Skip to main content

Main menu

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • JNM Podcasts
    • Continuing Education
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Corporate & Special Sales
    • Journal Claims
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI

User menu

  • Subscribe
  • My alerts
  • Log in
  • My Cart

Search

  • Advanced search
Journal of Nuclear Medicine
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI
  • Subscribe
  • My alerts
  • Log in
  • My Cart
Journal of Nuclear Medicine

Advanced Search

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • JNM Podcasts
    • Continuing Education
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Corporate & Special Sales
    • Journal Claims
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • Visit JNM on Facebook
  • Join JNM on LinkedIn
  • Follow JNM on Twitter
  • Subscribe to our RSS feeds
OtherBASIC SCIENCE INVESTIGATIONS

Precise Localization of Sentinel Lymph Nodes and Estimation of Their Depth Using a Prototype Intraoperative Mini γ-Camera in Patients with Breast Cancer

Carole Mathelin, Samuel Salvador, Daniel Huss and Jean-Louis Guyonnet
Journal of Nuclear Medicine April 2007, 48 (4) 623-629; DOI: https://doi.org/10.2967/jnumed.106.036574
Carole Mathelin
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Samuel Salvador
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Daniel Huss
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Jean-Louis Guyonnet
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Info & Metrics
  • PDF
Loading

Abstract

The aim of this study was to evaluate the ability of a prototype intraoperative mini γ-camera, the CarolIReS, with a 50 × 50 mm field of view, to precisely localize sentinel lymph nodes (SLNs) and to determine their depth in a series of patients with infiltrative breast cancer requiring SLN excision. Methods: With the use of phantoms, the broadness of the signal of an acquired projection was shown to linearly depend on its distance from the collimator. A preclinical ex vivo study of 25 excised SLNs demonstrated that SLN size did not influence depth estimation. The minimum activity threshold for successful use of the proposed method was determined. After a preoperative radioisotope injection and lymphoscintigraphy, the SLN was localized in a series of 11 patients using both the mini γ-camera and a γ-probe. During surgery, a ruler was used to measure the depth of all SLNs before their excision. Results: Using the measured linear dependence of image broadness, we found that the expected SLN anatomic depth was compatible with its measured depth during surgery. Conclusion: This study showed that the mini γ-camera efficiently estimated the location of SLNs in 3 dimensions.

  • breast cancer
  • lymphoscintigraphy
  • intraoperative gamma-camera
  • sentinel lymph node
  • CarolIReS

Most types of lymphophile cancers require assessment of regional lymph nodes for staging the tumor, determining prognosis, and devising therapeutic strategies. Complete lymph node dissections are frequently associated with significant morbidity. Combining diagnostic advantages with decreased morbidity (1,2), the sentinel lymph node (SLN) procedure represents one of the most recent advances in surgical oncology. It consists of the identification and removal of SLNs, which are lymph nodes that receive lymphatic drainage directly from the primary tumor. The SLN procedure is now a widely accepted method of lymph node staging for selected cases of invasive breast cancer (3,4) and cutaneous melanoma (5). The concept is also being evaluated in numerous other types of cancer, including pelvic gynecologic (vulval, cervical, and endometrial) (6–9), prostate (10), colon (11), gastric (12) and thyroid (13) cancer. In all cases, successful adaptation of the SLN procedure implies that the operator has a high identification rate and a low false-negative rate. For SLN identification, in addition to the peroperative injection of a blue dye, a 99mTc-radiomarked colloidal solution is generally injected the day before or the day of the surgery and 2 planar scintigraphic images are taken in a nuclear medicine department. In a recent metaanalysis (14) including 69 trials and 8,059 patients with low-risk breast cancer, the proportion of patients with successfully mapped SLNs ranged from 41% to 100%, with more than 50% of the studies reporting a rate under 90%. The false-negative rate ranged from 0% to 29%, averaging 7.3% overall. Significant inverse correlations were observed between the false-negative rate and the proportion of patients with successfully mapped SLNs (r = −0.32; P = 0.009).

