Abstract
Nuclear medicine has an important role in the care of newborns and children less than 1 y old. Patients in this age group present with a spectrum of diseases different from those of older children or adults. These patients can benefit from the full range of nuclear medicine studies. In these young children, nuclear medicine studies are more likely to be used to evaluate a wide range of congenital conditions but also can be helpful for evaluating acquired conditions such as infection, cancer, and trauma. This review first will cover the general aspects of nuclear medicine practice with these patients, including the special considerations that can help achieve successful diagnostic imaging. These topics will include clinical indications, imaging technology, instrumentation, software, positioning and immobilization, sedation, local and general anesthesia, radiopharmaceutical doses, radiation risk, and dose reduction. The review then will discuss the specific nuclear medicine studies that typically are obtained in patients in this age group. With extra care and attention to the special needs of this population, nuclear medicine departments can successfully study patients less than 1 y old.
Nuclear medicine plays a valuable role in the evaluation of pediatric patients in the first year of life. Evaluation of these patients differs from that of other children and must accommodate a different range of clinical problems, special imaging requirements, and extra care to acquire high-quality images that address the clinical question while delivering the lowest radiation dose possible. This review therefore will address special aspects associated with the imaging of these small children, including clinical indications, methods, patient handling, positioning, immobilization, sedation, administered radiopharmaceutical dosages, instrumentation, evaluation of radiation exposures and risk, and reduction of radiation exposure (1).
Clinical indications are broad and include the diagnosis and assessment of diseases of the central nervous, endocrine, cardiopulmonary, lymphatic, gastrointestinal, genitourinary, and musculoskeletal systems, as well as oncologic diseases. This is based on a retrospective review derived from the experience at our institution from 2000 to 2009. During this period, approximately 2,570 children under 1 y old underwent one or more diagnostic procedures, for a total of 3,330 studies. There were slightly more boys (male-to-female ratio, 1.14). Patient ages ranged from premature infants to 12 mo. Two thirds of the studies were of patients less than 6 mo old.
TECHNICAL CONSIDERATIONS
Imaging
Imaging of neonates and infants requires special care, patience, adaptation, experience, and skillful hands. Individual approaches are often needed that require deviation from routine protocols with respect to the administered radiopharmaceutical doses, imaging techniques, and types of images needed.
During the first year of life, infants naturally experience rapid growth and development, and radiopharmaceutical distribution and kinetics may vary from that in older children or young adults. For example, newborns and infants have a lower glomerular filtration rate, faster circulation, and faster pulmonary wash-in and washout of radioactive gases than older children. The uptake of pertechnetate in the gastric mucosa is lower and the background higher in infants than in older children. The distribution of skeletal imaging agents in this group of patients reveals relatively high tracer concentrations in growth centers and a lack of tracer uptake in the nonossified cartilaginous parts of the growing bone (Fig. 1). As a child matures, physiologic processes mature, and the kinetics and biodistribution of radiopharmaceuticals eventually reach adult levels.
Imaging of these patients must take into account the challenges of evaluating small organs and bodies with equipment designed for imaging adults. Newborns weigh an average of 3.4 kg (range, 2.5–4.5 kg). By 1 y old, the baby has tripled its birth weight to approximately 10 kg (range, 7.5–12 kg). Premature babies (<37 wk of gestation) usually weigh less than 2.5 kg, but some weigh as little as 800 g or even less.
Positioning, Immobilization, Sedation, and Local and General Anesthesia
Most nuclear medicine imaging requires the patient to remain still for a relatively long period. In babies, simple immobilization techniques (sandbags, safety belts, gently wrapping a bed sheet around the patient) and some holding by the gentle and skilled hands of nuclear medicine technologists help to ensure success. Newborns usually find swaddling comforting. In a quiet environment, with dim lights and care, many of these children fall asleep during longer examinations. Attenuation imaging artifacts resulting from holding patients by hand should be avoided if possible, but they should be anticipated and recognized on images. Allowing the child to hold a favorite toy or stuffed animal during some examinations may be permitted as long as it does not interfere with imaging. Our imaging rooms are equipped with television monitors that can be positioned within easy view by the patient. Video or audio programs often help to distract and relax patients (and parents) during the examination. Even if the patient moves during an examination, motion correction techniques can be applied effectively to aid in the qualitative and quantitative evaluation of images (2). Local skin anesthesia with cream containing lidocaine and prilocaine (e.g., EMLA cream; AstraZeneca LP) is helpful in reducing pain associated with venipuncture.
Infrequently, immobilization alone is not sufficient, and sedation or, much less frequently, general anesthesia is necessary. In our review, less than 3% of the children in this age group needed sedation, mainly for brain or whole-body SPECT or PET. Under optimal conditions, institutions have specialized pediatric sedation programs with dedicated teams of anesthesiologists and nurses who have particular experience with patients needing imaging. When sedatives are administered, potential side effects such as aspiration and respiratory arrest should be anticipated and appropriate means of treatment must be available. A physician or nurse must monitor the child from the time of sedation until full recovery. To support sedation, the nuclear medicine department must be equipped with age-appropriate medications and equipment. Sphygmomanometers and stethoscopes appropriate for these children should be within easy reach. Sedation or general anesthesia can affect physiology, such as cerebrospinal fluid flow, cardiovascular shunt flow, cardiac function, renal function, and cerebral function.
Instrumentation and Software
Optimal imaging of the newborn or infant requires special consideration of the appropriate choice of instrumentation, image acquisition technique, and processing technique. For planar imaging and SPECT in these small patients, choosing an appropriate imaging device is important and will depend on the particular study and the specific diagnostic task. A state-of-the-art dual-detector γ-camera system can acquire high-quality images of an infant. These cameras typically have a useful field of view of 50 × 30 cm. Both anterior and posterior images of the whole body can be obtained simultaneously in a reasonable period, thus reducing immobilization time. A fixed-geometry second detector may be in the way if the application requires direct access to the patient during imaging, and a dual-detector device that allows the second detector to be moved separately would therefore be preferable.
Choosing the appropriate collimator can make the difference between an excellent and a suboptimal study. The small size of the organs being imaged in the newborn or infant makes this choice even more critical. For some tasks, the ability to resolve a particular feature may be most important, and therefore an ultra-high-resolution collimator or pinhole collimator may be most appropriate even if requiring a slightly longer imaging time. In other instances, it may be necessary to acquire a less noisy image in a fixed amount of time based on the physiologic process to be evaluated, and therefore, a higher-sensitivity collimator may be a good choice. The ability to select from a variety of collimators (ultra-high-resolution, high-resolution, and high-sensitivity) provides flexibility to adapt to specific diagnostic needs. Pediatric nuclear medicine clinics also should have a high- or medium-energy collimator available for imaging 123I agents such as metaiodobenzylguanidine (MIBG). Although the primary γ-ray emitted by 123I has an energy of 159 keV, it also emits several higher-energy photons in low abundance (γ-rays with energies higher than 400 keV are emitted 2.6% of the time). Low-energy collimators are not effective at stopping these high-energy photons, leading to a significant amount of septal penetration and image degradation. When a low-energy collimator is used to image 123I, almost half the events measured within the 159-keV window result from these high-energy photons. Using a medium-energy collimator produces images of superior quality, compared with the high-sensitivity or even ultra-high-resolution collimators (3).
In some cases, even an ultra-high-resolution collimator may not resolve small features of importance. Pinhole collimator magnification scintigraphy provides the highest spatial resolution (a few millimeters) with γ-cameras and, in selected applications, is invaluable for imaging small body parts. Pinhole collimator magnification spreads the image over a larger area of the detector, effectively improving the overall system spatial resolution above what can be achieved with parallel-hole collimators (4). Pinhole images using an insert of a small aperture (2–3 mm) in infants undergoing 99mTc-dimercaptosuccinic acid (DMSA) renal scans or 99mTc-methylene diphosphonate (MDP) bone scans are of excellent quality in the smallest patients (Fig. 2). The pinhole collimator has the lowest sensitivity of any of the collimators routinely used, particularly with the smaller apertures, and images can take several minutes (10–15 min) to acquire. For rapid dynamic studies, a low-energy converging collimator provides a good combination of higher spatial resolution and sensitivity.
In newborns and infants, SPECT recording can be particularly challenging because of the limited spatial resolution due to the relatively large distance from the patient to the detector. In most devices, SPECT detectors cannot get close to the patient, because the imaging beds are typically around 38 cm wide. Specially designed narrow pediatric pallets allow the detectors to come closer to the patient's body, improving SPECT quality. The use of focusing collimators and new SPECT gantries with advanced robotics has made such approaches much simpler for routine use in cardiac imaging and, therefore, may optimize SPECT for imaging neonates and infants.
