems and optimized collimator and energy windows, an injected dose averaging 300 MBq (8.1 mCi) has been used while maintaining performance characteristics (12). Unlike mammography, radiation exposure during nuclear breast imaging is to the whole body, with the greatest accumulation of 99mTc-sestamibi seen in the colon, kidneys, bladder, and gallbladder. The effective dose from a 300-MBq (8.1-mCi) injection, considering the radiation risk to each organ, is 2.4 mSv, compared with about 0.4 mSv for mammography (13). Use of a 150-MBq (4-mCi) dose of 99mTc-sestamibi is under study and may be possible with optimal patient preparation (11) and the use of appropriate syringes that minimize the adherence of 99mTc-sestamibi (14).
The published literature on PEM has typically used an 18F-FDG dose of 370 MBq (10 mCi) for an effective dose of 6.2–7.1 mSv (15), with the highest dose to the bladder.
Importantly, such radiation exposures from nuclear breast imaging are at or below the 3- to 10-mSv background radiation dose from living on Earth for 1 y. According to the American Association of Physicists in Medicine, the risks of medical imaging at effective doses of less than 50 mSv for single procedures, or 100 mSv for multiple procedures over short periods, are too low to be detectable and may be nonexistent (16).
THE EVIDENCE
Many applications of nuclear breast imaging are for the same indications for which contrast-enhanced breast MRI would be used. Diagnostic breast imaging can include problem solving when vague abnormalities persist after additional mammographic views and ultrasound; but with current ultrasound equipment and tomosynthesis, it is rare to require further diagnostic imaging. Sensitivity is at least as high as ultrasound with BSGI, but specificity is higher with BSGI (17,18). In patients presenting with cancer of unknown primary cause (typically, a metastatic axillary node consistent with breast cancer), both 99mTc-sestamibi– and 18F-FDG–based imaging can identify the underlying breast cancer (Fig. 1); however, no series have been published on this issue. Although papillomas may show abnormal uptake of either radiotracer, MBI was negative in all 14 patients with bloody nipple discharge in one series, including 2 patients with papillomas and one with invasive cancer found 17 mo later (19).
EXTENT OF DISEASE
MRI has been widely studied for assessing extent of disease in women with newly diagnosed cancer. Metaanalyses show additional ipsilateral mammographically occult disease in 16% of women, with a positive predictive value (PPV) of additional MRI findings of 66% (20); and in the contralateral breast, occult disease is found in 4% of women on average, with a PPV of MRI findings of 48% (21).
A few retrospective studies have evaluated the use of 99mTc-sestamibi–based breast imaging to assess disease extent. In one series (22), 9 of 159 (6%) women had additional unsuspected ipsilateral foci, and 5 (3%) had contralateral foci seen on BSGI, with a PPV of suggestive findings on BSGI of 14 of 40 (35%). The PPV of additional suggestive findings was 15 of 25 (60%) in a series of 138 women studied by Zhou et al. (23), and 7 of 18 (39%) (including one ductal carcinoma in situ [DCIS] excised) in a series of 82 women reported by Killelea et al. (24). The PPV of additional suspicious findings of only 35%–60% underscores the importance of percutaneous biopsy of additional suspicious findings before converting a patient to mastectomy.
In a series of 66 women who underwent BSGI and MRI (25), specificity was higher for BSGI, at 90%, than for MRI, at 39% (P < 0.0001). In a series of 118 women anticipating breast-conserving therapy, 14 (11.9%) were appropriately converted to mastectomy on the basis of BSGI; another 15 of the 104 (14.4%) remaining women thought to be candidates for breast-conserving therapy required mastectomy (underestimated extent on BSGI), with more than half of those being underestimated because of a greater extent of DCIS than seen on BSGI (26). In a series of 286 patients with 390 invasive tumor foci imaged by MBI (6), 341 (87.4%) foci were seen on MBI, although ascertainment of missed cancers depends on detailed pathologic analysis exceeding that in routine practice.
