Abstract
15O-Water and dynamic PET allow noninvasive quantification of myocardial blood flow (MBF). However, complicated image analyzing procedures are required, which may limit the practicality of this approach. We have designed a new practical algorithm, which allows stable, rapid, and automated quantification of regional MBF (rMBF) using 15O-water PET. We designed an algorithm for setting the 3-dimensional (3D) region of interest (ROI) of the whole myocardium semiautomatically. Subsequently, a uniform input function was calculated for each subject using a time–activity curve in the 3D whole myocardial ROI. The uniform input function allows the mathematically simple and robust algorithm to estimate rMBF. Methods: Thirty-six volunteers were used in the static 15O-CO and dynamic 15O-water PET studies. To evaluate the reproducibility of the estimates, a repeated 15O-water scan was obtained under resting condition. In addition, to evaluate the stability of the new algorithm in the hyperemic state, a 15O-water scan was obtained with adenosine triphosphate. This algorithm includes a procedure for positioning a 3D ROI of the whole myocardium from 3D images and dividing it into 16 segments. Subsequently, the uniform input function was calculated using time–activity curves in the whole myocardial ROI and in the LV ROI. The uniform input function allowed this simple and robust algorithm to estimate the rMBF, perfusable tissue fraction (PTF), and spillover fraction (Va) according to a single tissue compartment model. These estimates were compared with those calculated using the original method. A simulation study was performed to compare the effects of errors in PTF or Va on the MBF using the 2 methods. Results: The average operating time for positioning a whole myocardial ROI and 16 regional myocardial ROIs was <5 min. The new method yielded less deviation in rMBF (0.876 ± 0.177 mL/min/g, coefficient of variation [CV] = 20.2%, n = 576) than those with the traditional method (0.898 ± 0.271 mL/min/g, CV = 30.1%, n = 576) (P < 0.01). In the hyperemic state, the new method yielded less deviation in rMBF (3.890 ± 1.250 mL/min/g, CV = 32.1%) than those with the traditional method (3.962 ± 1.762 mL/min/g, CV = 44.4%) (P < 0.05). This method yielded significantly higher reproducibility of rMBF (r = 0.806, n = 576) than the original method (r = 0.756, n = 576) (P < 0.05). Our new method yielded a better correlation in the repeated measurement values of rMBF and less variability among the regions in the myocardium than with the original theory of the 15O-water technique. The simulation study demonstrated fewer effects of error in the PTF or Va on the MBF value with the new method. Conclusion: We have developed a technique for an automated, simplified, and stable algorithm to quantify rMBF. This software is considered to be practical for clinical use in myocardial PET studies using 15O-water with a high reproducibility and a short processing time.
Previous animal and clinical studies with 15O-water and PET have shown the feasibility of quantitative estimates of regional myocardial blood flow (rMBF) (1–4). The use of 15O-water PET provides a noninvasive method for in vivo quantification of rMBF, perfusable tissue fraction (PTF), and spillover fraction (Va). However, the original method requires complicated image analyzing procedures, which may limit the practicality of this approach. A crucial limitation of the original method is the manual positioning of regional myocardial regions of interest (ROIs). It is a very time-consuming procedure, especially when the operator is required to set the same myocardial ROIs on the paired rest and stress studies. The manual approach remains operator dependent. Moreover, in the original algorithm, the input function is estimated using the time–activity curves in the regional myocardial ROI and in the left ventricular (LV) ROI (5,6). When the input function is derived from the regional myocardial time–activity curve, it would not be uniform for each regional myocardium in each subject (5). We accessed the new algorithm using a uniform input function, which may improve the stability and reproducibility of the estimates.
Several automatic algorithms are reported for setting the myocardial ROI in myocardial SPECT studies (7–9); The algorithms permit short operating time as well as operator-independent and robust results. For 15O-water myocardial PET studies, we attempted the same approach. In this study, we developed a software algorithm, which allows rapid, stable, and automated estimation of the rMBF. We put particular emphasis on the performance of the software to achieve practical analysis times for clinical deployment.
MATERIALS AND METHODS
We designed an algorithm for setting the 3-dimensional (3D) ROI of the whole myocardium and dividing it into 16 segments automatically, overcoming the difficulty of the process, which is significantly time consuming if it is performed manually. Subsequently, a uniform input function was calculated for each subject using time–activity curves in the whole myocardial ROI and in the LV ROI. The uniform input function allows the mathematically simple and robust algorithm to estimate rMBF, PTF, and Va. These estimates were compared with those calculated with the original method.