Image fusion of complementary anatomic and functional 3-dimensional (3D) image sets is increasingly regarded as a necessary part of imaging. Image fusion registers 2 or more 3D image sets of the same or different imaging techniques in the same orientation in the same space. PET/CT and SPECT/CT devices provide hardware image fusion. However, patient motion can still cause misregistration that needs to be corrected. Software fusion has expanded the application of image fusion to many combinations of 3D image sets, including PET, SPECT, CT, and MRI. For example, a useful approach in young patients with neuroblastoma is to fuse 123I-MIBG SPECT and MRI. Fusion can also be used on images of the same modality acquired under different circumstances or at different times. In the assessment of patients with epilepsy, image fusion is used to compare 2 functional states, such as ictal and interictal brain perfusion. Similarly, registration of SPECT or PET with MRI is a useful approach to assist in the localization of epileptogenic cortex. Most methods allow rigid fusion that works well for brain studies. Nonrigid registration and fusion of organs in the chest and abdomen is possible but more challenging to implement because of deformations due to body motion between 2 image acquisitions.
Radiopharmaceuticals
An important advantage of nuclear medicine in the first year of life is that it is extremely safe. Radiopharmaceuticals typically are nontoxic and nonallergenic. The total mass of material that is administered is small and does not elicit an osmotic effect. Similarly, the administered mass is low enough that no pharmacologic or toxic effect is produced. For example, a patient who is allergic to iodine can safely receive 123I-NaI or 123I-MIBG because the administered mass of iodine is well below the threshold needed to trigger an allergic effect. The volume and mass of a radiopharmaceutical that is administered is significantly lower than that of MRI and CT contrast agents. For example, in a 10-kg 1-y-old patient having a bone scan, the administered volume of 99mTc-MDP solution is 0.06 mL containing only 0.64 mg of radiopharmaceutical. In comparison, for the same 10-kg patient undergoing MRI, 940 mg of gadopentetate dimeglumine (Magnevist; Bayer) may be administered in a volume of 2.0 mL. For a contrast-enhanced CT scan, 6,400 mg of ioversol (Optiray 320; Covidien) may be administered in a volume of 20 mL. Thus, the administered volume for 99mTc-MDP is 20- to 200-fold less than that for MRI or CT contrast agents, and the administered mass is more than a thousandfold less (Table 1).
Pediatric radiopharmaceutical doses in neonates and infants should be determined by the minimum amount of activity necessary to ensure a satisfactory diagnostic examination in the shortest time. Administered doses that are too high lead to increased radiation dose to the patient without resulting in improved diagnostic sensitivity or accuracy. Doses that are too low may not permit an adequate examination. In both cases, too-high or too-low doses lead to unnecessary radiation exposure. In adult patients, dosing is roughly standardized. However, in most pediatric patients this has not been the case. Pediatric activity doses are often scaled down from the adult dose on the basis of body weight, surface area, size, or age. These dose estimations for pediatric patients based on adult dose corrected for body weight or body surface area are usually adequate for children over 1 y old. Presently, there is no universally accepted guideline for administered radiopharmaceutical doses in children (5). Different approaches to determining administered dose have accounted for substantial variations among centers. Two common methods use patient size as determined by the patient's weight or surface area: Clark's rule and the area rule. Other methods use age as the predominating factor in dose calculation: Young's rule, Webster's rule, and Fried's rule. The dosage method used at a particular clinic is largely dependent on the practitioner's preferences. Premature infants and newborns require special consideration, and one needs to consider the concept of minimum administered dose—the amount of administered radiopharmaceutical below which the study will be inadequate regardless of the patient's size (1).
We recently surveyed 13 dedicated pediatric hospitals in North America to determine radiopharmaceutical dose schedules for 16 commonly performed nuclear medical studies. The survey also inquired about minimum and maximum administered activity for each procedure. In our survey, we found that scaling the adult radiopharmaceutical doses according to the patient's weight is a common approach. This approach leads to similar radiation-absorbed doses to the patient irrespective of patient size because the radiation dose is defined as energy imparted per unit mass. When dosing is determined according to weight, the differences in patient geometric factors such as the size and shape of the individual organs and distance between organs actually lead to smaller patients receiving a slightly smaller radiation dose than larger patients (Table 2). From our survey, the average administered activity per body mass varied by a factor of 3, and the variation was by as much as a factor of 10 among the institutions. On the other hand, the variation in minimum administered activity was a factor of 10 on average and was as much as a factor of 20. Thus, the greatest variability in administered dose is in the smallest patients. This fact suggests that an opportunity exists to reexamine and redefine the lowest administered doses that ensure satisfactory studies in patients in this age group. The European Association of Nuclear Medicine has also developed a pediatric dose card (6).
As a result of that survey, during 2009 and 2010, a pediatric nuclear medicine group organized consensus conferences and workshops that included many pediatric nuclear medicine physicians, physicists, and technologists from North America. This group included representatives from the Society of Nuclear Medicine, the Society for Pediatric Radiology, the American College of Radiology, and the Image Gently Campaign. This group's activity has recently produced a new consensus guideline for pediatric radiopharmaceutical administered doses. This guideline, which has been approved by the Society of Nuclear Medicine and the Society for Pediatric Radiology, narrowed the variation in pediatric radiopharmaceutical doses, addressing the minimum administered doses for the smallest patients, as well as doses per kilogram of body weight (7).
Reduction of Radiation Exposure and Risk
In general, nuclear medicine procedures involve exposure to relatively low levels of ionizing radiation. The diagnostic value of these procedures, when used appropriately, far outweighs potential radiation exposure and thus potential risk. However, younger individuals may be at a higher risk than adults for long-term effects from radiation, partly because of the higher sensitivity of growing tissues in young individuals and because they have more potential years of life ahead during which the risk could be realized. When the administered doses are scaled by body weight, the range of effective doses for nuclear medicine procedures is 0.6–5.5 mSv and comparable to that from other common imaging procedures (8,9). A precise estimate of radiation dose requires knowledge of the administered dose, the size or geometry of the patient, and the biokinetics of the radiopharmaceutical. When radiopharmaceutical doses are scaled by body weight, the increasing radiation sensitivity for younger patients is partly offset by the lower radiation dose, and thus the total risk is reasonably constant across pediatric age groups (Table 3).
One way to achieve the lowest radiation exposures in pediatric nuclear medicine is to ensure that the requested procedure is the most appropriate for the specific diagnostic question and that the administered dose is appropriate for the specific diagnostic imaging task. Therefore, it is important to obtain sufficient clinical information on each case before proceeding with a study. The minimum administered dose of activity to be given to a child is determined by the type of study: dynamic or static. Rapid dynamic studies require a higher dose of tracer than static studies for sufficient image count statistics.
As an example, certain studies (e.g., hepatobiliary studies or 99mTc-mercaptoacetyltriglycine [MAG3] renal scans) have traditionally included an initial radionuclide angiogram at 1–3 frames/s during the first 60 s of the study, followed by a longer acquisition with approximately 1–4 frames/min. One should consider the clinical need to routinely acquire a radionuclide angiogram at rapid framing rates, or if the study interpretation could be adequately achieved with images acquired for 1–2 min each. Therefore, slow dynamic studies (e.g., hepatobiliary scintigraphy or dynamic renal scintigraphy) that do not require high temporal resolution can be obtained satisfactorily with a lower administered activity than studies requiring rapid framing rates, thus resulting in relatively lower patient radiation dose. Therefore, appropriate adjustments of acquisition protocols may allow the administration of a lower radiopharmaceutical dose without degradation of image quality or diagnostic value (Fig. 3).
Because the anteroposterior cross-section of an infant is small (8–10 cm), another option that can be considered for reducing administered doses is the use of a low-energy high-sensitivity collimator, which may provide an appropriate resolution at these short depths. Another approach to reducing the dose is the application of advanced image processing techniques. Iterative reconstruction makes SPECT more feasible in these small patients. Unlike filtered backprojection techniques, iterative reconstruction can incorporate corrections for the noise properties of the projection data, as well as changes in collimator resolution with distance, leading to a more accurate reconstruction. These iterative reconstruction techniques provide similar, and often improved, image quality with significantly fewer counts needed to produce diagnostic quality images. Therefore, high-quality studies can be obtained with lower radiation exposure to the patient, shorter imaging times, or a combination of both. In addition, advanced planar processing allows a reduction in administered radiopharmaceutical doses without loss of diagnostic information. Performing SPECT reconstruction with 3D ordered-subset expectation maximization and with resolution recovery methods can significantly reduce the image acquisition time and make it possible to acquire a high-quality SPECT study of an infant in only a few minutes (Fig. 4) (10–14). The utility of iterative reconstruction and sophisticated filtering in the newborn and infant is under evaluation.
