Abstract
The lower detection rate of 18F-FDG PET/MRI than 18F-FDG PET/CT regarding small lung nodules should be considered in the staging of malignant tumors. The purpose of this study was to evaluate the outcome of these small lung nodules missed by 18F-FDG PET/MRI. Methods: Fifty-one oncologic patients (mean age ± SD, 56.6 ± 14.0 y; 29 women, 22 men; tumor stages, I [n = 7], II [n = 7], III [n = 9], IV [n = 28]) who underwent 18F-FDG PET/CT and subsequent 18F-FDG PET/MRI on the same day were retrospectively enrolled. Images were analyzed by 2 interpreters in random order and separate sessions with a minimum of 4 wk apart. A maximum of 10 lung nodules was identified for each patient on baseline imaging. The presence, size, and presence of focal tracer uptake was noted for each lung nodule detected on 18F-FDG PET/CT and 18F-FDG PET/MRI using a postcontrast T1-weighted 3-dimensional gradient echo volume-interpolated breath-hold examination sequence with fat suppression as morphologic dataset. Follow-up CT or 18F-FDG PET/CT (mean time to follow-up, 11 mo; range, 3–35 mo) was used as a reference standard to define each missed nodule as benign or malignant based on changes in size and potential new tracer uptake. Nodule-to-nodule comparison between baseline and follow-up was performed using descriptive statistics. Results: Out of 134 lung nodules found on 18F-FDG PET/CT, 18F-FDG PET/MRI detected 92 nodules. Accordingly, 42 lung nodules (average size ± SD, 3.9 ± 1.3 mm; range, 2–7 mm) were missed by 18F-FDG PET/MRI. None of the missed lung nodules presented with focal tracer uptake on baseline imaging or follow-up 18F-FDG PET/CT. Thirty-three out of 42 missed lung nodules (78.6%) in 26 patients were rated benign, whereas 9 nodules (21.4%) in 4 patients were rated malignant. As a result, 1 patient required upstaging from tumor stage I to IV. Conclusion: Although most small lung nodules missed on 18F-FDG PET/MRI were found to be benign, there was a relevant number of undetected metastases. However, in patients with advanced tumor stages the clinical impact remains controversial as upstaging is usually more relevant in lower stages.
Metastatic spread to the lungs commonly implies a higher tumor stage in patients with malignant diseases often requiring a change of therapy regimen and ultimately decreasing chances of survival (1). Hence, the early detection of potential pulmonary metastases is essential. Because of a higher accuracy in TNM staging than either PET or CT alone (2), integrated 18F-FDG PET/CT has been implemented in the staging routine of a growing number of tumor entities and is regarded as the standard of reference by many authors (3–5). MRI, on the other hand, offers a sensitivity superior to that of CT concerning infiltration of the primary tumor into adjacent organs and detection of metastasis to parenchymatous organs such as the liver, brain, or kidneys (6–8). In addition, MRI provides valuable functional information from quantitative and multiparametric imaging (9). With the introduction of fully integrated PET/MRI systems, PET and MRI can be used synergistically in a single modality, taking advantage of an exact correlation between 18F-FDG–avid lesions and the anatomic details obtained by MR images (10–12). Yet, regarding lung imaging, even when using high spatial resolution T1-weighted (T1w) 3-dimensional (3D) gradient echo (GRE) sequences (e.g., volume-interpolated breathhold-examination [VIBE]) recommended for the identification of small pulmonary nodules (13–15), PET/MRI seems to be outmatched by PET/CT (16). Underlying reasons are found in PET/MRI’s lower sensitivity regarding 18F-FDG–negative lesions (17,18) because of a low proton density in aerated lungs, fast decay of signal caused by susceptibility artifacts at air–tissue boundaries, and motion artifacts caused by breathing and cardiac pulsation. Bearing in mind the increasing use of whole-body PET/MRI, its lower detection rate of small lung nodules potentially representing metastases should be considered in the staging of malignant tumors. Hence, the purpose of this study was to evaluate the outcome of small lung nodules detected on 18F-FDG PET/CT but missed by 18F-FDG PET/MRI in terms of their presumed dignity.
