Journal of Nuclear Medicine
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI
  • Subscribe
  • My alerts
  • Log in
  • My Cart
Journal of Nuclear Medicine

Advanced Search

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • Continuing Education
    • JNM Podcasts
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Journal Claims
    • Corporate & Special Sales
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • View or Listen to JNM Podcast
  • Visit JNM on Facebook
  • Join JNM on LinkedIn
  • Follow JNM on Twitter
  • Subscribe to our RSS feeds
Research ArticleOncology

Qualification of National Cancer Institute–Designated Cancer Centers for Quantitative PET/CT Imaging in Clinical Trials

Joshua S. Scheuermann, Janet S. Reddin, Adam Opanowski, Paul E. Kinahan, Barry A. Siegel, Lalitha K. Shankar and Joel S. Karp
Journal of Nuclear Medicine July 2017, 58 (7) 1065-1071; DOI: https://doi.org/10.2967/jnumed.116.186759
Joshua S. Scheuermann
1Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Janet S. Reddin
1Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Adam Opanowski
2American College of Radiology Imaging Network, Philadelphia, Pennsylvania
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Paul E. Kinahan
3Department of Radiology, University of Washington, Seattle, Washington
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Barry A. Siegel
4Mallinckrodt Institute of Radiology and Alvin J. Siteman Cancer Center, School of Medicine, Washington University, St. Louis, Missouri; and
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Lalitha K. Shankar
5Cancer Imaging Program, National Cancer Institute, Rockville, Maryland
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Joel S. Karp
1Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Info & Metrics
  • PDF
Loading

Abstract

The National Cancer Institute developed the Centers for Quantitative Imaging Excellence (CQIE) initiative in 2010 to prequalify imaging facilities at all of the National Cancer Institute–designated comprehensive and clinical cancer centers for oncology trials using advanced imaging techniques, including PET. Here we review the CQIE PET/CT scanner qualification process and results in detail. Methods: Over a period of approximately 5 y, sites were requested to submit a variety of phantoms, including uniform and American College of Radiology–approved phantoms, PET/CT images, and examples of clinical images. Submissions were divided into 3 distinct time periods: initial submission (T0) and 2 requalification submissions (T1 and T2). Images were analyzed using standardized procedures, and scanners received a pass or fail designation. Sites had the opportunity to submit new data for scanners that failed. Quantitative results were compared across scanners within a given time period and across time periods for a given scanner. Results: Data from 65 unique PET/CT scanners across 56 sites were submitted for CQIE T0 qualification; 64 scanners passed the qualification. Data from 44 (68%) of those 65 scanners were submitted for T2. From T0 to T2, the percentage of scanners passing the CQIE qualification on the first attempt rose from 38% for T1 to 67% for T2. The most common reasons for failure were SUV outside specifications, incomplete submission, and uniformity issues. Uniform phantom and American College of Radiology–approved phantom results between scanner manufacturers were similar. Conclusion: The results of the CQIE process showed that periodic requalification may decrease the frequency of deficient data submissions. The CQIE project also highlighted the concern within imaging facilities about the burden of maintaining different qualifications and accreditations. Finally, for quantitative imaging–based trials, further evaluation of the relationships between the level of the qualification (e.g., bias or precision) and the quality of the image data, accrual rates, and study power is needed.

  • CQIE
  • PET qualification
  • quantitative imaging

Increasingly, 18F-FDG and other radiopharmaceuticals are being used as PET/CT quantitative imaging biomarkers in oncology to assess treatment efficacy (1–8). Although many factors, both physiologic and instrumental (9–12), influence quantitative accuracy, the ability to determine treatment efficacy on the basis of PET is predicated on the ability of PET scanners to provide stable measurements of radiotracer concentrations—thus allowing the treatment response to be tracked over months or years. Short-term scanner variability is expected to be low, and long-term scanner variability likely can be minimized with standardized quality control procedures (13,14). Recently, however, it was shown that long-term stability cannot be taken for granted and should be checked (15,16).

An additional complication arises when accrual for research trials is accelerated by expansion of imaging to multiple centers. Because of differences in procedures at imaging sites, some PET studies may not be quantitatively reliable or the data may not be usable in a pooled analysis. It has been reported that approximately one-third of PET studies acquired at community-based imaging facilities may lack the necessary information to obtain quantitative imaging data (17). In addition, quantitative variability in multicenter trials can be expected to be larger than that in single-center trials (18,19). In a study at a single institution with multiple PET scanners, which were clinically accredited and maintained in accordance with manufacturer standards by qualified staff, the variance of PET measurements was shown to be greater in clinical practice than under ideal study settings (20).