PEM has been compared with MRI for assessing local disease extent in a prospective multicenter series of 388 women with newly diagnosed breast cancer (10,27). PEM showed comparable but complementary sensitivity in the ipsilateral breast and higher specificity than MRI, with a PPV of additional suspicious ipsilateral findings on PEM of 47 of 71 (66%) compared with 61 of 116 (53%) for MRI (P = 0.016). In the same series of 388 women, 15 (3.9%) had contralateral cancer identified after study entry, with 14 of 15 (93%) seen on MRI, only 3 of 15 (20%) prospectively seen on PEM, and only 11 of 15 (73%) visible even in retrospect on PEM. The PPV of PEM-prompted contralateral biopsies (3 of 14, or 21%) was not different from that of MRI (15 of 54, or 28%) (P = 0.58) (27).
The dilemma with imaging the extent of disease is that, despite extensive study, MRI generally has not been shown to reduce the risk of local or distant recurrence (28), and no studies have been conducted to show such an impact with nuclear breast imaging, but similar issues likely apply. Further studies with molecular subtypes may prove more fruitful and are ongoing. In breast cancers lacking estrogen, progesterone, and HER2-neu (ERB-B2) receptors—that is, triple negative—lack of exposure to pretreatment breast MRI has been shown to independently correlate with an increased risk of recurrence at a median follow-up of 6.1 y, with a hazard ratio of 2.66 (95% confidence interval, 1.49–4.75), comparable in magnitude to dense breasts and lymphovascular invasion (29). Unlike series across all tumor types (30,31), MRI did not increase the rate of mastectomy in women with triple-negative disease (29). Reexcision rates are generally not reduced by preoperative MRI (31,32), although MRI has been shown to be beneficial in presurgical planning for invasive lobular carcinoma (ILC) (33,34). Size of unifocal carcinoma is better assessed by imaging than is detailed mapping of multiple foci of disease (10), in which it can be difficult to translate imaging findings to intraoperative positioning and excision.
Neither 99mTc-sestamibi– nor 18F-FDG–based imaging is reliable at identifying metastatic axillary adenopathy; only 10 of 78 (13%) metastatic axillae were prospectively identified by PEM (10). The axilla cannot be reliably included on devices that use positioning similar to that of mammography. Normal nodes can show mild radiotracer accumulation, but small metastatic deposits may not. Extravasation of radiotracer during injection of 99mTc-sestamibi can produce abnormal uptake in ipsilateral nodes, which is more intense than that due to metastasis (35).
The use of nuclear medicine methods to assess residual disease before reexcision in patients with initial breast-conserving surgery for cancer with positive margins has not been studied. 18F-FDG uptake is increased by prior biopsy or surgery; thus, the area near the surgical site cannot be acutely evaluated with breast PET. Scars lacking inflammation from fat necrosis typically do not show uptake after about 2 y, and PET can be used to help distinguish scarring from recurrence (36).
SCREENING
The perfect screening test would detect clinically important breast cancers only when they can be easily treated, would not detect any other lesions or induce any malignancies, and would be widely available and well tolerated. Because mammography is the only screening method that has been studied in randomized controlled trials and shown to reduce mortality from breast cancer (37), any other methods of screening are still viewed as supplements to mammography. Annual screening with MRI is recommended for women at a high risk for breast cancer because of known or suspected pathogenic mutations in the BRCA1 or BRAC2 gene or other less common mutations (for which screening may begin at age 25–30 y); because of a history of chest radiation therapy at least 8 years earlier and before age 30 y; or, based on models that predict the risk of being a mutation carrier, because of a strong family history that carries at least a 20%–25% lifetime risk of breast cancer (38). Dense breast tissue reduces mammographic sensitivity and specificity (39,40) but does not reduce the sensitivity of 99mTc-sestamibi–based imaging (41). Most women in the United States (42) are being notified of their breast density: there is increasing discussion of supplemental screening in women with dense breasts (heterogeneously dense or extremely dense by BI-RADS [Breast Imaging Reporting and Data System] categories) (43).