Subjects
Thirty-six healthy volunteers (34 men, 2 women; mean age, 29.9 ± 8.6 y) were studied in Hokkaido University Medical Hospital. Data acquisitions were performed as a prospective study. All subjects had normal laboratory tests and no symptoms or history of cardiovascular disease. Heart rates and end-systolic and end-diastolic blood pressures were continuously monitored and recorded. The nature, purpose, and potential risks of the study were explained to all subjects before they gave their voluntary consent to participate. The study was approved by the ethical committee of Hokkaido University Medical Hospital.
Production of 15O-CO and 15O-Water
For the production of 15O-labeled compounds, a low-energy deuteron accelerator was used (CYPRIS-HM18; Sumitomo Heavy Industries). 15O-Water was produced with a dialysis technique in a continuously working water module. Sterility and pyrogen tests were performed daily to verify the purity of the product. Gas chromatographic analysis was performed to verify the purity of the product before each study.
Image Acquisition
PET scans were obtained using an ECAT HR+ scanner (Siemens/CTI Corp.) equipped with 68Ge/68Ga retractable line sources for transmission scans. All emissions and transmissions were acquired in the 2-dimensional mode and reconstructed using filtered backprojection with a Hanning filter (cutoff, 0.4). The in-plane resolution was 6.2-mm full width half maximum in a 128 × 128 matrix image. All data were corrected for dead time, decay, and measured photon attenuation.
The optimal imaging position was determined by a 1-min rectilinear scan. A 5-min transmission scan was then acquired for the purpose of attenuation correction of all subsequent emission scans.
A blood-pool scan was obtained in the following manner. Each subject’s nostrils were closed and he or she inhaled 15O-CO (0.14% CO mixed with room air) for 1 min. After inhalation of the tracer, a period of 3 min was allowed for the CO to combine with hemoglobin before a 5-min static scan was started. During the 5-min scan period, venous blood samples were drawn every 2 min, and the concentration of radioactivity in whole blood was measured with an automatic γ-counter (ARC-400; ALOKA Inc.). The inhaled dose in the CO examination was 2,000 MBq.
15O radioactivity returned to background levels 10 min after the blood-pool scan. 15O-Water was infused into an antecubital vein as a slow (2 min) infusion. The administrated dose of 15O-water was 500 MBq/min. A 20-frame dynamic PET scan, consisting of 6 × 5 s, 6 × 15 s, and 8 × 30 s frames, was acquired for 6 min.
After allowing a further 10 min for decay, another dynamic scan using 15O-water was acquired. Fifteen minutes later, adenosine triphosphate (ATP) was infused for 7 min at 0.16 mg/kg/min. A 20-frame dynamic 15O-water PET scan was started 2 min after the beginning of ATP infusion. Blood pressure and heart rate were recorded before each scanning.
Data Processing
Original coincidence data were stored as sinograms and then normalized, corrected for attenuation, and backprojected in a dedicated array processor to obtain the final images. The analysis of PET images was accomplished using an image analysis package (Dr. View; Asahi-Kasei) and software developed in our institute by use of programming language C.
Processing Blood Volume Images.
Functional blood volume images were created by dividing the 15O-CO PET counts by the blood sample counts measured in the cross-calibrated well counter. 15O-Water washout images were calculated by subtracting the blood volume images from the sum of the 15O-water dynamic images (10).
Calculation of Extravascular Density Images.
The reconstructed transmission data were normalized on a pixel-by-pixel basis to the activity pixel counts in a ROI in the LV chamber of the transmission image and the density of blood. The extravascular images were calculated by subtracting the vascular components from the normalized transmission images using the blood volume images (10).
ROI Definition
Short axial blood volume images were generated semiautomatically, rotating horizontal and vertical long axial blood volume images (Fig. 1). This process was performed to define the heart long axis. This program generates short axial 15O-water dynamic images and extravascular density images automatically, with the same rotation angles used in the blood volume images.
The short axial blood volume images were used to define the ROI in the left ventricle, showing 85% recovery of the true blood tracer concentration (11) (Fig. 2). Early-phase short axial 15O-water images were calculated by summing first the 2-min dynamic images. A linear 6-parameter registration (12) was performed for registration of the 15O-water images to the 15O-CO images. Short axial 15O-water washout images were generated calculating the mean counts from 180- to 300-s frame images.