These reports of PET scanner bias and variability highlight the need for standardized qualification processes to minimize variability in multicenter research trials. However, the need to qualify imaging systems before participation in research trials increases the time needed to accrue trial data, because it takes time to acquire qualification data, send it to the study sponsor for analysis, and obtain approval to participate in the trial. Methods for the quantitative qualification of PET scanners include testing accuracy and basic image quality (21), prospectively assessing accuracy and contrast recovery, and developing scanner-specific reconstruction parameters to unify contrast recovery coefficients across scanners (22,23). These efforts have resulted in guidelines for tumor imaging using 18F-FDG PET in research trials (24–26).

On the basis of this understanding of the importance of the quantitative accuracy of imaging biomarkers in clinical trials, the National Cancer Institute developed the Centers for Quantitative Imaging Excellence (CQIE) initiative in 2010 to prequalify imaging facilities at all of the National Cancer Institute–designated comprehensive and clinical cancer centers for oncology trials using CT, MRI, and PET. The intention of the CQIE project was to establish a group of “trial-ready” cancer imaging facilities to minimize the time between a multicenter research trial being developed and sites beginning accrual. An overview of the CQIE program for all 3 imaging modalities is provided elsewhere (27). Here we review the CQIE PET/CT scanner qualification process and results in more detail.

MATERIALS AND METHODS

Study Time Frame

The CQIE data submissions occurred from June 2010 to March 2014. They were divided into 3 time periods: T0, T1, and T2. T0 was the baseline qualification submission and ran from June 2010 to December 2011. T1 and T2 were subsequent, approximately annual requalification submissions. T1 ran from January 2012 to December 2012, and T2 ran from January 2013 to March 2014.

PET Qualification Procedure

Sites were provided with detailed instructions and training modules. If necessary, test phantoms were provided. Sites were required to submit a series of 7 phantoms and sample patient images for review: a uniform phantom using a static body protocol, a uniform phantom using a static brain protocol, a uniform phantom using a dynamic body protocol, an American College of Radiology (ACR) phantom using a static body protocol, an ACR-approved phantom using a static brain protocol, 2 anonymized patient brain scan samples, and 2 anonymized patient body scan samples. In an effort to remain consistent with existing ACR Imaging Network (ACRIN) qualification and ACR accreditation programs, sites were requested to reconstruct all static body and brain images using their standard clinical reconstruction protocols.

Each site was requested to select 1 scanner for qualification. For the initial T0 qualification tests, imaging centers were given the option of an on-site visit by a member of the CQIE qualification team to facilitate scanning and qualification. Phantoms (if needed) and CQIE standard operating procedure materials were then forwarded to the site within 2 wk of the planned site visit. Methods for image transfer via secure file transfer protocol were also established at this time. Sites were encouraged to go online to complete a review of learning modules describing the importance of the program before their site visit. For the T1 and T2 qualification tests, there were no visits by members of the CQIE qualification team.

Uniform Phantom Data Acquisition and Analysis

The uniform phantom datasets were based on a standard fillable cylinder without features, nominally 20 cm in diameter and length. The uniform phantom was filled with a dilute 18F solution to a concentration of approximately 5.00–6.11 kBq/mL (135–165 nCi/mL). For the brain dataset, the phantom was scanned at 1 bed position using the site’s standard clinical brain acquisition and reconstruction protocols. For the body dataset, the phantom was scanned at 2 bed positions using the site’s standard clinical body acquisition and reconstruction protocols. The sites were asked to measure and report the SUV for a large central region of interest (ROI).

For the dynamic phantom study, because many sites do not routinely perform dynamic imaging, the site was required to scan the phantom using 1 bed position and a specific timing protocol. The data were then reconstructed with a protocol that likely would be used for dynamic scanning. The results from this component are reported elsewhere and are not further discussed here (27,28).

The uniform phantom images were transferred to a central core laboratory and imported into an OsiriX display platform (29) for analysis. For each phantom dataset, several fields in the DICOM (Digital Imaging and Communications in Medicine) headers were compared with the site-reported data forms to verify the accuracy of the SUV calculations. These fields included AcquisitionTime, Weight, RadiopharmaceuticalStartTime, and RadionuclideTotalDose. If these fields were found to be inconsistent, then the site was contacted and asked to clarify the discrepancy. Once any DICOM header discrepancies were resolved, a 200-cm2 (approximately 16 cm in diameter) circular ROI was placed on every axial slice in the phantom, and the mean and SD of the SUV for each voxel within the ROI were recorded. For the static brain and body phantoms, axial slices up to 1.5 cm from the axial edge of the field of view (FOV) or the edge of the phantom were excluded from the analysis because of the typical falloff at the axial edge of the FOV and potential edge effects near the ends of the phantom.

