Skip to main content

Main menu

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • Continuing Education
    • JNM Podcasts
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Journal Claims
    • Corporate & Special Sales
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI

User menu

  • Subscribe
  • My alerts
  • Log in
  • My Cart

Search

  • Advanced search
Journal of Nuclear Medicine
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI
  • Subscribe
  • My alerts
  • Log in
  • My Cart
Journal of Nuclear Medicine

Advanced Search

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • Continuing Education
    • JNM Podcasts
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Journal Claims
    • Corporate & Special Sales
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • View or Listen to JNM Podcast
  • Visit JNM on Facebook
  • Join JNM on LinkedIn
  • Follow JNM on Twitter
  • Subscribe to our RSS feeds
Research ArticleBasic Science Investigation
Open Access

ISIT-QA: In Silico Imaging Trial to Evaluate a Low-Count Quantitative SPECT Method Across Multiple Scanner–Collimator Configurations for 223Ra-Based Radiopharmaceutical Therapies

Zekun Li, Nadia Benabdallah, Jingqin Luo, Richard L. Wahl, Daniel L.J. Thorek and Abhinav K. Jha
Journal of Nuclear Medicine May 2024, 65 (5) 810-817; DOI: https://doi.org/10.2967/jnumed.123.266719
Zekun Li
1Department of Biomedical Engineering, Washington University, St. Louis, Missouri;
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Nadia Benabdallah
2Mallinckrodt Institute of Radiology, Washington University, St. Louis, Missouri;
3Program in Quantitative Molecular Therapeutics, Washington University, St. Louis, Missouri;
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Jingqin Luo
4Siteman Cancer Center, Washington University, St. Louis, Missouri;
5Division of Public Health Sciences, Department of Surgery, Washington University, St. Louis, Missouri; and
6Division of Biostatistics, Washington University, St. Louis, Missouri
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Richard L. Wahl
2Mallinckrodt Institute of Radiology, Washington University, St. Louis, Missouri;
4Siteman Cancer Center, Washington University, St. Louis, Missouri;
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Daniel L.J. Thorek
1Department of Biomedical Engineering, Washington University, St. Louis, Missouri;
2Mallinckrodt Institute of Radiology, Washington University, St. Louis, Missouri;
3Program in Quantitative Molecular Therapeutics, Washington University, St. Louis, Missouri;
4Siteman Cancer Center, Washington University, St. Louis, Missouri;
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Abhinav K. Jha
1Department of Biomedical Engineering, Washington University, St. Louis, Missouri;
2Mallinckrodt Institute of Radiology, Washington University, St. Louis, Missouri;
4Siteman Cancer Center, Washington University, St. Louis, Missouri;
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Supplemental
  • Info & Metrics
  • PDF
Loading

Visual Abstract

Figure
  • Download figure
  • Open in new tab
  • Download powerpoint

Abstract

Personalized dose-based treatment planning requires accurate and reproducible noninvasive measurements to ensure safety and effectiveness. Dose estimation using SPECT is possible but challenging for alpha (α)-particle–emitting radiopharmaceutical therapy (α-RPT) because of complex γ-emission spectra, extremely low counts, and various image-degrading artifacts across a plethora of scanner–collimator configurations. Through the incorporation of physics-based considerations and skipping of the potentially lossy voxel-based reconstruction step, a recently developed projection-domain low-count quantitative SPECT (LC-QSPECT) method has the potential to provide reproducible, accurate, and precise activity concentration and dose measures across multiple scanners, as is typically the case in multicenter settings. To assess this potential, we conducted an in silico imaging trial to evaluate the LC-QSPECT method for a 223Ra-based α-RPT, with the trial recapitulating patient and imaging system variabilities. Methods: A virtual imaging trial titled In Silico Imaging Trial for Quantitation Accuracy (ISIT-QA) was designed with the objectives of evaluating the performance of the LC-QSPECT method across multiple scanner–collimator configurations and comparing performance with a conventional reconstruction–based quantification method. In this trial, we simulated 280 realistic virtual patients with bone-metastatic castration-resistant prostate cancer treated with 223Ra-based α-RPT. The trial was conducted with 9 simulated SPECT scanner–collimator configurations. The primary objective of this trial was to evaluate the reproducibility of dose estimates across multiple scanner–collimator configurations using LC-QSPECT by calculating the intraclass correlation coefficient. Additionally, we compared the reproducibility and evaluated the accuracy of both considered quantification methods across multiple scanner–collimator configurations. Finally, the repeatability of the methods was evaluated in a test–retest study. Results: In this trial, data from 268 223RaCl2 treated virtual prostate cancer patients, with a total of 2,903 lesions, were used to evaluate LC-QSPECT. LC-QSPECT provided dose estimates with good reproducibility across the 9 scanner–collimator configurations (intraclass correlation coefficient > 0.75) and high accuracy (ensemble average values of recovery coefficients ranged from 1.00 to 1.02). Compared with conventional reconstruction-based quantification, LC-QSPECT yielded significantly improved reproducibility across scanner–collimator configurations, accuracy, and test–retest repeatability (Embedded Image. Conclusion: LC-QSPECT provides reproducible, accurate, and repeatable dose estimations in 223Ra-based α-RPT as evaluated in ISIT-QA. These findings provide a strong impetus for multicenter clinical evaluations of LC-QSPECT in dose quantification for α-RPTs.

  • in silico imaging trial
  • low-count quantitative SPECT
  • theranostics
  • reproducibility
  • α-particle–emitting radiopharmaceutical therapy

Targeted α-therapy is gaining increasing clinical significance (1–3). Because of their short emission range and high linear energy transfer, α-particles can effectively ablate the regions where they are deposited, with minimal damage to adjacent tissues (3). However, the systemically administered radiopharmaceuticals distribute throughout the patient, accumulating to unknown levels at sites of disease and in radiosensitive vital organs. Thus, it is important to quantify the absorbed doses of the lesions and different organs of the patient treated with these potent agents. Absorbed dose measurement to diseased and healthy tissues is the standard of care for conventional external beam radiotherapy. Such dose quantification in radiopharmaceutical therapies can be beneficial by allowing for adaption of treatment regimens, prediction of therapy outcomes, and monitoring of adverse events (4).

