Abstract
We proposed and tested a novel geometry for PET system design analogous to pinhole SPECT called the virtual-pinhole PET (VP-PET) geometry to determine whether it could provide high-resolution images. Methods: We analyzed the effects of photon acolinearity and detector sizes on system resolution and extended the empiric formula for reconstructed image resolution of conventional PET proposed earlier to predict the resolutions of VP-PET. To measure the system resolution of VP-PET, we recorded coincidence events as a 22Na point source was stepped across the coincidence line of response between 2 detectors made from identical arrays of 12 × 12 lutetium oxyorthosilicate crystals (each measuring 1.51 × 1.51 × 10 mm3) separated by 565 mm. To measure reconstructed image resolution, we built 4 VP-PET systems using 4 types of detectors (width, 1.51–6.4 mm) and imaged 4 point sources of 64Cu (half-life = 12.7 h to allow a long acquisition time). Tangential and radial resolutions were measured and averaged for each source and each system. We then imaged a polystyrene plastic phantom representing a 2.5-cm-thick cross-section of isolated breast volume. The phantom was filled with an aqueous solution of 64Cu (713 kBq/mL) in which the following were imbedded: 4 spheric tumors ranging from 1.8 to 12.6 mm in inner diameter (ID), 6 micropipettes (0.7- or 1.1-mm ID filled with 64Cu at 5×, 20×, or 50× background), and a 10.0-mm outer-diameter cold lesion. Results: The shape and measured full width at half maximum of the line spread functions agree well with the predicted values. Measured reconstructed image resolution (2.40–3.24 mm) was ±6% of the predicted value for 3 of the 4 systems. In one case, the difference was 12.6%, possibly due to underestimation of the block effect from the low-resolution detector. In phantom experiments, all spheric tumors were detected. Small line sources were detected if the activity concentration is at least 20× background. Conclusion: We have developed and characterized a novel geometry for PET. A PET system following the VP-PET geometry provides high-resolution images for objects near the system's high-resolution detectors. This geometry may lead to the development of special-purpose PET systems or resolution-enhancing insert devices for conventional PET scanners.
PET has evolved from a research tool for studying neurologic and cardiac functions of humans (1) to a clinical diagnostic tool for cancer patients (2), particularly since the introduction of PET/CT technology (3). With the introduction of high-resolution animal PET scanners in the mid-1990s (4,5), PET became a driving force behind molecular imaging through in vivo imaging of small animals using positron-emitting radionuclide-labeled biomolecules (6).
Resolution of PET is limited by the positron range of the radionuclide, acolinearity of the annihilation γ-rays, and intrinsic spatial resolution of the detectors. For whole-body PET scanners with large diameters, the blurring of image resolution due to the acolinearity effect is approximately 2-mm full width at half maximum (FWHM) (7,8), whereas for animal PET scanners this effect may be as small as 0.2-mm FWHM. When radionuclides of small positron range (such as 18F and 64Cu) are used, the image resolution of most animal PET scanners is predominantly limited by the detector's intrinsic spatial resolution (9).
Although there have been significant improvements in the performance of PET scanners through new scintillation material and better detector design (10), fundamental PET scanner geometry remains unchanged—that is, a group of γ-ray detectors of the same size is arranged into multiple rings or planes to detect annihilation γ-rays along different angles through coincidence detection. There is maximal sampling density and minimal parallax error at the center of the field of view (FOV)—hence, the best image resolution is there. Several groups have demonstrated that sampling uniformity and the reconstructed image resolution can be improved by wobbling the system or the object being imaged (11,12) or by introducing a small offset between the center of detector pairs and the center of rotation (COR) of the system (13). However, the geometry of these systems does not deviate from the conventional PET geometry significantly.
This work describes a new approach to PET system geometry, the virtual-pinhole PET (VP-PET), derived from the well-known pinhole SPECT geometry. We first describe the theory of operation for VP-PET. We then analyze the intrinsic spatial resolution of the VP-PET system and validate the formula through experiments. We further measure image resolution of 4 VP-PET systems through experiments. Finally, we imaged a phantom to demonstrate the potential of the VP-PET geometry in future development of special-purpose PET devices.