For the static body and brain protocol phantom acquisitions, a volume average SUV was computed from the average of the mean SUVs from each axial slice. The maximum axial deviation also was calculated by finding the difference between the maximum mean slice SUV and the minimum mean slice SUV and dividing that value by the volume average SUV. For a static uniform phantom dataset to pass the quantitative analysis, the volume average SUV had to be between 0.90 and 1.10 and the maximum axial deviation had to be less than 10%. If the results were outside these specifications, then the site was contacted to try to determine the reason for the failure and to resolve the problem.

ACR-Approved Phantom Data Acquisition and Analysis

The ACR-approved PET phantom (30) contains a series of 4 hot contrast cylinders with diameters of 8, 13, 17, and 25 mm, a length of 25 mm, and a nominal cylinder-to-background ratio of 4:1. The phantom was filled in accordance with standard instructions (30), assuming a patient dose of 444 MBq (12 mCi), and scanned using the same body and brain protocols as those used for the uniform phantom. The ACR-approved phantom was chosen for this study because it is a common phantom in use by many clinical centers, thus obviating the purchase of a more complicated phantom.

In addition to the phantom datasets, sites were required to submit 2 anonymized patient brain scans and 2 anonymized patient body scans. The patient test cases were acquired with the sites’ standard clinical protocols. In addition, the image acquisition and reconstruction parameters used by the sites were recorded.

The ACR-approved phantom data were transferred to a central core laboratory and imported into OsiriX for analysis. The same comparison of DICOM headers as that done for each uniform phantom dataset was done for each ACR-approved phantom dataset. The SUV analysis was done in accordance with ACR instructions. First, image planes were summed together to form images that were between 9 and 12 mm thick. The slice that best showed the 4 hot cylinders was selected, and circular background and small cylinder ROIs were drawn; the background ROI was about 6–7 cm in diameter in the center of the image, and the small cylinder ROI was drawn just inside the largest hot cylinder. Copies of the small ROI (<25 mm in diameter) were drawn over the other hot cylinders and over the air, water, and bone cylinders. As part of the core laboratory analysis, we also recorded the SUVpeak measurement for each hot cylinder. The SUVpeak was defined as the average SUV in a 1.0-cm-diameter circular ROI centered on the maximum pixel in each cylinder. The recovery coefficients (ratio of measured SUV to true SUV) for the hot cylinders were plotted as a function of the cylinder diameters.

Patient Test Case Analysis

For the patient brain and body test cases, the DICOM headers were reviewed for accuracy and compared with the data forms. Any discrepancies were investigated with the site to determine the source of the discrepancies. Once any discrepancies were resolved, a qualitative review of overall image quality was performed. The fusion between the PET and CT images was checked, the patient positioning in the FOV was evaluated, and the appropriateness of the acquisition and reconstruction settings was evaluated on the basis of the overall qualitative smoothness of the PET images. For the brain test cases, SUVs were not recorded, but the ability to measure SUVs was verified. For the body test cases, SUV analysis of the liver was performed. A large, 2-dimensional, elliptic ROI was drawn on 7 consecutive transverse slices through the middle of the liver, as shown in Figure 1. The SUVmean for each 2-dimensional ROI was recorded, and an area-weighted average of the means was computed to determine the average liver SUV for each test case.

FIGURE 1.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 1.

Image of 1 of 7 adjacent ROIs in liver regions used in body test cases.

Qualification Summary Comparisons of T0 and T2

For all of the T0 scanners reviewed, we determined how many scanners passed the qualification review without any failure or intervention from the core laboratory, with a single failure, or with multiple failures. A failure was considered any issue that prevented the scanner from passing the qualification review without involvement of the CQIE. The reasons for failures also were cataloged. They were divided into the following categories: uniformity problem, SUV outside specifications, phantom filling issue, reconstruction problem, improper acquisition, incomplete submission, and problem with data forms.

The same analysis of qualification results was performed for T2 scanners that had also submitted data at T0.

RESULTS

Accrual

During the T0 period, a total of 65 PET scanners underwent CQIE testing. Most of the sites opted for on-site visits by the PET CQIE team.

During the T1 period, sites that participated in the T0 period were sent reminders that requalification was needed. No specific follow-up was undertaken. In addition, no on-site visits were provided, and sites were not given access to funds to defray costs associated with scanner qualification. Site participation decreased dramatically in year 2 relative to year 1. The year 2 participation rate dropped to 39 scanners; for some of these, data were not submitted for the T0 time period. The T2 participation rate increased to 52 scanners; for 44 of these, data were submitted for the T0 time period. Because of poor data accrual during the T1 period, only data from the T0 and T2 periods are reported.

Data from 65 unique PET/CT scanners at 56 sites were analyzed for the T0 period. Data for 1 of the 65 scanners could not be reviewed because the images could not be submitted in the DICOM format. For the T2 period, data from the 44 PET/CT scanners that had also submitted data for the T0 period were analyzed.