Therapeutic α-particle-emitting isotopes often also emit γ-ray photons detectable by γ-cameras. This provides a mechanism to quantify the absorbed dose by measuring the activity uptake in organs and lesions using quantitative SPECT methods. A major challenge in α-particle–emitting radiopharmaceutical therapy (α-RPT) SPECT is the extremely low number of detected photon counts. This number can be several orders of magnitude lower than in conventional diagnostic SPECT (5). Consequently, conventional quantitative SPECT (QSPECT) methods yield low precision and accuracy in estimating uptake, even with fine-tuned protocols (6–9). Another challenge lies in modeling and compensating for the complicated SPECT physics. α-emitting isotopes, such as 223Ra, 225Ac, and 227Th, have a complex decay chain with multiple progenies (5,10). The γ-spectra of those isotopes usually consist of multiple photopeaks, which further complicates modeling and compensation of SPECT physics including attenuation, scatter, collimator–detector response, and system resolution (11). Inaccurate modeling of the SPECT physics results in systematic variability of uptake quantification across different SPECT scanner–collimator configurations, including from different vendors, hardware specifications, and collimators (12). Previous investigations indicate that even for conventional (high-count) SPECT, differences in uptake estimates between scanner–collimator configurations may reach up to 41% of the actual value (12–14). However, such systematic variability of quantifications remains unexplored in the context of α-RPTs.

Unreliable activity quantification results in a lack of reliability in dose quantification. This significantly impedes the advancement of personalized α-RPT regimens (15) and hinders the computing of robust dose–response relationships for treatment optimization and drug development. Further, variability across SPECT scanner–collimator configurations complicates the combining and comparing of data across instruments and across centers (12–14,16). Critically these inconsistencies hinder multiinstitutional data comparison, a hallmark of the scientific approach to medical advancement. Finally, errors in quantifying absorbed doses in lesions and at-risk organs are undesirable given the high potential tissue toxicity of α-particles (17). This can also lead to an erroneous diagnosis and therapeutic outcome evaluation using radiobiologic models (18).

A recently developed low-count QSPECT (LC-QSPECT) method has demonstrated the ability to yield precise and accurate estimates of regional uptake in α-RPTs. The method estimates the uptake directly from projection data, avoiding information losses inherent in tomographic voxel-based reconstruction processes. In evaluations based on a single scanner–collimator configuration, the method yielded nearly unbiased absolute quantification of regional uptake, with estimate variance being close to the theoretically lowest possible values, outperforming conventional QSPECT methods (9). Notably, this LC-QSPECT method compensates for the physics of given SPECT scanner–collimator configurations directly and accurately in the quantification process. Therefore, we hypothesized that LC-QSPECT could deliver reproducible, repeatable, and accurate dose measurements across multiple SPECT scanner–collimator configurations, making it a compelling candidate for further multiple-center clinical evaluations and potential applications.

Clinical evaluation of the reproducibility across scanner–collimator configurations, accuracy, and repeatability of dose measurements with QSPECT methods requires that a patient administered α-RPT be imaged across multiple SPECT/CT scanner–collimator configurations or multiple times with the same configuration. This process in practice would lead to a substantial CT radiation dose, has ethical issues, is logistically challenging, is time-consuming, and, even if possible, would suffer from the limitation of lack of ground truth. The emerging virtual imaging trial paradigm provides a mechanism to overcome these limitations, enabling rigorous, objective evaluation of imaging technology performance in simulated clinical scenarios that account for patient population variability and system physics (19–21). For instance, the seminal VICTRE in silico trial demonstrated the utility of virtual imaging trials for regulatory evidence of imaging technologies (20). In this study, we conducted a virtual imaging trial (In Silico Imaging Trial for Quantitation Accuracy [ISIT-QA]) that recapitulates patient and imaging system variabilities to evaluate LC-QSPECT across multiple scanner–collimator configurations on the task of dose estimation in α-RPTs. Additionally, the method was compared to a conventional ordered subset expectation maximization (OSEM) method that reconstructed the images over a voxelized grid. We refer to this method as the OSEM method in the rest of the article.

MATERIALS AND METHODS

Study Design

ISIT-QA aimed to assess the performance of the SPECT methods for estimating tissue- and lesion-absorbed dose from 223RaCl2, the first α-RPT approved by the U.S. Food and Drug Administration for the treatment of bone-metastatic castration-resistant prostate cancer (1). Figure 1 shows the overall trial design. This institutional review board–exempt study (study identification no. 202303134) was conducted at Washington University in St. Louis.

FIGURE 1.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 1.

Overall design of ISIT-QA.

In this trial, we simulated 280 realistic and diverse bone-metastatic castration-resistant prostate cancer patients intravenously administered standard 223Ra activities. Virtual patients were excluded from the imaging cohort if errors arose during phantom generations. As in the VICTRE trial, exclusion criteria were enforced by software error messages, computer-aided checks, and visual inspections (20). Eligible virtual patients were imaged with 9 scanner–collimator configurations in silico, followed by a rescan using one of the scanner–collimator configurations. We then applied LC-QSPECT and the OSEM method to estimate the uptake and then the absorbed doses of different volumes of interest (VOIs) of the patients.

The primary objective of this trial was to evaluate the reproducibility of estimated doses obtained with LC-QSPECT across multiple scanner–collimator configurations. The secondary objectives included comparing the reproducibility of LC-QSPECT across scanner–collimator configurations with the OSEM method, evaluating the accuracy of the QSPECT methods across multiple scanner–collimator configurations, and assessing the test–retest repeatability of the methods.

Trial Population

The process to generate the trial population is shown in Figure 2 and briefly summarized here, with details provided in Supplemental Section 1 (supplemental materials are available at http://jnm.snmjournals.org) (7,9,22–30). To simulate virtual patients with bone-metastatic castration-resistant prostate cancer, we first generated 280 healthy virtual U.S. male phantoms using the Extended Cardiac-Torso (XCAT) (23,28). The heights and weights of the virtual patients were directly sampled from data on 2,718 U.S. men published by the Centers for Disease Control and Prevention. Then, we inserted lesions into each phantom, simulating realistic disease characteristics including lesion count, lesion diameters, and the spatial distribution of the lesions, all of which were derived from multiple clinical studies (22,24–27). A rigorous process was followed to insert the lesions, as outlined in the flowchart in Figure 2. Informed by distribution data from preclinical and clinical evaluations of 223Ra (26,27), we considered 5 categories of VOIs in the patients: lesions, bone, small intestine, large intestine, and background. Following a previously published approach (9), we modeled homogeneous activity distribution within each VOI, with mean uptake values determined as illustrated in Figure 2.

