JNM
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH
 QUICK SEARCH:   [advanced]


     


First published online December 12, 2007
J Nucl Med 2007, doi:10.2967/jnumed.107.044149
© 2007 by Society of Nuclear Medicine
This Article
Right arrow Full Text (Publish Ahead of Print[PDF])
Right arrow All Versions of this Article:
jnumed.107.044149v1
49/1/79    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Wu, H.
Right arrow Articles by Tai, Y.-C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Wu, H.
Right arrow Articles by Tai, Y.-C.


A Feasibility Study of a Prototype PET Insert Device to Convert a General-Purpose Animal PET Scanner to Higher Resolution

Heyu Wu 1, Debashish Pal 2, Joseph A. O'Sullivan 3, and Yuan-Chuan Tai 4*

1 Mallinckrodt Institute of Radiology, Washington University, St. Louis, Missouri; Institute of Modern Physics, Chinese Academy of Science, Lanzhou, China
2 Department of Biomedical Engineering, Washington University, St. Louis, Missouri
3 Department of Electrical and System Engineering, Washington University, St. Louis, Missouri
4 Mallinckrodt Institute of Radiology, Washington University, St. Louis, Missouri

* To whom correspondence should be addressed. E-mail: taiy{at}wustl.edu.


   Abstract

We developed a prototype system to evaluate the feasibility of using a PET insert device to achieve higher resolution from a general-purpose animal PET scanner. Methods: The system consists of a high-resolution PET detector, a computer-controlled rotation stage, and a custom mounting plate. The detector consists of a cerium-doped lutetium oxyorthosilicate array (12 x 12 crystals, 0.8 x 1.66 x 3.75 mm3 each) directly coupled to a position-sensitive photomultiplier tube (PS-PMT). The detector signals were fed into the scanner electronics to establish coincidences between the 2 systems. The detector was mounted to a rotation stage that is attached to the scanner via the custom mounting plate after removing the transmission source holder. The rotation stage was concentric with the center of the scanner. The angular offset of the insert detector was calibrated via optimizing point-source images. In all imaging experiments, coincidence data were collected from 9 angles to provide 180° sampling. A 22Na point source was imaged at different offsets from the center to characterize the in-plane resolution of the insert system. A 68Ge point source was stepped across the axial field of view to measure the sensitivity of the system. A 23.2-g mouse was injected with 38.5 MBq of 18F-fluoride and imaged at 3 h after injection for 2 h. Results: The transverse image resolution of the PET insert device ranges from 1.1- to 1.4-mm full width at half maximum (FWHM) without correction for the point-source dimension. This corresponds to approximately 33% improvement over the resolution of the original scanner (1.7- to 1.8-mm FWHM) in 2 of the 3 directions. The sensitivity of the device is 0.064% at the center of the field, 46-fold lower than the sensitivity of an existing animal PET scanner. The mouse bone scan had improved image resolution using the PET insert device over that of the existing animal PET scanner alone. Conclusion: We have demonstrated the feasibility of using a high-resolution insert device in an existing PET scanner to provide high-resolution PET. A PET insert device with more detector modules will improve sensitivity and may become an alternative to special-purpose PET systems for high-resolution PET.

Key Words: small-animal PET, high-resolution PET, PET insert, animal imaging




This article has been cited by other articles:


Home page
JNMHome page
H. Wu, D. Pal, T. Y. Song, J. A. O'Sullivan, and Y.-C. Tai
Micro Insert: A Prototype Full-Ring PET Device for Improving the Image Resolution of a Small-Animal PET Scanner
J. Nucl. Med., October 1, 2008; 49(10): 1668 - 1676.