Uniform Phantom Results

An example of a plot of the SUVmean for each image plane is shown in Figure 2, illustrating the calculation of the maximum axial deviation (MAD). The MAD is a surrogate for evaluating the quality of system normalization. In our experience at the University of Pennsylvania, we have found that renormalization decreases the MAD and improves the flatness of the “axial” profile. Also shown are a typical falloff of the SUVmean per image plane at the axial ends of the FOV and the region included in calculation of the MAD.

FIGURE 2.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 2.

(Left) Typical ROI used for uniformity analysis. (Right) Example of typical plot of mean SUVs per plane in ROI and calculation of maximum (max.) axial deviation (dev.) (MAD).

For the uniform phantom, the average SUV and MAD results obtained for static brain and body protocols during the T0 time period are summarized in Tables 1 and 2.

View this table:
  • View inline
  • View popup
TABLE 1

Uniform Phantom Results for Static Brain Protocol During T0 Period

View this table:
  • View inline
  • View popup
TABLE 2

Uniform Phantom Results for Static Body Protocol During T0 Period

ACR-Approved Phantom Results

Figure 3 shows a typical image of the ACR-approved PET phantom showing the ROIs used for analysis. The contrast recovery coefficients plotted as a function of the cylinder diameters for static brain and body acquisitions with each of the 3 PET scanner manufacturers are shown in Figure 4. The error bars represent the SD for each data point.

FIGURE 3.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 3.

Image of ACR-approved PET phantom. Red ROIs are used for SUVmax, and blue ROIs are used for SUVpeak. Also shown are large background ROI (purple) and smaller green ROIs for cold cylinders. Latter ROIs were not used in this analysis.

FIGURE 4.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 4.

Recovery coefficients (defined as ratio of measured SUV to true SUV) as function of cylinder diameters for static brain (A) and body (B) acquisitions.

Brain Test Case Results

In some cases, a field in the DICOM header was not properly populated; this problem prevented SUV calculation. The most common reasons for this problem were that the anonymization routine removed a required field, the operator did not enter a required data field in the acquisition interface, and a required DICOM field was changed or removed at some point in the processing.

Body Test Case Liver SUVs

Table 3 shows the liver SUVs for the body test cases during the T0 period; there appeared to be a divergence of average liver values between manufacturers. Despite careful evaluation of all aspects of the image acquisition protocols and processing chain, no systematic cause was found.

View this table:
  • View inline
  • View popup
TABLE 3

Average Liver SUVs by Manufacturer

Comparisons of T0 and T2

For the T0 period, 25 scanners (38%) passed without any core laboratory intervention, 30 scanners (46%) passed after the second submission, and 9 scanners (14%) required more than 2 submissions to pass (Table 4). The most common problems were SUV outside specifications and incomplete submission (Table 5). For the 50 issues that were cataloged, 21 likely were linked to system calibration problems (uniformity problem and SUV outside specifications). The remaining 29 issues were related to operator error. The total number of issues was not the same as the number of scanner disqualifications because some qualification attempts had multiple issues.

View this table:
  • View inline
  • View popup
TABLE 4

Differences in Scanner Qualifications for 3 Time Periods

View this table:
  • View inline
  • View popup
TABLE 5

Frequencies of Specific Issues During Scanner Qualification*

For the T2 period, data were submitted for 44 scanners that had also submitted data during the T0 period. All 44 scanners eventually passed the qualification review; 31 passed without any CQIE core laboratory intervention. Table 6 shows a comparison of T0 and T2 results after the initial review in the core laboratory. Eleven of 17 scanners that passed initially during T0 passed without any intervention during T2. Twenty of 27 scanners that initially did not pass during T0 passed without any intervention during T2.

View this table:
  • View inline
  • View popup
TABLE 6

Cross-Comparison of Passing Status on Initial Review for T0 and T2 Periods for 44 Scanners That Had Submitted Data for Both Qualification Reviews*

DISCUSSION

Our primary finding was that there was a lower frequency of qualification issues during the T2 period than during the T0 period (Tables 4–6). These data indicate that a consistent scanner qualification process helps to ensure standardized scanner performance throughout the entirety of a trial. In the T0 period, 50 issues were identified with the data submissions for the 65 scanners. Quantification problems, which can be due to system calibration problems, accounted for 21 of the 50 issues (42%). The other 29 issues (58%) were attributed to user error, which should be reduced by training. In the T2 period, only 14 issues were identified with data submissions for 44 scanners; 3 of the 14 (21%) were due to quantification problems, and the remaining 11 (79%) were attributed to user error. The lower overall rate of issues with submissions in the T2 period likely indicates that the sites better understood the submission process and were more comfortable with the requirements, so that fewer mistakes and omissions occurred. The lower rate of quantification problems may also indicate that the sites were more familiar with the analysis performed in the core laboratory and the passing criteria used, so that they were less likely to submit data that failed their internal analyses. Understanding the data analysis and passing criteria may also have made the sites more sensitive to changes in performance and more likely to address potential problems with quantitative imaging sooner. However, the fact that approximately 7% of the scanners (3/44) had quantification issues that required recalibration suggests the need for periodic requalification.