FIGURE 2.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 2.

Virtual patient generation process.

Imaging Protocol

We simulated the clinical scenario in which patients received a single intravenous 55 kBq/kg injection of 223Ra, aligning with the recommended dosage (31). On the basis of pharmacokinetic imaging studies demonstrating rapid blood clearance, we assumed patients would exhibit a representative 223Ra biodistribution, as indicated by the activity maps of the virtual patients, 30 min after administration (26). Thus, this time point was selected for imaging the patients.

SPECT scans were generated using SIMIND, a well-validated Monte Carlo simulation software (9,32–34). We further validated our SIMIND-based approach for simulating the imaging of 223Ra across various scanner–collimator configurations using physical phantom studies (Supplemental Section 7; Supplemental Figs. 1–5; Supplemental Tables 1–2 (9,35)). The entire patient cohort was imaged using 9 simulated SPECT scanner–collimator configurations: 3 scanners (GE Healthcare Optima 640, Siemens Symbia, and Philips Precedence), each with 3 collimator configurations (high-energy general-purpose, medium-energy general-purpose, and low-energy high-resolution). Details on simulating the imaging processes are provided in Supplemental Section 2 (6,9,32–34).

After completion of the primary imaging study, patients underwent a second scan 30 min after the end of the first scan, using the simulated Optima 640 scanner with a high-energy general-purpose collimator. We simulated the physical decay of 223Ra (half-life, 11.4 d) and a constant biologic distribution of the isotope over the 30-min duration.

We assumed a CT scan was acquired before each SPECT scan, yielding the required VOI definitions for the QSPECT methods. To reduce unrelated confounding factors, these VOI definitions were assumed to be accurate. Supplemental Section 8 (36,37) explores the impact of potential inaccuracies in the used VOI definitions, caused by rigid or nonrigid patient body transformations, on the performance of the QSPECT methods (Supplemental Figs. 6–7). Additionally, Supplemental Section 9 evaluates the impact of misalignment of VOI definitions in a test–retest study scenario (Supplemental Fig. 8; Supplemental Table 3).

Activity and Dose Quantification: LC-QSPECT Method and Comparison Studies

We first estimated the uptake of 223Ra in lesions and organs of patients using both LC-QSPECT and OSEM methods. The theory and implementation of the LC-QSPECT method were detailed by Li et al. (9) and are presented in Supplemental Section 3 (6,8,9,13,32–34,38–40). Implementation of the OSEM approach (41) is described in Supplemental Section 4 (41–44) and is similar to that presented by Li et al. (9). Briefly, the LC-QSPECT method is a maximum-likelihood technique (38) to estimate the mean uptake in various VOIs directly from SPECT projections, accounting for various image-degrading effects in α-particle SPECT, including the complicated SPECT and isotope physics (9). The OSEM approach adopts the conventional route of first reconstructing the image over a voxelized grid, compensating for all the relevant image-degrading effects in SPECT, followed by averaging the uptake within the VOIs defined over the reconstructed image.

Subsequently, the regional absorbed doses were calculated using output obtained from both methods on the basis of the estimated uptake using the MIRD schema, adopting a single-time-point dosimetry (45) model and assuming no redistribution of 223Ra or its progeny. A detailed description of the dose calculation process is in Supplemental Section 5 (45,46).

Statistical Considerations

For the primary objective, that is, to evaluate the reproducibility of the LC-QSPECT method, we calculated the intraclass correlation coefficient (ICC) of the estimated doses across the 9 scanner–collimator configurations. The ICC was calculated using the single-measurement, absolute-agreement, 2-way mixed-effects model and interpreted according to the guidelines proposed by Koo and Li (47). We calculated the sample size to achieve sufficient precision in estimating the interscanner–collimator-configuration ICC with a 2-sided 95% CI width of 0.1. We targeted an ICC of at least 0.5, ranging from 0.5 to 0.9. The total patient count was calculated as 267, 230, 168, 94, and 30 for ICCs of 0.5, 0.6, 0.7, 0.8, and 0.9, respectively. To account for an approximately 5% drop-out rate, we simulated 280 virtual patients. In addition, we also examined the interscanner- and intercollimator-configuration reproducibility of the method by assessing the ICC of the estimates across the 3 scanners with a high-energy general-purpose collimator and the ICC of estimates across the 3 collimator configurations with the Optima 640 scanner, respectively.

For our secondary objectives, we quantified and compared the accuracy and test–retest repeatability of both QSPECT methods. We used the recovery coefficient (RC)—defined as the ratio of the estimated value to the ground truth—for measuring accuracy. For quantifying test–retest repeatability, we used the within-subject coefficient of variation (CV). To assess the agreement of the estimated doses in the test–retest study, we conducted a Bland–Altman analysis using the percentage difference (defined as the difference between test and retest estimates normalized by their mean (48)), considering the heteroscedasticity of the estimates in this trial. Additionally, we compared the reproducibility of the LC-QSPECT and OSEM methods across all considered scanner–collimator configurations, across the 3 scanners with a high-energy general-purpose collimator, and across the 3 collimator configurations with an Optima 640 scanner, using the ensemble-averaged CV. Details on computing the figures of merit are provided in Supplemental Section 6. Paired-sample t tests were used to determine whether there were statistically significant differences between the performance of LC-QSPECT and OSEM methods.

RESULTS

Trial Accrual

In ISIT-QA, 12 of 280 virtual patients were excluded, leading to a final cohort of 268 virtual patients with a total of 2,903 lesions. The demographics and disease characteristics of these patients are shown in Table 1. Activity and attenuation maps for a representative virtual patient are shown in Figure 3A. Figure 3B presents estimated uptake of an index lesion (indicated in Fig. 3A) using both QSPECT methods. The corresponding OSEM-reconstructed images with different scanner–collimator configurations are shown in Figure 3C. Data generated as part of this virtual imaging trial are available at https://wustl.box.com/v/ISITQA-VIT-data.

View this table:
  • View inline
  • View popup
TABLE 1.