Another recent study compared preoperative lymphoscintigraphy and intraoperative γ-probe detection in a series of 823 breast cancer patients who underwent the SLN procedure based on lymphoscintigraphy and blue dye, followed by complete axillary dissection. Preoperative lymphoscintigraphy revealed the SLN in 72% of the 823 patients imaged. In this case, the SLN was identified intraoperatively in 98% of patients using an intraoperative probe and blue dye, with a false-negative rate of 7%. Of patients for whom the SLN had not been visualized on preoperative lymphoscintigraphy, the SLN was identified during surgery in 90%, with the same false-negative rate of 7% (15).

However, even though SLNs can successfully be identified in most patients for whom the SLN is not visualized on lymphoscintigraphy, the use of lymphoscintigraphic images and skin markings during the SLN procedure appears to improve morbidity outcomes over the use of either the probe or the dye alone (16). These results are extremely pertinent, because the main reason for performing the SLN procedure is to reduce morbidity (16).

It is well known that the mean detection efficiency of a conventional γ-camera is about 85% and that additional SLNs are found during surgery with the help of a handheld γ-probe. Nevertheless, there are no well-defined criteria to prove that the SLN detection was complete.

Because of the relatively low SLN detection efficiency and poor localization information given by conventional lymphoscintigraphy, a prototype intraoperative mini γ-camera, the CarolIReS, with a field of view (FOV) of 50 × 50 mm2, was designed and built at the Institut Pluridisciplinaire Hubert Curien to improve both detection efficiency and localization of SLNs before surgery (17). This device is intended for intraoperative use as an additional tool to facilitate the search for SLNs, thus decreasing the false-negative rate and postoperative morbidity. It will not replace intraoperative probes. This clinical report focuses on the ability of the detector to estimate the anatomic depth of SLNs, in addition to precisely localizing them with respect to the axillary area.

MATERIALS AND METHODS

Materials

A detailed technical description of the mini γ-camera has been published previously (17). Figure 1A shows how the device is used in the surgical room. The device is based on the use of a lead parallel collimator, a thin cerium-doped gadolinium orthosilicate (Gd2SiO5:Ce) scintillating crystal (Hitachi Chemical Co. Ltd.), and an H8500 multianode photomultiplier tube (Hamamatsu) and its associated electronics (18). The main characteristics of this device are given in Table 1. The measured intrinsic spatial resolution is 3 mm, and the spatial resolution is 10 mm for a source located 30 mm from the entrance of the collimator. The energy resolution is 45% of the full width at half maximum (FWHM), at 122 keV (17).

FIGURE 1. 
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 1. 

Photographs of intraoperative mini γ-camera (A) and of the clear acrylic blocks placed for preclinical study of SLN depth (B).

View this table:
  • View inline
  • View popup
TABLE 1

Main Characteristics of Mini γ-Camera Prototype

Taking into account the geometric efficiency of the collimator (2.3 × 10−3) and the 85% conversion probability of the 2-mm-thick scintillating crystal, the sensitivity of the mini γ-camera is equal to 1 cps·kBq−1 for the 99mTc 140-keV γ-ray.

Scintigraphic Method

To image an SLN at a known distance, we designed a phantom consisting of an acrylic glass block (95 × 95 × 60 mm3) with a centered hole 27 mm in diameter and 28 mm in length. This device was positioned in front of the γ-camera to center the hole in the FOV (Fig. 1B). Immediately after resection of the SLN, it was inserted into this hole and data were acquired for 2 min to image the isolated SLN at the bottom of the hole. The SLN dimensions were measured by the pathologist before examination of frozen sections.

Patients and SLN Method

Eleven patients with infiltrative breast cancer (diagnosed preoperatively by core biopsy) and a mean age of 55 y (range, 39–72 y) were enrolled in the study (19). The patients had no prior chemotherapy, locoregional radiotherapy, or prevalent axillary lymph node, and all patients gave informed consent.