The quantitative phantom results were mostly consistent between manufacturers. The uniform phantom results showed the average SUVs to be within 1 SD of the expected value of 1. The MAD for the body FOV cylinders was consistent between manufacturers, but for the brain FOV cylinder, the MAD for the Philips system was higher than that for the other 2 manufacturers. This finding may have been related to the lack of postprocessing smoothing, which is used by GE and Siemens but not by Philips. The ACR-approved phantom results for both brain and body were consistent between manufacturers. There were some differences between manufacturers, as shown in Figure 4, but the differences were within the error bars. In general, the quantitative differences between manufacturers were small.

For the current project, the inclusion of the ACR-approved phantom did not appear to add value to the qualification process, because all scanners that passed the uniform phantom analysis passed the ACR-approved phantom analysis, unless there was a phantom filling problem. This finding could have been due to the relatively wide acceptance criteria currently used for standard ACR submissions of clinical PET scanners, which were adopted for CQIE qualification. It may be appropriate to use narrower acceptance criteria for the ACR-approved phantom to better assess differences in contrast recovery that may arise from different reconstruction parameters. This discussion highlights an unresolved issue: Although it is clear that some level of qualification and routine quality assurance/quality control (QA/QC) should be included in clinical trials using quantitative imaging, the relationships between the type and degree of the qualification (and QA/QC procedures) and the quality (e.g., bias or precision) of the image data have not been established. Thus, if a higher degree of variance or bias in imaging data could be tolerated without study power being affected, then imaging-based trials might use excess resources for qualification and QA/QC to improve data quality unnecessarily or, alternatively, sites might be excluded unnecessarily—in turn, slowing down accrual. However, the opposite scenario is also possible; less rigorous qualification and QA/QC policies might allow for increased accrual rates, but at the cost of underpowering a study because of increased signal variation. Optimizing QA/QC procedures for imaging trials can, in theory, shift the power or accrual rate–versus–cost curve substantially toward better optimization of imaging trials. More evaluation of these trade-offs affecting accrual and study power is needed.

The CQIE project was affected by a lack of accrual during the T1 period. This lack of accrual may have been due to some imaging facilities giving the CQIE qualification a low priority, leading to an increased number of incomplete or faulty submissions and difficulty resolving issues. During the T2 period, the CQIE program intensified the requests for qualifications, leading to increased accrual. These issues also applied to the MRI and CT CQIE qualifications and are discussed in more detail in the CQIE overview (27).

Some sites also expressed hesitation to commit to another qualification regimen because of the time required to maintain other qualifications and accreditations, such as those from the ACR, Eastern Cooperative Oncology Group (ECOG)–ACRIN, Society of Nuclear Medicine and Molecular Imaging (SNMMI), and individual study sponsors. Committing to another qualification program was seen as overly burdensome.

The ACR, ECOG-ACRIN, SNMMI, and CQIE programs all require sites to submit phantom and patient image data to a core laboratory for analysis. The ACR and SNMMI programs require a fee to be paid for accreditation/qualification, but the ECOG-ACRIN and CQIE do not require a fee for qualification. The ECOG-ACRIN uses a uniform phantom, which should be provided by the manufacturers of all scanners, for qualification. The ACR and SNMMI use more complicated resolution phantoms; the ACR uses a cylindric phantom with fillable cylinders attached to the lid, and the SNMMI uses a chest simulator phantom that is more anthropomorphic and has spheres embedded throughout.

Unifying qualification and accreditation criteria between qualifying agencies and clinical research organizations would likely require the adoption of different levels of qualification. Depending on the specific aims of a given trial, more or less variability might be acceptable and would require more or less rigorous scanner qualification. Unifying qualification programs would require the adoption of standard uniformity and contrast phantoms industry-wide and agreement on filling and scanning procedures, analysis methodology, and passing criteria. Given the time and resources already invested into various qualification and accreditation programs, it will be difficult to develop a single qualification methodology acceptable to all organizations. However, if research organizations were more transparent about their qualification programs, including specifics about the analyses being performed and passing criteria, other organizations could more easily evaluate the qualification needs for their trials, compare the rigorousness of various qualification programs, and choose to accept specific organizations’ qualification processes.