99mTc-sestamibi–based breast imaging using BSGI has been explored for screening high-risk women who cannot tolerate MRI (44), although radiation exposure is of greater concern in young women and in BRCA1-, BRCA2-, or TP53-mutation carriers with impaired DNA repair. More extensively, a Mayo Clinic group has explored using 99mTc-sestamibi and dual-head cameras for supplemental screening in women with dense breasts (45,46). As shown in Table 1, the sensitivity and specificity of MBI compare quite favorably with other methods of supplemental screening. In an analysis of induced costs from such screening, the Mayo Clinic group concluded that the cost per cancer diagnosis for mammography plus MBI was actually lower than that of mammography alone (47). As a result of these studies, across its multistate system, the Mayo Clinic is now offering biennial screening with MBI to women with dense breasts to supplement annual mammography or tomosynthesis. Monitoring outcomes will be important.
In addition to concerns about radiation exposure, patient preparation, licensing requirements, and patient throughput when a minimum of 40 min is required for the examination, there are other current limitations to nuclear medicine approaches to screening. The cancer detection rates from annual or biennial 99mTc-sestamibi–based screening have not been reported; and there is no method of direct biopsy for findings detected on MBI, although stereotactic BSGI-guided biopsy is available. The logistics and radiation dose incurred when using 18F-FDG preclude interest in breast PET for screening, although direct PEM-guided biopsy is readily performed (48).
STANDARDIZED INTERPRETIVE CRITERIA AND QUANTIFICATION
Lexicons to describe findings on γ-camera imaging (49) and on breast PET (50) have been developed, and both have shown ease of use and substantial reproducibility for most terms (51,52). Whenever possible, the same terminology and format used in BI-RADS for mammography (43), ultrasound (53), and MRI (54) have been adopted for nuclear breast imaging, with assessments being scored on a 5-point scale: 1, negative; 2, benign; 3, probably benign; 4, suspicious; and 5, highly suggestive of malignancy. Nuclear breast imaging must be interpreted together with current mammography, any other prior breast imaging, and a thorough biopsy history; as a result, interpretation should be performed by, or at least include, radiologists who meet experience requirements for breast imaging. Indeed, in the multicenter PEM–MRI trial (10), 7 of 82 (9%) breasts with additional tumor foci were identified only on review of mammography and ultrasound. Any focal radiotracer uptake should be viewed with suspicion except when there is a known benign correlate such as a mammographically stable intramammary or axillary lymph node; a normal nipple; or a known area of fat necrosis, fibroadenoma, abscess, or other benign process. The use of a “probably benign” assessment on nuclear breast imaging is discouraged because the rate of malignancy approaches 5% for such findings (50), exceeding the accepted 2% threshold used in other breast imaging (55).
For 99mTc-sestamibi–based imaging, although a method of quantification with dual-head systems has been described (56), intensity of uptake is not routinely quantified with current software. Subcutaneous fat is used as the reference standard, with background-parenchyma uptake and lesion uptake described qualitatively as photopenic (less than subcutaneous fat), mild, moderate, or marked. Background uptake of 99mTc-sestamibi can be greater in the luteal phase of the menstrual cycle (57) and in postmenopausal women using hormonal replacement therapy (58) and tends to be greater in dense breasts (57,58), although 72 of 164 (44%) women with extremely dense parenchyma showed photopenic background in the Mayo series (58). As with MRI, in which increased background-parenchyma enhancement in the contralateral breast is independently associated with breast cancer (59,60), it is possible that increased background-parenchyma uptake of 99mTc-sestamibi could predict an increased risk of breast cancer, but further study is warranted.