Demographics and Disease Characteristics of Patients in Cohort

FIGURE 3.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 3.

(A) Three-dimensional visualization of activity and attenuation maps of 78th patient in cohort and representative slice from each map. Index lesion is indicated by red arrow. (B) Estimated uptake of index lesion using LC-QSPECT and OSEM methods for this patient. Ground truth uptake also provided. (C) OSEM-reconstructed images of representative slice with each of 9 SPECT scanner–collimator configurations. Lesion boundaries used for OSEM estimations are provided, and they are derived from same lesion masks used in LC-QSPECT method. HEGP = high-energy general-purpose; LEHR = low-energy high-resolution; MEGP = medium-energy general-purpose.

Reproducibility Across Scanner–Collimator Configurations

Table 2 shows the ICCs evaluating the reproducibility of the LC-QSPECT method across all considered scanner–collimator configurations, across the 3 scanners with a high-energy general-purpose collimator, and across the 3 collimator configurations with an Optima 640 scanner. The CIs of all the ICC values were between 0.76 and 0.82, demonstrating that the LC-QSPECT method has good reproducibility (47).

View this table:
  • View inline
  • View popup
TABLE 2.

ICC of Estimated Doses Using LC-QSPECT Method Across All Considered Scanner–Collimator Configurations, Across 3 Scanners with High-Energy General-Purpose Collimator, and Across 3 Collimator Configurations with Optima 640 Scanner

Figures 4A–4C compare the ensemble-averaged CVs of the LC-QSPECT and OSEM methods, evaluating the reproducibility across the scanner–collimator configurations, scanners, and collimator configurations, respectively. Overall, for all groups of VOIs, the LC-QSPECT method yielded ensemble-averaged CVs below 0.4 and consistently outperformed the OSEM method (Embedded Image.

FIGURE 4.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 4.

Ensemble-averaged (Ens.) CV for each group of VOIs evaluating reproducibility of estimated doses across all considered scanner–collimator configurations (A), 3 scanners with high-energy general-purpose collimator (B), and 3 collimators with Optima 640 scanner (C). Error bars show standard error (SE) of ensemble-averaged CVs. *P < 0.01. BKGD = background; LI = large intestine; SI = small intestine.

Accuracy

The distributions of the RCs of the estimated doses in all 2,903 lesions in the cohort, across the 9 SPECT scanner–collimator configurations using the LC-QSPECT and OSEM methods, are shown in Figure 5 as violin plots. Figure 5 also presents the ensemble-averaged RCs with each scanner–collimator configuration. The estimated doses from the LC-QSPECT method were nearly ensemble-unbiased, with ensemble-averaged RCs close to one for all considered scanner–collimator configurations. By comparison, the OSEM approach yielded ensemble RCs of less than one across all acquisition configurations. LC-QSPECT outperformed the OSEM method in accuracy assessment (Embedded Image.

FIGURE 5.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 5.

Violin plots of and ensemble-averaged RCs for estimated doses of all 2,903 lesions in trial cohort, obtained using LC-QSPECT and OSEM methods, with 9 SPECT scanner–collimator configurations. SE of each ensemble-averaged (Ens.) RC is also presented. HEGP = high-energy general-purpose; LCQ = LC-QSPECT; LEHR = low-energy high-resolution; MEGP = medium-energy general-purpose.

Test–Retest Repeatability

Figure 6 shows the Bland–Altman analyses of estimated doses for lesions in 3 diameter ranges using both LC-QSPECT and OSEM methods in the test–retest study. For both methods, the mean percentage differences showed no significant deviation from zero across all lesion diameter ranges. However, the LC-QSPECT method had a lower SD in the percentage difference between the test and retest estimates than did the OSEM method, for lesions with all diameter ranges. Furthermore, as shown in Table 3, the LC-QSPECT method yielded a significantly lower within-subject CV for all groups of VOIs (Embedded Image.

FIGURE 6.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 6.

Bland–Altman plots of estimated doses for lesions with different diameter ranges: >35 mm (A), 25–35 mm (B), and <25 mm (C) in test–retest experiment. Average estimated dose was calculated by mean of first and second estimates. Vertical axis represents percentage difference, defined by difference between first and second estimates normalized by their mean. Means (±95% CI) and SDs of distributions were calculated among all data points in each group.

View this table:
  • View inline
  • View popup
TABLE 3.

Within-Subject CV Calculated for Estimated Doses of VOIs in Each Group in Test–Retest Experiment

DISCUSSION

Accurate and precise estimation of absorbed dose delivered by α-RPTs is important for their preclinical development, clinical translation, and optimal clinical application. The low number of detected photons and complex emission spectra with these therapies pose considerable challenges in using noninvasive methods to make these measures. Further, the wide variety of detector system configurations limits the comparability across data acquired from multiple centers and even within the same institution. In this ISIT-QA trial, we have studied the reproducibility across scanner–collimator configurations, accuracy, and test–retest repeatability in estimating the dose of the Food and Drug Administration–approved 223RaCl2 with LC-QSPECT and OSEM methods.

Figures 4⇑–6 and Tables 2 and 3 demonstrate the high reproducibility across scanner–collimator configurations, accuracy, and repeatability of the LC-QSPECT method in estimating 223Ra-based α-RPT dose. As shown in Table 2 and Figure 4, the LC-QSPECT method yielded highly reproducible dose estimates across the ranges of acquisition configurations as demonstrated by the ICCs and outperformed the conventional OSEM method on this criterion. Across all scanner–collimator configurations, the average ensemble-averaged CVs across the 3 groups of lesions were 0.20 and 0.36 for the LC-QSPECT and OSEM methods, respectively (Fig. 4). These findings are further validated by our physical-phantom study (National Electrical Manufacturers Association phantom) detailed in Supplemental Section 7 and Supplemental Figures 4 and 5. The significance of these results is underscored by our use of a vendor-neutral OSEM reconstruction algorithm that has been demonstrated to reduce interscanner variability in estimates compared with vendor-specific algorithms (12). Comparing the reproducibility of LC-QSPECT across scanner–collimator configurations with vendor-specific OSEM methods warrants future research given their clinical relevance. The ensemble-averaged CVs for the OSEM method were slightly higher than those reported values in previous studies that were not with α-RPTs (12–14). This is expected given the higher noise in α-SPECT due to the lower counts.