CONCLUSION

The results of the CQIE process showed that periodic requalification may decrease the frequency of deficient data submissions. The results suggested that, as sites become more aware of the qualification process and passing criteria, they will be more likely to address problems before submitting qualification data to a core laboratory or study sponsor.

The CQIE project also highlighted the concern within imaging facilities about the burden of maintaining different qualifications and accreditations. Discussions with personnel at various facilities emphasized the need to develop a common set of qualification criteria across various research organizations to reduce the burden on imaging facilities of participating in many different clinical trials. Accomplishing this goal may encourage facilities to participate in a larger number of multicenter clinical trials.

Finally, for quantitative imaging–based trials, further evaluation of the relationships between the level of the qualification (e.g., bias or precision) and the quality of the image data, accrual rates, and study power is needed.

DISCLOSURE

This work was supported in part by NCI–SAIC subcontract 10XS070; NIH grants U01CA148131, U01CA190254, and U10CA180820; and NIH–NCI contract HHSN261200800001E. No other potential conflict of interest relevant to this article was reported.

Acknowledgments

We gratefully acknowledge the following individuals for their time and effort in support of this publication: Mark Rosen, Deborah Harbison, Joseph Maffei, Mark Muzi, Finbarr O’Sullivan, Brian Elston, Darin Byrd, and Mattie Lloyd.

Footnotes

  • Published online Mar. 2, 2017.

  • © 2017 by the Society of Nuclear Medicine and Molecular Imaging.