MATERIALS AND METHODS
A typical pinhole SPECT system consists of a planar γ-camera, a pinhole collimator, and a rotation mechanism that allows projection data to be collected from multiple angles. Figure 1A illustrates a simplified pinhole SPECT system without its rotation mechanism. Radioactivity distribution in an object is projected through the pinhole collimation onto the camera surface. Magnification (minification) can be achieved if the object-to-pinhole distance (b) is smaller (larger) than the collimator cone length (f), with the magnification factor M = f/b. The image resolution of a pinhole SPECT system can be approximated by the following equations (14,15):Eq. 1Eq. 2where RColli is the collimator resolution, d is the diameter of the pinhole aperture, RDet is the detector intrinsic spatial resolution, and RSystem is the system resolution. When the object-to-pinhole distance (b) is significantly smaller than the collimator cone length (f), the collimator resolution approaches the size of the pinhole aperture, whereas the system resolution becomes less dependent on the detector intrinsic spatial resolution. The use of pinhole collimator(s) permits high-resolution animal SPECT using γ-cameras of low intrinsic spatial resolution (16).
A PET system contains one or more pairs of detectors for coincidence detection. Coincidence detection circuits replace the physical collimator to perform electronic collimation. Although most PET scanners were designed to use multiple detectors of the same dimension in a system, one can also use 2 completely different γ-ray detectors for coincidence detection. Figure 1B illustrates a system with 2 detector arrays of different sizes. Detector array 1 consists of a single detector element of width w1. Detector array 2 consists of multiple detector elements of width w2. The coincidence lines of response between the detector in array 1 and all detector elements in array 2 form a fanbeam geometry. If the detector pair in Figure 1B is rotated around the object located at the COR, coincidence events can be collected from multiple angles and reconstructed using fanbeam reconstruction algorithms to obtain tomographic images.
If one considers the detector array 1 in Figure 1B as an x-ray source and the array 2 as part of a full ring of x-ray detectors, this system approximates the geometry of a fourth-generation (4G) CT scanner (17). The fundamental difference between a VP-PET system and a CT system is that the signal carrier in PET is the annihilation γ-rays originating from the object, whereas for CT the signal carrier is the x-ray flux originating from the x-ray tube. If one considers the fanbeam geometry in Figure 1B, the closer the object is to the detector array 1, the larger the object is projected onto the surface of detector array 2. The magnification factor based solely on geometry is M = (d1 + d2)/d1, which is the same as for a CT system. However, unlike a CT system where the focal spot of an x-ray tube is typically considered to be a point source, the detector in array 1 has a finite width, w1. Therefore, the image resolution of such a PET system is a function of detector widths (w1 and w2) and object-to-detector distances (d1 and d2). This is similar to a pinhole SPECT system whose image resolution depends on several system parameters as described in Equations 1 and 2. Hence, we have named the geometry in Figure 1B the “Virtual-Pinhole” PET geometry and systems that follow this geometry VP-PET systems. We characterize the resolution of a VP-PET system through analyses of the intrinsic spatial resolution between 2 detectors of different dimensions, followed by measurement of image resolution of VP-PET systems of 4 different configurations.
Resolution for VP-PET Geometry
Among the 3 factors (positron range, acolinearity, and detector intrinsic spatial resolution) that limit the spatial resolution of an annihilation coincidence detection system (or a PET system), the latter 2 can depend on the system design and, therefore, so were analyzed for a VP-PET system.
Theoretic Derivation.
