Abstract
Today, PET and SPECT tracers cannot be imaged simultaneously at high resolutions but require 2 separate imaging systems. This paper introduces a Versatile Emission Computed Tomography system (VECTor) for radionuclides that enables simultaneous submillimeter imaging of single-photon and positron-emitting radiolabeled molecules. Methods: γ-photons produced both by electron–positron annihilation and by single-photon emitters are projected onto the same detectors by means of a novel cylindric high-energy collimator containing 162 narrow pinholes that are grouped in clusters. This collimator is placed in an existing SPECT system (U-SPECT-II) with 3 large-field-of-view γ-detectors. From the acquired projections, PET and SPECT images are obtained using statistical image reconstruction that corrects for energy-dependent system blurring. Results: For PET tracers, the tomographic resolution obtained with a Jaszczak hot rod phantom was less than 0.8 mm, and 0.5-mm resolution images of SPECT tracers were acquired simultaneously. SPECT images were barely degraded by the simultaneous presence of a PET tracer, even when the activity concentration of the PET tracer exceeded that of the SPECT tracer by up to a factor of 2.5. Furthermore, we simultaneously acquired fully registered 3- and 4-dimensional multiple functional images from living mice that, in the past, could be obtained only sequentially. Conclusion: High-resolution complementary information about tissue function contained in SPECT and PET tracer distributions can now be obtained simultaneously using a fully integrated imaging device. These combined unique capabilities pave the way for new perspectives in imaging the biologic systems of rodents.
Imaging the distributions of radiolabeled molecules in small animals facilitates in vivo quantitative assessments of molecular mechanisms and the development of new pharmaceuticals and diagnostic tracers (1,2). In molecular imaging, SPECT and PET are key modalities, offering the ability to track and quantify radiolabeled biomarkers (radiotracers or radiopharmaceuticals) with detection sensitivity below the nano- or even picomolar range. Current small-animal SPECT scanners can be used to image one or multiple SPECT tracers in a single scan with image resolutions that have improved enormously over the last decade (3–7), well below 0.5 mm in 3 dimensions (8,9). In contrast, high-end commercial small-animal PET attains image resolutions of approximately 1 mm (10–15) and cannot distinguish between simultaneously injected tracers, because positron emission from different tracers results in γ-photon pairs with equal energy (511 keV). Recently, it was also shown that fast dynamic (total body) imaging is possible with small-animal SPECT based on stationary detectors (16,17). This ability makes SPECT, besides PET, an important tool in, for example, kinetic modeling.
Using complementary sets of radiolabeled molecules, PET and SPECT make it possible to visualize and quantify different aspects of biologic function. However, today’s scanners cannot take advantage of the entire complement of available radiotracers because, until now, these 2 modalities could not be combined into a single integrated system capable of imaging both classes of tracers simultaneously at high resolutions. Here, we present the first truly integrated submillimeter SPECT/PET platform that is designed to use a single set of detectors and a single user interface. In contrast to 2 stand-alone systems or 2 separate systems attached to one another in a series, simultaneous SPECT/PET acquisition provides perfectly aligned images in space and time of different tissue functions that can therefore be directly correlated. Furthermore, such a fully integrated SPECT/PET scanner not only saves costs, imaging time, training time, and floor space but also enables the imaging of PET and SPECT probes with a single dose of anesthesia.
Traditionally, PET has been based on a ring detector using dedicated electronics for the coincidence detection of pairs of annihilation photons. In contrast, SPECT systems use collimators to detect single γ-photons emitted by SPECT tracers. In several studies over the last decade, it has been shown that dedicated SPECT provides higher-resolution images of mice than dedicated PET does, because several image-degrading effects that play a role in small-animal PET are not present or are dramatically reduced with pinhole imaging (18). Therefore, we developed a novel collimation technique to enable separate imaging of the 2 high-energy γ-photons emitted by PET tracers simultaneously with photons emitted by SPECT tracers. This technique exploits clusters of pinholes (18–20), where each cluster samples the same field of view as a single traditional pinhole (Fig. 1A). Pinhole edge penetration, which would be a major problem were standard SPECT collimators to be used for high-energy annihilation photons, is dramatically reduced because each of the pinholes in a cluster has a smaller opening angle, thereby increasing the path length that a photon has to travel to penetrate the pinhole at a certain distance from the edge of the hole.