Although iterative reconstruction has been used in PET, incorporation of the spatial resolution variation within the field of view into the algorithm is relatively new. In addition, the use of 3D PET is most appropriate to very small children. 3D PET leads to a large gain in sensitivity relative to 2D PET. Thereby, the use of 3D PET in these patients should lead to PET studies of outstanding image quality in either a shorter imaging time or with less administered activity, depending on the diagnostic task. Newer PET systems that offer high spatial resolution across the field of view, improved field uniformity, faster imaging times, and improved imaging processing have obvious advantages for imaging small children. The use of these systems may allow a reduction in administered doses and thus a reduction of patient radiation exposure or of the need for sedation. An approach leading to reduction in radiation exposure suggests that it is possible to obtain adequate 18F-FDG brain PET images of term and preterm babies using 3.7 MBq (0.1 mCi)/kg (15). This is lower than the current usual administered doses for this study (minimum, 5.18 MBq [0.14 mCi]/kg, and maximum, 7.4 MBq [0.20 mCi]/kg, with a mean of 5.66 MBq [0.153 mCi]/kg) (7,16). Another way to reduce radiation exposure in children who are candidates for PET/CT or SPECT/CT is to consider the use of previously acquired diagnostic CT for image fusion and not repeat the CT at diagnostic levels of radiation during the PET acquisition. CT during PET/CT or SPECT/CT can then be obtained at much lower doses for attenuation correction or anatomic correlation only. Software can then be used to fuse the 3D image sets.
CLINICAL APPLICATIONS
The second part of this review will discuss specific nuclear medicine studies that typically are obtained for patients in this age group. These include a full range of nuclear medicine studies, including oncologic studies and studies of the brain, thyroid, heart, lungs, liver, gastrointestinal tract, kidneys and genitourinary tract, spleen, lymphatic system, and skeleton.
Brain and Cerebrospinal Fluid
Frequent indications for brain SPECT, PET, and planar scintigraphy during the first year of life include seizures, tumors, effect of hypothermia and hypoxia, evaluation of cerebrospinal fluid flow, and assistance in the determination of brain death. An important consideration in this age group is brain immaturity and development. As the brain matures, regional cerebral blood flow and metabolic patterns change with age and follow the order of anatomic, evolutional, and behavioral development of the child. These changes are reflected on 99mTc-ethylcysteinate dimer (ECD) or exametazime perfusion brain SPECT and 18F-FDG brain PET (Fig. 5). In the newborn, regional cerebral perfusion and regional metabolism are initially more intense in the sensorimotor cortex, thalamus, brain stem, and cerebellar vermis. Later, the parietal, temporal, and occipital cortex; basal ganglia; cerebellar cortex; and finally the frontal cortex are included. The level of cerebral glucose metabolism is lower in the newborn than in adults but increases with age. The pattern of 18F-FDG uptake in the pediatric brain becomes similar to that of young adults by 1 y old. However, further brain development continues over several years. At about 3 y old, a child's brain metabolic activity exceeds that of an adult, and by approximately the age of 7 a plateau is reached with a value 1.3 times higher than in adults. The pattern of uptake and distribution of 18F-FDG in the brain is much lower in the thalamus and basal ganglia, cerebellum, sensorimotor cortex, frontotemporal, and occipital regions in preterm infants than in term infants (17–19).
In infants with intractable seizures, ictal and interictal perfusion brain and 99mTc-ECD SPECT aid in the localization of epileptogenic cortex. Ictal SPECT usually reveals increased local uptake in the epileptogenic region, whereas interictal SPECT shows regional reduction in cerebral perfusion within the same region. Ictal-minus-interictal subtraction of perfusion SPECT assists further in the localization of epileptogenic cortex.
99mTc-ECD is a lipophilic agent that after intravenous injection rapidly concentrates in the brain at approximately 6% of the administered dose. This agent is a first-pass indicator and does not redistribute after its initial brain localization. Imaging can begin 5–10 min after the intravenous injection of the tracer or can be obtained a few hours (3–5 h) afterward, allowing the patient to be stabilized before imaging if needed. Typical administered doses are 7.4–11.1 MBq (0.2–0.3 mCi)/kg, with a minimum dose of 37 MBq (1 mCi). In the interictal state, care should be taken to reduce environmental stimulation during the intravenous injection of the tracer. Ictal perfusion SPECT requires that the patient be given the tracer during a seizure. Successful ictal perfusion SPECT requires meticulous attention to detail and coordination (20). If the tracer is injected after the seizure activity has subsided, the resultant SPECT image depicts postictal regional cerebral blood flow and makes localization of a primary epileptogenic focus difficult.
18F-FDG PET is a useful method to localize epileptogenic cortex and to assess tumor viability. Active tumors show increased tracer uptake. However, it is well known that uptake of 18F-FDG within normal brain tissue can interfere with the detection of tumor activity. Usual administered doses for 18F-FDG for brain PET are 5.18 MBq (0.14 mCi)/kg at a minimum, with a mean of 5.66 MBq (0.153 mCi)/kg. However, it has been shown possible to obtain adequate 18F-FDG brain PET images in term and preterm babies using 3.7 MBq (0.1 mCi)/kg (15). 18F-FDG PET is performed 1 h after intravenous administration of the tracer. Images can be obtained in 10 min or less using a 3D PET acquisition mode.
Interictal 18F-FDG PET usually reveals regional decreased tracer uptake. Fusion of ictal–interictal subtraction SPECT images to MRI is helpful in localizing the seizure onset and in guiding the intracranial recording in childhood epilepsy (Figs. 5 and 6). Image fusion improves sensitivity and makes interpretation easier (21,22). In the event that the patient has a seizure during the 18F-FDG uptake period, the PET scan will demonstrate increased tracer uptake in the epileptogenic zone. This is rarely seen, as it is not possible to predict seizure activity during the hour or so of the 18F-FDG brain uptake period. Therefore, most 18F-FDG PET studies in patients with epilepsy are interictal.
For SPECT, it is important to ensure that the patient lies still for the examination and that the γ-camera detectors are positioned as close to the head as possible. We recommend the use of a dual-detector system equipped with ultra-high-resolution collimators. Each detector rotates 360° around the patient's head and records 120 images of 15 s each in a 128 × 128 matrix. In our experience, sedation is usually needed for brain SPECT or PET in many of these very young patients. Cerebrospinal fluid flow can be evaluated effectively with radionuclide cisternography in case of diversionary shunts and in the prechemotherapy assessment of cerebrospinal fluid flow (23).
Thyroid
Thyroid scintigraphy is useful in the evaluation of hypothyroidism in infants. Thyroid hormone is essential for the normal development of the central nervous system. Untreated congenital hypothyroidism is one of the most common preventable causes of mental retardation (24). Causes of neonatal hypothyroidism include thyroid agenesis, dysgenesis, or ectopia (25); dyshormonogenesis due to inborn errors in thyroid hormone synthesis (26,27); and transient infantile hypothyroidism caused by maternal thyroid-blocking antibodies, maternal antithyroid medications, or iodine exposure (28). In North America, congenital hypothyroidism has an incidence of approximately 1 in every 3,000–4,000 births, with approximately a 2:1 female predominance (24,27). Nearly all new cases of infantile hypothyroidism in North America and Europe are now detected by neonatal laboratory screening for hypothyroidism, which facilitates prompt evaluation and early institution of thyroid hormone replacement therapy to prevent neurocognitive damage.
During embryonal development, the thyroid gland arises from endodermal tissue at the floor of the oropharynx and descends along the thyroglossal duct from the pharynx to the neck. A remnant thyroglossal duct may be seen as a pyramidal lobe, persistent duct, or thyroglossal cyst. With disrupted migration, ectopic thyroid tissue may lie anywhere along the migratory pathway from the base of the tongue to the thyroid bed, or even into the mediastinum (29). The relationship between abnormal thyroid migration and dysfunction is poorly understood. In some individuals an ectopic thyroid may produce thyroid hormone; in others, an ectopic thyroid may be dysgenic or show diminished function. Thyroid dyshormonogenesis can result from many different inborn errors of metabolism of thyroid hormone synthesis. Transient hypothyroidism is a clinical diagnosis, and the affected infant may not even be referred for imaging studies. Patients may be referred for nuclear medicine evaluation to exclude concurrent thyroid dysgenesis or ectopia. Because the neonatal brain is dependent on thyroid hormone for normal development, thyroid hormone replacement should not be discontinued during nuclear medicine evaluation of the thyroid (28–30).
Nuclear medicine and ultrasound are complementary for the evaluation of infantile hypothyroidism. Thyroid scintigraphy is highly sensitive for detection of ectopic thyroid tissue, and ultrasound is useful for evaluation of the gland size and morphology. Ultrasound may be helpful for confirming the presence of nonfunctioning thyroid tissue of patients with thyroid suppression due to hormone replacement therapy, recent nonradioactive iodine overload, thyroid-stimulating hormone deficiency, or maternal thyroid-blocking antibodies (31).