The effect on spatial resolution due to acolinearity of the annihilation γ-rays is commonly expressed as R180° = 0.0022·D, where D is the diameter of a PET scanner or the distance between the detectors. This formula represents the FWHM of the blurring profile at the center between 2 detectors due to angular distribution of the annihilation photon acolinearity, which is approximately 0.5° (0.0088 rad) FWHM (18). For the geometry in Figure 1B, the acolinearity effect on spatial resolution can be expressed as (in FWHM):Eq. 3This term can be significantly smaller than the commonly accepted expression, R180° = 0.0022·D, when d1 is significantly smaller than d2—that is, when the object is very close to one of the detectors.
Intrinsic spatial resolution of a pair of detectors of the same dimension (Fig. 2A) has been thoroughly explored (15,19). The detection profile is symmetric around the center between the 2 detectors with its shape approaching a triangle at the center and a rectangle near both ends. For locations other than these 2 positions, the resolution profile approaches a trapezoid function. The width of the base of the detection profiles is defined by the width of the detectors, w, whereas the width of the plateau is defined by the 2 dashed lines connecting the 2 detectors—that is, the width of the plateau changes from zero near the center of the system to w near the edge of the system.
For a coincidence detection system that uses detectors of 2 different sizes (Fig. 2B), the detection profile is asymmetric. The width of the base of the profiles is still defined by the detector widths, but is no longer constant, instead approaching w1 and w2 at the 2 ends. The width of the plateau is still defined by the 2 dashed lines connecting the 2 detectors, but the location where plateau width equals zero is not at the center and is instead near the smaller detector. Given separation D between 2 detectors, it can be shown that the width of the base, wb, the width of the plateau, wp, and the FWHM of the detection profile, Rdet, at distance x from detector 1 are:Eq. 4Eq. 5Eq. 6For a VP-PET system following the geometry in Figure 1B with its COR located at d1 from detector array 1 and d2 from detector array 2, the system resolution near the center of the FOV can be expressed as:Eq. 7where Rsrc is the effective source dimension that includes the positron range effect. Because Equation 7 is derived from 2 individual detectors in Figure 2B, an additional term similar to the “block effect” that was suggested by Moses and Derenzo (7) and Derenzo et al. (20) needs to be included in the quadratic sum in Equation 7 to describe the blurring effect due to the use of block detectors in a VP-PET system. Combining the system resolution in Equation 7 and the empiric formula proposed by Moses and Derenzo that describes the reconstructed image resolution of a conventional PET scanner using cylindric geometry, we obtain the following expression for the reconstructed image resolution of a VP-PET system near its COR:Eq. 8where BE is the block effect and the factor 1.25 accounts for the difference between system resolution and reconstructed image resolution due to insufficient sampling and the use of the filtered-backprojection algorithm.
Intrinsic Spatial Resolution Measurements.
To investigate the intrinsic spatial resolution of a VP-PET system, we set up a coincidence detection system using 2 identical PET detectors separated by 565 mm. This distance represents a potential VP-PET application when high-resolution detectors are positioned at 159 mm from the center of the FOV inside a clinical whole-body PET scanner of 406-mm radius. Each detector contains a lutetium oxyorthosilicate (LSO) array directly coupled to a position-sensitive photomultiplier tube ([PS-PMT] R5900-C12; Hamamatsu Photonics, Japan). The LSO array has 12 × 12 crystals each measuring 1.51 × 1.51 × 10 mm3 with a pitch of 1.6 mm in both directions. The position of an event was determined by a ratiometric positioning algorithm commonly used in microPET detectors (21,22). A sealed 22Na point source was positioned at various distances between the 2 detectors and stepped across the coincidence lines of response with a step size of 0.177 mm. Coincidence events were recorded at each source location. In postprocessing, coincidence events between a single crystal of the left detector module and a group of 4-by-4 crystals of the right detector module were counted and plotted as a function of the point source location. This effectively measured the line spread function of a pair of annihilation coincidence detection detectors of different dimensions (1.51 × 1.51 × 10 mm3 vs. 6.31 × 6.31 × 10 mm3) at different locations along the coincidence line of response. The FWHM and full width at tenth maximum (FWTM) were calculated from each line spread function by linear interpolation between 2 nearest measured values bracketing the half-values and one-tenth values of the peak, respectively.