Figure 5 shows that LC-QSPECT yielded close-to-zero ensemble bias dose estimates for all considered SPECT scanner–collimator configurations, with average RCs close to 1 (range, 0.999–1.019). In contrast, ensemble-averaged RCs for the OSEM method were around 0.75 (range, 0.726–0.781). A counterintuitive observation in this figure is the narrow ranges of the ensemble-averaged RCs for both LC-QSPECT and OSEM methods, since previous research (49) indicates that the use of low-energy or medium-energy collimators rather than high-energy collimators could result in lower lesion-to-background ratios. This observation, which warrants further investigation, suggests that for regional uptake estimation, the choice of collimator may not have a significant impact on ensemble accuracy. The ability of the LC-QSPECT method to accurately and precisely quantify regional uptake in 223Ra-based α-RPT has been shown using a single SPECT scanner–collimator configuration in our previous work (9). Results from ISIT-QA advance these abilities to a multicenter setup by demonstrating the reliability of the LC-QSPECT method in quantifying doses across multiple SPECT scanner–collimator configurations.

In the test–retest experiment (Fig. 6; Table 3), LC-QSPECT yielded a low within-subject CV of 0.28, on average across all lesions, with very low values for large lesions and soft tissues. The SD of the percentage differences between the first and second estimates calculated across all 2,903 lesions was 39.6%. These values are considerably low given the high noise levels in the SPECT projection and are lower than those from the OSEM method (within-subject CV, 0.37; SD, 52.6%).

The reliable performance of the LC-QSPECT method on the task of measuring dose, as demonstrated in ISIT-QA, is of high clinical significance. Personalized dosimetry, as observed with conventional radiopharmaceutical therapies, has been shown to enhance patient outcomes by optimizing therapeutic dose levels to balance treatment efficacy and minimize adverse events (15). However, the application of personalized dosimetry and dosimetry-guided therapy in α-RPTs has been limited, primarily because of the lack of a reliable dose quantification method (50). The LC-QSPECT method, with its ease of implementation and reliable performance, provides a promising solution for reliable regional dose quantification. In this context, we recognize that microdosimetry is important in α-RPTs (51), but LC-QSPECT (and OSEM SPECT reconstruction methods in α-RPTs) cannot directly estimate uptake distribution at submillimeter resolution. However, the reliable estimates of regional uptake given by the LC-QSPECT method can serve as input to microdosimetry approaches that use pharmacokinetic models, which can model the microscale kinetics based on macrolevel measurements (51,52). Further, the high reproducibility across scanner–collimator configurations and repeatability of the LC-QSPECT method are also important in the development and evaluation of dosimetry-guided therapy models in clinical trials. These properties of the method enable comparing and combining data from multiple centers, thus increasing the robustness of the dose–response relationship prediction (14,16).

We took great care while designing ISIT-QA to ensure clinical realism at each step of the trial pipeline, including assembling the patient cohort and simulating SPECT systems. The realism of the trial is supported by the dose quantification results. As illustrated in Figure 6, the 95% interval of the mean estimated time-integrated doses for all lesions using LC-QSPECT in the test–retest study was 1.4–7.2 Gy, reflecting absorbed doses reported in clinical studies with 223Ra (25).

This trial did not evaluate the impact of multiple readers, nor did it evaluate the variability and inaccuracy in the segmentation of the CT images on the LC-QSPECT approach; both of these issues may be nontrivial (14) and need to be further evaluated. Furthermore, in this trial and that of Li et al. (9), virtual patients had homogeneous uptake within the VOIs, which aligns with the assumption of the LC-QSPECT. However, real-world patients may present with a heterogeneous intraregional uptake distribution. Even though the LC-QSPECT method was observed to be relatively insensitive to certain degrees of intralesional heterogeneous uptake (9) and we can intentionally segment VOIs with relatively uniform uptake, further research could investigate the implications of potential nonuniform uptake within VOIs. Additionally, because of the limited dataset of patient-specific 223Ra radiopharmacokinetics, we assumed a static distribution of the isotope across acquisitions and considered only its physical decay in virtual patients. Therefore, we used a simplified, single-time-point dosimetry estimate focusing primarily on assessing the impact of QSPECT methods on dose. A promising direction for future research is to incorporate models of radiopharmacokinetics into virtual patient simulations and consider this in the dose calculation, which could further enhance the realism of patient simulations and enable more advanced development and evaluation of QSPECT dose estimation techniques. Data from the ongoing clinical trial evaluating the distributions of 223Ra in α-RPTs (e.g., NCT04521361) might be used for this purpose in the future.

The encouraging outcomes observed in ISIT-QA provide motivation and inputs toward designing a multicenter patient trial for clinical evaluation of LC-QPSECT. On the basis of the high reproducibility across scanner–collimator configurations, accuracy, and test–retest repeatability of the LC-QSPECT seen in the ISIT-QA trial, we anticipate that the patient trial will indicate similar performance for LC-QPSECT and thus provide stronger evidence for the use of LC-QSPECT in modeling the radiopharmacokinetics of 223Ra and strengthening the robustness of dose–response relationship models, thereby advancing the personalization of treatment in 223Ra-based α-RPTs.

CONCLUSION

Results from ISIT-QA demonstrate that a projection-domain LC-QSPECT method yielded both high reproducibility and high accuracy across multiple scanner–collimator configurations and high test–retest repeatability on the task of estimating dose for α-RPTs. Further, LC-QSPECT outperformed the conventional OSEM method on this estimation task. These results demonstrate the potential of, and motivate the multicenter clinical assessment of, dose quantification using LC-QSPECT in α-RPTs.

DISCLOSURE

Financial support was received from National Institute of Biomedical Imaging and Bioengineering grants R01-EB031962 and R01-EB031051 and from a Society of Nuclear Medicine and Molecular Imaging 2020 Student Research grant. No other potential conflict of interest relevant to this article was reported.

KEY POINTS

QUESTION: How well does the projection-domain LC-QSPECT method perform in dose quantification for 223Ra-based α-RPTs across multiple SPECT scanner–collimator configurations within an in silico clinical trial setting?

PERTINENT FINDINGS: ISIT-QA showed that LC-QSPECT provided high reproducibility and accuracy across multiple scanner–collimator configurations and high test–retest repeatability on the dose estimation task, outperforming the conventional reconstruction-based quantification method.