Functional imaging of the thyroid gland can be obtained with either 99mTc-pertechnetate or radioiodine. 123I is the radioisotope of choice for imaging the infant's thyroid. For pediatric thyroid imaging, a typical dose of 123I is 0.9 MBq (0.025 mCi). In an infant, the average thyroid dose is predicted to be less than 5 mSv/MBq (18.5 rem/mCi) (30,32). Although 99mTc-pertechnetate accumulates in the thyroid gland, it is not organified and can wash out of the thyroid. This method of thyroid imaging has been popular because of the ready availability of 99mTc-pertechnetate and the ability to image the thyroid immediately after intravenous administration of the tracer. For pediatric thyroid imaging, a typical dose of 99mTc-pertechnetate is 1.1 MBq (0.03 mCi)/kg, with a minimum dose of 7.5 MBq (0.2 mCi). However, compared with 123I, 99mTc assesses only the trapping function (30). Thyroid images are best obtained with a pinhole collimator, either to provide a magnified view of the thyroid by placing the collimator as close as possible to the neck or to provide a birds-eye view of a larger region to survey for ectopic thyroid tissue that may be anywhere between the oropharynx and mediastinum (30). Anatomic localization of functional thyroid tissue is facilitated by acquiring a transmission scan of the upper body to overlay with the diagnostic scan. Small radioactive sources are used to mark anatomic landmarks.
Lungs
Pulmonary ventilation and perfusion studies are indicated in several conditions in patients less than 1 y old: pneumonia, inflammation, or infection; cyanosis; lobar emphysema; assessment of pulmonary artery angioplasty or surgery; congenital diaphragmatic hernia and repair; bronchopulmonary dysplasia; foreign bodies; aspiration; asthma; pulmonary hypoplasia, stenosis, aplasia, or agenesis; pulmonary sequestration; pulmonary embolism; pulmonary valve stenosis; arteriovenous malformation; and evaluation of pulmonary transplants. Pulmonary perfusion scintigraphy is useful as part of the overall cardiopulmonary assessment of patients before and after catheter interventions, intravascular stent placement, occlusion of unwanted vascular communications, surgical arterioplasty, and other surgical interventions in patients with congenital or acquired heart disorders (Fig. 7). Frequently, patients with congenital heart disease referred for assessment of pulmonary perfusion have a right-to-left shunt of various levels of severity, as demonstrated by systemic penetration of macroaggregated particles. This condition is the result of microembolization of all the systemic organs and is most prominently appreciated in organs with high relative blood flow (e.g., the brain and kidneys). The pulmonary-to-systemic flow ratio is lower than 1:1 and can be calculated by determination of the relative pulmonary and systemic activity on a whole-body scan. The biologic half-life of 99mTc-macroaggregated albumin in the lungs is approximately 6 h. Once trapped in the pulmonary arterioles, 99mTc-macroaggregated albumin particles are degraded into smaller particles and polypeptides, which are taken up by the liver and ultimately eliminated in bile. Although macroaggregated particles are rapidly cleared and are considered safe in patients with right-to-left shunts, it seems prudent to minimize systemic microembolization by reducing the number of administered particles without compromising diagnostic yield.
The pulmonary vasculature of infants shows a rapid increase from approximately a tenth to a third of the adult values during the first year of life and reaches approximately half the adult number by the age of 3 y. Although administration of ±500,000 99mTc-macroaggregated albumin particles is considered safe in an adult, the number of injected particles should not exceed 50,000 or 165,000 in a 1-y-old child (33). In our routine practice, we use fewer than 10,000 particles in neonates and infants, as well as in all patients with right-to-left shunts and in patients with pulmonary hypertension or severe pulmonary disease (Table 4). Because pulmonary ventilation studies can be obtained rapidly and effectively using 133Xe gas even in the smallest patients, air trapping can be documented and ventilation perfusion ratios can be calculated (34).
Heart
Although nuclear cardiology is not frequently performed on children, it is useful in several cases. Indications for pediatric nuclear cardiology in children during the first year of life include the following: Kawasaki disease; repair of tetralogy of Fallot; an anomalous left coronary artery departing from the pulmonary artery; a postarterial switch operation; cardiomyopathies; cardiac transplantation; estimation of right ventricular pressure overload or right ventricular hypertension; left-to-right shunts; right-to-left shunts; and determination of left and right ventricular ejection fractions. Myocardial perfusion imaging is used in the evaluation of myocardial ischemia, infarction, and viability in young children with Kawasaki disease, an anomalous origin of the left coronary artery from the pulmonary artery, or coronary ectasia, and patients after the arterial switch operation for transposition of the great arteries can be evaluated noninvasively with myocardial perfusion imaging (Fig. 8) (35).
99mTc-methoxyisobutylisonitrile (MIBI) is a first-pass indicator that accumulates in the myocardium according to regional myocardial perfusion. After intravenous injection, this agent concentrates in several organs, including the thyroid, myocardium, kidneys, and striated muscle. The agent clears rapidly from the blood, with less than 8% of the administered activity remaining in blood at 5 min, and less than 1% of the tracer is protein-bound in the plasma. The principal route of elimination of 99mTc-MIBI is the hepatobiliary system. Tracer activity appears in the intestine within the first hour after injection. At rest, approximately 1.5% of the injected dose is taken up in the myocardium, where it remains fixed, with no redistribution over time. With 99mTc-MIBI, both resting and pharmacologic stress evaluations can be performed on children (35,36).
In preparation for myocardial perfusion scintigraphy, the patient should fast for 2 h before administration of the tracer. A short intravenous catheter is placed and secured to the skin with tape, and the line is kept viable with normal saline. The intravenous line is maintained in place so it can be used to inject the tracer during pharmacologic stress. Because 99mTc-MIBI has no significant redistribution over a 4- to 6-h period, 2 injections of the radiopharmaceutical are used to obtain rest and stress myocardial perfusion imaging within the same day. For a single study (rest or stress) done alone, a dose of 37 MBq (1 mCi) is used. If rest and stress studies are done on separate days, the same dose of 99mTc-MIBI can be used for the second study. For rest and stress studies performed on the same day, a 37-MBq (1-mCi) dose is suggested for the rest 99mTc-MIBI SPECT study; 2–4 h afterward, the stress study is performed using a dose of 74 MBq (2 mCi) given at peak stress. Imaging is obtained 5 min after tracer administration. Acquisition protocols should be adapted to individual SPECT systems. SPECT may be acquired using the following parameters: 120 total projections (60 stops with a dual-detector system) over 180° with a 128 × 128 matrix for a total of 20–30 min. Electronic magnification is used according to the patient's heart size. After acquisition and reconstruction, the study is reoriented along the long axis of the left ventricle. Gated myocardial perfusion SPECT provides information on myocardial contractility, wall motion, end-diastolic and end-systolic ventricular volumes, ejection fraction, regurgitant fraction, and cardiac output.
Cardiac transplantation has become a treatment option for neonates, infants, and children with end-stage cardiomyopathy or those with congenital heart disease that cannot be treated by surgical palliation or repair. Early recognition of accelerated transplant vasculopathy could be achieved with myocardial perfusion scintigraphy. Collateral circulation occurs with some frequency and may be a factor affecting the detectability of myocardial perfusion abnormalities at rest. Most young children show adequate perfusion during the early postoperative period. Along with coronary angiography, myocardial perfusion SPECT helps in the diagnosis of coronary artery disease and myocardial viability (35).
Patients with an anomalous left coronary artery arising from the pulmonary artery have severe myocardial dysfunction and ischemia during early infancy. The left ventricle becomes perfused with desaturated blood and at coronary pressures that rapidly fall below systemic pressures. Classic complications include ischemia and infarction of the anterolateral left ventricular free wall, followed by mitral valve incompetence due to infarction of the anterior papillary muscle. This leads to congestive heart failure in the first year of life. Several surgical techniques have been used to transfer the anomalous coronary artery back to the aortic cusp. Myocardial perfusion scintigraphy can assess ventricular wall damage and evaluate recovery of function after repair (35). Children with right ventricular hypertension or right ventricular overload show greater tracer uptake of myocardial perfusion agents in the right ventricular myocardium. The increased uptake correlates with peak right ventricular pressures (37).
Left-to-Right Shunts
The recommended procedures for detection and anatomic characterization of left-to-right shunts are ultrasound and MRI. Radionuclide methods have played a major role in the past and can be used to quantitate shunts (38,39). 99mTc-pertechnetate is the most commonly used radiopharmaceutical. The patient should be premedicated with sodium perchlorate (intravenously) or potassium perchlorate (orally) to reduce thyroid uptake of the tracer. The recommended administered dose of 99mTc is 74 MBq (2 mCi). The total volume of radiopharmaceutical solution should not exceed 0.2 mL so that a small rapid intravenous bolus can be administered. Patients do not need sedation for this procedure. An intravenous needle (butterfly type, 23–25 gauge) or a short intravenous catheter is inserted. Once one is assured that the needle or catheter is properly placed and that there is no possibility of extravasation, it should be secured to the patient's skin. If possible, an antecubital vein is selected for the injection. Other venous sites can be used as long as they are patent and can tolerate a rapid 1- to 3-mL saline flush. The patient is placed supine on the imaging table. The γ-camera, equipped with a parallel-hole, high-sensitivity collimator, is positioned over the patient's chest. The field of view extends from the suprasternal notch to just below the xyphoid and covers both pulmonary fields. Radionuclide angiocardiography for the assessment of left-to-right shunting is recorded at 2 or 4 frames/s for 25 s on a 128 × 128 matrix or in list mode. Injection technique is of utmost importance for quantitation of left-to-right shunts. The volume of saline flush should be approximately 1–3 mL in infants. The tracer should not be injected while the patient is crying or producing Valsalva maneuvers. Increase in intrathoracic pressure can cause bolus fragmentation and render the study inadequate. Injecting the tracer with the saline simultaneously allows rapid, uniform delivery of the tracer with uninterrupted saline flushing. If the radioactive bolus is fragmented, deconvolution analysis ensures a high percentage of successful studies.