The SLN procedure was initiated the day before surgery using a preoperative 0.1-mL intradermal injection of rhenium sulfur (19–24 MBq of colloidal 99mTc, Nanocis; Oris) at the 4 cardinal points in the subareolar region. Lymphoscintigraphy with a conventional γ-camera (anterior and 45° oblique views) was performed in the nuclear medicine department 3 h after the injections. The patient was on the radiologic bed, with the arm in abduction at 90°.

Afterward, the axillary area was explored a second time using the mini γ-camera. Because of its 50 × 50 mm2 FOV, 4 images were required to map the entire axillary area (100 × 100 mm2) with the patient in the operative position (the arm in abduction at 80°). Data acquisition for each image required 2 min.

During surgery, the SLN was located with the aid of the CarolIReS intraoperative probe (20), and its depth was measured using a surgical ruler (Aspen Surgical) before resection. The depth was defined as the perpendicular distance from the center of the SLN to the skin surface, with an estimated error of 5 mm. The activity of the SLN was measured using a γ-counter immediately after its resection (20), and standard frozen sections were then prepared for histopathologic analysis. The axillary node was completely cleared only if the SLN had metastatic involvement.

RESULTS

The initial studies were performed using 5 phantoms with activities ranging from 10 to 36 kBq of 57Co (122-keV γ-rays) to test the validity of the method. These phantoms were positioned at 7 distances (dc) from the collimator by putting various numbers of 10-mm-thick clear acrylic plates in front of the mini γ-camera. Clear acrylic plates were chosen to simulate the diffusion of γ-rays in human tissue. The x- and y-coordinates of the interaction point of the γ-ray in the crystal were calculated as the center of gravity of the fired multianode photomultiplier tube cells (17). For each distance dc, 2 profiles were obtained by integrating the data in both directions (x and y). Each profile was fitted by the sum of a constant value representing the flat background contribution and a gaussian curve as illustrated in Figure 2 for 4 different distances. The variation in the FWHM (calculated using the SD of the gaussian distribution) in both profiles is presented in Figure 3, as well as the least squared fit of a straight line. The relationship between the FWHM and the distance dc, where the x and y data are combined, is given by FWHM (mm) = (0.15 ± 0.02)·dc (mm) + (6.5 ± 0.5 mm), with an r value of 0.98. As expected, the relationship did not depend on the phantom activity and was on the order of 1.5 mm per 10 mm of distance from the collimator, with an offset of 6.5 mm, which is the spatial resolution of the system for a dc of 0 mm.

FIGURE 2. 
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 2. 

Profiles on 1 coordinate axis of image of 35.7-kBq 57Co pointlike source at 4 collimator distances (dc = 5, 25, 45, and 65 mm). Data are fitted by sum of constant and gaussian function (superimposed curve).

FIGURE 3. 
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 3. 

Linear variation of FWHM (obtained by fitting procedure defined in Fig. 2) vs. distance dc, from collimator, of 5 different activity pointlike 57Co sources.

Preclinical Study

This study was performed to evaluate the effects of SLN size and activity on the determination of its depth. The method was applied to a sample of 56 fresh SLNs positioned 32 mm from the collimator at the bottom of the hole of the device (Fig. 1B). The SLN size ((π·long axis·short axis)/4) is given as the ellipsoidal cross-section surface in square millimeters calculated with the 2 measured axes (12.4 ± 6.5 mm and 7.4 ± 3.6 mm on average for the long axis and the short axis, respectively). SLN activity and size did not correlate, as is shown in Figure 4A (r = 0.12). Figures 4B and 4C present the distributions of the measured activity and the size of the 56 SLN samples. No correlation was found between the FWHM calculated from the profiles and the SLN size (r = 0.09). In other words, it was impossible to extract a contribution to the width of the profiles coming from a distribution of the radioactivity inside the SLN. The estimated distance dp obtained by the method defined by the phantom study is shown in Figure 5. This was possible only for a subsample of 25 SLNs (23 for both profiles and 2 for one profile only). The data gave a gaussian curve with mean values of 39.9 mm for dp and 11.5 mm for σdp. The predicted distance dp appears an overestimation, compared with the experimentally fixed value of 32 mm. Taking into account the 10-mm mean value of 2 SLN axis lengths, this result is compatible with the distance from the SLN center to the collimator. The red part of Figures 4B and 4C represents, respectively, the activity and the size of the 25-SLN subsample. At SLN activity values below 2 kBq (Fig. 4B), the calculation fails mainly because of a poor signal-to-noise ratio. Above a 2-kBq activity threshold, the efficiency of the method is evaluated to be 86% ± 23%, independent of the SLN size (Fig. 4C).