IMPLICATIONS FOR PATIENT CARE: The LC-QSPECT method holds promise for application in multicenter clinical trials and for helping personalize α-RPTs.

Footnotes

  • Published online Apr. 4, 2024.

  • © 2024 by the Society of Nuclear Medicine and Molecular Imaging.

Immediate Open Access: Creative Commons Attribution 4.0 International License (CC BY) allows users to share and adapt with attribution, excluding materials credited to previous publications. License: https://creativecommons.org/licenses/by/4.0/. Details: http://jnm.snmjournals.org/site/misc/permission.xhtml.

REFERENCES

  1. 1.↵
    1. Kluetz PG,
    2. Pierce W,
    3. Maher VE,
    4. et al
    . Radium Ra 223 dichloride injection: U.S. Food and Drug Administration drug approval summary. Clin Cancer Res. 2014;20:9–14.
    OpenUrlAbstract/FREE Full Text
  2. 2.
    1. McDevitt MR,
    2. Thorek DL,
    3. Hashimoto T,
    4. et al
    . Feed-forward alpha particle radiotherapy ablates androgen receptor-addicted prostate cancer. Nat Commun. 2018;9:1629.
    OpenUrl
  3. 3.↵
    1. Tafreshi NK,
    2. Doligalski ML,
    3. Tichacek CJ,
    4. et al
    . Development of targeted alpha particle therapy for solid tumors. Molecules. 2019;24:4314.
    OpenUrl
  4. 4.↵
    1. Brans B,
    2. Bodei L,
    3. Giammarile F,
    4. et al
    . Clinical radionuclide therapy dosimetry: the quest for the “Holy Gray.” Eur J Nucl Med Mol Imaging. 2007;34:772–786.
    OpenUrlCrossRefPubMed
  5. 5.↵
    1. Benabdallah N,
    2. Scheve W,
    3. Dunn N,
    4. et al
    . Practical considerations for quantitative clinical SPECT/CT imaging of alpha particle emitting radioisotopes. Theranostics. 2021;11:9721–9737.
    OpenUrl
  6. 6.↵
    1. Benabdallah N,
    2. Bernardini M,
    3. Bianciardi M,
    4. de Labriolle-Vaylet C,
    5. Franck D,
    6. Desbrée A
    . 223Ra-dichloride therapy of bone metastasis: optimization of SPECT images for quantification. EJNMMI Res. 2019;9:20.
    OpenUrl
  7. 7.↵
    1. Gustafsson J,
    2. Rodeño E,
    3. Mínguez P
    . Feasibility and limitations of quantitative SPECT for 223Ra. Phys Med Biol. 2020;65:085012.
    OpenUrl
  8. 8.↵
    1. Yue J,
    2. Hobbs R,
    3. Sgouros G,
    4. Frey E
    . SU‐F‐J‐08: quantitative SPECT imaging of Ra‐223 in a phantom. Med Phys. 2016;43:3407.
    OpenUrl
  9. 9.↵
    1. Li Z,
    2. Benabdallah N,
    3. Abou DS,
    4. et al
    . A projection-domain low-count quantitative SPECT method for ɑ-particle-emitting radiopharmaceutical therapy. IEEE Trans Radiat Plasma Med Sci. 2023;7:62–74.
    OpenUrl
  10. 10.↵
    1. Abou DS,
    2. Pickett J,
    3. Mattson JE,
    4. Thorek DLA
    . Radium-223 microgenerator from cyclotron-produced trace actinium-227. Appl Radiat Isot. 2017;119:36–42.
    OpenUrl
  11. 11.↵
    1. Sgouros G,
    2. Frey E,
    3. Du Y,
    4. Hobbs R,
    5. Bolch W
    . Imaging and dosimetry for alpha-particle emitter radiopharmaceutical therapy: improving radiopharmaceutical therapy by looking into the black box. Eur J Nucl Med Mol Imaging. 2021;49:18–29.
    OpenUrl
  12. 12.↵
    1. Peters SM,
    2. Meyer Viol SL,
    3. van der Werf NR,
    4. et al
    . Variability in lutetium-177 SPECT quantification between different state-of-the-art SPECT/CT systems. EJNMMI Phys. 2020;7:9.
    OpenUrl
  13. 13.↵
    1. Lassmann M,
    2. Eberlein U,
    3. Tran-Gia J
    . Multicentre trials on standardised quantitative imaging and dosimetry for radionuclide therapies. Clin Oncol (R Coll Radiol). 2021;33:125–130.
    OpenUrl
  14. 14.↵
    1. Uribe C,
    2. Peterson A,
    3. Van B,
    4. et al
    . An international study of factors affecting variability of dosimetry calculations, part 1: design and early results of the SNMMI dosimetry challenge. J Nucl Med. 2021;62(suppl 3):36S–47S.
    OpenUrl
  15. 15.↵
    1. Beauregard J-M
    . Use of imaging-based dosimetry for personalising radiopharmaceutical therapy of cancer. Cancer Imaging. 2022;22:67.
    OpenUrl
  16. 16.↵
    1. Gregory RA,
    2. Murray I,
    3. Gear J,
    4. et al
    . Standardised quantitative radioiodine SPECT/CT imaging for multicentre dosimetry trials in molecular radiotherapy. Phys Med Biol. 2019;64:245013.
    OpenUrl
  17. 17.↵
    1. Humm JL,
    2. Sartor O,
    3. Parker C,
    4. Bruland OS,
    5. Macklis R
    . Radium-223 in the treatment of osteoblastic metastases: a critical clinical review. Int J Radiat Oncol Biol Phys. 2015;91:898–906.
    OpenUrlPubMed
  18. 18.↵
    1. Sgouros G,
    2. Hobbs RF
    . Dosimetry for radiopharmaceutical therapy. Semin Nucl Med. 2014;44:172–178.
  19. 19.↵
    1. Abadi E,
    2. Segars WP,
    3. Tsui BM,
    4. et al
    . Virtual clinical trials in medical imaging: a review. J Med Imaging (Bellingham). 2020;7:042805.
    OpenUrl
  20. 20.↵
    1. Badano A,
    2. Graff CG,
    3. Badal A,
    4. et al
    . Evaluation of digital breast tomosynthesis as replacement of full-field digital mammography using an in silico imaging trial. JAMA Netw Open. 2018;1:e185474.
    OpenUrl
  21. 21.↵
    1. Jha AK,
    2. Myers KJ,
    3. Obuchowski NA,
    4. et al
    . Objective task-based evaluation of artificial intelligence-based medical imaging methods: framework, strategies, and role of the physician. PET Clin. 2021;16:493–511.
    OpenUrl
  22. 22.↵
    1. Parker C,
    2. Nilsson S,
    3. Heinrich D,
    4. et al
    . Alpha emitter radium-223 and survival in metastatic prostate cancer. N Engl J Med. 2013;369:213–223.
    OpenUrlCrossRefPubMed
  23. 23.↵
    1. Segars WP,
    2. Sturgeon G,
    3. Mendonca S,
    4. Grimes J,
    5. Tsui BM
    . 4D XCAT phantom for multimodality imaging research. Med Phys. 2010;37:4902–4915.
    OpenUrlCrossRefPubMed
  24. 24.↵
    1. Smith M,
    2. Parker C,
    3. Saad F,
    4. et al
    . Addition of radium-223 to abiraterone acetate and prednisone or prednisolone in patients with castration-resistant prostate cancer and bone metastases (ERA 223): a randomised, double-blind, placebo-controlled, phase 3 trial. Lancet Oncol. 2019;20:408–419.
    OpenUrlCrossRefPubMed
  25. 25.↵
    1. Murray I,
    2. Chittenden SJ,
    3. Denis-Bacelar AM,
    4. et al