Integration of clustered-pinhole collimator into existing SPECT/CT platform. (A) Traditional pinhole with opening angle α and cluster of 4 pinholes with approximately same field of view and opening angle α/2. (B) Clustered-pinhole collimator optimized for imaging SPECT and PET tracers, into which mouse is placed. (C) Collimator mounted in SPECT/CT system.
Here we present and apply the experimental realization of submillimeter simultaneous SPECT/PET based on clustered pinholes. This prototype Versatile Emission Computed Tomography system (dubbed VECTor, and available through MILabs) is based on a specially designed cylindric collimator containing 48 clusters of 4 pinholes each (Fig. 1B) that was placed into an existing SPECT system with integrated CT (U-SPECT-II/CT [MILabs], Fig. 1C). Similar to the standard multipinhole SPECT collimators of this system, the clustered-pinhole collimator applies a focusing geometry; all pinhole clusters focus on a central scan volume. This design allows the count yield from a specific organ or a tumor to be increased (21) but also allows imaging of larger regions—up to (dynamic) whole-body mouse scans (17,22). We present a quantitative performance evaluation of VECTor using phantom data and show the results of simultaneously acquired multiple functional data obtained with positron and single-photon emitters in mice.
MATERIALS AND METHODS
Pinhole Geometry and System Description
The collimator that enables imaging of single-photon and positron emitters was integrated in the U-SPECT-II/CT system (8), which uses 3 large-field-of-view γ-detectors. The imaging platform has been dubbed VECTor. The tungsten collimator (shown in Fig. 1B) contains 48 clusters of four 0.7-mm-diameter pinholes (illustrated in Fig. 1A) placed in 4 rings. The pinholes in the inner 2 rings of the collimator have opening angles of 18°, whereas in the outer rings opening angles of 16° were chosen in order to achieve a more uniform coverage of the detectors with pinhole projections. Of all 192 clustered pinholes, 162 were being used for image reconstruction. The collimator has an inner diameter of 48 mm, the pinhole centers are placed at a diameter of 64 mm, and the collimator has a wall thickness of 43 mm.
All the pinhole clusters together observe a field of view that extends over the entire collimator tube diameter (48 mm) and has the shape of an hourglass (21). The average longitudinal length of the collimator field of view is 36 mm. A portion of the field of view, referred to as the central field of view, is sampled by all pinhole clusters simultaneously, and in that portion complete data acquisition is obtained without any translation of the bed. For the clustered-pinhole collimator used here, the central field of view had a diameter of approximately 12 mm and a longitudinal length of approximately 9 mm.
Image Reconstruction and Calibration
The scanning focus method was used for data acquisition (22). A pixel-based ordered-subsets expectation maximization algorithm (23) with 16 subsets was used for image reconstruction. The system matrix for the SPECT tracers was obtained using the method described by Van der Have et al. (24). The system matrix for 18F-labeled tracers was calculated using a ray-tracing code (25). Inputs to this code were the precise pinhole and detector positions and detector orientations, determined via point-source measurements (24), whereas pinhole axis orientation and the exact pinhole opening angle were estimated from a detector image of an extended source.
Scatter Correction
In all experiments, scatter was corrected using a triple-energy-window technique (26). All photopeak windows were set to a width of 15%. For 18F-labeled tracers, background windows of 400–450 keV and 550–600 keV were chosen; for 99mTc-labeled tracers, these windows were set to 100–115 keV and 155–170 keV; and for 123I-labeled tracers, background windows were set to 105–120 keV and 260–275 keV.