MATERIALS AND METHODS
Patients and Inclusion Criteria
Patients with a known malignancy (Table 1) who underwent 18F-FDG PET/MRI including a postcontrast, fat-suppressed (fs) T1w VIBE sequence of the thorax after clinically indicated 18F-FDG PET/CT for tumor staging on the same day were retrospectively enrolled in this study. A follow-up CT or 18F-FDG PET/CT not less than 3 mo after the initial examination had to be available for outcome evaluation. According to these criteria, 51 patients (mean age ± SD, 56.6 ± 14.0 y; 29 women, 22 men) between May 2012 and December 2014 were eligible for retrospective analysis. The study cohort comprised patients with tumor stages I (n = 7), II (n = 7), III (n = 9), and IV (n = 28). Eight out of 51 patients had a recurrent malignancy. The study was approved by the local ethics committee, and all subjects signed an informed consent form.
PET/CT Imaging
Patients underwent whole-body (i.e., head to upper thighs) 18F-FDG PET/CT on a Biograph mCT or Biograph Duo (Siemens Healthcare GmbH) 61 ± 8 min after intravenous injection of a mean activity of 260 ± 60 MBq of 18F-FDG, depending on their body weights. At injection time, blood glucose levels needed to be below 150 mg/dL. Twenty-three patients were examined using the low-dose CT technique. In 28 patients, full-dose CT scans were obtained. Full-dose CT scans were acquired after intravenous injection of a contrast agent (Imeron 300; Bracco Imaging Deutschland GmbH). In all full-dose PET/CT scans, an additionally acquired dedicated lung scan applying a sharp b 70 or b 90 kernel in deep inspiration was used for the detection of lung nodules. CT images were displayed on lung window setting and using a slice thickness of 2 mm. For dose-reduction purposes, the manufacturer-supplied solutions CareKV and CareDose 4D (Siemens Healthcare GmbH) were used for both full- and low-dose PET/CT scans (presets: 120 kV, 210mAs and 120 kV, 40 mAs, respectively). PET acquisition time using static frames varied from 2 to 3.5 min per bed position. Iterative reconstruction (3 iterations, 21 subsets) with a gaussian filter of 4.0 mm was applied. In general, the slice thickness of reconstructed PET images was 3 mm. For attenuation correction of the PET dataset, the portal venous phase of full-dose CT scans and low-dose CT data in low-dose scans were used.
PET/MR Imaging
Whole-body (i.e., head to upper thighs) 18F-FDG PET/MRI using Flex coils (Siemens Healthcare AG GmbH) was performed on a 3-T Biograph mMR (Siemens Healthcare AG). The average delay after intravenous tracer injection was 117 ± 29 min. For morphologic assessment of the lungs, a transverse T1w fs VIBE sequence (repetition time, 4.08 ms; echo time, 1.51 ms; slice thickness, 3.5 mm; field of view, 400 × 300 mm; matrix size, 512 × 307.2; voxel size, 1.3 × 0.8 × 3.5 mm) after contrast administration (Dotarem; Guerbet GmbH) was acquired. Attenuation correction was based on a coronal 3D Dixon-VIBE sequence (repetition time, 3.6 ms; echo time 1, 1.23 ms; echo time 2, 2.46 ms; slice thickness, 3.12 mm; field of view, 500 × 328 mm; matrix size, 192 × 121 mm; voxel size, 4.1 × 2.6 × 3.1 mm). In general, PET acquisition time was 3 min per bed position. PET images were acquired in list-mode and reconstructed using an iterative algorithm of ordered-subsets expectation maximization (3 iterations and 21 subsets). A gaussian filter of 4.0 mm was applied. In general, the slice thickness of reconstructed PET images was 3 mm.