    . The potential of 223Ra and 18F-fluoride imaging to predict bone lesion response to treatment with 223Ra-dichloride in castration-resistant prostate cancer. Eur J Nucl Med Mol Imaging. 2017;44:1832–1844.
    OpenUrl
  26. 26.↵
    1. Chittenden SJ,
    2. Hindorf C,
    3. Parker CC,
    4. et al
    . A phase 1, open-label study of the biodistribution, pharmacokinetics, and dosimetry of 223Ra-dichloride in patients with hormone-refractory prostate cancer and skeletal metastases. J Nucl Med. 2015;56:1304–1309.
    OpenUrlAbstract/FREE Full Text
  27. 27.↵
    1. Yoshida K,
    2. Kaneta T,
    3. Takano S,
    4. et al
    . Pharmacokinetics of single dose radium-223 dichloride (BAY 88-8223) in Japanese patients with castration-resistant prostate cancer and bone metastases. Ann Nucl Med. 2016;30:453–460.
    OpenUrl
  28. 28.↵
    1. Segars WP,
    2. Bond J,
    3. Frush J,
    4. et al
    . Population of anatomically variable 4D XCAT adult phantoms for imaging research and optimization. Med Phys. 2013;40:043701.
    OpenUrlCrossRefPubMed
  29. 29.
    1. Song N,
    2. Du Y,
    3. He B,
    4. Frey EC
    . Development and evaluation of a model‐based downscatter compensation method for quantitative I‐131 SPECT. Med Phys. 2011;38:3193–3204.
    OpenUrlCrossRefPubMed
  30. 30.↵
    1. Jha AK,
    2. Caffo B,
    3. Frey EC
    . A no-gold-standard technique for objective assessment of quantitative nuclear-medicine imaging methods. Phys Med Biol. 2016;61:2780.
    OpenUrl
  31. 31.↵
    1. Zimmerman BE,
    2. Bergeron D,
    3. Cessna J,
    4. Fitzgerald R,
    5. Pibida L
    . Revision of the NIST standard for 223Ra: new measurements and review of 2008 data. J Res Natl Inst Stand Technol. 2015;120:37–57.
    OpenUrl
  32. 32.↵
    1. Ljungberg M,
    2. Strand S-E
    . A Monte Carlo program for the simulation of scintillation camera characteristics. Comput Methods Programs Biomed. 1989;29:257–272.
    OpenUrlCrossRefPubMed
  33. 33.
    1. Bahreyni Toossi MT,
    2. Islamian JP,
    3. Momennezhad M,
    4. Ljungberg M,
    5. Naseri S
    . SIMIND Monte Carlo simulation of a single photon emission CT. J Med Phys. 2010;35:42–47.
    OpenUrl
  34. 34.↵
    1. Morphis M,
    2. van Staden JA,
    3. du Raan H,
    4. Ljungberg M
    . Validation of a SIMIND Monte Carlo modelled gamma camera for iodine-123 and iodine-131 imaging. Heliyon. 2021;7:e07196.
    OpenUrl
  35. 35.↵
    1. Jaszczak RJ,
    2. Floyd CE,
    3. Coleman RE
    . Scatter compensation techniques for SPECT. IEEE Trans Nucl Sci. 1985;32:786–793.
    OpenUrlCrossRef
  36. 36.↵
    1. Song N,
    2. He B,
    3. Frey E
    . The effect of volume-of-interest misregistration on quantitative planar activity and dose estimation. Phys Med Biol. 2010;55:5483–5497.
    OpenUrl
  37. 37.↵
    1. Tang L,
    2. Hamarneh G,
    3. Celler A
    . Co-registration of bone CT and SPECT images using mutual information. IEEE Xplore website. https://ieeexplore.ieee.org/document/4042223. Published January 29, 2007. Accessed March 4, 2024.
  38. 38.↵
    1. Shepp LA,
    2. Vardi Y
    . Maximum likelihood reconstruction for emission tomography. IEEE Trans Med Imaging. 1982;1:113–122.
    OpenUrlCrossRefPubMed
  39. 39.
    1. Soret M,
    2. Bacharach SL,
    3. Buvat I
    . Partial-volume effect in PET tumor imaging. J Nucl Med. 2007;48:932–945.
    OpenUrlAbstract/FREE Full Text
  40. 40.↵
    1. Cloquet C,
    2. Defrise M
    . MLEM and OSEM deviate from the Cramer-Rao bound at low counts. IEEE Trans Nucl Sci. 2012;60:134–143.
    OpenUrl
  41. 41.↵
    1. Hudson HM,
    2. Larkin RS
    . Accelerated image reconstruction using ordered subsets of projection data. IEEE Trans Med Imaging. 1994;13:601–609.
    OpenUrlCrossRefPubMed
  42. 42.
    1. Merlin T,
    2. Stute S,
    3. Benoit D,
    4. et al
    . CASToR: a generic data organization and processing code framework for multi-modal and multi-dimensional tomographic reconstruction. Phys Med Biol. 2018;63:185005.
    OpenUrl
  43. 43.
    1. Ichihara T,
    2. Ogawa K,
    3. Motomura N,
    4. Kubo A,
    5. Hashimoto S
    . Compton scatter compensation using the triple-energy window method for single-and dual-isotope SPECT. J Nucl Med. 1993;34:2216–2221.
    OpenUrlAbstract/FREE Full Text
  44. 44.↵
    1. King M,
    2. Devries D,
    3. Pan T-S,
    4. Pretorius P,
    5. Case J
    . An investigation of the filtering of TEW scatter estimates used to compensate for scatter with ordered subset reconstructions. IEEE Trans Nucl Sci. 1997;44:1140–1145.
    OpenUrlCrossRef
  45. 45.↵
    1. Hou X,
    2. Brosch J,
    3. Uribe C,
    4. et al
    . Feasibility of single-time-point dosimetry for radiopharmaceutical therapies. J Nucl Med. 2021;62:1006–1011.
    OpenUrlAbstract/FREE Full Text
  46. 46.↵
    1. Schumann S,
    2. Eberlein U,
    3. Muhtadi R,
    4. Lassmann M,
    5. Scherthan H
    . DNA damage in leukocytes after internal ex-vivo irradiation of blood with the α-emitter Ra-223. Sci Rep. 2018;8:2286.
    OpenUrl
  47. 47.↵
    1. Koo TK,
    2. Li MY
    . A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med. 2016;15:155–163.
    OpenUrlCrossRefPubMed
  48. 48.↵
    1. Giavarina D
    . Understanding Bland Altman analysis. Biochem Med (Zagreb). 2015;25:141–151.
    OpenUrlCrossRef
  49. 49.↵
    1. Owaki Y,
    2. Nakahara T,
    3. Kosaka T,
    4. et al
    . Ra-223 SPECT for semi-quantitative analysis in comparison with Tc-99m HMDP SPECT: phantom study and initial clinical experience. EJNMMI Res. 2017;7:81.
    OpenUrl
  50. 50.↵
    1. Pandit-Taskar N,
    2. Iravani A,
    3. Lee D,
    4. et al
    . Dosimetry in clinical radiopharmaceutical therapy of cancer: practicality versus perfection in current practice. J Nucl Med. 2021;62(suppl 3):60S–72S.
    OpenUrl
  51. 51.↵
    1. Sgouros G,
    2. Hobbs R,
    3. Song H
    . Modelling and dosimetry for alpha-particle therapy. Curr Radiopharm. 2011;4:261–265.
    OpenUrlCrossRefPubMed
  52. 52.↵
    1. Hobbs RF,
    2. Song H,
    3. Huso DL,
    4. Sundel MH,
    5. Sgouros G
    . A nephron-based model of the kidneys for macro-to-micro α-particle dosimetry. Phys Med Biol. 2012;57:4403–4424.
    OpenUrl
  • Received for publication September 19, 2023.
  • Revision received February 13, 2024.
PreviousNext
Back to top