Left-to-right shunts can be easily quantitated by analysis of the pulmonary time–activity curve in first-pass radionuclide angiography (Fig. 9). This method can accurately quantify left-to-right shunts of pulmonary-to-systemic flow ratios between 1.2 and 3.0 (38,39). For determination of right and left ventricular ejection fraction, radionuclide angiocardiography should be recorded at a minimum rate of 25 frames per second for 25 s on a 128 × 128 matrix or in list mode. Normal values of right and left ventricular ejection fractions using first-pass radionuclide angiocardiography have been determined in children (40).
Hepatobiliary Indications
The most frequent indications for hepatobiliary scintigraphy during the first year of life are the differentiation of biliary atresia from hepatocellular disease and the diagnosis of choledochal cysts. Hepatobiliary scintigraphy can exclude the diagnosis of biliary atresia by demonstrating biliary passage of the radiopharmaceutical into bowel. When this passage cannot be demonstrated, the diagnosis of biliary atresia is highly suspected (Fig. 10). The definitive diagnosis of biliary atresia requires percutaneous or intraoperative transhepatic cholangiography. Hepatobiliary scintigraphy repeated after a few days may be helpful in newborns in whom excretion of tracer in the bowel is not demonstrated on an early single examination.
In jaundiced infants, pretreatment with phenobarbital increases biliary secretion and facilitates the differentiation of neonatal hepatitis from biliary atresia. Phenobarbital is given for 3–5 d before hepatobiliary scintigraphy in doses of 5 mg/kg/d (41). In neonatal jaundice, hepatobiliary scintigraphy after phenobarbital therapy is highly accurate for differentiating biliary atresia from other causes of neonatal jaundice. Hepatobiliary scintigraphy has 97% sensitivity and 82% specificity (91% accuracy) in the diagnosis of biliary atresia. A simple visual grading that includes the assessment of hepatocyte clearance and timing of radiotracer appearance in the intestine or extrahepatic biliary system for the evaluation of biliary atresia is helpful. In a normal study, radiotracer appears in the biliary tract or intestine within 15 min after injection. The scintigraphic diagnosis of biliary atresia is made when there is no intestinal activity through 24 h and hepatocyte uptake is high or relatively preserved. Studies are interpreted as compatible with hepatocellular damage when there is impaired hepatocyte uptake with a delayed hepatobiliary transit time, and radiotracer reaches the intestine (42).
Gastric Emptying, Gastroesophageal Reflux, Aspiration, and Esophageal Transit
Gastroesophageal reflux is common in babies. Some infants with gastroesophageal reflux present with respiratory symptoms including apnea, recurrent pneumonia, wheezing, recurrent stridor, and a chronic cough. Radionuclide assessment of gastric emptying is physiologic, noninvasive, and safe. Using a liquid meal, this method is useful both for assessment of gastric emptying and for diagnosis of gastroesophageal reflux. Radionuclide assessment of gastroesophageal reflux has advantages over conventional upper gastrointestinal series, as it is more sensitive, delivers a lower radiation exposure, monitors events continuously, does not involve administration of contrast agents, and does not require abdominal compressions.
In newborns and infants, gastric emptying is evaluated using milk or formula mixed with 99mTc-sulfur colloid. The patient should have been fasting for approximately 4 h before the study. Barium studies should not be obtained within 48 h before scintigraphy. The total volume of milk or formula to be administered should be approximately the same as the child usually takes in a regular meal. The tracer is mixed with approximately a third of the total expected volume to be administered, followed by the nonradioactive volume. Oral feeding is preferred, but depending on the patient's condition, labeled formula or milk can be administered by nasogastric or gastrostomy tube.
The patient is placed supine on the imaging table, and the γ-camera, equipped with a low-energy high-resolution collimator, is placed under the table. An initial image is used to determine the amount of tracer that has already left the stomach. Dynamic imaging is obtained for 60 min at 1 frame/min. Regions of interest are marked over the stomach, the esophagus, the rest of the abdomen, and the background, and time–activity curves, corrected for decay, are obtained. The gastric time–activity curve after placement of the child on the examining table provides another measure of gastric emptying time. Results are expressed as the percentage of the initial activity in the stomach (43–45). There are no widely accepted standards of normal gastric emptying times in this group. A study on healthy infants revealed a 1-h normal gastric residual of 48%–70%. Another study showed that gastric emptying at 1 h in children younger than 2 y was 27%–81%. Gastric emptying time is affected by the position of the patient and other factors such as type and volume of food, patient anxiety, and pain. Children that show slow gastric emptying when supine can show rapid emptying by a simple change of position (45).
Pulmonary Aspiration
Pulmonary aspiration is a frequent cause of lung disease in children. Common imaging studies have traditionally included barium swallows, upper gastrointestinal series, and gastroesophageal scintigraphy, but these approaches have shown low diagnostic yield for aspiration. The infrequent demonstration of aspiration with radionuclide gastroesophageal studies suggests that aspiration of gastric contents is not a common mechanism of recurrent pulmonary infections in children.
The radionuclide salivagram has proven to be a sensitive and simple method for detecting pulmonary aspiration of oral contents (46–48). No patient preparation is required for the salivagram. A single drop of saline containing 3.7–7.4 MBq (0.100–0.200 mCi) of 99mTc-sulfur colloid is delivered sublingually. Typically, the patient is imaged for 60 min at 1 frame/min in the supine position with a γ-camera viewing the posterior aspect of the head, thorax, and abdomen. It is important to identify any contamination of tracer outside the patient's body that could be confused with aspiration.
Images are inspected for tracer aspiration into the proximal or distal airways (Fig. 11). Tracer deposited in the proximal airways tends to be cleared rapidly by the mucociliary epithelium, whereas tracer reaching the distal airways (where there is less ciliary epithelium) persists longer. By assessing the ability to clear aspirated secretions, the radionuclide salivagram evaluates the functional integrity of airway-protective mechanisms. Aspiration into the distal airways suggests a more worrisome outcome than proximal aspiration. The whole-body radiation dose to the patient is approximately 0.05 mSv (5 mrem), whereas barium swallow delivers a higher radiation dose (46–48).
Gastrointestinal Bleeding
The most common cause of lower gastrointestinal bleeding in previously healthy infants is from a Meckel's diverticulum. This is a congenital abnormality of the antimesenteric duct that is the result of incomplete closure of the embryonic vitelline or omphalomesenteric duct. Bleeding from a Meckel's diverticulum is a result of peptic ulceration of ileal mucosa either in the diverticulum or in the adjacent ileum. The typical presentation of a bleeding Meckel's diverticulum is painless rectal bleeding of bright red blood. Most infants presenting with bright red bleeding from a Meckel's diverticulum contain functioning ectopic gastric mucosa (49–52).
Premedication with perchlorate should be avoided because it suppresses tracer uptake in gastric mucosa. Scintigraphy is performed with 99mTc-pertechnetate injected intravenously (9.25 MBq [0.25 mCi]). The patient lies supine with the γ-camera viewing the lower thorax, abdomen, and pelvis. If indicated, a radionuclide angiogram should be obtained to evaluate for vascular malformations. Recording is set at 1- to 5-s frames for 60 s, followed immediately with a series of 1-min images for 30 min (52).
The scintigraphic appearance of a Meckel's diverticulum is typical. A well-defined area of increased radiopertechnetate uptake, usually in the right lower quadrant, is seen. This area can move and be located anywhere in the abdomen. Pertechnetate uptake in the ectopic gastric mucosa appears at the same time as activity in the stomach and shows increasing intensity over time. Lateral and posterior images are helpful to differentiate this activity from tracer within the ureters and the bladder. Because a Meckel's diverticulum could lie near the urinary bladder, preventing detection, obtaining other views or ensuring that the bladder is empty helps detection (50). If the initial 30-min study appears normal in newborns or young infants and a Meckel's diverticulum is strongly suspected, the physician may consider obtaining additional images 1–2 h later to achieve higher lesion-to-background ratios. If the index of suspicion is high, a scan repeated a few months later may reveal functioning ectopic mucosa.
The accuracy of pertechnetate for the diagnosis of Meckel's diverticulum in surgical patients is 90%, the overall sensitivity is 85%, and the specificity is 95% (51). Enteric duplication, duplication cysts, and gastrogenic cysts are other abnormalities that may contain ectopic gastric mucosa. Other conditions, not containing gastric mucosa, that may accumulate pertechnetate include intestinal obstruction or intussusception, inflammation, vascular malformations, ulcers, some tumors, and various urinary tract abnormalities, but these are usually easy to recognize (52).
Spleen
Splenic scintigraphy using 99mTc-heat-damaged red blood cells is an effective method in the diagnosis of functional asplenia or polysplenia in these small patients (53).