FIGURE 4. 
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 4. 

(A) Plot of activity vs. size for ex vivo measurements of 56 SLNs. (B and C) Distribution of activity (B) and size (C) for sample of 56 SLNs with successful (red) and unsuccessful (blue) determination of their depth.

FIGURE 5. 
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 5. 

Distribution of estimated distance dp between SLN and collimator for subsample of 25 SLNs. Data are fitted by gaussian function (superimposed curve).

Clinical Study

The final analysis of the removed breast lesions revealed an infiltrative ductal carcinoma in 7 cases and an infiltrative lobular carcinoma in 4 cases, with an average size of 15 mm (range, 2–27 mm). Fourteen SLNs were detected by conventional lymphoscintigraphy and 16 by the mini γ-camera. The complete SLN procedure allowed the detection of 20 SLNs with a mean of 1.8 (range, 1–3) per patient. Four involved SLNs were detected in 3 patients who subsequently underwent complete axillary clearance (Table 2).

View this table:
  • View inline
  • View popup
TABLE 2

Summary of Patient Characteristics and SLN Analysis (Size, Activity, Depth, Pathologic Analysis)

Figure 6 compares a conventional lymphoscintigraphy image with an image from the mini γ-camera.

FIGURE 6. 
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 6. 

(A) Anterior view of conventional lymphoscintigraphy. Yellow arrow indicates SLN. Blue one indicates subareolar injection site. (B) For same patient, mini γ-camera image shows 2 closer SLNs indicated by 2 arrows.

The linear dependence of the FWHM on the source distance was applied to data obtained with the mini γ-camera. By measuring the broadness of the image, which depends on the distance of the SLN from the collimator entrance, we could predict the depth of the SLN in the axillary area. The results concerning the expected depth (given, respectively, by the x and y profiles) of 12 SLNs are summarized in Table 2. Because the required information was not fully contained in the 50 × 50 mm2 FOV of the current prototype of the mini γ-camera, our technique could not be applied to the total imaged sample of SLNs. For 4 of them, the information required to obtain useful profiles was only partially contained in the image (at a corner of the FOV). When a difference existed between the 2 expected depths, the mean value was used to determine the correlation between the predicted (dp ± 11.5 mm) and the measured (dm ± 5 mm) SLN depths, as presented in Figure 7. A straight line fit gives the following result: dp (mm) = (0.5 ± 0.2)·dm (mm) + (24 ± 9 mm), with an r value of 0.76.

FIGURE 7. 
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 7. 

Correlation between predicted depth dp of SLN and its measurement dm by ruler during surgery. Linear squared fit to data is also represented.

DISCUSSION

In our clinical series, the proportion of patients with successfully mapped SLNs was 64% (7/11). In 4 cases, conventional lymphoscintigraphy did not show the totality of the SLNs. Only 14 of 20 SLNs were visualized on conventional lymphoscintigraphy (70% efficiency). On the day of the injection, use of the mini γ-camera allowed detection of 2 additional SLNs. In 3 cases, the total number of SLNs was not detected preoperatively by conventional lymphoscintigraphy or by the mini γ-camera. There are probably several explanations. First, the pictures were obtained the day before surgery, and lymphatic drainage had perhaps not been completed. In some cases, SLNs were too closely situated for the spatial resolution of the imaging device. The low SLN detection efficiency of the mini γ-camera (80%) could also be explained by the limited FOV. The 4 data acquisitions obtained by moving the mini γ-camera did not ensure that the entire axillary area was covered (it was difficult to take precisely consecutive and joint images to obtain the 100 × 100 mm2 full image).