REFERENCES

  1. 1.↵
    1. Pantel AR,
    2. Mankoff DA
    . Molecular imaging to guide systemic cancer therapy: illustrative examples of PET imaging cancer biomarkers. Cancer Lett. 2017;387:25–31.
    OpenUrl
  2. 2.
    1. Bengtsson T,
    2. Hicks RJ,
    3. Peterson A,
    4. Port RE
    . 18F-FDG PET as a surrogate biomarker in non–small cell lung cancer treated with erlotinib: newly identified lesions are more informative than standardized uptake value. J Nucl Med. 2012;53:530–537.
    OpenUrlAbstract/FREE Full Text
  3. 3.
    1. Cornelis F,
    2. Storchios V,
    3. Violari E,
    4. et al
    . 18F-FDG PET/CT is an immediate imaging biomarker of treatment success after liver metastasis ablation. J Nucl Med. 2016;57:1052–1057.
    OpenUrlAbstract/FREE Full Text
  4. 4.
    1. Humbert O,
    2. Riedinger JM,
    3. Charon-Barra C,
    4. et al
    . Identification of biomarkers including 18FDG-PET/CT for early prediction of response to neoadjuvant chemotherapy in triple-negative breast cancer. Clin Cancer Res. 2015;21:5460–5468.
    OpenUrlAbstract/FREE Full Text
  5. 5.
    1. Kim JW,
    2. Oh JS,
    3. Roh JL,
    4. et al
    . Prognostic significance of standardized uptake value and metabolic tumour volume on 18F-FDG PET/CT in oropharyngeal squamous cell carcinoma. Eur J Nucl Med Mol Imaging. 2015;42:1353–1361.
    OpenUrl
  6. 6.
    1. Kostakoglu L,
    2. Duan F,
    3. Idowu MO,
    4. et al
    . A phase II study of 3′-deoxy-3′-18F-fluorothymidine PET in the assessment of early response of breast cancer to neoadjuvant chemotherapy: results from ACRIN 6688. J Nucl Med. 2015;56:1681–1689.
    OpenUrlAbstract/FREE Full Text
  7. 7.
    1. Pimiento JM,
    2. Davis-Yadley AH,
    3. Kim RD,
    4. et al
    . Metabolic activity by 18F-FDG-PET/CT is prognostic for stage I and II pancreatic cancer. Clin Nucl Med. 2016;41:177–181.
    OpenUrl
  8. 8.↵
    1. Siva S,
    2. Deb S,
    3. Young RJ,
    4. et al
    . 18F-FDG PET/CT following chemoradiation of uterine cervix cancer provides powerful prognostic stratification independent of HPV status: a prospective cohort of 105 women with mature survival data. Eur J Nucl Med Mol Imaging. 2015;42:1825–1832.
    OpenUrl
  9. 9.↵
    1. Adams MC,
    2. Turkington TG,
    3. Wilson JM,
    4. Wong TZ
    . A systematic review of the factors affecting accuracy of SUV measurements. AJR. 2010;195:310–320.
    OpenUrlCrossRefPubMed
  10. 10.
    1. Boellaard R
    . Standards for PET image acquisition and quantitative data analysis. J Nucl Med. 2009;50(suppl 1):11S–20S.
    OpenUrlAbstract/FREE Full Text
  11. 11.
    1. Kinahan PE,
    2. Fletcher JW
    . Positron emission tomography-computed tomography standardized uptake values in clinical practice and assessing response to therapy. Semin Ultrasound CT MR. 2010;31:496–505.
    OpenUrlCrossRefPubMed
  12. 12.↵
    1. Shankar LK,
    2. Hoffman JM,
    3. Bacharach S,
    4. et al
    . Consensus recommendations for the use of 18F-FDG PET as an indicator of therapeutic response in patients in National Cancer Institute trials. J Nucl Med. 2006;47:1059–1066.
    OpenUrlFREE Full Text
  13. 13.↵
    1. Doot RK,
    2. Scheuermann JS,
    3. Christian PE,
    4. Karp JS,
    5. Kinahan PE
    . Instrumentation factors affecting variance and bias of quantifying tracer uptake with PET/CT. Med Phys. 2010;37:6035–6046.
    OpenUrlCrossRefPubMed
  14. 14.↵
    1. Doot RK,
    2. Thompson T,
    3. Greer BE,
    4. et al
    . Early experiences in establishing a regional quantitative imaging network for PET/CT clinical trials. Magn Reson Imaging. 2012;30:1291–1300.
    OpenUrlCrossRefPubMed
  15. 15.↵
    1. Byrd DW,
    2. Doot RK,
    3. Allberg KC,
    4. et al
    . Evaluation of cross-calibrated 68Ge/68Ga phantoms for assessing PET/CT measurement bias in oncology imaging for single- and multicenter trials. Tomography. 2016;2:353–360.
    OpenUrl
  16. 16.↵
    1. Doot RK,
    2. Pierce LA,
    3. Byrd D,
    4. Elston B,
    5. Allberg KC,
    6. Kinahan PE
    . Biases in multicenter longitudinal PET standardized uptake value measurements. Transl Oncol. 2014;7:48–54.
    OpenUrlCrossRefPubMed
  17. 17.↵
    1. Tahari AK,
    2. Wahl RL
    . Quantitative FDG PET/CT in the community: experience from interpretation of outside oncologic PET/CT exams in referred cancer patients. J Med Imaging Radiat Oncol. 2014;58:183–188.
    OpenUrl
  18. 18.↵
    1. Fahey FH,
    2. Kinahan PE,
    3. Doot RK,
    4. Kocak M,
    5. Thurston H,
    6. Poussaint TY
    . Variability in PET quantitation within a multicenter consortium. Med Phys. 2010;37:3660–3666.
    OpenUrlCrossRefPubMed
  19. 19.↵
    1. Weber WA,
    2. Gatsonis CA,
    3. Mozley PD,
    4. et al
    . Repeatability of 18F-FDG PET/CT in advanced non–small cell lung cancer: prospective assessment in 2 multicenter trials. J Nucl Med. 2015;56:1137–1143.
    OpenUrlAbstract/FREE Full Text
  20. 20.↵
    1. Kumar V,
    2. Nath K,
    3. Berman CG,
    4. et al
    . Variance of SUVs for FDG-PET/CT is greater in clinical practice than under ideal study settings. Clin Nucl Med. 2013;38:175–182.
    OpenUrlCrossRefPubMed
  21. 21.↵
    1. Scheuermann JS,
    2. Saffer JR,
    3. Karp JS,
    4. Levering AM,
    5. Siegel BA
    . Qualification of PET scanners for use in multicenter cancer clinical trials: the American College of Radiology Imaging Network experience. J Nucl Med. 2009;50:1187–1193.
    OpenUrlAbstract/FREE Full Text
  22. 22.↵
    1. Barrington SF,
    2. Mackewn JE,
    3. Schleyer P,
    4. et al
    . Establishment of a UK-wide network to facilitate the acquisition of quality assured FDG-PET data for clinical trials in lymphoma. Ann Oncol. 2011;22:739–745.
    OpenUrlAbstract/FREE Full Text
  23. 23.↵
    1. Boellaard R,
    2. Oyen WJ,
    3. Hoekstra CJ,
    4. et al
    . The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials. Eur J Nucl Med Mol Imaging. 2008;35:2320–2333.
    OpenUrlCrossRefPubMed
  24. 24.↵
    1. Boellaard R,
    2. Delgado-Bolton R,
    3. Oyen WJ,
    4. et al
    . FDG PET/CT: EANM procedure guidelines for tumour imaging—version 2.0. Eur J Nucl Med Mol Imaging. 2015;42:328–354.
    OpenUrlCrossRefPubMed
  25. 25.
    QIBA profile: FDG-PET/CT as an imaging biomarker measuring response to cancer therapy—version 1.13, technically confirmed version. RSNA website. http://qibawiki.rsna.org/images/1/1f/QIBA_FDG-PET_Profile_v113.pdf. Published November 18, 2016. Accessed March 22, 2017.
  26. 26.↵
    1. Graham MM,
    2. Wahl RL,
    3. Hoffman JM,
    4. et al
    . Summary of the UPICT protocol for 18F-FDG PET/CT imaging in oncology clinical trials. J Nucl Med. 2015;56:955–961.
    OpenUrlAbstract/FREE Full Text
  27. 27.↵
    1. Rosen M,
    2. Kinahan PE,
    3. Gimpel JF,
    4. et al
    . Performance observations of scanner qualification of NCI-designated cancer centers: results from the Centers of Quantitative Imaging Excellence (CQIE) program. Acad Radiol. 2017;24:232–245.
    OpenUrl
  28. 28.↵
    1. Mou T,
    2. Huang J,
    3. Zhang Y,
    4. et al
    . Spatial covariance characteristics in a collection of 3-D PET scanners used in clinical imaging trials. In: Proceedings of the 2014 IEEE Nuclear Science Symposium and Medical Imaging Conference. Seattle, WA: IEEE; 2014.
  29. 29.↵
    1. Rosset A,
    2. Spadola L,
    3. Ratib O
    . OsiriX: an open-source software for navigating in multidimensional DICOM images. J Digit Imaging. 2004;17:205–216.
    OpenUrlCrossRefPubMed
  30. 30.↵
    1. MacFarlane CR
    ; American College of Radiologists. ACR accreditation of nuclear medicine and PET imaging departments. J Nucl Med Technol. 2006;34:18–24.
    OpenUrlAbstract/FREE Full Text
  • Received for publication November 9, 2016.
  • Accepted for publication February 8, 2017.
PreviousNext
Back to top