Evaluation of System Characteristics
Resolution for SPECT and PET Tracers
To test the resolution of VECTor for the most common PET and SPECT isotopes (18F and 99mTc) in a simultaneous acquisition, a miniature resolution phantom (Jaszczak phantom; Vanderwilt Techniques) with 6 sectors was manufactured. Each sector contains a set of equally sized capillaries with diameters of 0.85, 0.8, 0.75, 0.55, 0.5, and 0.45 mm (Fig. 2). The distance between capillary centers was equal to twice the capillary diameter. To test if the presence of a PET tracer influenced the resolution of SPECT images, we covered a wide range of SPECT-to-PET tracer concentration ratios in the resolution phantom scan by performing a scan over more than 10 half-lives of 18F (20 h of total scanning time), with the acquired data split into time frames of 1 h. At the beginning of the scan series, capillaries contained 29.8 MBq of 18F solution mixed with 10.4 MBq of 99mTc solution. In this paper, we show SPECT/PET images of the scan starting at 2 h (14 MBq of 18F and 8.3 MBq of 99mTc solution); other images from this scan series are provided in the supplemental data (available online at http://jnm.snmjournals.org).
Simultaneously acquired images of SPECT and PET tracers in phantom with 6 segments with capillary diameters of 0.85, 0.80, 0.75, 0.55, 0.50, and 0.45 mm. At start of this scan, capillaries contained 14 MBq of 18F solution and 8.3 MBq of 99mTc solution. Phantom was imaged for 60 min. All images shown had slice thickness of 3.2 mm.
Phantom with Cylindric Compartments
Some amount of PET tracer may be expected to “leak” into the SPECT images, because the 511-keV annihilation photons produced by the PET tracer can down-scatter into the SPECT photopeak. This effect can lead to overestimation of the amount of SPECT tracer in a certain volume of interest where PET tracer is also present. To test if such an effect occurred, we used a cylindric phantom (25 mm in diameter, 25 mm in length) containing 2 smaller cylindric compartments (7 mm in diameter, 12 mm in length) (Supplemental Fig. 2A). The large cylindric volume was filled with water. The left small cylinder was filled with a mix of approximately 20 MBq of 99mTc and 50 MBq of 18F solution, and the right one was filled with approximately 19 MBq of 99mTc solution. By comparing estimated amounts of SPECT tracers in the left and right compartments over time, we investigated if the PET tracer in the left compartment hindered accurate estimation of the amount of SPECT tracer in that same compartment. In reality, this SPECT-to-SPECT tracer ratio remains constant during the scan, but if the presence of PET tracer influences quantification of SPECT tracer in the left compartment this ratio is expected to become time-dependent. (Because the PET tracer decays much more rapidly than the SPECT tracer, its effect on SPECT quantification will also diminish over time.) To this end, we scanned for 14 h with data split into 60-min scans. Uptake of tracer in each compartment was determined by drawing a cylinder around the compartments with a diameter of 10 mm, larger than the cylinders themselves, to avoid partial-volume effects. SPECT and PET images were scaled to the same relative intensities using calibration factors for 99mTc and 18F that were determined by scanning a small source filled with a known quantity of 99mTc or 18F and then applying attenuation correction (27).
Sensitivity
A small source was made by enclosing 6 μL of 18F or 99mTc solution in a cylinder with a diameter of 2 mm. Sensitivity profiles were acquired by moving this source over 3 mutually perpendicular axes that crossed in the center of the collimator. The sensitivity was defined as the ratio of the number of detected γ-photons to the number of (positron or γ) emissions. Sensitivity was determined for photopeak windows with widths of 30%, 25%, 20%, and 15%.
Animal Studies
Animal studies were performed in accordance with the Dutch Law on Animal Experimentation and protocols approved by the Animal Research Committee of the University Medical Center Utrecht.
Mouse Heart Scan
A 29-g C57Bl6 mouse was anesthetized with isoflurane and then injected intravenously with 130 MBq of 99mTc-tetrofosmin and 24 MBq of 18F-FDG. A 60-min SPECT/PET acquisition was obtained, starting 10 min after injection.
Mouse Brain Scan
A 23-g C57Bl6 mouse, which had been kept fasting for 12 h, was injected intraperitoneally with 40 MBq of 18F-FDG and 30 MBq of 123I-N-ω-fluoropropyl-2β-carbomethoxy-3β-(4-iodophenyl)nortropane (123I-FP-CIT). The mouse was anesthetized with isoflurane 45 min after radionuclide injection. A 60-min SPECT/PET acquisition was obtained, starting 105 min after injection.