Follow-up Imaging
Chest CT (n = 32) and whole-body 18F-FDG PET/CT (n = 19) scans were considered appropriate for follow-up imaging and categorization of dignity of the missed lung nodules. Follow-up CT examinations were conducted on 1 of the following CT scanners: Definition AS+, Definition Flash, and Definition Force (Siemens Healthcare GmbH). Follow-up 18F-FDG PET/CT examinations were performed on the aforementioned Biograph mCT or Biograph Duo (Siemens Healthcare AG). Baseline and follow-up examinations had to be a minimum of 3 mo apart. For follow-up imaging, the most recent CT or 18F-FDG PET/CT scan was selected. The average follow-up interval was 11 ± 8 mo (range, 3–35 mo). Table 2 gives an overview of patient numbers and tumor types in different follow-up intervals.
Image Analysis
Images were evaluated on a dedicated OsiriX Workstation (Pixmeo SARL). Baseline 18F-FDG PET/CT and 18F-FDG PET/MRI datasets as well as follow-up CT or 18F-FDG PET/CT scans were analyzed by 2 independent radiologists with 3 and 5 y of experience in hybrid imaging. Interpreters were aware of patients’ diagnosis. Any discrepancies between the 2 interpreters were resolved in a subsequent consensus reading. PET images were reviewed with and without attenuation correction of the PET data to prevent false-positive findings caused by attenuation-correction artifacts. Baseline 18F-FDG PET/CT and 18F-FDG PET/MRI were assessed in random order and in separate sessions with a minimum of 4 wk apart to avoid recognition bias. Morphologic T1w fs VIBE and CT images and PET from PET/CT and PET from PET/MRI were analyzed separately and as fused datasets. The presence, size (i.e., longitudinal axis diameter on transverse images) on the CT component of 18F-FDG PET/CT, and presence of focal tracer uptake above surrounding background was noted for each lung nodule detected on 18F-FDG PET/CT and on 18F-FDG PET/MRI. Initially, a maximum of 10 lung nodules was identified for each patient on baseline imaging, beginning with the right upper lobe proceeding to the left lower lobe. Each lung nodule found on 18F-FDG PET/CT but not detected on 18F-FDG PET/MRI was rated a missed nodule. In a separate session, the size of each missed lung nodule was reassessed on follow-up CT or 18F-FDG PET/CT, respectively. In cases, in which a PET/CT scan was available for follow-up, missed nodules were also analyzed in regard to a potential new 18F-FDG uptake.
Standard of Reference
Because histopathologic correlation could not be obtained for any of the missed lung nodules, the standard of reference to determine the outcome of a missed lung nodule was based on change in nodule size between baseline and follow-up imaging (chest CT or 18F-FDG PET/CT) and presence of new 18F-FDG avidity of a lung nodule on follow-up PET/CT.
Potential systemic therapy was addressed as follows: nodules presenting smaller on follow-up were rated malignant if the patient underwent systemic cancer therapy during the time of follow-up (i.e., therapy effect) or rated benign provided no cancer therapy had been administered. Nodules presenting constant in size were rated benign, regardless of whether the patient received cancer therapy or not. This procedure for nodules constant in size was decided on to avoid overestimation of the number of metastases while acknowledging that the overall number of metastases may have been underestimated by this approach. Nodules with an increased size on follow-up or a new 18F-FDG avidity on follow-up PET/CT were rated malignant (Fig. 1).
Statistics
SPSS Statistics 22 (IBM) was used for statistical analysis. All data are mean ± SD. Descriptive analysis was used for the calculation of the patients’ characteristics and for follow-up nodule-to-nodule comparison. Similar to previous reports (19), reliability between interpreters was performed in a descriptive way.