In this issue

Journal of Nuclear Medicine: 58 (7)
Journal of Nuclear Medicine
Vol. 58, Issue 7
July 1, 2017
  • Table of Contents
  • Table of Contents (PDF)
  • About the Cover
  • Index by author
Print
Download PDF
Article Alerts
Sign In to Email Alerts with your Email Address
Email Article

Thank you for your interest in spreading the word on Journal of Nuclear Medicine.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Qualification of National Cancer Institute–Designated Cancer Centers for Quantitative PET/CT Imaging in Clinical Trials
(Your Name) has sent you a message from Journal of Nuclear Medicine
(Your Name) thought you would like to see the Journal of Nuclear Medicine web site.
Citation Tools
Qualification of National Cancer Institute–Designated Cancer Centers for Quantitative PET/CT Imaging in Clinical Trials
Joshua S. Scheuermann, Janet S. Reddin, Adam Opanowski, Paul E. Kinahan, Barry A. Siegel, Lalitha K. Shankar, Joel S. Karp
Journal of Nuclear Medicine Jul 2017, 58 (7) 1065-1071; DOI: 10.2967/jnumed.116.186759

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Share
Qualification of National Cancer Institute–Designated Cancer Centers for Quantitative PET/CT Imaging in Clinical Trials
Joshua S. Scheuermann, Janet S. Reddin, Adam Opanowski, Paul E. Kinahan, Barry A. Siegel, Lalitha K. Shankar, Joel S. Karp
Journal of Nuclear Medicine Jul 2017, 58 (7) 1065-1071; DOI: 10.2967/jnumed.116.186759
Twitter logo Facebook logo LinkedIn logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One
Bookmark this article

Jump to section

  • Article
    • Abstract
    • MATERIALS AND METHODS
    • RESULTS
    • DISCUSSION
    • CONCLUSION
    • DISCLOSURE
    • Acknowledgments
    • Footnotes
    • REFERENCES
  • Figures & Data
  • Info & Metrics
  • PDF

Related Articles

  • This Month in JNM
  • PubMed
  • Google Scholar

Cited By...

  • Conceptualising centres of excellence: a scoping review of global evidence
  • Conceptualizing centers of excellence: A global evidence
  • Google Scholar

More in this TOC Section

Oncology

  • The value of 18-FDG PET/CT as a supplemental screening method to overcome the challenges of dense breast imaging
  • False positive radioactive iodine uptake in Nabothian cyst in a patient with thyroid cancer.
  • Rediscovering the indispensable role of FDG PET in kidney cancer imaging
Show more Oncology

Clinical

  • TauIQ: A Canonical Image Based Algorithm to Quantify Tau PET Scans
  • Dual PET Imaging in Bronchial Neuroendocrine Neoplasms: The NETPET Score as a Prognostic Biomarker
  • Addition of 131I-MIBG to PRRT (90Y-DOTATOC) for Personalized Treatment of Selected Patients with Neuroendocrine Tumors
Show more Clinical

Similar Articles

Keywords

  • CQIE
  • PET Qualification
  • quantitative imaging
SNMMI

© 2025 SNMMI

Powered by HighWire