When the 15O-water washout images were superimposed on extravascular images and analyzed with a semiautomated edge-detection routine, a whole myocardial ROI was determined. The program is designed to extract a 3D region enclosing the whole myocardial wall. The process is performed in the 3D matrices. At first, the apical and basal portion (valve plane) are pointed manually. The activity distribution of the midventricular and basal portions is expressed in circular cylindric coordinates relative to the heart long axis, assuming a shape of the heart that resembles a hemisphere in the apex and a cylindroid in the midventricular and basal portions. For each possible ray in a radial direction from the heart long axis, the first crossing of a given activity threshold (e.g., 60%) is detected and every point farther than the maximal possible wall thickness (e.g., 2 cm) is excluded from calculation (13). Then, a 3 × 3 × 3 3D smoothing filter is applied to the edge of the ROI, and a bullet-shaped ROI that encloses the whole myocardium is generated (Fig. 3A). Regional myocardial ROIs were positioned automatically in 16 anatomic segments according to the rule in the American Society of Echocardiography report (14) (Fig. 3B).
The LV time–activity curve was generated with the LV ROI and dynamic 15O-water data. The whole and regional myocardial tissue time–activity curves were generated with these myocardial ROIs and dynamic 15O-water data.
Original Method for Estimation of rMBF
A conventional theory to estimate rMBF using 15O-water is described by the following equation for a single tissue compartment model (Fig. 4A): Eq. 1 where Ca(t) is the concentration of 15O-water in arterial whole blood (cps/mL), Ct(t) is the concentration of 15O-water in myocardial tissue (cps/g), F is rMBF (mL/g/min), and p is the partition coefficient of water in the myocardium (0.91 mL/g), respectively.
The concentration of radioactivity in the LV ROI, LV(t), is expressed as follows: Eq. 2 where ρ is a myocardial tissue density (1.04 g/mL), and β corresponds to a recovery coefficient of the LV ROI concentration. The value of β is calculated using the 15O-CO blood volume scan as follows (6): β = LV ROI counts in the 15O-CO blood volume scan (cps/mL)/sampled blood counts during the 15O-CO scan (cps/g)/whole blood density (1.06 g/mL)
The measured tissue time–activity curve in a regional myocardial ROI, R(t) is described by: Eq. 3 where PTF is the grams of perfusable tissue per 1 mL of myocardial ROI (g/mL) and Va is the arterial blood volume in the myocardial ROI and spillover from the left ventricle into the myocardial ROI (mL/mL).
The values of PTF, Va, and F are estimated by Equations 1–3 with a nonlinear least-squares regression analysis.
Modified Method for Estimation of rMBF
In the modified method in this study, we have developed a simplified algorithm (Fig. 4B). First, a uniform input function Cawhole(t) was derived using a whole myocardial ROI curve with the original method. When the values of PTF, Va, and F for a whole myocardial ROI were determined by the non–least-squares method, the input function curve Cawhole(t) was calculated by Equations 1–3. Then, the following equations were used to determine the values of PTF, Va, and F for the regional myocardium with a nonlinear least-squares method: Eq. 4 Eq. 5
The following 4 sets of rMBF in each segment in the myocardial wall were estimated for each subject:
Original rMBF1 was calculated using the first 15O-water dynamic data with a conventional method.
Original rMBF2 was calculated using the second 15O-water dynamic data with a conventional method.
Modified rMBF1 was calculated using the first 15O-water dynamic data with a modified method.
Modified rMBF2 was calculated using the second 15O-water dynamic data with a modified method.
In addition, stability of rMBF in the hyperemic state was estimated for each subject:
Original rMBF was calculated using hyperemic 15O-water dynamic data with a conventional method.
Modified rMBF was calculated using hyperemic 15O-water dynamic data with a modified method.
Estimates (MBF, PTF, and Va) in the apical, anterior, septal, inferior, and lateral wall were calculated from apical, middle, and basal regions as shown in Figure 5A. Our new program presents rMBF on the polar map and on 15O-water images (Fig. 5B). When paired rest and stress studies are performed, our program presents rMBF at rest, rMBF in the hyperemic state, and coronary flow reserve in each region on the polar maps (Fig. 6).