In this issue

Journal of Nuclear Medicine: 65 (5)
Journal of Nuclear Medicine
Vol. 65, Issue 5
May 1, 2024
  • Table of Contents
  • Table of Contents (PDF)
  • About the Cover
  • Index by author
  • Complete Issue (PDF)
Print
Download PDF
Article Alerts
Sign In to Email Alerts with your Email Address
Email Article

Thank you for your interest in spreading the word on Journal of Nuclear Medicine.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
ISIT-QA: In Silico Imaging Trial to Evaluate a Low-Count Quantitative SPECT Method Across Multiple Scanner–Collimator Configurations for 223Ra-Based Radiopharmaceutical Therapies
(Your Name) has sent you a message from Journal of Nuclear Medicine
(Your Name) thought you would like to see the Journal of Nuclear Medicine web site.
Citation Tools
ISIT-QA: In Silico Imaging Trial to Evaluate a Low-Count Quantitative SPECT Method Across Multiple Scanner–Collimator Configurations for 223Ra-Based Radiopharmaceutical Therapies
Zekun Li, Nadia Benabdallah, Jingqin Luo, Richard L. Wahl, Daniel L.J. Thorek, Abhinav K. Jha
Journal of Nuclear Medicine May 2024, 65 (5) 810-817; DOI: 10.2967/jnumed.123.266719

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Share
ISIT-QA: In Silico Imaging Trial to Evaluate a Low-Count Quantitative SPECT Method Across Multiple Scanner–Collimator Configurations for 223Ra-Based Radiopharmaceutical Therapies
Zekun Li, Nadia Benabdallah, Jingqin Luo, Richard L. Wahl, Daniel L.J. Thorek, Abhinav K. Jha
Journal of Nuclear Medicine May 2024, 65 (5) 810-817; DOI: 10.2967/jnumed.123.266719
Twitter logo Facebook logo LinkedIn logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One
Bookmark this article

Jump to section

  • Article
    • Visual Abstract
    • Abstract
    • MATERIALS AND METHODS
    • RESULTS
    • DISCUSSION
    • CONCLUSION
    • DISCLOSURE
    • Footnotes
    • REFERENCES
  • Figures & Data
  • Supplemental
  • Info & Metrics
  • PDF

Related Articles

  • PubMed
  • Google Scholar

Cited By...

  • No citing articles found.
  • Google Scholar

More in this TOC Section

  • 225Ac α-Pretargeted Radioimmunotherapy of Human Epidermal Growth Factor Receptor 2–Expressing Breast Cancer
  • 212Bi-Macroaggregated Albumin Inhibited Mouse Melanoma Growth by Regulating Cell Cycle Checkpoint Markers Without Promoting Living Cell Repopulation
  • In Vivo Visualization and Quantification of Brain Heat Shock Protein 90 with [11C]HSP990 in Healthy Aging and Neurodegeneration
Show more Basic Science Investigation

Similar Articles

Keywords

  • in silico imaging trial
  • low-count quantitative SPECT
  • theranostics
  • reproducibility
  • α-particle–emitting radiopharmaceutical therapy
SNMMI

© 2025 SNMMI

Powered by HighWire