Mouse Bone Scans
A 33-g C57Bl6 mouse was anesthetized with isoflurane and injected intravenously with 220 MBq of 99mTc-hydroxymethylenediphosphonate (99mTc-HDP) and 60 MBq of 18F-fluoride. A 60-min whole-body SPECT/PET acquisition was obtained, starting 30 min after injection.
A 27-g C57B16 mouse was anesthetized with isoflurane and injected intravenously with 130 MBq of 99mTc-HDP and 85 MBq of 18F-fluoride. A 30-min SPECT/PET acquisition of the lumbar spine and pelvis was obtained, starting 60 min after injection.
RESULTS
Evaluation of System Characteristics
Resolution for SPECT and PET Tracers
SPECT and PET images of the resolution phantom filled with a mix of 99mTc and 18F solution (SPECT-to-PET concentration ratio, ∼1:1.7) are shown in Figure 2. Rods of 0.5 mm can be distinguished in the 99mTc image, and 0.8-mm rods are still clearly visible for 18F.
Supplemental Figure 1 shows SPECT images of the same phantom with different SPECT-to-PET tracer concentration ratios. Even when the concentration of the PET tracer exceeded that of the SPECT tracer by a factor of 2.9, 0.5-mm rods were still visible in the SPECT images, demonstrating that the resolution of SPECT images was hardly affected by the simultaneous presence of a PET tracer. Such a degrading effect might be expected to occur because of down-scatter of PET annihilation photons into the SPECT photopeak.
Phantom with Cylindric Compartments
The ratio of the amount of SPECT tracer (99mTc) in the left and right compartments of the water-filled mouse-sized phantom (Supplemental Fig. 2A) is shown in Supplemental Figure 2B as a function of the relative amount of PET tracer (18F) that was also present in the left compartment. Because all data points result from multiple time frames of a single scan, the true SPECT-to-SPECT tracer ratio is constant. However, leakage of PET tracer distribution into the SPECT images would affect this ratio, an effect that is expected to decrease over time because of the relatively rapid decay of the PET tracer. The experimental results demonstrated that, even though the SPECT-to-PET tracer concentration ratio in the left compartment varied widely during the scan (from 1:2.5 to 12:1), the SPECT-to-SPECT tracer ratio of the 2 cylinders did not vary much. In fact, this ratio differed by only 3.8% between the first and last time points of the scan. Furthermore, SPECT image slices and profiles showed no activity in the water-filled area (Supplemental Fig. 2B).
Sensitivity
The peak sensitivity of the system for 99mTc and 18F for different photopeak window widths is provided in Table 1. For a photopeak window width of 30%, 99mTc and 18F peak sensitivities were 0.25% and 0.31%, respectively. The system sensitivity differed only slightly for photopeak windows of 25%, 20%, and 15%. In Supplemental Figure 3, sensitivity profiles are provided.
Measured Peak Sensitivity for 99mTc and 18F
Taken together, these system tests demonstrated that PET and SPECT tracers could be imaged simultaneously with excellent resolutions and minimal cross-talk.
Simultaneous SPECT/PET Images in Living Mice
To determine whether it would be possible to obtain highly detailed, simultaneous SPECT and PET images of complex biologic systems in mice, we acquired in vivo heart, brain, and bone images with SPECT and PET tracer molecules. All the tracer combinations we tested were chosen with specific biologic applications in mind—applications in which a direct correlation of multiple processes is of importance. Naturally, these combinations represent only a small subset of the imaging studies that clustered-pinhole SPECT/PET now makes possible.