The FWHM was plotted against the source-to-detector distance and compared with the resolutions predicted by Equation 7. In the calculation, the effective source dimension was assumed to be 0.8-mm FWHM based on its nominal diameter of 0.5 mm and the relatively small positron range for 22Na. Because the dimension and the charge-division readout of our detectors are similar to those used in a commercial animal PET scanner (microPET R4; Siemens Medical Solutions, Inc.), we expect the block effect of these detectors to be similar to that of the microPET scanner. Tomic et al. (23) recently showed that the block effect is approximately 1.7-mm FWHM for a clinical PET scanner (ECAT HR+; Siemens Medical Solutions, Inc.) and 0.8-mm FWHM for the microPET R4 system. Therefore, we assumed a block effect of 0.8-mm FWHM for our detectors and calculated the predicted resolution with and without the inclusion of block effect in Equation 7.
Reconstructed Image Resolution Measurements.
We evaluated the image resolution of VP-PET systems via the experimental setup in Figure 3A to mimic the VP-PET geometry in Figure 1B. Detectors of different sizes were mounted on 2 concentric rotation stages that can be rotated independently for a full 360° to acquire coincidence events from all angles. On the basis of a 2-dimensional coordinate system with its origin located at the COR, a high-resolution detector (array 1) was positioned with its front surface centered at (−147, 0) (in mm), whereas a low-resolution detector (array 2) was positioned with its front surface centered at (0, 400). Detector array 1 (left) was either an array of 8 × 8 LSO crystals each measuring 2.2 × 2.2 × 10 mm3 or an array of 12 × 12 LSO crystals each measuring 1.51 × 1.51 × 10 mm3. Scintillation light was detected by the same type of PS-PMT as already described. These types of detectors are commonly used in high-resolution animal PET systems (22,24). Detector array 2 (right) was either an array of 8 × 8 LSO crystals each measuring 6.4 × 6.4 × 25 mm3 or an array of 13 × 13 LSO crystals each measuring 4.0 × 4.0 × 20 mm3. Scintillation light was detected by 4 units of single-channel PMT. These types of detectors are commonly used in clinical PET/CT scanners.
We constructed 4 point sources using micropipettes of 0.5-mm inner diameter (ID) and 1.0-mm outer diameter (OD). Each source contained approximately 18.5 MBq of 64Cu solution. 64Cu has a half-life of 12.7 h, permitting data to be collected from multiple angles over an extended time. The presence of prompt γ-rays from 64Cu, however, could be in coincidence with the annihilation γ-rays and result in effects that are similar to scatter coincidences that could degrade the FWTM of the resolution measurement slightly. The sources were positioned at (−9, 0), (11, 0), (1, 4), and (1, −6) (in mm). In each of the 4 configurations, the sources were imaged through step-and-acquire motion (Fig. 3B). Detector array 1 was rotated to 48 positions to cover 360° and provide full 2π sinograms. Each time detector array 1 was moved, detector array 2 was moved to 3 positions to cover all 4 point sources. Coincidence events collected at different angles were sorted to form fanbeam sinograms assuming the high- and low-resolution crystals were uniformly distributed along the 2 circles in Figure 3B without considering the depth of interaction.
To reconstruct the images, we implemented a filtered-backprojection algorithm based on the work by Besson (25) that was originally developed for the 4G CT geometry. In contrast to the original algorithm that assumes the x-ray source to be an infinitely small point source, the refined algorithm incorporates the width of the detectors in both arrays 1 and 2 in the calculation of the system matrix to better represent the geometry of VP-PET systems (26). Data were not normalized for detector efficiency nor were corrected for attenuation. The reconstruction filter was a ramp filter cutoff at the Nyquist frequency. From the reconstructed images, tangential and radial resolution were measured for the 4 line sources and then averaged to obtain the average image resolution of VP-PET systems near their centers of FOV.