Our preclinical study demonstrated that SLN depth was estimated efficiently only when the SLN activity was greater than 2 kBq for a 2-min data acquisition. This limitation is significant if the images are to be obtained just before surgery, when an injection on the order of 20 MBq of radiocolloid takes place the day before the surgery. The efficiency of the method can be improved if the image acquired by the mini γ-camera is obtained after injection, with a time delay that has to be experimentally determined.

A new clinical trial to compare the performance of conventional lymphoscintigraphy with that of a mini γ-camera having a FOV of 100 × 100 mm will begin this year. It is still too early to propose replacing conventional γ-cameras with intraoperative mini γ-cameras.

A simple comparison between probe and mini γ-camera and between conventional γ-camera and mini γ-camera in terms of sensitivity is not relevant to their classification because of their different goals. For example, the sensitivities of the devices used in our study are 0.1, 1.0, and 10.0 cps·kBq−1 for the conventional γ-camera, the mini γ-camera, and the γ-probe, respectively. In addition, those numbers have to be related to the respective FOVs: 400 × 500, 50 × 50, and 12 mm2. Because of the small number of patients in our series, we could not significantly demonstrate that the mini γ-camera was better than the conventional γ-camera, but the possibility of using the mini γ-camera before, during, and after surgery is a huge advantage.

Our clinical series demonstrates a strong correlation between the predicted and measured depths of SLNs. It is crucial to know the precise number of SLNs present and to locate them in 3 dimensions. This knowledge improves SLN identification, avoids the removal of other lymph nodes, and may reduce hospitalization time.

CONCLUSION

Our study showed that the mini γ-camera is well adapted to SLN localization and efficiently estimates SLN depth. With the 100 × 100 mm2 FOV device under construction, the full axillary area will be covered, improving the efficiency of SLN detection. As a means of quality control and to help decrease the false-negative rate, a postoperative scintigraphic image can be acquired to prove that no SLNs remain after surgery.

Acknowledgments

We thank Kadour Annane, Jean-Pierre Bellocq, Jean-Philippe Brettes, Marie-Pierre Chenard, Doris Frankhauser, Philippe Liégeois, Jean-Louis Mehl, and Roland Schaeffer for their collaboration and David Brasse and Shanti Natarajan-Amé for their helpful discussions. This work was supported by funds from the Centre National de la Recherche Scientifique, the Hôpitaux Universitaires de Strasbourg, the University Louis Pasteur, the Région Alsace, and the Pfizer Laboratory. We are indebted to the Ligue Nationale Française contre le Cancer and the Comités du Haut-Rhin et du Bas-Rhin for their constant and considerable financial support.

Footnotes

  • COPYRIGHT © 2007 by the Society of Nuclear Medicine, Inc.