Mouse Heart Scan
First, we performed a dynamic (4-dimensional) heart scan on a mouse using a combination of 99mTc-tetrofosmin (a clinically used SPECT myocardial perfusion tracer) and 18F-FDG (used clinically to distinguish between dysfunctional/hibernating but viable myocardial tissue and scar tissue). This SPECT/PET tracer combination is of great practical interest in experimental cardiology, because it might be used to accurately distinguish between viable and scarred myocardium (28). Three perpendicular slices through the mouse heart are shown in Figure 3. These images were averaged over the cardiac cycle. Supplemental Video 1 (4-dimensional) of the beating mouse heart was reconstructed from the same scan with electrocardiography gating). The SPECT components of the cardiac images show somewhat more detail than do the images of the PET tracer. We attribute this mainly to the degree of γ-photon penetration through the pinhole edges. Furthermore, because 99mTc-tetrofosmin—in contrast to 18F-FDG—accumulates in the gallbladder, this organ is visible only in the SPECT images.
Simultaneous SPECT/PET cardiac scan. We imaged mouse, injected with 130 MBq of 99mTc-tetrofosmin and 24 MBq of 18F-FDG, for 60 min. Here, we show 3 mutually perpendicular slices through mouse’s heart. Images represent average over cardiac cycle. Supplemental Video 1, of beating heart, was reconstructed from same scan.
Mouse Brain Scan
We also acquired simultaneous SPECT/PET brain images, combining a dopamine transporter ligand (123I-FP-CIT) and a glucose analog (18F-FDG), both merged with a CT scan (Fig. 4). Specific binding of 123I-FP-CIT to dopamine transporters can be resolved in subregions of the brain (e.g., the striatum and the olfactory tubercle), whereas uptake in the Harderian glands may reflect nonspecific uptake. The images obtained for 18F-FDG showed an overall distribution with high concentrations in the olfactory bulbs and the cortical, thalamic, and midbrain areas. Earlier, we imaged the effects of cocaine (a dopamine transporter blocker) on the temporal characteristics of striatal 123I-FP-CIT binding in the mouse brain (16), whereas 18F-FDG has been used as a brain activity marker (29). Imaging this mix of SPECT and PET tracer molecules combined with the administration of a drug of abuse (e.g., cocaine) may create opportunities to directly relate molecular and functional processes in the brain by relating the binding location of the drug (revealed in the case of cocaine by displacement of 123I-FP-CIT) to metabolic activity changes (revealed by 18F-FDG) in the brain.
Simultaneously acquired SPECT/PET mouse brain images (color) overlayed on CT (gray). Coronal (A), sagittal (B), and horizontal (C) slices at identical levels are shown. Mouse was injected with 30 MBq of 123I-FP-CIT and 40 MBq of 18F-FDG and imaged for 60 min starting 105 min after injection. Uptake of 123I-FP-CIT (SPECT, top) in small brain structures such as striatum (1), olfactory tubercle (2), and Harderian glands (3) can be resolved. 18F-FDG (PET scan, bottom) is seen in multiple structures, including olfactory bulbs (4), cerebral cortex (5), and thalamic and midbrain areas (6).
Mouse Bone Scan
Finally, to illustrate a typical example of how SPECT and PET tracers suitable for a similar imaging task can be compared one-to-one under exactly the same biologic circumstances, we performed parallel skeletal SPECT/PET with the bone tracers 18F-fluoride and 99mTc-HDP. Figure 5 shows maximum-intensity projections (MIPs) from a simultaneously acquired whole-body bone scan. Figure 6 shows MIPs from a second bone scan with the same tracer combination in which we focused on the lumbar spine and pelvis. In these focused images, uptake of 99mTc-HDP and 18F-fluoride in tiny parts of the individual processes and other tiny vertebral parts is visualized. Like the cardiac image in Figure 3, the images of the SPECT tracer reveal a bit more detail than do the PET components. Rotating MIPs from the same bone scans are available as Supplemental Videos 2 and 3.
Example of simultaneous SPECT/PET scan obtained with 2 tracers that target same biologic function. Images represent MIPs of mouse injected with 2 bone tracers, 220 MBq of 99mTc-HDP (top) and 60 MBq of 18F-fluoride (bottom), and scanned for 60 min. Supplemental Video 2 shows rotating MIPs.