Statistical Analysis
To evaluate the reproducibility of the results, Pearson correlation coefficients were calculated and a Bland–Altman plot was applied for rMBF derived from a traditional method (Original rMBF1 vs. Original rMBF2) and from a modified method (Modified rMBF1 vs. Modified rMBF2). The differences in these correlation coefficients were compared using Fisher z-transformation. To evaluate the regional variability of the estimates (MBF, PTF, and Va), paired Student t tests were performed to examine the difference between the original and modified methods. The χ2 test was performed to evaluate the differences in the variation of each estimate. All data are presented as mean ± SD. P < 0.05 is considered to be statistically significant in assessing the differences of correlation coefficients and differences of variability.
Simulation Study
A simulation study was performed to assess the effects of errors in the measurement of PTF or Va on the calculated value of MBF in the original and modified methods.
The mean myocardial and LV time–activity curves were generated using data of the 36 subjects. The percentage error in MBF was calculated when the percentage error of the PTF or Va was changed from −10% to +10%.
RESULTS
The variables of systolic pressure, diastolic pressure, and heart rate in the first and second 15O-water dynamic scans were 105.3 ± 8.4 and 105.9 ± 9.1 mm Hg, 54.7 ± 7.7 and 53.4 ± 8.8 mm Hg, and 60.3 ± 8.5 and 60.8 ± 7.3 beats/min, respectively. No significant hemodynamic change was observed during the 2 dynamic acquisitions. In the hyperemic state, the variables of systolic pressure, diastolic pressure, and heart rate were 103.6 ± 9.5 mm Hg, 52.1 ± 8.7 mm Hg, and 87.2 ± 8.4 beats/min.
The average analysis time to define a whole myocardial ROI and 16 regional myocardial ROIs was <5 min. With the conventional manual ROI setting program, it took >30 min. The use of the algorithm for defining ROIs resulted in the decrease of the analysis time. After positioning a whole myocardial ROI and 16 regional myocardial ROIs, the procedure to estimate these rMBFs took only a few seconds with both the original and our modified methods.
Table 1 shows a comparison of the variables of rMBF, PTF, and Va at rest in the various regions of the myocardium among subjects and within a subject. The modified method yielded less deviation and a lower rMBF than those with the traditional method. The apical, anterior, and lateral regions showed a significant decrease in rMBF. The modified method yielded a lower PTF than those with the original method. The apical, anterior, and lateral regions showed a significant decrease in PTF. The simplified method yielded a higher Va than those with the traditional method. The apical, anterior, and lateral regions yielded a significant increase in Va. Significant differences and less deviation of these estimates in the resting condition were presented within a subject between the original and modified methods.
Table 2 presents a comparison of the variables of the measurement of rMBF in the hyperemic state among subjects. The new method yielded less deviation in rMBF (3.890 ± 1.250 mL/min/g, coefficient of variation [CV] = 32.1%) than those with the traditional method (3.962 ± 1.762 mL/min/g, CV = 44.4%) (P < 0.05) in the hyperemic condition. Significant differences and less deviation of these estimates in the hyperemic condition were presented within a subject between the original and modified methods.
Figure 7 shows plots of the correlations between rMBF calculated from the first and second 15O-water dynamic scans using the original method and the modified algorithm. The modified method yielded significantly higher reproducibility than the original method (P < 0.05). The Bland–Altman test shows better reproducibility with the modified method.
Figure 8 shows the results of the simulation studies, which demonstrated the error of the PTF or the Va propagated to the measured value of MBF with a simplified method or a modified method. Fewer errors in the estimated MBF were induced by errors in the PTF or the Va with a modified method. In the original algorithm, an overestimation of 10% in the PTF caused an overestimation of approximately 28% in MBF, whereas, in the modified algorithm, the overestimation was suppressed 20% in MBF. In the original algorithm, an underestimation of 10% in the Va caused an overestimation of approximately 26% in MBF, whereas in the modified algorithm, the overestimation was suppressed 20% in MBF.
DISCUSSION
In this study, we designed a software algorithm for setting the 3D ROI of the whole myocardium and dividing it into 16 regions automatically with a clinically acceptable processing time. At the same time, the software was equipped with a simplified algorithm to calculate MBF using a uniform input function for each region for stable estimation.
We designed an algorithm for setting the 3D ROI of the whole myocardium and dividing it into 16 segments automatically, overcoming the difficulty of the process, which is significantly time consuming if it is performed manually. At the present time, our ROI setting software requires manual handling of a few processes, but it enables ROI definition almost independently of the operator and allows stable rMBF quantification. Various approaches are used for automatic 3D delineation of the LV myocardium in myocardial SPECT studies (15–17). However, it might be difficult to apply these algorithms for 15O-water PET data, because the PET studies performed with 15O-labeled tracers yield less delineated and less homogeneous images than SPECT studies. We designed a simple method for positioning a 3D ROI, which covers the whole LV myocardium. For estimating rMBF with 15O-water PET, is not necessary to set a tight ROI tracing the contours of the myocardium (18).