Image resolution of the 4 VP-PET systems was also calculated using Equation 8. It was assumed that detectors 1 and 2 are separated by a fixed distance of 547 mm. The COR location was varied from the surface of detector 1 to the surface of detector 2. The effective source dimension was assumed to be 0.8-mm FWHM for the 64Cu point sources. The block effect was also assumed to be 0.8-mm FWHM. The calculated image resolution of the 4 VP-PET systems was plotted as a function of COR location between the 2 detectors.
Phantom Study
A piece of polystyrene plastic (20 × 18 × 2.5 cm3) was carved to form a cavity that has a shape that resembles the cross-section of an isolated breast volume. The cavity was filled with 170 mL of aqueous solution of 64Cu (713 kBq/mL) to mimic background activity in tissues. This unusually high-activity concentration was used because the experiment requires 720 detector positions to completely cover the phantom using only 2 detector arrays. The acquisition time at each detector position was very short and, thus, requires high-activity concentration to obtain a reasonable number of counts. Four fillable spheres of different IDs (12.6, 10.0, 5.6, and 1.8 mm) were filled with an aqueous solution of 64Cu at an activity concentration of 3.56 MBq/mL (5× background) and inserted into the cavity to mimic tumors with high uptake. Another fillable sphere of 10.0-mm OD was filled with water to mimic a cold lesion in warm background. Six micropipettes (3 of 1.1-mm ID and 2.0-mm OD, and 3 of 0.7-mm ID and 1.6-mm OD) were filled with an aqueous solution of 64Cu at 3 different activity concentrations (5×, 20×, and 50× background) to mimic 6 5-mm-long line sources. The locations of all tumor inserts and micropipettes in the phantom are shown in Figure 4.
The entire phantom was centered in a VP-PET system similar to Figure 3A. The detector array 1 was based on LSO crystals of 1.51 × 1.51 × 10 mm3. The detector array 2 was based on LSO crystals of 4.0 × 4.0 × 20 mm3. The distances from the COR to the 2 detectors remained 147 and 400 mm, respectively. Coincidence events were collected after the step-and-acquire motion. Detector array 1 was rotated to 48 positions to cover 360°. Each time detector array 1 rotated to a new position, detector array 2 was rotated to 15 positions to cover the entire phantom. It took 15 h to collect data from 720 positions, with the acquisition time at each location adjusted for radioactive decay. Coincidence events collected at different locations were sorted to form fanbeam sinograms. The same filtered-backprojection algorithm described previously was used to reconstruct 20 slices of tomographic images across the phantom. Data were not normalized for detector efficiency nor were attenuation corrected. The reconstruction filter was a ramp filter cutoff at the Nyquist frequency.
RESULTS
Intrinsic Spatial Resolution of VP-PET Geometry
Figure 5A shows the line spread functions measured at different locations between 2 detectors of different dimensions. The shapes agree well with those illustrated in Figure 2B. The profiles are no longer symmetric relative to the center between the 2 detector arrays. The FWHM of the profile approaches detector width w1 (or w2) when the point source was near detector array 1 (or 2).
Figure 5B shows good agreement between the FWHM measured from these line spread functions and those predicted by Equation 7 using various source-to-detector distances (d1 and d2). When the block effect is included in Equation 7, there is little change for the right half of the plot (Fig. 5B) in which system resolution is dominated by the intrinsic resolution of array 2. However, near detector array 1, where the system resolution is roughly 2-mm FWHM, inclusion of the block effect better predicts measured system resolution.
Reconstructed Image Resolution
Figure 6A shows images of four 64Cu point sources obtained from the 4 VP-PET systems. The measured image resolution of the 4 VP-PET configurations ranged from 2.40- to 3.24-mm FWHM (Table 1).
Figure 6B shows the image resolution (FWHM) calculated by Equation 8 for 4 VP-PET systems as a function of the COR location. The block effect of the detector was assumed to be 0.8-mm FWHM and included in the calculation. The solid vertical line denotes the COR that corresponds to our 4 imaging experiments (i.e., d1 = 147 mm, d2 = 400 mm). The predicted image resolution of the 4 VP-PET systems ranges from 2.30- to 2.83-mm FWHM (Table 1).