On breast PET, uptake can be quantified by drawing a region of interest with comparison typically to a region of normal background parenchyma. Background-parenchyma uptake of 18F-FDG increases with increasing breast density, with about a twofold average difference between extremely dense breasts and fatty breasts (61). In one series, the uptake ratios of tumor to normal tissue on PEM were higher than those seen on PET/CT for small tumors (9), but quantification is generally similar for PEM and PET/CT (8,9) despite the lack of attenuation correction on PEM.
TUMOR SUBTYPES
ILC can be notoriously difficult to identify on mammography because it tends to grow in single-file columns of cells rather than forming a discrete mass or causing distortion, and calcifications are quite rare with ILC (62). ILC is generally seen well on 99mTc-sestamibi–based breast imaging (63). Uptake is less intense in ILC than in invasive ductal carcinoma (IDC); and in one large MBI series (6) sensitivity to ILC foci, at 38 of 55 (69%), was lower than that to IDC, at 210 of 227 (92.5%; P = 0.0062). The sensitivity of 99mTc-sestamibi–based imaging to DCIS is estimated at 60%–91% on the basis of small series (64) and unpublished results (Deborah Rhodes, written communication, 2014).
Quantitative uptake of 18F-FDG is lower in ILC and DCIS than in IDC (3,8). PEM sensitivity to newly detected DCIS foci has been found to be 23 of 56 (41%) in women with cancer elsewhere who underwent mammography and MRI (10).
High-grade IDC is seen well on nuclear breast imaging. Quantitatively higher 18F-FDG uptake has been observed with estrogen receptor–negative, HER2-receptor–positive tumors and triple-negative tumors than with estrogen receptor–positive disease (65–67). Higher 18F-FDG uptake in the primary breast cancer is predictive of an increased risk of recurrence (68).
FUTURE
For γ-camera breast imaging to be widely adopted for screening, at a minimum there should be proof that women receiving MBI have lower rates of clinically detected cancers in the interval between screenings (interval cancers) and lower rates of advanced-stage breast cancer than a control group not receiving MBI. Ideally, there also would not be increased rates of low- and intermediate-grade DCIS in the MBI group. Incidence screening results are unknown. Direct MBI-guided breast biopsy is needed so that MRI is not required to sort out abnormalities seen only on MBI; this process is under development and does exist for BSGI systems.
One area of great, but largely unrecognized, potential for nuclear breast imaging is in predicting and assessing response to primary (neoadjuvant) chemotherapy (Figs. 2 and 3). In a pilot study of 19 patients imaged sequentially with MBI, a decrease in the tumor-to-background ratio 3–5 wk after the initial round of chemotherapy was shown to predict response to treatment; larger studies are in progress (69). Lack of a substantial decrease in 18F-FDG uptake after 2 cycles of chemotherapy has been shown to predict residual tumor at the conclusion of chemotherapy and a high risk of recurrence for triple-negative breast cancer (70).
Promising work is ongoing with many other radiotracers not currently approved for clinical use in the United States, including 99mTc-tetrofosmin (71,72) and a 99mTc-labeled angiogenesis agent (NC100692, a peptide that binds to receptors of the integrin class) (73). 18F-fluoroestradiol binds to estrogen receptors and is being explored for characterization of metastatic lesions outside the liver in women with estrogen receptor–positive breast cancer (74); 18F-fluoroestradiol may have a role in predicting early response to treatment in the breast, with initial flare in responders (75). 3′-deoxy-3′-18F-fluorothymidine (18F-FLT) is an analog of thymidine that is phosphorylated by thymidine kinase, the activity of which is increased in rapidly dividing cells. 18F-FLT cannot be incorporated into DNA but rather is trapped in the cytoplasm. 18F-FLT shows high liver and bone uptake but low brain uptake and can be used to assess for brain metastases. Use of 18F-FLT in early studies assessing treatment response has been reviewed by Kong et al. (76).