To improve the reproducibility and stability of quantitative measurements of the rMBF, we applied a uniform input function for each region. Our modified method yielded a better correlation in the repeated measurement values of rMBF and less variability among the regions in the myocardium than with a conventional theory of the 15O-water technique. Also, in the hyperemic state, the new method yielded less deviation in rMBF than those with the original method. A simulation study also showed that the modified algorithm yielded fewer errors in the estimated MBF.
The main reason for the better reproducibility and stability in the modified method is explicable. In the original method, a nonlinear least-squares method is used to determine 3 unknown values (PTF, Va, and MBF) and 2 unknown curves (Ca(t), Ct(t)) by fitting 2 curves (PTF Ct(t) + VaCa(t) and βCa(t) + (1 − β)ρCt(t)) to the 2 known curves (R(t) and LV(t)), respectively. The value (1 − β) is the contribution ratio of the whole myocardial radioactivity (6); it is an anatomic parameter for a whole LV myocardium. The equation LV(t) = βCa(t) + (1 − β)ρCt(t) should be applied to the whole myocardial data. When the value (1 − β) is applied to estimate regional myocardial data, the contribution ratio must be overestimated. Then the original method might overestimate rMBF and underestimate the regional spillover fraction (Va).
Our present study demonstrated that the original method yielded significantly higher rMBF and a lower Va than the modified method, especially in the apical, anterior, and lateral regions. In these regions, Va is overestimated in the conventional method.
To improve the stability of quantitative measurement of rMBF, we designed a new modified algorithm. In the modified method, using the known curve Cawhole(t), the parameters (PTF, Va, and MBF) are derived by fitting only 1 curve (PTF Ct(t) + Va Cawhole(t)) to the known curve R(t). This fitting is mathematically simplified and more stable to estimate the parameters than that in the original algorithm.
The simulation study indicated that the modified method showed fewer effects of error in the PTF and the Va on the MBF compared with the original method.
Theoretically, whole myocardial ROI data are recommended to derive an input function (11); it was difficult to determine a whole myocardial ROI because the procedure requires a 3D ROI setting technique. In this study, we have developed a new program for the procedure that overcame the difficulty of positioning the whole myocardial ROI. We developed a new algorithm to improve the stability of the estimates calculated from 15O-water PET; however, it is also essential to improve the reconstruction method for achieving better image quality using 15O-water with PET (19).
The principal difficulty in validating the new approach is the lack of a gold standard to determine the success of the algorithm. Therefore, we relied primarily on the improvement of reproducibility and the decrease in the variability of MBF among the regions in the myocardium. We confirmed these results and appropriate values of estimates (20).
In this study, we performed 15O-CO scanning once in the protocol, which was not repeated before each 15O-water PET scan. The 15O-CO image is used only for positioning the LV ROI in our program, and each 15O-water PET image is transferred appropriately with the registration algorithm. It is ideal to acquire an 15O-CO image before every 15O-water scan for precise ROI setting in the left ventricle. However, the algorithm to derive MBF in this study is equipped for estimating appropriate spillover in each myocardium (5); the estimated MBF with the present program might be robust enough to detect the subtle difference in the size or positioning of the LV ROI.
The limitation of this study is that we could not assess the reproducibility of measured rMBF in the hyperemic state. However, it was impossible to perform repeated 15O-water PET studies with ATP because of ethical concerns. In the present study, we did not apply our new program to the patients with ischemic heart disease. Further clinical validations are needed.
At present, our program requires a 15O-CO scan for estimation of MBF because this program is designed not only for calculating MBF but also for estimating extravascular tissue density and the PTF index in the myocardium (11). These parameters require 15O-CO images. Recently, however, a new technique has been proposed for calculating MBF images directly from dynamic 15O-water scans (21–23). When performing a PET study only for estimating MBF, our program should be equipped with the algorithm free from 15O-CO scan.
CONCLUSION
We have developed a practical and automatic algorithm to quantify rMBF using 15O-water with PET. This method includes a modified method, which is mathematically simplified and more stable than that in the original algorithm. This software yields desirable performance to achieve high reproducibility and practical computational times for clinical use in myocardial PET studies with 15O-water.