Example of simultaneous SPECT/PET bone scan of lumbar spine and pelvis. MIPs of mouse injected with 130 MBq of 99mTc-HDP (top) and 85 MBq of 18F-fluoride (bottom) and scanned for 30 min are shown. Supplemental Video 3 shows rotating MIPs.
DISCUSSION
We have shown that SPECT images are hardly degraded by the simultaneous presence of a PET tracer—even when the activity concentration of the PET tracer exceeds that of the SPECT tracer by up to a factor of 2.5 (Supplemental Figs. 1 and 2). Typically, activities used in biologic SPECT experiments are higher than those in PET experiments (30). To confirm this, we performed a search for the 20 most recent papers on mouse SPECT and on mouse PET published in The Journal of Nuclear Medicine (search terms mice SPECT or mice PET). These papers report average injected activities of 52 MBq and 8.1 MBq for mouse SPECT and mouse PET, respectively. As a result, we predict that in most biologic experiments, spill-over from the PET tracer into the SPECT images will be negligible, allowing for significant flexibility in the use of tracers and consequently a greater versatility in their application to biologic systems.
We were able to obtain high-resolution PET images despite the relatively low sensitivity (peak sensitivity of 0.31%) of pinhole PET compared with coincidence PET (peak sensitivities typically in the 1%–10% range). The level of detail visible in the images shows that sensitivity alone is insufficient as a predictor of image quality. In fact, it is the combination of sensitivity and the accuracy with which the direction of a detected γ-photon is known that determines how much image detail can be obtained.
In this paper we have presented focused scans, restricted to a specific organ or part of the mouse, as well as whole-body images. Because all clusters in the VECTor collimator focus on a central scan volume that is smaller than a mouse, focusing on a restricted volume of interest can increase the number of detected counts from that region, resulting in improved noise-resolution trade-offs (21). The benefits of focusing are well illustrated by the mouse bone scans in Figures 5 and 6; the visibility of small details is enhanced in the focused scan, compared with the whole-body scan. For this scan, simulations indicated that focusing increased the average sensitivity over the volume of interest by a factor of 5.
We obtained the images presented in this paper by placing the novel collimator into an existing SPECT system. This system has γ-detectors that are optimized for imaging SPECT tracers. The crystals have a thickness of 9 mm, which means that only about 10% of the 511-keV γ-photons that reach the detector are detected in the photopeak (25). The implication is that there is ample room for improving the sensitivity of the system. Increases of up to a factor of 8 might be achieved simply by, for example, using an 80-mm-thick NaI crystal. We estimated that down-scatter of 511-keV γ-photons entering the NaI crystal and detected at 140 keV would then decrease from 0.9% to 0.6%, thus reducing contamination of SPECT images. If, in addition, larger pinholes (e.g., 1.0 instead of 0.7 mm) were to be used, peak sensitivities similar to those in coincidence small-animal PET could readily be obtained. Future research on pinhole SPECT/PET will certainly include the optimization of detectors and geometry (e.g., size and shape of pinholes and number of pinholes per cluster).
Clustered-pinhole SPECT/PET has not been designed with a single imaging application in mind but rather as an enabling technology for use in a variety of research areas in which a direct correlation of multiple biologic functions will lead to enhanced insights. With simultaneous SPECT/PET, we enable a whole new range of applications in biomedical research, since 2 tremendously large and important classes of tracers can now be combined for the first time.
In cardiovascular research, a SPECT/PET tracer combination has already attracted considerable attention over the last 2 decades. It involves the use of 18F-FDG together with a blood-flow tracer (such as 99mTc-tetrofosmin or 201Tl). Such a tracer combination permits the assessment of both myocardial glucose utilization and flow in a single study (28). However, until now, the lack of suitable technology has hampered its high-resolution application. Small-animal models of disease play an important role in the optimization of imaging procedures and in the development of tracers and therapies. The high-resolution mouse heart images presented in Figure 3 show that the proposed cardiovascular tracer combinations can now be applied in a preclinical setting.