Kidneys
Radionuclide evaluation of renal function is one the most frequent indications in children during the first year of life. Clinical applications of these methods include obstruction, postoperative assessment, pyelonephritis, duplication, ectopic kidney or nonvisualization of a kidney, renal vein thrombosis, acute renal failure, multicystic dysplastic kidney, reflux, trauma, transplant, and tumors (abdominal or renal) (54). The 2 principal radiopharmaceuticals used in this group of patients are 99mTc-MAG3 to evaluate tubular function, split renal function, and collecting system drainage and 99mTc-DMSA to image the renal cortex to assess split renal function and cortical defects.
99mTc-MAG3 is the agent of choice for dynamic renal scintigraphy and can effectively evaluate renal function and urine drainage. After intravenous injection, 99mTc-MAG3 is eliminated rapidly by the kidneys, mainly by active tubular cell transport. The plasma clearance of this tracer is approximately 300 mL/min, and after 3 h approximately 90% of the injected dose can be recovered in urine. 99mTc-MAG3 has a high initial renal uptake and provides a high kidney-to-background ratio and high temporal resolution. An intravenous line is secured and connected to a saline infusion at 10 mL/kg, which is maintained throughout the study. A bladder catheter is inserted if urinary tract obstruction is suspected in patients with vesicoureteric reflux, ureteropelvic obstruction, or ureterovesical obstruction or in young patients who may not be able to void on command. The γ-camera, equipped with a parallel-hole high- or ultra-high-resolution collimator is placed under the examination table viewing the kidneys and the bladder. The usual administered dose of 99mTc-MAG3 is 3.7 MBq (0.1 mCi)/kg in approximately 0.1 mL given intravenously. The study is recorded for 20 min at 1–4 frames/min on a 128 × 128 matrix format. If, at the end of the 20-min acquisition period, spontaneous urine drainage from the pelvicalyceal system is not seen, the infant is picked up and held upright for a few minutes to promote drainage. Then, a static image is obtained and is compared with the previous images to determine postural drainage. If drainage is adequate, the study is ended. If the image still shows tracer retained in the pelvicalyceal system, a diuretic challenge may be considered. Furosemide (1 mg/kg) is administered intravenously while an additional 30-min dynamic study is recorded at 1–4 frames/min. The study is typically viewed at 1 frame/min, and time–activity curves are generated from regions over the kidneys (whole kidney, cortex) and background. A normal dynamic renal study reveals rapid and intense tracer concentration in the renal cortex at approximately 2 min after injection. Passage of the tracer into the renal pelvicalyceal system normally occurs at 2–4 min. The parenchymal phase (60–120 s) provides information about total renal function, relative renal size, position, configuration, and split renal function. The parenchymal image can reveal hydronephrosis, a duplex kidney, or other relatively large anatomic abnormalities. During the drainage phase, the tracer is gradually eliminated into the bladder through the pelvicalyceal system and ureters. By 20 min, a large proportion of the tracer has left the kidneys and less than 30% of the peak activity may remain in the renal region. If the patient moved during the study, motion correction is applied. Images and time–activity curves are also generated after the administration of furosemide. These curves are evaluated to determine whether there is obstruction, intermediate drainage, or no obstruction (Figs. 12 and 13).
During the first months of life, renal immaturity affects the handling of 99mTc-MAG3, and interpretation of studies must take this factor into account. Immature kidneys show relatively low uptake of 99mTc-MAG3 and slower cortical transit times, with prolonged time to peak and increased residual tracer at 20 min. As renal maturation takes place, function matures and handling of 99mTc-MAG3 approaches adult levels. In renal immaturity, 99mTc-MAG3 studies may over- or underestimate the true contribution of a hydronephrotic kidney to total renal function. In these cases, a 99mTc-DMSA scan provides a better assessment of differential renal parenchymal function.
99mTc-DMSA is the radiopharmaceutical of choice for renal cortical scintigraphy. It is 90% bound to plasma proteins, and 0%–5% is bound to red cells. Its blood disappearance follows a single exponential with a mean half-time of 60 min, and 6%–9% of the dose is present in the blood at 14 h after injection. Renal uptake of 99mTc-DMSA is 40%–50% of the injected activity at 3–4 h after injection. Activity is found principally in the proximal convoluted tubules, with minimal activity elsewhere in the kidney. Renal uptake of 99mTc-DMSA is closely correlated with regional renal blood flow and renal mass (55–57). 99mTc-DMSA is administered intravenously 3–4 h before imaging. The usual administered dose is 1.85 MBq (0.05 mCi)/kg. In children younger than 1 y, planar and pinhole images are obtained. The planar images are useful for evaluating the position and overall configuration of the kidneys and for calculating split renal function. The pinhole images provide higher resolution to detect focal cortical defects. The patient is examined supine, and a posterior image containing 300,000–500,000 counts on a 256 × 256 matrix is recorded using a high-resolution or an ultra-high-resolution collimator. Posterior images with the patient prone are obtained with the pinhole collimator (2- to 4-mm aperture), which provides resolution of approximately 2–3 mm. Cortical defects in pyelonephritis, infarction, scarring, duplication, and fetal lobations can be discerned better with pinhole magnification than with parallel-hole collimators. Each pinhole image is obtained for approximately 150,000 counts in a 256 × 256 matrix and requires approximately 10 min. The normal differential renal uptake of 99mTc-DMSA is 50% ± 5% (54). A small amount of 99mTc-DMSA is eliminated in the urine, and in children with hydronephrosis unrecognized pooling of tracer in the pelvicalyceal system could lead to a misleading assessment of differential renal function. In these cases, an image obtained 12–24 h after tracer administration, allowing for a more complete drainage of retained tracer, should permit a better estimate of differential renal function. Also, a 99mTc-DMSA image is useful when 99mTc-MAG3 does not reveal much renal function.
After prenatal diagnosis of hydronephrosis, there are important clinical questions that should be answered so that appropriate treatment can be instituted. These questions include how much function the hydronephrotic kidney has, whether renal obstruction is present, and whether the problem is unilateral or bilateral. 99mTc-MAG3 renography can help answer these questions rapidly and efficiently. Early 99mTc-MAG3 studies on the newborn or infant with hydronephrosis may reveal obstruction, an intermediate pattern, or a dilatation of the pelvicalyceal system without obstruction. Early findings are not generally predictive of the ultimate outcome but are considered a useful snapshot in time during renal maturation (54–56,58–64). If immediate surgical intervention is not indicated after the initial study, reassessment in 3–6 mo should be considered. Subsequent studies may indicate spontaneous resolution or progression of the disorder. Patients with hydronephrosis and normal or nearly normal cortical transit times have a high likelihood of recovery of renal function postoperatively, whereas prolonged cortical transit times are associated with a poorer prognosis for functional improvement. Postoperative dynamic renal studies are useful for patients who have undergone pyeloplasty or another intervention for the relief of obstruction. Postoperative studies should be performed approximately 2 wk after surgery to reduce the effects of postoperative swelling and edema. Postoperative 99mTc-MAG3 studies may reveal a reduction in the size of the previously hydronephrotic kidney with return to more normal function. However, the dilated system may not be completely resolved. Although postoperative 99mTc-MAG3 studies may show residual obstruction, a diuretic challenge may help separate those patients in whom the surgery successfully relieved the obstruction from those with residual obstruction.
If prenatal ultrasonography demonstrates a single or solitary kidney, a 99mTc-DMSA scan can be useful in confirming the presence of a solitary kidney or in identifying a previously undetected ectopic kidney.
Pyelonephritis in young children may present with symptoms that vary and can be confusing. Differentiation of upper from lower urinary tract infection on clinical grounds alone is not easy. Neonates and infants, in particular, present with nonspecific clinical findings such as fever, abdominal pain or tenderness, malaise, irritability, or leukocytosis with or without bacteriuria, but there may be no indication of renal infection. 99mTc-DMSA scintigraphy is the most reliable, simplest, and most sensitive method for the early diagnosis of focal or diffuse functional cortical damage. This test can identify the degree of renal damage and assess recovery of function after treatment. The high sensitivity of 99mTc-DMSA scintigraphy for the diagnosis of pyelonephritis is well established. Sensitivity is 96%, and specificity is better than 98% (62–64).
Renal duplication can be detected by 99mTc-DMSA scintigraphy. Duplication may not easily be seen with high-resolution collimators, but pinhole magnification can clearly reveal this abnormality and evaluate function of the different moieties.
Multicystic dysplasia of the kidney is the most frequent cystic disorder of infants. In approximately 75% of the cases, 99mTc-DMSA scintigraphy does not demonstrate tracer uptake in the involved kidney. However, in some instances some variable tracer uptake can be seen in regions of distorted renal cortex. Often this uptake can be demonstrated only with computer enhancement. The scintigraphic appearance of multicystic dysplasia of the kidney can occasionally be similar to neonatal hydronephrosis.
In anuric newborns during the first day of life, radionuclide studies can demonstrate the presence of functional renal units even before the baby is able to urinate.