RESULTS
There was 97% agreement between both interpreters on all lung nodules detected by 18F-FDG PET/CT and 18F-FDG PET/MRI, with only 3% needing a consensus reading, in which agreement was found in all cases. In 51 patients, 134 lung nodules were found on 18F-FDG PET/CT (mean size, 12.9 ± 15.7 mm; range, 2–98 mm; lung nodules per patient, 0–10). 18F-FDG PET/MRI detected 92 of these lung nodules, signifying a detection rate of 68.7%. 18F-FDG PET/MRI and 18F-FDG PET/CT detected concordant numbers of lung nodules in 21 patients, 9 of which did not exhibit any lung nodules. In the other 30 patients, 42 lung nodules were missed on 18F-FDG PET/MRI but detected on 18F-FDG PET/CT (Fig. 2). The percentage of missed nodules with respect to the total number of lung nodules was 31.3%. The mean size of the missed nodules was 3.9 ± 1.3 mm (range, 2–7 mm) (Fig. 3). On average, 0.8 pulmonary nodules per patient were missed on 18F-FDG PET/MRI compared with 18F-FDG PET/CT. None of the missed lung nodules presented with focal 18F-FDG uptake above the surrounding background on baseline attenuation-corrected or non–attenuation-corrected PET images. Nodule-to-nodule follow-up assessment revealed that 71.4% (30/42) of missed lung nodules remained constant in size, 26.2% (11/42) of missed lung nodules decreased in size or completely resolved, and 2.4% (1/42) of lung nodules increased in size on follow-up CT or 18F-FDG PET/CT compared with baseline imaging. None of the missed lung nodules demonstrated a new tracer uptake on follow-up 18F-FDG PET/CT. Seventy percent (21/30) of the patients with missed lung nodules received systemic or local cancer therapy in the interval of baseline PET/MRI or PET/CT and follow-up CT or 18F-FDG PET/CT, whereas 30% (9/30) of the patients were not subjected to cancer therapy. According to the reference standard, 78.6% (33/42) of the missed pulmonary nodules in 26 out of 30 (86.7%) patients were rated benign, whereas 21.4% (9/42) of the missed pulmonary nodules in 4 out of 30 (13.3%) patients were rated malignant (Figs. 4 and 5). Because of the occurrence of a new metastatic spread to the right lung, upstaging from tumor stage I to tumor stage IV was required in 1 of the 4 patients. This patient was initially diagnosed with a T1 N0 M0 non–small cell lung cancer in the contralateral lung. As a consequence, the contralateral lung metastasis demanded restaging to M1a (20). Because distant metastases had already been diagnosed in the other 3 patients, no adjustment of TNM staging was necessary in these cases.
DISCUSSION
The intention of this study was to evaluate the outcome of small lung nodules detected on 18F-FDG PET/CT, which were not identifiable on 18F-FDG PET/MRI in oncologic patients. According to the reference standard, most (78.6%) of these missed lung nodules were rated as benign but there was a small but relevant number (21.4%) of undetected lung nodules that were suggestive of metastases.
The detection of small lung nodules is a key clinical demand in cancer staging because potential metastatic spread can have far-reaching effects on therapy and patient survival (1,13,20). Today, whole-body 18F-FDG PET/CT is widely available and used not only to identify lung nodules but also to discriminate malignant from benign pulmonary masses, enabling a comprehensive tumor staging in a 1-stop shop examination (21–23). On the other hand, the latest transition from a mere research modality into clinical practice raised issues regarding the eligibility of 18F-FDG PET/MRI to detect small pulmonary nodules compared with CT or PET/CT as the modality of choice for lung imaging.
None of the missed lung nodules, including those that were suggestive of metastasis, presented with a focal tracer uptake on the PET component of 18F-FDG PET/CT or 18F-FDG PET/MRI. However, given the fact that 21.4% of non-18F-FDG–avid nodules were likely malignant, our results indicate that PET negativity is not suitable to rule out malignancy in small lung lesions. Studies by Yilmaz et al., Farid et al., and Khalaf et al. demonstrating a high prevalence of PET negativity in malignant nodules less than 1 cm have corroborated these results (24–26). False-negative findings might have been attributed to breathing motion, low metabolic activity (27), small nodule size, and the limited spatial resolution of PET, leading to a substantial underestimation of the true activity within the lesion (8). Moreover, differences between 18F-FDG PET/MRI and 18F-FDG PET/CT regarding PET reconstruction parameters—for instance, slice thickness—can influence nodule detectability of the PET component. With a slice thickness of around 3 mm for PET from 18F-FDG PET/MRI and from 18F-FDG PET/CT, both PET components can be regarded equally prone to partial-volume effects in our study. However, considering that a relevant proportion of nodules were 3 mm or less in size partial-volume effects might have been a relevant factor.