Equation 8 predicted within 6% the image resolution for 3 of the 4 configurations (cases a, b, and c in Table 1). For case d, the predicted image resolution deviated from the measured value by 12.6% (2.83 vs. 3.24 mm). In this configuration, both w1 and w2 are at their maximum (2.2 vs. 6.4 mm, respectively) compared with the other configurations. Therefore, the 0.8-mm FWHM block effect used in the image resolution calculation was likely to be an underestimation of the actual block effect and may have caused this discrepancy.
Phantom Study
Figure 7 shows 2 image planes through the phantom. Although these 2 images exhibit artifacts due to the lack of attenuation correction, the high-resolution imaging capability of the VP-PET system detected many of the small lesions. All spheric sources, including the 10.0-mm OD cold spheric lesion and the 1.8-mm ID hot lesion are visible. The 1.1- and 0.7-mm ID line sources are visible when the tumor-to-background ratio is 20 or higher.
DISCUSSION
The derivation of Equations 7 and 8 was based on a VP-PET system that uses detectors of 2 different widths, with their front surfaces directly facing each other. It did not consider detectors facing each other at an oblique angle, nor did it factor in the effective depth of γ-ray interactions that depends on the crystal lengths. Furthermore, the derivation did not consider the type of detector material whose photofraction may impact the positioning accuracy in a detector. Therefore, it is not suggested that Equations 7 and 8 can precisely predict the spatial resolution of a VP-PET system. Rather, these equations represent a logical extension of the empiric formula by Moses and Derenzo (7) that estimates the image resolution near the center of FOV of a conventional PET system. With Equation 8, one may estimate the image resolution of a VP-PET system near its center of the FOV (also its COR) located distance d1 from detector 1 and distance d2 from detector 2. This should not be confused with the image resolution of a VP-PET system at an arbitrary location within its FOV because the latter largely depends on the crystal length and the associated parallax error, both of which were not modeled here.
It is difficult to predict the block effect theoretically because it would require accurate modeling of all physical processes involved in the γ-ray detection and electronic processing. Tomic et al. (23) have demonstrated that the block effect can be estimated through a series of measurements and data fitting. However, this approach was developed for conventional PET geometry and is only applicable to sources equidistant between a pair of identical detectors. Because a VP-PET system uses detector blocks of different crystal dimensions, fully characterizing the block effect would require a few thousand measurements at different locations between 2 detector blocks and using detector blocks of different sizes, and doing so is beyond the scope of this work. It is possible that the block effect of individual detectors can be measured conventionally and incorporated into the last term of Equation 7 through the use of “effective crystal width” that combines the physical crystal width with the block effect. However, this can be validated only after extensive measurements and will be left for future investigations. In the Intrinsic Spatial Resolution Measurements section, we used 2 identical detector blocks as array 1 and array 2 and, subsequently, grouped multiple crystals in detector 2 to mimic crystals of larger dimension. We expected the 2 detector blocks to have similar block effects, which later led to good agreement between the predicted resolution and the measured resolution in Figure 5B. In the Reconstructed Image Resolution Measurements section, we assumed that the block effect of detector 1 (0.8-mm FWHM) is the dominant factor because the COR of the 4 VP-PET systems was relatively close to detector 1. This is likely to be an underestimation of the actual block effect, particularly for systems that use 6.4-mm-wide crystals in detector block 2.