Vesicoureteral Reflux (VUR)
VUR is due to a failure of the ureterovesical valve mechanism caused by a congenital variation, a pathologic process, an infection, or immaturity that distorts the anatomy or function (or both) of the ureterovesical junction. The incidence of VUR in apparently healthy neonates and infants is approximately 1%. The diagnosis, follow-up, and treatment of VUR is based on the premise that in children with a history of urinary tract infection, persistent and untreated VUR may lead to renal damage. Most VUR tends to disappear with maturation. Spontaneous cessation of reflux occurs in approximately 70% of children and 79% of ureters. Spontaneous resolution of reflux is more frequent with low-grade reflux and is less frequent with high grades of reflux, although spontaneous resolution can occur even with high grades of reflux (65,66). Radionuclide cystography (RNC) can be used effectively in children younger than 1 y (67).
The indications for RNC in children less than 1 y old include the following: an initial diagnosis of reflux in patients with signs and symptoms of urinary tract infection, follow-up of an earlier diagnosis of reflux to assess spontaneous cessation, diagnosis of familial reflux, and postoperative assessment of antireflux surgery.
RNC is sensitive, because it provides continuous monitoring both during filling of the bladder and during voiding and has a high contrast, enabling detection of reflux as slight as 0.2 mL (68,69). There is no need for patient preparation or sedation. Sterile urethral catheterization trays containing all needed materials are prepared for each study in advance. The patient is positioned supine on the examination table, which is covered with absorbent plastic-lined material to collect any contamination. After appropriate cleansing of the perineum and genitals with antiseptic solution, a small-diameter catheter covered with anesthetic jelly is introduced into the patient's bladder. Careful and gentle catheterization technique by experienced personnel is essential to ensure consistently successful studies. 99mTc-pertechnetate (37 MBq [1 mCi]) is introduced into the bladder, followed by a saline infusion with the saline bag kept 70–90 cm above the bladder. Serial posterior images are obtained with the γ-camera (0.25-s frames at a 128 × 128 matrix) until the patient's functional bladder capacity is reached and the patient is able to void either around the catheter or immediately after catheter removal (69). If the patient cannot void or does not void after complete bladder filling, a postvoid image is obtained. This step is important because approximately 20% of reflux is detected only during voiding. Residual bladder volume, voided volume, bladder capacity, and any problems during the catheterization should be documented.
Reflux can be detected during one or all phases: early, mid, or late filling; voiding; and after voiding. The international classification of reflux by vesicoureterogram includes 5 grades (I–V) of severity (65). RNC classification includes 3° of increasing severity: RNC grade 1 refers to reflux within the distal ureter only and not reaching the renal pelvis (equivalent to vesicoureterogram grade I). Typically, 0.2 mL of reflux that projects at least 2 cm away from the projected border of the bladder can be detected. RNC grade 2 refers to reflux that reaches the renal pelvis, without pelvic dilatation and without a dilated or redundant ureter (equivalent to vesicoureterogram grades II and III). RNC grade 3, the most severe, shows reflux into the renal pelvis with significant dilatation of the pelvis or tortuosity of the ureter (equivalent to vesicoureterogram grades IV and V) (68).
Approximately two thirds of all VUR resolves spontaneously as a result of growth and maturation. Low-grade VUR varies from examination to examination. VUR varies also with bladder volume, voiding or filling, patient position, and level of anxiety. Thus, a single normal examination may not be sufficient to ensure that reflux has ceased. Serial cystograms over several (6–12) months can be more reassuring, and in the small number of patients with apparently resolved VUR, another urinary infection may be an indication for a repeated RNC.
The incidence of VUR in siblings of children with reflux is approximately 40%. We have found that reflux severity in these asymptomatic children varies widely from mild to severe (68–70). The incidence of renal cortical damage in these siblings with VUR is high (68,71–74). Thus, when VUR is diagnosed in a child, evaluation for VUR is appropriate in the siblings (74,75).
Radiation exposure in children undergoing RNC was estimated at 0.18 mGy (18 mrad) to the bladder wall and at 0.01 mGy (1 mrad) to the ovaries. The testicular dose is lower than the dose to the ovaries. The dose to the ovaries is much less with RNC than with a conventional vesicoureterogram, even with the use of pulsed fluoroscopy.
Bone
Indications for skeletal scintigraphy in children under 1 y old include osteomyelitis, trauma, avascular necrosis, Langerhans cell histiocytosis, fever of unknown origin, and neoplastic disease (76). Skeletal planar scintigraphy and SPECT with 99mTc-MDP, as well as 18F-NaF PET, are useful imaging tools. The usual administered dose of 99mTc-MDP is 37 MBq (1 mCi). The usual 18F-NaF PET dose is 18.5 MBq (0.5 mCi). PET is acquired 15–30 min after the intravenous injection. The effective dose for 18F-NaF PET for a 1-y-old patient is estimated at 3.1 mSv (310 mrem) (77).
Image interpretation in infants requires special attention due to the differences in radiopharmaceutical biodistribution between immature children and adults. In infants, an important consideration is the absence of 99mTc-MDP or 18F-NaF uptake in structures that have not yet ossified. For example, the femoral capital epiphysis ossifies between the ages of 2 and 7 mo, and absence of uptake should not be confused with avascular necrosis. The tarsal navicular bone usually ossifies after the age of 1 y, and radiotracer uptake is absent in infants. The cartilaginous junction between the inferior pubic ramus and the ischium (ischiopubic synchondrosis) ossifies after the age of 4 y. Therefore, in infants, there is physiologic discontinuity of the inferior pubic ramus. In contrast, there is high radiotracer uptake in the physeal growth centers of long bones and the apophyseal growth centers of flat and irregular bones. Infants have physiologically increased uptake in the apophyses at the inferior tip of the scapula and the vertebral transverse process. Normally, there is high tracer uptake in the skull base, orbits, nasal region, temporomandibular joints, and cranial sutures, whereas absence of uptake may be seen in the region of the anterior fontanelle. As growth centers close, the biodistribution of 99mTc-MDP gradually reaches the adult pattern (76–84). One special indication in this age group is evaluation for skeletal injury in suspected cases of child abuse (85).
Infection and Inflammation
Infection is a common indication for skeletal scintigraphy in infants. Early diagnosis and treatment of osteomyelitis is important to avoid complications such as sepsis, growth arrest, and skeletal injury. Skeletal scintigraphy is a reliable tool for the diagnosis of osteomyelitis even in the youngest patients (79). Three-phase skeletal scintigraphy can demonstrate abnormalities within 24–72 h of symptom onset and is both sensitive and specific for osteomyelitis (Fig. 14) (76). Whole-body imaging is helpful because symptoms may not be localized and findings may be multifocal. Diagnostic aspiration and antibiotic treatment do not need to be delayed, because the appearance on scintigraphy does not change rapidly. Photopenic defects may indicate regional ischemia.
Septic arthritis or inflammation of a joint typically involves the hip or knee in children less than 3 y old. Timely diagnosis is important to avoid complications. Skeletal-phase images may be normal or demonstrate mildly increased juxtaarticular uptake. Reduced tracer uptake may occasionally be seen, such as in the presence of a tense joint effusion. In cellulitis, skeletal scintigraphy reveals increased regional blood flow and increased tracer accumulation in the soft tissue in the region involved and a normal or slightly and diffusely increased tracer concentration in the adjoining bones (Fig. 15).
Fever of Unknown Origin
The term fever of unknown origin is defined as recurrent fever of 101°F (38.3°C) or higher, lasting 2–3 wk or longer with no specific diagnosis after 1 wk of inpatient investigation. The differential diagnosis of fever of unknown origin in infants is broad and includes infection, connective tissue disease, and neoplasm. Skeletal scintigraphy is a sensitive tool for the detection of osseous disease and is often requested in the evaluation of infants with fever of unknown origin. However, in our experience, skeletal scintigraphy usually adds little diagnostic information, without localizing findings, in the assessment of infants presenting to the emergency department with fever.
If skeletal scintigraphy is performed, the physician should pay close attention to subtle soft-tissue uptake, which may identify the location of an occult inflammatory or neoplastic process. 18F-FDG PET has the potential to identify soft-tissue lesions (inflammation, tumors).
Trauma and Nonaccidental Injury
In infants with suspected trauma, the history of injury is not always available and clinical localizing findings may not be present. Furthermore, trauma from suspected nonaccidental injury often results in multiple skeletal fractures at different stages of healing. Skeletal scintigraphy offers a highly sensitive imaging tool for evaluation of the entire skeleton (84). Osseous stress or fracture may be identified on skeletal scintigraphy by increased radiotracer uptake within a few hours of symptom onset and before radiographic changes appear (80,81). Although MRI is sensitive for the detection of osseous disease and does not expose the infant to ionizing radiation, it has a limited field of view and usually requires sedation. Skeletal trauma and fractures are a common presentation in infants suspected of child abuse. The detection of all fractures is essential in this clinical setting because the discovery of new fractures has the potential to alter the medicolegal management of the patient. Skeletal scintigraphy is established as a study complementary to the radiographic skeletal survey, and some authors have recommended that both studies be done (82,83). The sensitivity of standard 99mTc-MDP skeletal scintigraphy for the detection of fractures varies greatly depending on the location of the lesion (Fig. 16) (78,83).