SUMMARY
Standardized technique and interpretive criteria are in place for 99mTc-sestamibi– and 18F-FDG–based dedicated breast imaging. Evidence from women with cancer and screening studies to date suggests that 99mTc-sestamibi–based breast imaging using dedicated dual-head detector systems has sensitivity comparable to that of MRI but higher specificity. Results to date may be biased by the lack of direct MBI-guided biopsy and the resulting need for MRI of equivocal findings. Importantly, direct stereotactic biopsy is available for BSGI. Efforts are ongoing to reduce radiotracer dose and time to complete imaging, but 99mTc-sestamibi–based imaging has been shown to be cost-effective for screening.
Breast PET also shows comparable sensitivity to MRI but higher specificity, and direct PEM-guided biopsy is available. Both 99mTc-sestamibi– and 18F-FDG–based breast imaging benefit from the patient fasting and sitting in warm conditions after the injection of a radiotracer, but 99mTc-sestamibi imaging can begin essentially immediately after injection, whereas breast PET requires a minimum of a 1-h circulation time. PET tracers are high-energy and require substantial shielding. Breast PET can be used in assessing the local extent of breast cancer, and single-center results show some benefit from 99mTc-sestamibi–based imaging, but neither has been shown to reduce positive margins or recurrence; further study is needed. Both methods show promise in assessing the response to primary chemotherapy. Comparison of both methods with contrast-enhanced mammography (77,78) is needed.
Importantly, although specificity appears favorable with nuclear breast imaging, false-positives are not uncommon and can include fat necrosis, fibroadenomas, papillomas, normal lymph nodes, abscesses, phyllodes tumors, lobular carcinoma in situ, atypical ductal hyperplasia, and sometimes even fibrocystic change (61,79). Therefore, it is important to correlate findings with mammography, other breast imaging, and prior biopsy history and therefore to include breast imaging specialists in interpreting nuclear breast imaging studies. Percutaneous biopsy should be performed of any suspicious findings on nuclear breast imaging that would change patient management, and a “probably benign” assessment should be avoided because of a malignancy rate exceeding 2%.
DISCLOSURE
No potential conflict of interest relevant to this article was reported.
Acknowledgments
I thank Carrie Hruska, PhD, Michael O’Connor, PhD, and the rest of Team Mayo at the Mayo Clinic for many helpful discussions, images, and comments regarding MBI. I also thank Mary K. Hayes, MD, for supplying PEM images.
- © 2016 by the Society of Nuclear Medicine and Molecular Imaging, Inc.
REFERENCES
- 1.↵
- 2.↵
- 3.↵
- 4.↵
- 5.↵
- 6.↵
- 7.↵
- 8.↵
- 9.↵
- 10.↵
- 11.↵
- 12.↵
- 13.↵
- 14.↵
- 15.↵
- 16.↵
- 17.↵
- 18.↵
- 19.↵
- 20.↵
- 21.↵
- 22.↵
- 23.↵
- 24.↵
- 25.↵
- 26.↵
- 27.↵
- 28.↵
- 29.↵
- 30.↵
- 31.↵
- 32.↵
- 33.↵
- 34.↵
- 35.↵
- 36.↵
- 37.↵
- 38.↵
- 39.↵
- 40.↵
- 41.↵
- 42.↵
- 43.↵
- 44.↵
- 45.↵
- 46.↵
- 47.↵
- 48.↵
- 49.↵
- 50.↵
- 51.↵
- 52.↵
- 53.↵
- 54.↵
- 55.↵
- 56.↵
- 57.↵
- 58.↵
- 59.↵
- 60.↵
- 61.↵
- 62.↵
- 63.↵
- 64.↵
- 65.↵
- 66.
- 67.↵
- 68.↵
- 69.↵
- 70.↵
- 71.↵
- 72.↵
- 73.↵
- 74.↵
- 75.↵
- 76.↵
- 77.↵
- 78.↵
- 79.↵
- Received for publication August 25, 2015.
- Accepted for publication October 28, 2015.