Figure 5B shows that the intrinsic spatial resolution of a VP-PET system is dominated by w1 when the source is near detector 1 and dominated by w2 when the source is near detector 2. The transition point is at a location where d1/d2 = w1/w2. This can be explained by the term related to detector widths in Equation 7. When d1/d2 < w1/w2, the system resolution depends primarily on w1. When d1/d2 > w1/w2, the system resolution depends primarily on w2. If w1 = w2, then the transition point is at the location where d1 = d2, which corresponds to the center of a conventional PET system where the intrinsic spatial resolution is at its minimum. In a VP-PET system, the object should be placed near detector 1 to obtain the highest spatial resolution. To sample the object from all angles with high resolution, the COR should also be located near detector 1 rather than at the center between detectors 1 and 2. This is similar to pinhole SPECT where an object is placed near the pinhole collimator and the system is rotated around the object with a fixed object-to-pinhole distance to maintain the same magnification from all angles.
The background activity concentration in the phantom study shown in Figure 7 was 713 kBq/mL. Because the branching ratio for β+ decay is 18% and 97% for 64Cu and 18F, respectively, the effective activity concentration in the phantom was 132 kBq/mL if the radionuclide used was 18F. This activity concentration is approximately 32-fold higher than the activity concentration of 18F-FDG in a typical whole-body PET protocol, assuming 370 MBq of 18F-FDG are injected and evenly distributed in a 70-kg subject and a 40-min uptake period. In our experiment, the acquisition time at the initial angle was approximately 45 s, which would correspond to a 24-min acquisition time for the first angle with an activity concentration of a typical patient study. However, if one builds a VP-PET system that has large solid angle coverage such that it requires little or no rotation to sample the entire object (27), it is possible to obtain the same level of signal-to-noise ratio and image resolution as was shown in Figure 7 within a reasonable amount of acquisition time.
The potential of VP-PET geometry lies in the fact that one may construct a PET system with different types of detectors and configure the system such that the image resolution is minimally dependent on the low-resolution detectors. There are 2 main disadvantages to such a system. First, the system sensitivity is low compared with a typical PET scanner if only 2 detector blocks are used. Similar to a SPECT system, where one can use a multipinhole collimator to raise the overall system sensitivity, increasing the number of detectors in array 1 will improve the overall system sensitivity of VP-PET systems. Because collimation is performed electronically, the sensitivity of VP-PET is theoretically much higher than that of multipinhole SPECT.
The second drawback is that the system's high resolution is limited to a small imaging FOV. This is, again, very similar to a pinhole SPECT system, where a large magnification and a small usable FOV are used to gain high-resolution imaging capability. For certain imaging applications, such as brain or breast imaging, it may be acceptable to have a smaller imaging FOV. Compared with typical positron-emission mammography systems that use 2 planar PET detectors against the chest wall to image a breast, a breast imaging system using VP-PET geometry may allow detector 2 to be positioned farther away from the breast so that its trajectory clears the shoulder and allows the coincidence line of response to better sample tissues near the chest wall. However, this advantage may be offset by the reduced system sensitivity due to reduced solid angle coverage by detector 2. The value of VP-PET in such applications remains to be seen and will require further investigation.
Finally, the VP-PET geometry has inspired us to design high-resolution insert devices to locally enhance the image resolutions of existing PET scanners, an idea similar to the use of surface coils for MRI. We have studied the potential improvement in image resolution using a full-ring insert device in a clinical whole-body PET scanner through Monte Carlo simulations (28). We are currently developing 2 PET insert systems for human and small-animal imaging applications. The potential of such devices will be investigated when the systems are completed.
CONCLUSION
We have developed and characterized a novel geometry for PET. PET systems based on this virtual-pinhole geometry exhibit certain characteristics similar to those found in a pinhole SPECT system. The potential of VP-PET geometry in creating new PET system design and new PET applications need to be further explored.
Acknowledgments
The authors thank Richard Laforest, Ananya Ruangma, Martin Janecek, and Stefan Siegel for inspiring discussion; Aaron McAllister and Bill Margenau for technical support; and Dave Politte and Daniel Keesing for sharing code. This work was supported in part by the National Cancer Institute of the National Institutes of Health (grants R33-CA110011, R24-CA83060, and P30 CA91842) and by the Susan G. Komen for the Cure (grant BCTR0601279).