18F-NaF PET has recently been proposed as an alternative to 99mTc-MDP in the evaluation of child abuse. 18F-NaF has less protein binding in blood than 99mTc-MDP, allowing imaging at about 15 min after administration (77,85). Both 99mTc-MDP and 18F-NaF PET are effective in child abuse and show an overall higher sensitivity than an initial radiographic skeletal survey. 18F-NaF PET was found to have a sensitivity of 92% for the detection of rib fractures, versus 68% by skeletal survey (85). Comparative studies of the relative sensitivity and specificity of 99mTc-MDP, planar imaging, SPECT, and 18F-NaF PET in children less than 1 y old are not available.
Oncology
Oncologic disorders of infancy account for approximately 10% of all cancer diagnoses in children younger than 15 y (86). It is estimated that the peak age for pediatric cancer is the first year of life. Neuroblastoma is the most commonly diagnosed tumor in this age range, followed by leukemia and brain tumors, germ cell tumors, and other soft-tissue tumors (Fig. 17) (87). Nuclear medicine is an effective way to diagnose, stage, and survey cancer in infants (88). The radiopharmaceuticals used for planar oncologic scintigraphy include 99mTc-MDP, 123I-MIBG, and 99mTc-MIBI. SPECT provides increased contrast, and newer reconstruction methods provide improved spatial resolution (10–12). Furthermore, SPECT may be fused with MRI or CT for anatomic localization. 18F-FDG PET is increasingly being used in pediatric oncology, although the exact role of PET in infants has not yet been established.
Neuroblastoma
Neuroblastoma, ganglioneuroblastoma, and ganglioneuroma are tumors of the sympathetic nervous system. These tumors arise from the neural crest cells that form the sympathetic tissue. Neuroblastoma is the most frequently diagnosed malignancy within the first few months of life. Neuroblastoma most commonly involves the adrenal glands, or it may be associated with sympathetic nerves in the cervical, posterior mediastinal, retroperitoneal, and presacral regions. Less common sites of disease include the thymus, lung, kidney, anterior mediastinum, stomach, and cauda equina (88). Disseminated disease is seen in approximately 70% of neuroblastoma cases at diagnosis (89).
123I-MIBG and 99mTc-MDP skeletal scintigraphy play a key role in the diagnosis, staging, and follow-up evaluation of infants with neuroblastoma (90,91). 123I-MIBG scintigraphy and SPECT are both sensitive and specific for the detection of primary and metastatic neuroblastoma. 123I-MIBG is highly specific in the diagnosis of neuroendocrine tumors. Physiologic 123I-MIBG uptake can include the salivary glands, myocardium, liver, spleen, adrenal glands, skeletal muscle, brown adipose tissue, bladder, and colon (91,92).
Although most neuroblastoma tumors are 123I-MIBG–positive, a normal 123I-MIBG scintigraphy result does not completely exclude malignancy. 99mTc-MDP scintigraphy provides additional helpful findings in the evaluation of children with neuroblastoma. 99mTc-MDP is taken up by soft-tissue disease but is predominantly taken up by cortical metastases (91,93). 18F-FDG PET may be used in the evaluation of infants with neuroblastoma. 18F-FDG is avidly taken up at sites of disease, and recent reports suggest that 18F-FDG PET and 123I-MIBG scintigraphy may provide complementary information. 123I-MIBG is specific and superior to 18F-FDG PET for the detection of neuroblastoma, but 18F-FDG PET may be helpful in the evaluation of tumors that weakly accumulate 123I-MIBG (92). The role of 18F-FDG PET or 18F-FDG PET/CT in the evaluation of infants with neuroblastoma has not been established.
Antihypertensive medication, tricyclic antidepressants, and sympathomimetics should be discontinued, if possible, because they interfere with 123I-MIBG uptake. A supersaturated solution of potassium iodide should be prescribed to reduce 123I uptake by the thyroid gland. Specifically, the infant is given 1 drop of a solution of potassium iodide by mouth 3 times a day starting the day before 123I-MIBG injection and continuing for 3 d. If the infant is allergic to iodine, 200 mg of orally administered potassium perchlorate are recommended, starting the day before 123I-MIBG injection and continuing for 3 d. The usual dose of 123I-MIBG is 37 MBq (1 mCi). 123I-MIBG is administered by slow intravenous injection over 2 min. Whole-body planar scintigraphy is normally obtained 24 h after tracer injection. SPECT of the neck, thorax, abdomen, and pelvis is routinely obtained at our institution. It is estimated that the effective dose for a 123I-MIBG scan in a 1-y-old child is 2.5 mSv (250 mrem) with the above-administered dose (94).
Standard treatment of neuroblastoma may include surgery, chemotherapy, and radiation depending on the stage and biologic factors of the tumor (88,89). 131I-MIBG radiotherapy has been used as palliative therapy in children with neuroblastoma.
Other tumors of infancy include Wilms tumors, germ cell tumors, soft-tissue tumors, lymphoma, and hepatic tumors, among others. Wilms tumor (or nephroblastoma) is the most common pediatric renal malignancy (95). The typical clinical presentation for these tumors is an asymptomatic abdominal mass. Anatomic imaging is commonly performed first in the evaluation of these patients. Scintigraphy and PET are infrequently used (89) but may identify sites of disease (96,97). The role of 18F-FDG PET in infants with soft-tissue tumors has not yet been established.
Acknowledgments
We acknowledge the following individuals: Chris Durall for tireless work on the multiple versions of the manuscript; Nancy Drinan for impeccable editorial work; Michael Stabin for dose estimates; S. James Adelstein for radiation risk estimates; Alan Packard, Pat Gallaher, and Justin Paolino for radiopharmaceutical and contrast agent calculations of volume and mass; Chuck Hornberger for database searches; Joanne Louis, Vicky Kourmouzi, and the technologist staff for their dedication and meticulous technical expertise; and our nurses and child life specialists for their expert care.
Footnotes
Learning Objectives: On successful completion of this activity, participants should be able to describe, for children less than 1 y old, (1) the wide variety of nuclear medicine indications; (2) the nuclear medicine studies, relative to other imaging methods, that can provide the highest diagnostic yield; (3) how to obtain high-quality nuclear medicine examinations at the lowest radiation exposures; (4) how growth and development affect the appearance, and therefore the interpretation, of nuclear medicine images; and (5) special care needed, such as patient preparation, immobilization, distraction techniques, environment, sedation or general anesthesia, imaging techniques, image processing, and presentation of results.
Financial Disclosure: The authors of this article have indicated no relevant relationships that could be perceived as a real or apparent conflict of interest.
CME Credit: SNM is accredited by the Accreditation Council for Continuing Medical Education (ACCME) to sponsor continuing education for physicians. SNM designates each JNM continuing education article for a maximum of 1.0 AMA PRA Category 1 Credit. Physicians should claim only credit commensurate with the extent of their participation in the activity.
For CE credit, participants can access this activity through the SNM Web site (http://www.snm.org/ce_online) through June 2012.
- © 2011 by Society of Nuclear Medicine
REFERENCES
- 1.↵
- 2.↵
- 3.↵
- 4.↵
- 5.↵
- 6.↵
- 7.↵
- 8.↵
- 9.↵
- 10.↵
- 11.
- 12.↵
- 13.
- 14.↵
- 15.↵
- 16.↵
- 17.↵
- 18.
- 19.↵
- 20.↵
- 21.↵
- 22.↵
- 23.↵
- 24.↵
- 25.↵
- 26.↵
- 27.↵
- 28.↵
- 29.↵
- 30.↵
- 31.↵
- 32.↵
- 33.↵
- 34.↵
- 35.↵
- 36.↵
- 37.↵
- 38.↵
- 39.↵
- 40.↵
- 41.↵
- 42.↵
- 43.↵
- 44.
- 45.↵
- 46.↵
- 47.
- 48.↵
- 49.↵
- 50.↵
- 51.↵
- 52.↵
- 53.↵
- 54.↵
- 55.↵
- 56.↵
- 57.↵
- 58.↵
- 59.
- 60.
- 61.
- 62.↵
- 63.
- 64.↵
- 65.↵
- 66.↵
- 67.↵
- 68.↵
- 69.↵
- 70.↵
- 71.↵
- 72.
- 73.
- 74.↵
- 75.↵
- 76.↵
- 77.↵
- 78.↵
- 79.↵
- 80.↵
- 81.↵
- 82.↵
- 83.↵
- 84.↵
- 85.↵
- 86.↵
- 87.↵
- 88.↵
- 89.↵
- 90.↵
- 91.↵
- 92.↵
- 93.↵
- 94.↵
- 95.↵
- 96.↵
- 97.↵
- Received for publication March 16, 2010.
- Accepted for publication April 5, 2011.