High-resolution SPECT/PET may also play a significant role in simultaneous and possibly dynamic imaging of different brain processes, such as metabolic activity in defined brain regions together with the functioning of the neurotransmitter system (Fig. 4)—or, alternatively, in monitoring various aspects of the neurotransmitter system simultaneously. Specifically, many research groups are currently investigating the effect of an intervention (e.g., amphetamine or cocaine administration) on metabolic brain activity (monitored with 18F-FDG) or on the neurotransmitter system (imaged with SPECT tracers such as IBZM, FP-CIT, or TRODAT). Because the biologic system does not react in the same way to repeated interventions (31) (an effect called sensitization), these 2 effects can be directly correlated only through simultaneous SPECT/PET. Furthermore, in research related to Parkinson disease, the possibility of using simultaneous SPECT/PET to image multiple sites or processes of interest—thereby enabling a direct investigation of interactions between different processes—has already attracted interest (32).
In tumor imaging, 18F-FDG is widely used as a marker of tumor response, whereas in other studies, long-living SPECT tracers such as 111In or 125I are attached to antibodies used for anticancer therapy to determine their distribution. With simultaneous SPECT/PET, these different aspects of tumor imaging can now be combined in a single longitudinal study. In this way, a repeated read-out of therapy effects can be directly related to the amount and distribution of SPECT-labeled pharmaceutical within the tumor.
In addition, simultaneous SPECT/PET can be used in tracer research to compare a SPECT and a PET tracer that target the same biologic function (Figs. 5 and 6). In a simultaneous SPECT/PET scan, the targeting properties of a molecule labeled with a heavy metal—such as is commonly used for SPECT tracers—could be compared directly with the behavior of an identical molecule labeled with a small positron-emitting radioactive atom. Such a comparison, at the same time and at exactly the same position in a single animal, rules out the effects of varying physiology and spatiotemporal misregistration—effects that today severely complicate tracer comparisons. This, in turn, can aid in the modification of a PET tracer into a SPECT tracer. The modification of PET tracers into SPECT tracers is of practical interest because of the wider availability of SPECT and may also be necessary when a process needs to be monitored for a long time—something prohibited by the relatively short lifetimes of most PET labels.
Previously, we showed that mouse organs and entire bodies can be imaged with half-minute time frames using high-resolution focused multipinhole SPECT (16,17). Because VECTor has a better sensitivity than these SPECT systems (both for SPECT and PET tracers), we expect that similar time resolutions will be feasible, allowing, for example, tracer kinetic modeling in simultaneous SPECT/PET.
The clustered-pinhole technology presented here may also lend itself to other novel applications in biomedical imaging. For example, suppose that the clusters were positioned closer to one other. Such a set-up would result in a severe overlap of pinhole projections—which in most situations would be undesirable (33,34). However, by adding small shutters that block a subset of the pinholes within each cluster, one could create imaging devices capable of focusing rapidly and flexibly and that could quickly be adapted to specific imaging needs—for example, for highly specific imaging of regions of the brain containing certain receptors or plaques, or for focusing on a tumor. Adaptive imaging of radionuclides has recently drawn much attention (35), and many new developments can be expected in this emerging area of research.
CONCLUSION
We have shown that collimated simultaneous imaging of SPECT and PET tracers with submillimeter image resolutions is possible even when detectors are used that are suboptimal for detecting 511-keV γ-photons resulting from PET tracers. From the large number of specific SPECT/PET tracer combinations that are enabled by our new technology, we have selected 3 prominent examples, and we have demonstrated our ability to simultaneously acquire separate images of these probes with excellent resolutions. This evidence leads us to believe that clustered-pinhole SPECT/PET may create unique imaging opportunities in a large number of biologic research applications.
DISCLOSURE STATEMENT
The costs of publication of this article were defrayed in part by the payment of page charges. Therefore, and solely to indicate this fact, this article is hereby marked “advertisement” in accordance with 18 USC section 1734.
Acknowledgments
We thank René Gommers and René den Oudsten for their technical assistance in the construction of the collimator. This research was cofunded by grant PID06015, forming part of the “Pieken in de Delta Zuidvleugel” program of the Netherlands Ministry of Economic Affairs and the Province of Zuid-Holland. No other potential conflict of interest relevant to this article was reported.