References

  1. 1.↵
    Schulze T, Mucke J, Markwardt J, Schlag PM, Bembenek A. Long-term morbidity of patients with early breast cancer after sentinel lymph node biopsy compared to axillary lymph node dissection. J Surg Oncol. 2006;93:109–119.
    OpenUrlCrossRefPubMed
  2. 2.↵
    Wilke LG, McCall LM, Posther KE, et al. Surgical complications associated with sentinel lymph node biopsy: results from a prospective international cooperative group trial. Ann Surg Oncol. 2006;13:491–500.
    OpenUrlCrossRefPubMed
  3. 3.↵
    Lyman GH, Giuliano AE, Somerfield MR, et al. American Society of Clinical Oncology guideline recommendations for sentinel lymph node biopsy in early-stage breast cancer. J Clin Oncol. 2005;23:7703–7720.
    OpenUrlAbstract/FREE Full Text
  4. 4.↵
    James TA, Edge SB. Sentinel lymph node in breast cancer. Curr Opin Obstet Gynecol. 2006;18:53–58.
    OpenUrlCrossRefPubMed
  5. 5.↵
    Vaquerano J, Kraybill WG, Driscoll DL, Cheney R, Kane JM III. American Joint Committee on Cancer clinical stage as a selection criterion for sentinel lymph node biopsy in thin melanoma. Ann Surg Oncol. 2006;13:198–204.
    OpenUrlCrossRefPubMed
  6. 6.↵
    Zhang WJ, Zheng R, Wu LY, Li XG, Li B, Chen SZ. Clinical application of sentinel lymph node detection to early stage cervical cancer [in Chinese]. Ai Zheng. 2006;25:224–228.
    OpenUrlPubMed
  7. 7.
    Robison K, Steinhoff MM, Granai CO, Brard L, Gajewski W, Moore RG. Inguinal sentinel node dissection versus standard inguinal node dissection in patients with vulvar cancer: a comparison of the size of metastasis detected in inguinal lymph nodes. Gynecol Oncol. 2006;101:24–27.
    OpenUrlCrossRefPubMed
  8. 8.
    Rob L, Charvata M, Robova H, et al. Sentinel lymph node identification (SLNI) in the management of conservative surgery in early cervical cancer: is it acceptable? Gynecol Oncol. 2005;99(3, suppl 1)S147–S148.
    OpenUrlCrossRefPubMed
  9. 9.↵
    Carcopino X, Houvenaeghel G, Buttarelli M, Charaffe-Jauffret E, Gonzague L, Rossi I. Feasibility and morbidity of sentinel lymph node detection in patients with vulvar carcinoma [in French]. Bull Cancer. 2005;92:489–497.
    OpenUrlPubMed
  10. 10.↵
    Petros JA. Sentinel lymph node mapping in prostate cancer: opportunities remain to identify truly positive nodes. J Urol. 2006;175:421–422.
    OpenUrlCrossRefPubMed
  11. 11.↵
    Saha S, Seghal R, Patel M, et al. A multicenter trial of sentinel lymph node mapping in colorectal cancer: prognostic implications for nodal staging and recurrence. Am J Surg. 2006;191:305–310.
    OpenUrlCrossRefPubMed
  12. 12.↵
    Mochiki E, Kuwano H, Kamiyama Y, et al. Sentinel lymph node mapping with technetium-99m colloidal rhenium sulfide in patients with gastric carcinoma. Am J Surg. 2006;191:465–469.
    OpenUrlCrossRefPubMed
  13. 13.↵
    Pelizzo MR, Merante Boschin I, Piotto A, et al. Sentinel lymph node procedure in thyroid carcinoma patients: our experience. Minerva Chir. 2006;61:25–29.
    OpenUrlPubMed
  14. 14.↵
    Kim T, Giuliano AE, Lyman GH. Lymphatic mapping and sentinel lymph node biopsy in early-stage breast carcinoma: a metaanalysis. Cancer. 2006;106:4–16.
    OpenUrlCrossRefPubMed
  15. 15.↵
    Goyal A, Newcombe RG, Mansel RE, et al. Role of routine preoperative lymphoscintigraphy in sentinel node biopsy for breast cancer. Eur J Cancer. 2005;41:238–243.
    OpenUrlCrossRefPubMed
  16. 16.↵
    Kim SC, Kim DW, Moadel RM, et al. Using the intraoperative hand held probe without lymphoscintigraphy or using only dye correlates with higher sensory morbidity following sentinel lymph node biopsy in breast cancer: a review of the literature. World J Surg Oncol. 2005;3:64.
    OpenUrlCrossRefPubMed
  17. 17.↵
    Piqueras I, Gizard B, Huss D, Brasse D, Mathelin C, Guyonnet J. A new mini gamma camera for sentinel lymph node identification in patients with breast cancer. In: Proceedings of SCINT 2005 (International conference on inorganic scintillators and their industrial application); September 19–23, 2005; Alushta, Ukraine.
  18. 18.↵
    Bard P, Avner S, Bekaert V, et al. Electronics dedicated to multi-anode photomultiplier tubes for imaging scintigraphic devices. In: Proceedings of the 14th IEEE-NPSS Real Time Conference; 2005; Stockholm, Sweden.
  19. 19.↵
    ClinicalTrials.gov Web site (a service of the U.S. National Institutes of Health). Available at: http://www.clinicaltrials.gov/ct/show/NCT00357487;jsessionid=1B2EA82778105B140C998A953959497?order=29. Accessed January 29, 2007
  20. 20.↵
    Mathelin C, Piqueras I, Guyonnet JL. Performances of a new prototype handheld scintillating probe for the search of sentinel lymph nodes in breast cancer. Eur J Surg Oncol. 2006;32:24–28.
    OpenUrlCrossRefPubMed
  • Received for publication January 26, 2006.
  • Accepted for publication December 6, 2006.
PreviousNext
Back to top