From the variety of clinically used MR sequences of the thorax, T1w 3D GRE sequences such as VIBE offer the highest detection rates of small pulmonary nodules and thus served as the only morphologic dataset in this study (13–15,28–30). Performed in a comparatively short imaging time (∼20 s), they allow for an acquisition during breath-hold with accurate fusion of simultaneously acquired PET signals (31). Further advantages encompass high spatial resolution offering high-quality depiction of lung anatomy and a lower rate of artifacts compared with 2-dimensional GRE sequences (32). However, comparative studies examining the sensitivity of MRI and CT for lung nodules less than 1 cm found that MRI still lags behind CT, with detection rates ranging between 80% and 90% (14,17,30,33). Signal loss because of cardiac pulsation and respiration, susceptibility artifacts arising from multiple air–tissue interfaces, and low proton density in aerated lungs are known drawbacks hampering the identification of small pulmonary nodules (34). With a detection rate of approximately 70% for lung nodules less than 0.7 cm, our results seem to support these prior investigations.
Out of 4 patients with lung metastases missed by 18F-FDG PET/MRI, 2 patients had lung cancer and 2 had breast cancer, entities renowned for a high potential of developing lung metastases and thus further substantiating the suspicion of malignancy (35,36). However, because in 3 of the 4 patients distant metastases had already been diagnosed, a missed metastatic spread to the lungs should not be expected to have high therapeutic significance. A clinically relevant upstaging from tumor stage I to IV was required in only 1 of the patients. For all other patients with lung nodules missed by 18F-FDG PET/MRI, no adjustment of TNM staging was required. Given that most missed lung nodules proved benign and upstaging seemed to be necessary only in exceptional cases, from a clinical standpoint our results may generally endorse the use of whole-body 18F-FDG PET/MRI in oncologic patients. Nevertheless, this is partly due to study-specific patient characteristics, which mainly included patients with very advanced tumor stages. The evaluation of consequences on treatment decisions and prognosis might be addressed in future projects.
This study has some limitations. One limitation was the small and heterogeneous patient cohort comprising different tumor entities with a varying degree of metastatic potential and 18F-FDG avidity. The use of 2 different PET/CT protocols (1 with additional low-dose CT of the chest in maximal inspiration, 1 without) may have led to underestimation of the overall number of pulmonary nodules on the reference standard. Adding a low-dose CT in maximal inspiration should be a prerequisite when designing a follow-up study addressing a potential effect of missed pulmonary metastases on patient management. Also, the slice thickness of the CT component of baseline 18F-FDG PET/CT was 2 mm, whereas the slice thickness of T1w fs VIBE of 18F-FDG PET/MRI was 3.5 mm. Bearing in mind that missed lung nodules measured around 4 mm, the larger slice thickness of T1w fs VIBE might have contributed to the inferior detection rate of 18F-FDG PET/MRI. However, T1w VIBE images of the chest are not yet feasible with a lower slice thickness than 3.5 mm. It could be assumed that the MRI component might have performed better using a lower slice thickness. However, technical requirements are yet to be introduced into clinical routine. It also has to be considered that a minor part of lung nodules presenting smaller in patients undergoing cancer therapy might still have been of infectious origin. In this case, the number of missed metastases may have been overestimated. Furthermore, some of the nodules that remained constant in size on follow-up might have represented metastases with partial response to systemic or local cancer therapy. Although the mean follow-up interval was 11 mo, many patients had follow-up intervals of less than 6 mo. In slowly growing lesions, one might argue that it was difficult to exclude false-negative readings. Lacking a histopathologic reference standard, these limitations have to be kept in mind when interpreting the results of this study.
CONCLUSION
Because of its lower sensitivity in detecting PET-negative lung nodules, whole-body 18F-FDG PET/MRI using a T1w VIBE sequence misses a relevant proportion of small lung nodules in oncologic patients. Although most missed lung nodules proved to be benign, there are a considerable number of metastases among those missed nodules. However, in patients with very advanced tumor stages the clinical impact remains controversial because potential upstaging is usually more relevant in lower tumor stages.