In this issue

Journal of Nuclear Medicine: 48 (4)
Journal of Nuclear Medicine
Vol. 48, Issue 4
April 2007
  • Table of Contents
  • Table of Contents (PDF)
  • About the Cover
  • Index by author
Print
Download PDF
Article Alerts
Sign In to Email Alerts with your Email Address
Email Article

Thank you for your interest in spreading the word on Journal of Nuclear Medicine.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Precise Localization of Sentinel Lymph Nodes and Estimation of Their Depth Using a Prototype Intraoperative Mini γ-Camera in Patients with Breast Cancer
(Your Name) has sent you a message from Journal of Nuclear Medicine
(Your Name) thought you would like to see the Journal of Nuclear Medicine web site.
Citation Tools
Precise Localization of Sentinel Lymph Nodes and Estimation of Their Depth Using a Prototype Intraoperative Mini γ-Camera in Patients with Breast Cancer
Carole Mathelin, Samuel Salvador, Daniel Huss, Jean-Louis Guyonnet
Journal of Nuclear Medicine Apr 2007, 48 (4) 623-629; DOI: 10.2967/jnumed.106.036574

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Share
Precise Localization of Sentinel Lymph Nodes and Estimation of Their Depth Using a Prototype Intraoperative Mini γ-Camera in Patients with Breast Cancer
Carole Mathelin, Samuel Salvador, Daniel Huss, Jean-Louis Guyonnet
Journal of Nuclear Medicine Apr 2007, 48 (4) 623-629; DOI: 10.2967/jnumed.106.036574
Twitter logo Facebook logo LinkedIn logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One
Bookmark this article

Jump to section

  • Article
    • Abstract
    • MATERIALS AND METHODS
    • RESULTS
    • DISCUSSION
    • CONCLUSION
    • Acknowledgments
    • Footnotes
    • References
  • Figures & Data
  • Info & Metrics
  • PDF

Related Articles

  • This Month in JNM
  • PubMed
  • Google Scholar

Cited By...

  • Added Value of Intraoperative Real-Time Imaging in Searches for Difficult-to-Locate Sentinel Nodes
  • A Portable {gamma}-Camera for Intraoperative Detection of Sentinel Nodes in the Head and Neck Region
  • Paraaortic Sentinel Lymph Nodes: Toward Optimal Detection and Intraoperative Localization Using SPECT/CT and Intraoperative Real-Time Imaging
  • Google Scholar

More in this TOC Section

  • How Sensitive Is the Upper Gastrointestinal Tract to 90Y Radioembolization? A Histologic and Dosimetric Analysis in a Porcine Model
  • 11C-Methionine PET of Myocardial Inflammation in a Rat Model of Experimental Autoimmune Myocarditis
  • Counting Rate Characteristics and Image Distortion in Preclinical PET Imaging During Radiopharmaceutical Therapy
Show more Basic Science Investigations

Similar Articles

SNMMI

© 2024 Journal of Nuclear Medicine

Powered by HighWire